PHYSIOFLOW ENDURO

K103283 · Vasocom, Inc. · DSB · Dec 3, 2010 · Cardiovascular

Device Facts

Record IDK103283
Device NamePHYSIOFLOW ENDURO
ApplicantVasocom, Inc.
Product CodeDSB · Cardiovascular
Decision DateDec 3, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.2770
Device ClassClass 2

Indications for Use

The PhysioFlow Enduro PF07 and PhysioFlow System Model PF05 have the same intended use/indications for use. Both devices are intended for use in adults only to noninvasively measures cardiac output and other related cardiac parameters, including the following: cardiac output, cardiac index, contractility index, maximum value dZ/dt, end diastolic volume, ejection fraction, heart rate, left cardiac work index, pre-ejection period, stroke volume, systemic vascular resistance, systemic vascular resistance index, thoracic fluid index, ventricular ejection time, and base impedance. Both devices are intended for use under the direct supervision of a licensed practitioner or personnel trained in its proper use within a hospital or facility providing healthcare.

Device Story

PhysioFlow Enduro PF07 is a portable, battery-powered, noninvasive hemodynamic monitor; utilizes thoracic electrical bioimpedance technology to measure cardiac output and related parameters. Device attaches to patient via six-lead cable and silver/silver chloride skin electrodes. Electronic unit captures/stores up to 24 hours of data; transmits via Bluetooth or USB to host computer running PhysioFlow PF107 software. Software performs signal processing, parameter computation, and report generation. Used in hospitals or healthcare facilities under supervision of licensed practitioners or trained personnel. Output allows clinicians to monitor cardiac function and hemodynamic status, aiding in clinical decision-making.

Clinical Evidence

Bench testing and simulated use validation performed. Design verification confirmed all pre-specified acceptance criteria met. Product validation in simulated use setting compared subject device to predicate; results demonstrated <4% difference across measured parameters, below 5% clinical variability threshold.

Technological Characteristics

Thoracic electrical bioimpedance technology; silver/silver chloride skin electrodes (Skintact FS-50). Portable electronic unit (<250g with batteries). Connectivity: Bluetooth wireless, USB wired. Software: PhysioFlow PF107 (MS Windows-based host). Standalone recorder capability (24-hour storage).

Indications for Use

Indicated for use in adults only to noninvasively measure cardiac output and related hemodynamic parameters (CI, CO, CTI, dZ/dt max, EDV, EF, HR, LCWI, PEP, SV, SVR, SVRI, TFI, VET, Z0). Intended for use under direct supervision of licensed healthcare practitioner or trained personnel in hospital or healthcare facility settings.

Regulatory Classification

Identification

An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.

Special Controls

*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Vasocom, Inc. # K103283 CONFIDENTIAL # 510(k) SUMMARY This 510(k) summary is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.87 Establishment Registration Number: 3005467960 DEC - 3 2010 Address of Manufacturer: 2014 Ford Road, Unit G Bristol, PA 19007 Contact Person: James Gunnerson President Date Prepared: Oct. 26, 2010 PhysioFlow Enduro Model PF07 Common or Usual Name: Trade or Proprietary Name: Classification Name: Noninvasive hemodynamic monitor Impedance plethysmograph (21 CFR 870.2770) Product code: DBS (plethysmograph, impedance) Class II Predicate Device Identification: PhysioFlow System Model PF05 (K060387) #### Device Description: The PhysioFlow Enduro Model PF07 is a noninvasive, hemodynamic monitor that utilizes thoracic electrical bioimpedance technology to measure cardiac output and related cardiac parameters. It consists of a small, portable, lightweight (less than 250 gm with batteries) {1}------------------------------------------------ ### CONFIDENTIAL #### Vasocom, Inc. electronic unit that attaches to the patient via a six lead, patient cable using commerciallyavailable silver/silver chloride skin electrodes. The electronic unit incorporates recorder capabilities, enabling it to capture and store up to 24 hours of monitor data for later transmission to a host computer running the PhysioFlow PF107 software. Communication options include Bluetooth wireless and USB wired technology. The PhysioFlow PF107 software performs multiple tasks including signal processing and analysis, measured parameter computation, user interface display and control, measurement process control, event marker management, and output display/report generation. Available accessories include the patient cable, a Bluetooth antenna. USB cable, and electronic unit carrying case. #### Intended use and comparison to predicate devices: The PhysioFlow Enduro PF07 and PhysioFlow System Model PF05 have the same intended use/indications for use. Both devices are intended for use in adults only to noninvasively measures cardiac output and other related cardiac parameters, including the following: cardiac output, cardiac index, contractility index, maximum value dZ/dt, end diastolic volume, ejection fraction, heart rate, left cardiac work index, pre-ejection period, stroke volume, systemic vascular resistance, systemic vascular resistance index, thoracic fluid index, ventricular ejection time, and base impedance. Both devices are intended for use under the direct supervision of a licensed practitioner or personnel trained in its proper use within a hospital or facility providing healthcare. # Technological characteristics and comparison to predicate devices: The PhysioFlow Enduro Model PF07 is a modification of the PhysioFlow System Model PF05. It utilizes the same thoracic electrical bioimpedance technology to measure the same cardiac output and related cardiac parameters, but its size has been reduced to make it portable and lightweight (less than 250 gm with batteries). Compared with the Model PF05, technological changes also now enable the use of battery power and communication using Bluetooth wireless and USB wired technology with a MS Windows based host computer running an upgraded version of the PF05 software (i.e., PhysioFlow PF107). In addition, software and hardware changes now enable a standalone recorder capability, enabling of up to 24 hours capture of monitor data for later transmission to the host PC. Use of the same silver/silver chloride skin electrodes (Skintact FS-50 Pre-gelled Electrodes) is recommended. The differences in technological characteristics between the subject and predicate devices do not raise new types of safety or effectiveness questions. Accepted scientific methods, such as performance (bench) testing, do exist for assessing the effect of the differences in characteristics. {2}------------------------------------------------ #### CONFIDENTIAL # Vasocom, Inc. # Summary of performance data: Design verification tests were performed on representative samples of the PhysioFlow Enduro Model PF07 and PhysioFlow PF107 software as a result of a comprehensive product development and risk management plan. All tests were verified to meet the pre-specified acceptance criteria. Product validation tests were performed in a simulated use setting and compared subject and predicate device performance. The results demonstrate a difference between the subject and predicate devices of less than 4% across the measured parameters, which is below the acceptable clinical variability of 5%. As such, the PhysioFlow Enduro Model PF07 is considered substantially equivalent to the PhysioFlow System Model PF05. James A. Gunnerson Special 510(k) - PhysioFlow Enduro PF07 October 26, 2010 {3}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 VasoCOM, Inc. c/o Mr. James Gunnerson President, 2014 Ford Road, Unit G Bristol, PA 19007 DEC - 3 2010 Re: K103283 Trade/Device Name: PhysioFlow® Enduro Model PF07 Regulatory Number: 21 CFR 870.2770 Regulation Name: Impedance Plethysmograph Regulatory Class: II (two) Product Code: 74 DSB Dated: October 27, 2010 Received: November 5, 2010 Dear Mr. Gunnerson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {4}------------------------------------------------ # Page 2 - Mr. James Gunnerson comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ CONFIDENTIAL Vasocom, Inc. | Indications for Use | | |---------------------|--| | K103283 | | 510(k) Number (if known): Device Name: PhysioFlow Enduro Model PF-07 DEC - 3 2010 Indications for Use: Indicated for use in adults only. The PhysioFlow Enduro PF07 noninvasively measures cardiac output and other related cardiac parameters. These parameters include: | CI | Cardiac Index | |-------------|------------------------------------| | CO | Cardiac Output | | CTI | Contractility Index | | dZ / dt max | Maximum value dZ / dt | | EDV | End Diastolic Volume | | EF | Ejection Fraction | | HR | Heart Rate | | LCWI | Left Cardiac Work Index | | PEP | Pre-Ejection Period | | SV | Stroke Volume | | SVR | Systemic Vascular Resistance | | SVRI | Systemic Vascular Resistance Index | | TFI | Thoracic Fluid Index | | VET | Ventricular Ejection Time | | ZO | Base Impedance | The Enduro System Model PF-07 is intended for use under the direct supervision of a licensed healthcare practitioner or personnel trained in its proper use within a hospital or facility providing healthcare. (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) # Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use: ____________________________________________________________________________________________________________________________________________________________ X (21 CFR 801 Subpart D) AND/OR Over-the-Counter Use: (21 CFR 807 Subpart e) W 25 Special 510(k) - PhysioFlow Enduro PF07 October 26, 2010 slon Slan-Off) of Cardiovascular Dev 10(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...