← Product Code KJJ · K103244

# QMIX 2IN1 ENDODONTIC IRRIGATING SOLUTION (K103244)

_Dentsply Intl., Inc. · KJJ · Feb 14, 2011 · DE · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K103244

## Device Facts

- **Applicant:** Dentsply Intl., Inc.
- **Product Code:** KJJ
- **Decision Date:** Feb 14, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** DE
- **Attributes:** Therapeutic

## Intended Use

QMix™ 2in1 Endodontic Irrigating Solution is a premixed dual-action device that cleanses and disinfects the root canal system after endodontic instrumentation.

## Device Story

QMix™ 2in1 is a premixed chemical irrigating solution used by dentists during endodontic procedures. Applied after mechanical instrumentation of the root canal, the solution acts to remove the smear layer and disinfect the canal system. It functions as a chemical cleanser to prepare the canal for subsequent obturation. The device is intended for professional use in a clinical dental setting.

## Clinical Evidence

No clinical studies were conducted. Efficacy and biocompatibility were demonstrated via non-clinical in vitro and ex vivo testing.

## Technological Characteristics

Premixed chemical irrigating solution. Components consist of materials previously used in legally marketed dental devices. No electronic, software, or mechanical components.

## Predicate Devices

- BioPure MTAD Root Canal Cleanser ([K053167](/device/K053167.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

FEB 1 4 2011

K103244

Image /page/0/Picture/2 description: The image shows the word "DENTSPLY" in a bold, sans-serif font. The letters are all capitalized and black. The letters are tightly spaced, and the overall appearance is clean and modern. The image is likely a logo or branding element for a company named DENTSPLY.

## SECTION 5. 510(k) SUMMARY

for

#### DENTSPLY International

World Headquarters Susquehanna Commerce Center 221 West Philadelphia Street ork. PA 17405-0872 717) 845-7511 (voice) 717) 849-4343 (fax) ww.dentsply.com

# QMix™ 2in1 Endodontic Irrigating Solution

- 1. Submitter Information: DENTSPLY International Susquehanna Commerce Center 221 West Philadelphia Street York, PA 17405

| Contact Person:   | Helen Lewis  |
|-------------------|--------------|
| Telephone Number: | 717-849-4229 |
| Fax Number:       | 717-849-4343 |

01 February 2011 Date Prepared:

- 2. Device Name:
	- Proprietary Name: .

QMix™ 2in1 Endodontic Irrigating Solution

- Cleanser, Root Canal
- CFR Number: .

Classification Name:

N/A Unclassified

- Device Class: . Product Code: KJJ .
- 3. Predicate Device: BioPure MTAD Root Canal Cleanser, DENTSPLY International, K053167
- Description of Device: 4.

.

QMix™ 2in1 Endodontic Irrigating Solution is a premixed dual-action device that cleanses and disinfects the root canal system by removing the smear layer and killing bacteria after endodontic instrumentation.

- റ് Indications for Use: QMix™ 2in1 Endodontic Irrigating Solution is a premixed dual-action device that cleanses and disinfects the root canal system after endodontic instrumentation.
- Description of Safety and Substantial Equivalence: Q. Technological Characteristics.

All of the components found in QMix™ 2in1 Endodontic Irrigating Solution have been used in legally marketed devices and were found safe for dental use. We believe that prior use of components in legally marketed devices, the performance and biocompatibility data provided support the safety and effectiveness of QMix™ 2in1 Endodontic Irrigating Solution for the indicated uses.

{1}------------------------------------------------

### Non-Clinical Performance Data.

The efficacy and biocompatibility of QMix™ 2in1 Endodontic Irrigating Solution was demonstrated via non-clinical in vitro and ex vivo studies.

## Clinical Performance Data.

No clinical studies were conducted on this device.

# Conclusion as to Substantial Equivalence

QMix™ 2in1 Endodontic Irrigating Solution, to be manufactured by DENTSPL Y International, is substantially equivalent to the currently cleared and marketed BioPure MTAD root canal cleanser.

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows a logo for the Department of Health & Human Services USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the circumference. To the right of the circle is a stylized symbol that resembles three abstract human figures or waves, stacked vertically.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Helen Lewis Director of Corporate Compliance and Regulatory Affairs DENTSPLY International Susquehanna Commerce Center 221 West Philadelphia Street York, Pennsylvania 17405

FFB 1 4 201

Re: K103244

Trade/Device Name: QMix" 2in1 Endodontic Irrigating Solution Regulatory Class: Unclassified Product Code: KJJ Dated: February 1, 2011 Received: February 2, 2011

Dear Ms. Lewis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Ms. Lewis

Enclosure

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

N for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

## SECTION 4. INDICATIONS FOR USE STATEMENT

510(k) Number (if known): K103244

Device Name: QMix™ 2in1 Endodontic Irrigating Solution

Indications for Use:

QMix™ 2in1 Endodontic Irrigating Solution is a premixed dual-action device that cleanses and disinfects the root canal system after endodontic instrumentation.

Prescription Use ____X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Buenger

(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K102241

---

**Source:** [https://fda.innolitics.com/device/K103244](https://fda.innolitics.com/device/K103244)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
