← Product Code [KWP](/productcode/KWP) · K103205

# BRIDGEPOINT  SPINOUS PROCESS CAMP- POSTERIOR FIXATION SYSTEM (K103205)

_Alphatec Spine, Inc. · KWP · Jun 2, 2011 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K103205

## Device Facts

- **Applicant:** Alphatec Spine, Inc.
- **Product Code:** [KWP](/productcode/KWP.md)
- **Decision Date:** Jun 2, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3050
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The BridgePoint™ Spinous Process Clamp System is a posterior, non-pedicle supplemental fixation device, intended for use in non-cervical spine (T1-S1). It is intended for plate fixation/attachment to spinous process for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); and/or tumor. It is not intended for stand alone use.

## Device Story

Implantable posterior fixation system; clamps bilaterally to spinous processes. Device consists of two sets of plates (male/female components) coupled by two rigid posts that pivot at midline; variable plate length achieved via sliding mechanism; locked by lateral set screw. Used by surgeons in clinical settings to provide supplemental stability during spinal fusion procedures. Output is mechanical stabilization of the spinous process; intended to facilitate fusion in conjunction with other fixation methods. Benefits include supplemental support for spinal segments affected by degenerative disease, trauma, or tumors.

## Clinical Evidence

Bench testing only. Static and fatigue performance testing conducted per recognized consensus standards and FDA guidance document 'Spinal System 510(k)s - Guidance for Industry and FDA Staff' to demonstrate mechanical integrity and substantial equivalence.

## Technological Characteristics

Implantable spinous process clamp; multi-component design with sliding male/female plates and lateral set screw locking mechanism. Materials identical to previously cleared Alphatec Spine products. Mechanical fixation principle; non-pedicle posterior attachment.

## Regulatory Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

## Predicate Devices

- NuVasive® Spinous Process Plate System (k073278)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JUN - 2 2011

# BridgePoint™ Spinous Process Clamp- Posterior Fixation System

Alphatec Spine

# 510(k) SUMMARY September 2010

Submitter: Alphatec Spine, Inc. 5818 El Camino Real Carlsbad, CA 92008 Direct: (760) 494-6739 Fax: (760) 431-0289

Official Contact: Olga Lewis, Regulatory Affairs Specialist

Trade/Model Name: BridgePoint™ Spinous Process Clamp-Posterior Fixation System

Spinal Intervertebral Body Fixation Orthosis Common Name:

Classification Regulation: KWP - Spinal Interlaminal Fixation Orthosis

# Device Description:

The BridgePoint™ Spinous Process Clamp- Posterior Fixation System is an implantable device that clamps bilaterally to the spinous processes. It is a multi-component device consisting of two sets of plates. Each set is composed of a male and female plate. Two sets are coupled by two rigid posts that connect and pivot at the midline. The plates have a male and female component which slide and result in a variable plate length. These sliding plates are locked by a lateral set screw.

## Indications for Use

The BridgePoint™ Spinous Process Clamp System is a posterior, non-pedicle supplemental fixation device, intended for use in non-cervical spine (T1-S1). It is intended for plate fixation/attachment to spinous process for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); and/or tumor. It is not intended for stand alone use.

# Substantial Equivalence:

The BridgePoint™ Spinous Process Clamp System is substantially equivalent in intended use and function to NuVasive® Spinous Process Plate System (k073278)

Confidential and Proprietary

Page 23-2

{1}------------------------------------------------

# BridgePoint™ Spinous Process Clamp- Posterior Fixation System

Image /page/1/Picture/2 description: The image contains a logo or emblem with stylized text. The text "Alphatec Spine" is visible, with the word "Alphatec" appearing above "Spine". The font is somewhat stylized, giving it a unique appearance. The logo is simple and clean, with a focus on the text.

#### Technological Characteristics Comparison:

The BridgePoint™ Spinous Process Clamp System is substantially equivalent to the referenced device in that it is intended to be used for plate fixation/attachment to spinous process for the purpose of achieving supplemental fusion in non-cervical spine (T1-S1).

Material composition is identical to numerous other Alphatec Spine products that have been cleared via the 510(k) process.

#### Nonclinical Performance Data:

Mechanical testing for BridgePoint™ Spinous Process Clamp- Posterior Fixation System was performed that provides reasonable assurance of safety and effectiveness for device's Performance testing was performed per the recognized consensus intended use. standards and per the guidance document, Spinal System 510(k)s - Guidance for Industry This testing documented both static and fatigue performance and FDA Staff. characteristics. This testing clearly demonstrated that the performance characteristics satisfy the requirements of posterior non-pedicle supplemental fixation. As a result of this testing, the BridgePoint™ Spinous Process Clamp System is substantially equivalent to the predicate device.

Page 23-3

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around it. Inside the circle is an abstract symbol that resembles an eagle or bird with outstretched wings.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JUN - 2 2011

Alphatec Spine, Inc. % Ms. Olga Lewis Regulatory Affairs Specialist 5818 El Camino Real Carlsbad, California 92008

Re: K103205

> Trade Name: BridgePoint Spinous Process Clamp - Posterior Fixation System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: K WP Dated: May 09, 2011 Received: May 10, 2011

Dear Ms. Lewis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

Page 2 - Ms. Olga Lewis

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

For M. Thomas c.c.n

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): K103205

Device Name: BridgePoint™ Spinous Process Clamp - Posterior Fixation System

Indications For Use:

The BridgePoint™ Spinous Process Clamp System is a posterior, non-pedicle supplemental fixation device, intended for use in non-cervical spine (T1-S1). It is intended for plate fixation/attachment to spinous process for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); and/or tumor. It is not intended for stand alone use.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

signature

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number_KIO3205

Page 1 of

---

**Source:** [https://fda.innolitics.com/device/K103205](https://fda.innolitics.com/device/K103205)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
