XELIS 3D

K103130 · Infinitt Co., Ltd. · LLZ · Jul 28, 2011 · Radiology

Device Facts

Record IDK103130
Device NameXELIS 3D
ApplicantInfinitt Co., Ltd.
Product CodeLLZ · Radiology
Decision DateJul 28, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device, 3rd-Party Reviewed

Indications for Use

Xelis 3D™ is a software device intended for viewing of images acquired from various imaging modalities: CT. MR. CR. DR. US. PET and other DICOM compliant medical imaging systems when installed on suitable commercial standard hardware. Images and data can be captured, stored, communicated, processed, and displayed within the system and or across computer networks at distributed locations. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary diagnosis or image interpretation. For primary diagnosis, post process DICOM "for presentation" images must be used. Mammographic images should only be viewed with a monitor approved by FDA for viewing mammographic images. It is the User's responsibility to ensure monitor quality, ambient light conditions, and image compression ratios are consistent with the clinical application.

Device Story

XELIS 3D is a software-based PACS application for viewing, manipulating, and reporting DICOM-compliant medical images (CT, MRI, PET, CR, DR, US). Operates on standard commercial Windows-based workstations (32/64-bit). Enables real-time viewing, 3D volume rendering, segmentation, and registration. Connects to PACS servers, workstations, and imaging modalities via DICOM 3.0. Used by radiologists, clinicians, and physicians in clinical settings to interpret images for diagnosis. Does not perform automated detection or diagnosis; relies on human intervention for clinical decision-making. Benefits include centralized image management, enhanced visualization, and efficient workflow for diagnostic interpretation.

Clinical Evidence

Bench testing only. No clinical data provided. System validation testing evaluated all input/output functions and operational modes against predetermined acceptance criteria. Integration and installation verification tests were conducted to ensure performance consistency.

Technological Characteristics

Software-only device; no patient contact. Operates on standard commercial hardware (Windows 32/64-bit). DICOM 3.0 compliant for interoperability. Supports 3D volume rendering, segmentation, and registration. No automated detection algorithms. Not life-sustaining.

Indications for Use

Indicated for viewing, processing, and reporting of medical images (CT, MR, CR, DR, US, PET) from DICOM-compliant systems. Intended for use by radiologists, clinicians, and referring physicians for clinical image interpretation. Contraindicated for primary diagnosis using lossy compressed mammographic images or digitized film screen images.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ JUL 28 2011 103130 1 This 510(k) summary of safety information is being submitted in accordance with the requirements of SMDA 1990. Date Prepared: June 1, 2011 Submitter's Information: 21 CFR 807.92(a)(1) Mr. Tristan Choi, Program Manager INFINITT Co., Ltd. 12F Daerung Post Tower III, 182-4 Guro-dong, Guro-gu, Seoul, 152-050, Korea Tel: +82-2-2194-6673 Fax: +82-2-2194-1698 Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2) | Trade Name: | XELIS 3D | |------------------------|-----------------------------------------| | Common Name: | Picture Archiving Communications System | | Classification Name: | system, image processing, radiological | | Product code: | LLZ | | Device Classification: | 892.2050 | Predicate Device: 21 CFR 807. 92(a)(3) | 510(k) Number | K082990 | K070831 | |--------------------------------------|-------------------------------------------|-----------------------------------------------------------------| | Device Classification Name | System, image<br>processing, radiological | System, image<br>processing, radiological | | Device Name | INFINITT XELIS | Voxar 3D Enterprise with<br>ColonMetrix and PET/CT<br>Perfusion | | Applicant | INFINITT CO., LTD | BARCOVIEW MIS<br>EDINBURGH | | Regulation Number | 892.2050 | 892.2050 | | Classification Product Code | LLZ | LLZ | | Decision Date | 11/20/2008 | 05/22/2007 | | Classification Advisory<br>Committee | Radiology | Radiology | | Type | Traditional | Traditional | ### Device Description: 21 CFR 807 92(a)(4) XELIS 3D is a software package intended for viewing and manipulating DICOMcompliant medical images from CT (computerized tomography), PET (positron emission tomography) or MRI (magnetic resonance imaging) scanners and other imaging modalities. XELIS 3D can be used for real-time viewing, 3D volume rendering, segmentation, registration, and reporting. {2}------------------------------------------------ ## 510(k) Summary of Safety XELIS 3D application conforms to the DICOM 3.0 standard to allow interoperability with other DICOM compliant systems and is based upon the predicate device software used in Infinitt Xelis K082990. XELIS 3D does not include any automated or semi-automated process for the detection of nodules or other shapes. XELIS 3D is supplied to end users in both Windows 32 bit and 64 bit operating systems. The device software can connect to other workstations, PACS server/software, and modalities, through DICOM communication standard. #### Indications for Use: 21 CFR 807 92(a)(5) Xelis 3D™ is a software device intended for viewing of images acquired from various imaging modalities: CT. MR. CR. DR. US, PET and other DICOM compliant medical imaging systems when installed on suitable commercial standard hardware. Images and data can be captured, stored, communicated, processed, and displayed within the system and or across computer networks at distributed locations. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary diagnosis or image interpretation. For primary diagnosis, post process DICOM "for presentation" images must be used. Mammographic images should only be viewed with a monitor approved by FDA for viewing mammographic images. It is the User's responsibility to ensure monitor quality, ambient light conditions, and image compression ratios are consistent with the clinical application. #### Technological Characteristics: 21 CFR 807 92(a)(6) XELIS 3D is a software device that does not contact the patient, nor does it control any life sustaining devices. Diagnosis is not performed by the software but by Radiologists, Clinicians and referring Physicians. A physician, providing ample opportunity for competent human intervention interprets images and information being displayed and printed. The new device and predicate devices are substantially equivalent in the areas of technical characteristics, general function, application, and intended use. The new device does not raise any new potential safety risks and is equivalent in performance to the existing legally marketed devices. #### Nonclinical Testing: The complete system configuration has been assessed and tested at the factory and has passed all in-house testing criteria. The Validation Test Plan was designed to evaluate all input functions, output functions, and actions performed by the XELIS 3D software in each operational mode and followed the process documented in the System Validation Test Plan. Nonclinical testing results are provided in the 510(k). Validation testing indicated that as required by the risk analysis, designated individuals performed all verification and validation activities and that the results demonstrated that the predetermined acceptance criteria were met. {3}------------------------------------------------ # 510(k) Summary of Safety If the device is installed by INFINITT CO., LTD, integration and installations verification tests are conducted against acceptance criteria prior to release to the client. results are provided in the 510(k) Conclusion: 21 CFR 807 92(b)(1) The 510(k) Pre-Market Notification for XELIS 3D contains adequate information, data, and nonclinical test results to enable FDA - CDRH to determine substantial equivalence to the predicate device. The subject device will be manufactured in accordance with the voluntary standards listed in the enclosed voluntary standard survey. The new device and predicate devices are substantially equivalent in the areas of technical characteristics, general function, application, and intended use does not raise any new potential safety risks and is equivalent in performance to existing legally marketed devices. Nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as the predicate devices. Therefore, XELIS 3D is substantially equivalent to the predicate devices. {4}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined strands. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular pattern around the caduceus. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 Infinitt Co., Ltd. % Mr. Casey Conry Senior Project Engineer Underwriters Laboratories 1285 Walt Whitman Rd. MELVILLE NY 11747 JUL 2 8 2011 Re: K103130 Trade/Device Name: Xelis 3DTM Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: June 29, 2011 Received: July 1, 2011 Dear Mr. Conry: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {5}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance: You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, Mary Patel Mary S. Pastel. Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use 510(k) Number (if known): K / / 3 / 30 Device Name: Xelis 3D™ Indications for Use: Xelis 3D™ is a software device intended for viewing of images acquired from various imaging modalities: CT. MR. CR. DR. US. PET and other DICOM compliant medical imaging systems when installed on suitable commercial standard hardware. Images and data can be captured, stored, communicated, processed, and displayed within the system and or across computer networks at distributed locations. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary diagnosis or image interpretation. For primary diagnosis, post process DICOM "for presentation" images must be used. Mammographic images should only be viewed with a monitor approved by FDA for viewing mammographic images. It is the User's responsibility to ensure monitor quality, ambient light conditions, and image compression ratios are consistent with the clinical application. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Mohd. O'Khas Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) Number K103130 Page 1 of
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...