PRECICONTROL HBA1C NORM; PRECICONTROL HBA1C PATH

K103099 · Roche Diagnostics · JJX · Dec 6, 2010 · Clinical Chemistry

Device Facts

Record IDK103099
Device NamePRECICONTROL HBA1C NORM; PRECICONTROL HBA1C PATH
ApplicantRoche Diagnostics
Product CodeJJX · Clinical Chemistry
Decision DateDec 6, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Intended Use

PreciControl HbA1c norm and path are for use in quality control by monitoring accuracy and precision for the quantitative methods as specified in the value sheets.

Device Story

PreciControl HbA1c norm and path are liquid, ready-to-use, single-analyte quality control materials derived from hemolyzed human blood containing HbA1c. Used in clinical laboratory settings to monitor the accuracy and precision of HbA1c quantitative assays on COBAS INTEGRA and cobas c Systems analyzers. Two levels provided: one representing normal HbA1c concentrations, one representing pathological concentrations. Healthcare providers use these controls to verify instrument performance; results are compared against batch-specific value sheets provided by the manufacturer. The controls ensure the reliability of patient HbA1c testing, which is critical for diabetes management and monitoring.

Clinical Evidence

No clinical data. Bench testing only. Stability studies confirmed unopened vial stability at 2-8°C and opened vial stability at 2-8°C (28 days) and -15°C to -25°C (12 weeks) with recoveries within 10% of reference material.

Technological Characteristics

Liquid, ready-to-use quality control material. Matrix: hemolyzed human blood. Analyte: in vitro glycated HbA1c. Traceable to IFCC standards and DCCT/NGSP. Compatible with COBAS INTEGRA and cobas c Systems analyzers. No software or electronic components.

Indications for Use

Indicated for use as a quality control material to monitor accuracy and precision of quantitative HbA1c methods on COBAS INTEGRA and cobas c Systems analyzers. For prescription use only.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k103099 B. Purpose for Submission: New Device C. Measurand: Quality control material for HbA1c D. Type of Test: Not applicable E. Applicant: Roche Diagnostics F. Proprietary and Established Names: PreciControl HbA1c norm and PreciControl HbA1c path G. Regulatory Information: | Product Code | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | JJX | Class I, reserved | 21 CFR 862.1660, single (specified) analyte controls (assayed and unassayed) | Clinical Chemistry (75) | H. Intended Use: 1. Intended use(s): See indications for use below. 2. Indication(s) for use: {1} PreciControl HbA1c norm and path are for use in quality control by monitoring accuracy and precision for the quantitative methods as specified in the value sheets. 3. Special conditions for use statement(s): For prescription use only 4. Special instrument requirements: COBAS INTEGRA and cobas c Systems analyzers: Integra 400, Integra 800, cobas c 111, c311, c501, and c 502. I. Device Description: PreciControl HbA1c is a quality control with two levels; one level represents HbA1c concentrations in the normal range and one in the pathological range. The two levels are sold separately as PreciControl HbA1c norm and PreciControl HbA1c path. They are liquid, ready-to-use, single-analyte controls based on hemolyzed human blood that contains the component HbA1c. Concentrations are batch-specific and adjustable to ensure that their levels fall in the appropriate ranges. The exact values are provided in the batch-specific value sheets. They are for use with COBAS INTEGRA and cobas c Systems analyzers. Reagents – working solutions Reactive components in the liquid controls: Hemolyzed human blood, in vitro glycated HbA1c The human blood products used in the manufacture of these quality control materials has been tested using FDA approved methods and found to be non-reactive for HBsAg and antibodies to HCV and HIV-1/2. The testing methods applied were FDA-approved or cleared in compliance with the European Directive 98/79/EC, Annex II, List A. J. Substantial Equivalence Information: 1. Predicate device name: Roche Diagnostics HbA1c Control N and HbA1c Control P Predicate 510(k) number(s): k072714 {2} 2. Comparison with predicate: | Feature | Candidate Device: PreciControl HbA1c norm/PreciControl HbA1c path | Predicate Device: HbA1c Control N/HbA1c Control P | | --- | --- | --- | | Intended Use/Indications for Use | For use in quality control by monitoring accuracy and precision for the quantitative methods as specified in the value sheets. | Same | | Format | Liquid | Lyophilized | | Reagent Composition | Matrix The controls are based on hemolyzed human blood. Components HbA1c is isolated from a normal patient population then glycated in vitro. | Matrix Same Components The glycated HbA1c is isolated from a diabetic patient population. | | Stability | Unopened 2-8°C until expiration Stability after opening 2-8°C for 28 days (-15)-(-25)°C for 12 weeks (freeze only once) | Unopened Same Stability after reconstitution 2-8°C for 4 weeks (-15)-(-25)°C for 3 months (only freeze once in aliquots of no less than 0.1 mL) | | Traceability | Traceable to IFCC standards | Same | K. Standard/Guidance Document Referenced (if applicable): None were referenced. L. Test Principle: {3} Not applicable ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: a. Precision/Reproducibility: Not applicable b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): #### Traceability and Value Assignment: The assigned values for PreciControl HbA1c norm and path are determined using the hemolysate and whole blood applications of the Roche Tinaquant Hemoglobin A1c Gen 2 and Gen 3 assays which are traceable to IFCC and transferrable to results traceable by calculation (NGSP = [0.09148 * IFCC] + 2.152) to DCCT/NGSP. For lot-specific value assignment, each level of quality control is tested in singlicate in at least three separate runs on at least three separate master analyzers (Integra 800 and cobas c 501 analyzers). All master analyzers used in the value assignment process were calibrated using a master calibrator. The assigned target value of each control level is defined as the mean of all the runs for each assay and analyzer. The corresponding control range is calculated as the target value ± 3 standard deviations (with the standard deviation being the value obtained from several target value determinations). #### Stability: ##### Closed Vial Stability (Real-time/ shelf-life stability) Unopened vials of PreciControl norm and path stored at 2 to 8°C (test material) were tested against unopened vials of PreciControl norm and path stored at -80°C (reference material). Each material was tested in triplicate after timepoints of 0, 12, 18, 20, and 21 months. Percent recoveries at each timepoint were within 10% of the reference material concentration. The unopened control vials are stable until the stated expiration date (15 months) printed on the vials when stored at 2-8°C. ##### Opened Vial Stability 4 {4} Opened and subsequently closed vials of PreciControl norm and path stored at 2-8°C and at -15°C to -25°C were tested against unopened vials of PreciControl norm and path stored at -80°C (reference material). Each material was tested in triplicate after timepoints stated in the stability protocol. Percent recoveries of each material was within 10% of the reference material concentration. Data supports the package insert's open vial stability claim of 28 days when stored at 2-8°C or 12 weeks at -15°C - (-25)°C (when frozen once). d. Detection limit: Not Applicable e. Analytical specificity: Not Applicable f. Assay cut-off: Not Applicable 2. Comparison studies: a. Method comparison with predicate device: Not Applicable b. Matrix comparison: Not Applicable 3. Clinical studies: a. Clinical Sensitivity: Not Applicable b. Clinical specificity: Not Applicable c. Other clinical supportive data (when a. and b. are not applicable): Not Applicable 4. Clinical cut-off: {5} Not Applicable 5. Expected values/Reference range: The expected values for each analyzer are presented in the labeling. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 6
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...