← Product Code [JXG](/productcode/JXG) · K103003

# AESCULAP- MEITHKE PROGAV PROGRAMMABLE SHUNT SYSTEM (K103003)

_Aesculap, Inc. · JXG · Nov 9, 2010 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K103003

## Device Facts

- **Applicant:** Aesculap, Inc.
- **Product Code:** [JXG](/productcode/JXG.md)
- **Decision Date:** Nov 9, 2010
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.5550
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The Miethke proGAV Programmable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum. Adjustments of the proGAV shunt can be verified by using the verification instrument but must be confirmed by radiograph (X-ray).

## Device Story

Programmable hydrocephalus shunt system; diverts CSF from lateral ventricles to peritoneum. Valve utilizes leaf spring and ball mechanism; pressure settings controlled via internal magnets. Physician uses external manual instruments to adjust and verify pressure settings; setting confirmation requires X-ray. Titanium outer casing. Includes accessories: shunt assistants, catheters, connectors, deflectors, reservoirs. Modification involves internal magnet design. Used in clinical settings by neurosurgeons to manage intracranial pressure; benefits patient by providing adjustable CSF drainage.

## Clinical Evidence

Bench testing only. Testing included ASTM F2119 (MR Image Artifacts), ASTM F2182 (RF Induced Heating), ASTM F2213 (Magnetically Induced Torque), and ASTM F2052 (Magnetically Induced Displacement Force).

## Technological Characteristics

Titanium housing; leaf spring and ball valve mechanism; magnetically controlled pressure settings. Complies with ASTM F2119, F2182, F2213, and F2052 for MRI safety. Mechanical device; no software or electronic components.

## Regulatory Identification

A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.

## Predicate Devices

- Miethke proGAV Programmable Shunt System ([K062009](/device/K062009.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K103003

# Page 1 of 2

#### 510(k) SUMMARY (as required by 21 CFR 807.92) B.

Miethke proGAV Shunt November 9, 2010

NOV - 9 2010

| COMPANY:             | Aesculap®, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number:  2916714 |
|----------------------|---------------------------------------------------------------------------------------------------------------------|
| CONTACT:             | Kathy A. Racosky<br>610-984-9291 (phone)<br>610-791-6882 (fax)<br>kathy.racosky@aesculap.com                        |
| COMMON NAME:         | Hydrocephalus Shunt System                                                                                          |
| CLASSIFICATION NAME: | Aesculap Miethke proGAV Programmable Shunt System                                                                   |
| REGULATION NUMBER:   | 882.5550                                                                                                            |
| PRODUCT CODE:        | JXG                                                                                                                 |

#### DEVICE DESCRIPTION

The proGAV is a "programmable" shunt that can be set for a range of pressures. The valve in the proGAV is a leaf spring and ball mechanism that is mechanically controlled by internal magnets. The outer case for the device is made of titanium. The shunt comes with a manual device to verify the pressure setting and another to set or re-set the pressure. These manual accessories are both for external use by the physician. Once verified using the instrument the setting must be confirmed with an X-ray. Various Miethke shunt system accessories such as shunt assistants, catheters, connectors, deflectors and reservoirs are also offered with the proGAV . The proposed modification is to the internal magnet.

#### INDICATIONS FOR USE

The Miethke proGAV Programmable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum. Adjustments of the proGAV shunt can be verified by using the verification instrument but must be confirmed by radiograph (X-rav).

#### SUBSTANTIAL EQUIVALENCE

Aesculab®, Inc. believes that the modified proGAV Shunt is substantially equivalent to the proGAV Shunt of the Miethke proGAV Programmable Shunt System (K062009). The modified shunt is similar to the current proGAV Shunt. The only difference is the internal magnets. The materials, design, principle of operation and intended use is equivalent to the previously cleared proGAV Shunt.

### PURPOSE FOR PREMARKET NOTIFICATION

The purpose for this submission is to gain marketing clearance for the new Miethke proGAV Shunt.

{1}------------------------------------------------

# PERFORMANCE DATA

1

The following tests demonstrated the proGAV shunt is safe and effective.

- ASTM F2119 Evaluation of MR Image Artifacts �
- ASTM F2182 Measurement of Radio Frequency Induced Heating During Magnetic . Resonance Imaging
- ASTM F2213 Qualitative Measurement of Magnetically Induced Torque in the . Magnetic Resonance Environment
- ASTM F2052 Measurement of Magnetically Induced Displacement Force on the in . the Magnetic Resonance Environment

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of an eagle or bird-like figure with three tail feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the graphic.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Aesculap, Inc. c/o Ms. Kathy A. Racosky Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, PA 18034

NOV - 9 2010

Re: K103003

Trade/Device Name: Aesculap Miethke proGAV Programmable Shunt System Regulation Number: 21 CFR 882.5550 Regulation Name: Central Nervous System Fluid Shunt and Components Regulatory Class: Class II Product Code: JXG Dated: October 11, 2010 Received: October 12, 2010

Dear Ms. Racosky:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated-in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Kathy A. Racosky

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Qun. L. Qum. for

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

NOV - 9 2010

510(k) Number: K103003

Device Name: Miethke proGAV Programmable Shunt System

## Indications For Use:

The Miethke proGAV Programmable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum. Adjustments of the proGAV shunt can be verified by using the verification instrument but must be confirmed by radiograph (X-ray).

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

JOE HUTTER

Page I of 1

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

K103003

510(k) Number

---

**Source:** [https://fda.innolitics.com/device/K103003](https://fda.innolitics.com/device/K103003)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
