ELITECH CLINICAL SYSTEMS TRIGLYCERIDES SL, ELITECH CLINICAL SYSTEMS CHOLESTEROL SL
Applicant
Seppim S.A.S.
Product Code
CDT · Clinical Chemistry
Decision Date
May 19, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1705
Device Class
Class 1
Indications for Use
ELITech Clinical Systems TRIGLYCERIDES SL is intended for the quantitative in vitro diagnostic determination of triglycerides in human serum and plasma on ELITech Vital Scientific Selectra/Flexor analyzers. It is not intended for use in Point of Care settings. ELITech Clinical Systems CHOLESTEROL SL is intended for the quantitative in vitro diagnostic determination of cholesterol in human serum and plasma on ELITech Vital Scientific Selectra/Flexor analyzers. It is not intended for use in Point of Care settings. ELITech Clinical Systems ELICAL 2 is a multi-parametric calibrator for in vitro diagnostic use in the calibration of quantitative ELITech Clinical Systems methods on the ELITech Vital Scientific Selectra/Flexor Analyzers. ELITech Clinical Systems ELITROL I & ELITROL II are multiparametric control sera for use in quality control of ELITech Clinical Systems methods on ELITech Vital Scientific Selectra/Flexor Analyzers.
Device Story
System uses enzymatic colorimetric assays to measure cholesterol and triglycerides in human serum/plasma; performed on Vital Scientific Selectra/Flexor analyzers. Cholesterol assay: cholesterol esterase/oxidase/peroxidase reaction produces quinoneimine complex; absorbance measured at 505 nm. Triglyceride assay: lipoprotein lipase/glycerol kinase/glycerol-3-phosphate oxidase/peroxidase reaction produces quinoneimine complex; absorbance measured at 505 nm. System includes calibrator (ELICAL 2) and controls (ELITROL I/II). Operated by laboratory personnel in clinical settings. Output is quantitative concentration (mg/dL) used by physicians to diagnose/treat lipid metabolism disorders and monitor patients with diabetes, nephrosis, or liver obstruction.
Clinical Evidence
Bench testing only. Precision evaluated per CLSI EP5-A2 (n=80 per sample). Linearity per CLSI EP6-A (R^2 > 0.999). Interference testing per CLSI EP7-A. Method comparison against predicate (n=99) showed high correlation (r=0.999) for both assays. Traceability established via ID-MS reference method and CRMLN certification.
Indicated for quantitative measurement of triglycerides and cholesterol in human serum and plasma to aid in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, lipid metabolism disorders, and endocrine disorders. Calibrator and controls are indicated for use with ELITech Clinical Systems methods on Selectra/Flexor analyzers.
Regulatory Classification
Identification
A triglyceride test system is a device intended to measure triglyceride (neutral fat) in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders.
Predicate Devices
Horiba ABX PENTRA Cholesterol CP (k060854)
Horiba ABX PENTRA Triglycerides CP (k060854)
Roche Diagnostics Calibrator for Automated Systems (C.f.a.s) (k033501)
Roche Diagnostics Precinorm U and Precipath U (k041227)
Submission Summary (Full Text)
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k102993
B. Purpose for Submission:
New devices
C. Measurand:
Cholesterol and Triglyceride
D. Type of Test:
Quantitative, enzymatic colorimetric
E. Applicant:
SEPPIM S.A.S.
F. Proprietary and Established Names:
ELITech Clinical Systems CHOLESTEROL SL
ELITech Clinical Systems TRIGLYCERIDES SL
ELITech Clinical Systems ELICAL 2
ELITech Clinical Systems ELITROL I and II
G. Regulatory Information:
| Regulation Section | Product Code | Classification | Panel |
| --- | --- | --- | --- |
| 862.1175
(Cholesterol test system) | CHH | Class I meets the limitations to exemptions 21 CFR §862.9 (c) (4) | Chemistry (75) |
| 862.1705
(Triglyceride test system) | CDT | Class I meets the limitations to exemptions 21 CFR §862.9 (c) (4) | Chemistry (75) |
{2}
| 862.1150
(Calibrator) | JIX | Class II | Chemistry
(75) |
| --- | --- | --- | --- |
| 862.1660 (Quality control materials) | JJY | Class I, reserved | Chemistry
(75) |
H. Intended Use:
1. Intended use(s):
See Indications for Use below
2. Indication(s) for use:
ELITech Clinical Systems CHOLESTEROL SL is intended for the quantitative in vitro diagnostic determination of cholesterol in human serum and plasma on the ELITech Vital Scientific Selectra/Flexor analyzers. It is not intended for use in Point of Care settings. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
ELITech Clinical Systems TRIGLYCERIDES SL is intended for the quantitative in vitro diagnostic determination of triglycerides in human serum and plasma on the ELITech Vital Scientific Selectra/Flexor analyzers. It is not intended for use in Point of Care settings. Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders.
ELITech Clinical Systems ELICAL 2 is a multi-parameter calibrator for in vitro diagnostic use in the calibration of quantitative ELITech Clinical Systems methods on the ELITech Vital Scientific Selectra/Flexor analyzers.
ELITech Clinical Systems ELITROL I and II are multi-parameter control sera for use in quality control of ELITech Clinical Systems methods on the ELITech Vital Scientific Selectra/Flexor analyzers.
3. Special conditions for use statement(s):
For Prescription Use
It is not intended for use in Point of Care settings.
4. Special instrument requirements:
For use with the Vital Scientific Selectra Junior analyzer which is also trademarked as the Vital Scientific Flexor Junior analyzer.
{3}
I. Device Description:
ELITech Clinical Systems Cholesterol SL is a one (1) reagent system with reagent R. Reagent R is supplied in liquid ready-to-use form and contains Pipes buffer, 4-Aminoantipyrine (4-AAP), Cholesterol esterase (CHE bacterial), Cholesterol oxidase (CHO microorganisms), Peroxidase (POD horseradish), Sodium cholate, Phenol and Sodium azide.
ELITech Clinical Systems Triglycerides SL is a one (1) reagent system with reagent R. Reagent R is supplied in liquid ready-to use form and contains Pipes buffer, $p$-chlorophenol, ATP, Amino-4-antipyrine (4-AAP), Lipoprotein lipase (bacterial), Glycerol kinase (bacterial), Glycerol-3-phosphate oxidase (microorganisms), Peroxidase (horseradish), Potassium ferrocyanide, Magnesium $(\mathrm{Mg}^{2+})$ and Sodium azide.
ELITech Clinical Systems Elical 2 is a lyophilized calibrator prepared from human serum which contains chemical additives and material of biological origin, and stabilizers.
ELITech Clinical Systems Elitrol I and Elitrol II are lyophilized control serum prepared from human serum which contains chemical additives and material of biological origin.
All human source materials were tested with FDA-approved methods and found to be negative for HbsAG and to antibodies to HCV and HIV 1/2.
J. Substantial Equivalence Information:
1. Predicate device name(s)
Horiba ABX PENTRA Cholesterol CP
Horiba ABX PENTRA Triglycerides CP
Roche Diagnostics Calibrator for Automated Systems (C.f.a.s)
Roche Diagnostics Precinorm U and Precipath U
2. Predicate K number(s):
k060854 (cholesterol and triglyceride reagents)
k033501 (calibrator)
k041227 (control)
3. Comparison with predicate:
{4}
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Predicate Device ABX PENTRA cholesterol reagent (k060854) | Candidate device ELITech Clinical Systems Cholesterol SL |
| Intended Use | For in vitro diagnostic use in the quantitative determination of cholesterol in human serum and plasma. | Same |
| Indications for use | Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. | Same |
| Test Method | Enzymatic colorimetric method using cholesterol esterase/cholesterol oxidase coupled with peroxidase (Trinder method) | Same |
| Sample type | Serum and plasma | Same |
| Reagent | Liquid form, ready to use | Same |
| Instrument | ABX Pentra 400 Analyzer | The Vital Scientific Selectra Junior Analyzer (also trademarked as the Vital Scientific Flexor Junior Analyzer) |
| Calibration Frequency | 8 days | 28 days |
| Measuring range | 2.55 to 580 mg/dL | 20-600 mg/dL |
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Predicate Device ABX PENTRA triglyceride reagent (k060854) | Candidate device ELITech Clinical Systems Triglyceride SL |
| Intended Use | For in vitro diagnostic use in the quantitative determination of triglyceride in human serum and plasma. | Same |
| Indications for use | Measurements obtained by this device are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders. | Same |
| Test Method | Enzymatic colorimetric method | Same |
| Sample type | Serum and plasma | Same |
{5}
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Predicate Device ABX PENTRA triglyceride reagent (k060854) | Candidate device ELITech Clinical Systems Triglyceride SL |
| Reagent | Liquid form, ready to use | Same |
| Instrument | ABX Pentra 400 Analyzer | The Vital Scientific Selectra Junior Analyzer (also trademarked as the Vital Scientific Flexor Junior Analyzer) |
| Calibration Frequency | 14 days | same |
| Measuring range | 3.1-1470 mg/dL | 30-1000 mg/dL |
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Predicate device Roche Calibrator (C.f.a.s) (k033501) | Candidate device ELITech Clinical Systems Elical 2 |
| Intended Use/Indications for Use | For in vitro diagnostic use in the calibration of quantitative method | Same |
| Format | Lyophilized calibrator based on human serum with constituents added as required to obtain desired components levels. | Same |
| Level | Single level | Same |
| Stability | Lyophilized: store at 2-8°C and protect from light until the expiry date
After reconstitution*:8 hours between 15-25°C, 2 days between 2-8°C, 4 weeks between -25 and -15 °C (when frozen once) | Same |
| Instrument | Roche Analyzers | The Vital Scientific Selectra Junior Analyzer (also trademarked as the Vital Scientific Flexor Junior Analyzer) |
{6}
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Predicate device
Roche Precinorm U/ Precipath U
(k041227) | Candidate device
ELITech Clinical Systems
Elitrol I/Elitrol II |
| Intended Use/
Indications for Use | For in vitro diagnostic use in accuracy
control of quantitative method | Same |
| Format | Lyophilized human sera with
constituents added as required to obtain
defined components levels | Same |
| Levels | Two levels | Same |
| Stability | Lyophilized: Store at 2-8°C and
protected from light until the expiry
date. After Reconstitution*: 12 hours
between 15-25°C, 5 days between 2-
8°C, 4 weeks between -25 and -15°C
(when frozen once) | Same |
K. Standard/Guidance Document Referenced (if applicable):
EP05-A2-Evaluation of precision performance of quantitative measurement methods; Approved guideline Second-Edition
EP06-A-Evaluation of the linearity of the measurement of quantitative procedures: a statistical approach-First Edition
EP07-A2-Interference Testing in Clinical Chemistry-Second Edition
EP09-A2-Method comparison and bias estimate using patient samples-Second Edition
L. Test Principle:
Cholesterol: Cholesterol esters are hydrolyzed to cholesterol and fatty acids by cholesterol esterase (CHE). Then the cholesterol is oxidized by cholesterol oxidase (CHO) to produce cholest-4-en-3-one and hydrogen peroxide (H₂O₂). H₂O₂ reacts with 4-aminoantipyrine (4-AAP) and phenol under the catalytic action of peroxidase to form a colored quinoneimine complex. The absorbance of the quinoneimine complex at 505 nm is proportional to the concentration of cholesterol in the sample.
Triglycerides: Triglycerides are hydrolyzed to glycerol and fatty acids by a lipoprotein lipase. In the presence of ATP and glycerol kinase, the glycerol is phosphorylated to glycerol-3-phosphate, which is then oxidized by glycerol-3-phosphate-oxidase to produce hydrogen peroxide (H2O2). H2O2 reacts with 4-amino-antipyrine (4-AAP) and p-chlorophenol under
{7}
the catalytic action of peroxidase to form a colored quinoneimine complex. The absorbance of the quinoneimine complex at 505 nm is proportional to the concentration of triglycerides in the sample.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Precision was evaluated with one serum control and two human serum pool samples according to the CLSI EP5-A2 guideline. All samples were assayed on the Vital Scientific Selectra Junior analyzer in duplicate, twice a day for 20 days. The mean, SD, and %CV were calculated for within-run and total imprecision. Results are summarized in the table below:
Precision for cholesterol assay:
| Material | n | Mean (U/L) | Within-run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD | CV (%) | SD | CV (%) |
| Serum Pool (low) | 80 | 116 | 2.8 | 2.4 | 3.1 | 2.6 |
| Serum Control (medium) | 80 | 190 | 3.6 | 1.9 | 5.1 | 2.7 |
| Spiked Serum Pool (high) | 80 | 298 | 5.1 | 1.7 | 7.9 | 2.7 |
Precision for triglycerides assay:
| Material | n | Mean (U/L) | Within-run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD | CV (%) | SD | CV (%) |
| Serum Pool (low) | 80 | 48 | 0.7 | 1.5 | 1.9 | 3.9 |
| Serum Control (medium) | 80 | 142 | 1.5 | 1.0 | 3.8 | 2.7 |
| Spiked Serum Pool (high) | 80 | 273 | 2.0 | 0.7 | 12.2 | 4.5 |
{8}
b. Linearity/assay reportable range:
Cholesterol assay:
A linearity study was performed according to the CLSI EP6-A guideline. Linearity across the assay range was determined by using two pools of patient serum, one at a low concentration (20 mg/dL) and one at a high concentration (618 mg/dL). The low sample was prepared by diluting the serum sample pool with saline and the high sample was prepared by spiking the serum sample pool. Admixtures were prepared by mixing different proportions of low and high pool samples. All samples (11 samples) were assayed in triplicate. The expected values were plotted against the observed values and a linear regression line was fitted with the following regression equation: $Y = 0.9854X + 2.5152$, $R^2 = 0.9994$.
Results of the study support the sponsor’s claim that the cholesterol assay is linear from 20-600 mg/dL.
Triglycerides assay:
Similarly, linearity across the assay range was determined by using two pools of patient serum, one at a low concentration (26 mg/dL) and one at a high concentration (1072 mg/dL). The low sample was prepared by diluting the serum sample pool with saline and the high sample was prepared by spiking the serum sample pool. Admixtures were prepared by mixing different proportions of low and high pool samples. All samples (11 samples) were assayed in triplicate. The expected values were plotted against the observed values and a linear regression line was fitted with the following regression equation: $Y = 0.9932X - 1.4687$, $R^2 = 0.9997$.
Results of the study support the sponsor’s claim that the triglyceride assay is linear from is 30-1000 mg/dL
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
The ELITech Clinical Systems Elical 2 has both cholesterol and triglycerides assays traceable to the ID-MS (Isotope Dilution- Mass Spectrometry) reference method.
Traceability to the National Reference System for Cholesterol was established by performing a direct comparison with the cholesterol reference method using human specimens that cover the National Cholesterol Education Program (NCEP) medical decision points. The ability to meet the NCEP’s performance criteria for accuracy was demonstrated through testing and certification by the Cholesterol Reference Method Laboratory Network (CRMLN). The comparison study was performed using SEPPIM Cholesterol SL reagent and SEPPIM ELICAL 2 calibrator. A certificate from CRMLN was provided by the sponsor. The certificate will expire in 2 years and
8
{9}
a list of the certified methods can be found at the following link: http://www.cdc.gov/labstandards/crmln.htm
## Stability:
Calibrator material is purchased from a commercial vendor (previously cleared under k033501). Calibrator assigned value was verified by testing on one Selectra Junior analyzer. The sponsor claimed the following for stability: Elical 2 is stable until the expiration date printed on the label when stored at 2-8°C prior to reconstitution. After reconstitution the stability is 8 hours when stored at 15-25°C, 2 days at 2-8°C or 4 weeks (when frozen once) at -25° and -15°C.
Protocols and acceptance criteria were provided to support on-board calibration stability for 28 days for cholesterol and 14 days for triglyceride.
Control material is purchased from a commercial vendor (previously cleared under k041227). Control assigned value was verified by testing on one Selectra Junior analyzer. The sponsor claimed the following for stability: Before reconstitution, the shelf-life of the ELITech Clinical Systems Elitrol 1 and Elitrol II is 30 months at 2-8°C. After reconstitution the stability is 12 hours when stored at 15-25°C, 5 days when stored at 2-8°C or 4 weeks (when frozen once) at -25° and -15°C.
## Value assignment of Elical 2:
Elical 2 is tested against predetermined values on multiple Vital Scientific analyzers using the cholesterol and triglyceride reagents. The mean analyte value is calculated and a target value is assigned.
## Value assignment of Elitrol I and II controls:
For both cholesterol and triglyceride assays, assignments are based on two Vital Scientific analyzers with two lot of reagents. A total of 36 measurements were obtained and the mean, SD and %CV were calculated. The control ranges were established by calculating the mean ± 15% for cholesterol and mean ± 16% for triglyceride.
## d. Detection limit:
### Cholesterol assay:
A detection limit study was evaluated according to CLSI EP17-A guideline. LoB was determined by running a blank sample 60 times on the Vital Scientific Selectra Junior analyzer. The limit of detection (LoD) was determined by assaying 4 low (3.6 U/L) diluted sample pools 60 times on the Selectra Junior analyzer. Limit of Quantitation (LoQ) was determined by running 4 low (5 U/L) diluted sample pools 60 times on the Vital Scientific Selectra Junior analyzer. LoB is calculated to be 0.1 mg/dL, LoD is
9
{10}
0.2 mg/dL. LoQ is calculated to be 10 mg/dL based on a total error of 8.6%.
## Triglycerides assay:
A detection limit study was evaluated according to CLSI EP17-A guideline. LoB was determined by running a blank sample 60 times on the Vital Scientific Selectra Junior analyzer. The limit of detection (LoD) was determined by assaying 4 low (3.6 U/L) diluted sample pools 60 times on the Selectra Junior analyzer. Limit of Quantitation (LoQ) was determined by running 4 low (5 U/L) diluted sample pools 60 times on the Vital Scientific Selectra Junior analyzer. LoB is calculated to be 1 mg/dL, LoD is 2 mg/dL. LoQ is calculated to be 10 mg/dL based on a total error of 9.8%.
The sponsor claimed that the cholesterol's measuring range is 20-600 mg/dL and triglyceride's measuring range is 30-1000 mg/dL.
## e. Analytical specificity:
### Cholesterol assay:
Testing for interfering substances was based on CLSI EP-7A. Testing was performed on a minimum of five concentrations for each interfering substance. Two different concentrations of cholesterol (116 and 309 mg/dL) were used for evaluation. Samples with increasing amounts of triglycerides (Intralipid®), Unconjugated bilirubin, Conjugated bilirubin, Hemoglobin, Uric Acid, Methyl dopa and Ascorbic Acid were tested in triplicate and compared to the same sample without the interferent. The sponsor defined no significant interference as < 6.6% for cholesterol. Results are summarized in the table below:
| | Highest concentration tested that showing no significant interference (mg/dL) (≤6.6%) |
| --- | --- |
| Triglyceride | 614 |
| Unconjugated Bilirubin | 6.0 |
| Conjugated Bilirubin | 5.9 |
| Hemoglobin | 250 |
| Ascorbic Acid | 2.0 |
| Methyl dopa | 0.8 |
| Uric acid | 23.0 |
The sponsor has the following in their limitations in their labeling:
{11}
Ascorbic acid concentrations above therapeutic levels (>2.0 mg/dL) will interferent with the cholesterol level.
Do not use hemolyzed samples.
Do not use icteric samples.
## Triglycerides assay:
Testing for interfering substances was based on CLSI EP-7A. Testing was performed on a minimum of six concentrations for each interfering substances. Two different concentrations of triglycerides (133 and 266 mg/dL) were used for evaluation. Samples with increasing amounts of Unconjugated bilirubin, Conjugated bilirubin, Hemoglobin, Uric Acid, Methyl dopa and Ascorbic Acid were tested in triplicate and compared to the same sample without the interferent. The sponsor defined no significant interference as < 10% for triglycerides. Results are summarized in the table below:
| | Highest concentration tested that showing no significant interference (mg/dL) (≤10%) |
| --- | --- |
| Unconjugated Bilirubin | 15.0 |
| Conjugated Bilirubin | 5.9 |
| Hemoglobin | 250 |
| Ascorbic Acid | 2.0 |
| Methyl dopa | 1.0 |
| Uric acid | 23.0 |
The sponsor has the following in their limitations in their labeling:
Ascorbic acid concentrations above therapeutic levels (>2.0 mg/dL) will interfere with the triglyceride level.
Do not use hemolyzed samples.
Do not use icteric samples.
f. Assay cut-off:
Not applicable
2. Comparison studies:
{12}
a. Method comparison with predicate device:
**Cholesterol assay:**
A method comparison study was performed using the HORIBA ABX Cholesterol CP reagent on a PENTRA 400 chemistry analyzer and the ELITech cholesterol SL reagent on the Selectra Junior analyzer according to CLSI EP9-A2. A total of 99 serum samples covering the measuring range of the cholesterol assay were assayed. Of these, 10 were altered samples (2 diluted and 8 spiked). The samples had values ranging from $20\mathrm{mg/dL}$ to $585\mathrm{mg/dL}$. The comparison resulted in a slope of 1.006, an intercept of -1.734, correlation coefficient of $r=0.999$. The standard error of the estimate is equal to $4\mathrm{mg/dL}$.
**Triglycerides assay:**
A method comparison study was performed using the HORIBA ABX triglycerides CP reagent on a PENTRA 400 chemistry analyzer and the ELITech triglycerides SL reagent on the Selectra Junior analyzer according to CLSI EP9-A2. A total of 99 serum samples covering the measuring range of the triglyceride assay were assayed. Of these, 6 were altered samples (2 diluted and 4 spiked). The samples had values ranging from $31\mathrm{mg/dL}$ to $963\mathrm{mg/dL}$. The comparison resulted in a slope of 1.040, an intercept of 0.339, correlation coefficient of $r=0.999$. The standard error of the estimate is equal to $7\mathrm{mg/dL}$.
b. Matrix comparison:
**Cholesterol assay:**
45 paired serum and plasma (lithium heparin) samples, ranging from 24 to 580 mg/dL, were tested on the Selectra Junior analyzer. 3 samples were altered. The comparison resulted in a slope of 1.020, an intercept of -4.0, correlation coefficient of $r = 1.000$. The standard error of the estimate is equal to $5\mathrm{mg/dL}$.
**Triglycerides assay:**
47 paired serum and plasma (lithium heparin) samples, ranging from 50 to 951 mg/dL, were tested on the Selectra Junior analyzer. 4 samples were altered. The comparison resulted in a slope of 0.986, an intercept of 1.0, correlation coefficient of $r = 0.999$. The standard error of the estimate is equal to $11\mathrm{mg/dL}$.
Based on the data, the sponsor claims that lithium heparin is an acceptable anti-coagulant for the cholesterol and triglyceride assays.
3. Clinical studies:
a. Clinical Sensitivity:
{13}
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
The expected values are based on the NCEP ATP III *recommendations:
Cholesterol assay:
Desirable cholesterol: <200 mg/dL
Borderline high cholesterol: 200-239 mg/dL
High cholesterol: ≥240 mg/dL
Triglyceride assay:
Normal: <150 mg/dL
Borderline high: 150-199 mg/dL
High: 200-499 mg/dL
Very high: ≥500 mg/dL
*“Third Report of the NCEP Expert Panel on Detection, Evaluation, and Treatment of High Blood Cholesterol in Adults (Adult Treatment Panel III)-Final report”. NIH publication (2002), No. 02-5215
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.