K102938 · Acqpoint, Inc. · JFB · Sep 23, 2011 · Physical Medicine
Device Facts
Record ID
K102938
Device Name
ACQPOINT THERMAJADE MASSAGE TABLE
Applicant
Acqpoint, Inc.
Product Code
JFB · Physical Medicine
Decision Date
Sep 23, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5880
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the AcQPoint Thermajade Massage Table is to provide the user with muscle relaxation therapy by delivering have as a gentle massage. Additionally the heat elements provide topical heating for temporary relief of minor muscle and joint pain and stiffness; temporary relief of joint pain associated with arthritis; temporary increase in local circulation in the areas affected by the device and relaxation of muscles.
Device Story
Electrically powered, motorized multi-function physical therapy table; provides muscle relaxation therapy via heat and massage. Input: user-selected programs or manual mode via wired remote. Mechanism: two sets of jade rollers on carriages (torso and leg sections) travel along a curved guide rail track conforming to spine; powered by two DC motors. Heat: halogen lamps inside rollers and carbon/epoxy heating panels on platform. Output: mechanical massage and radiant/conductive heat. Used in clinic or home settings; operated by user. Microprocessor controls roller motion, temperature settings, and LED display. Benefits: temporary pain relief, increased local circulation, muscle relaxation.
Clinical Evidence
Bench testing only. Testing confirmed device operates within therapeutic skin temperature range (40°C to 42.5°C) and meets ISO 60601-1 (safety) and ISO 60601-1-2 (EMC) standards. Risk analysis conducted per ISO 14971:2007.
Technological Characteristics
Materials: jade rollers, carbon/epoxy heating panels, polyurethane foam mattress, synthetic leather cover. Energy: 120V AC, max 480W. Dimensions: 27" x 78" x 20". Sensing/Actuation: DC motors, cable/pulley drive, microprocessor-based control. Connectivity: wired remote control with LED display. Standards: ISO 60601-1, ISO 60601-1-2, ISO 14971:2007.
Indications for Use
Indicated for users seeking muscle relaxation therapy, temporary relief of minor muscle and joint pain/stiffness, temporary relief of arthritis-associated joint pain, and temporary increase in local circulation.
Regulatory Classification
Identification
A multi-function physical therapy table is a device intended for medical purposes that consists of a motorized table equipped to provide patients with heat, traction, and muscle relaxation therapy.
Predicate Devices
Ceragem RH-1 (K06276)
Submission Summary (Full Text)
{0}------------------------------------------------
# DIANE GREIF-SHEPPARD, PH.D.
K102938
SEP 2 3 2011
### SECTION 3
## 510 (K) PREMARKET NOTIFICATION SUMMARY
Date: September 20, 2010
### Owner And Submitter:
AcQPoint, Inc. Diane Greif-Sheppard, Ph.D. 79-825 Highway 111, Suite 101 Loa Quinta, CA 92253 Tel: (760) 775-7900; Fax: (760) 775-7911 Contact : Diane Greif-Sheppard or Morris Sheppard
### Proprietary Name:
AcQPoint Thermajade Massage Table Model BL-7600.
#### Common or Usual Name:
Thermal Massage Table
#### Classification Name:
Multi-Function Physical Therapy Table (21 CFR 890.5880, Product Code JFB).
#### Predicate Device:
Ceragem RH-1, 510 (K) number K06276
#### Device Description:
The AcQPoint Thermajade Massage Table is an electrically powered, motorized multi-function physical therapy table. It is intended to provide muscle relaxation therapy by delivering heat and gentle massage. The massage function is delivered by two sets of natural jade rollers mounted on carriages in both the torso and leg sections of the device, These roller carriages travel along a guide rail track which is curved to conform to the natural curve of the spine. They are moved by a cable and pulley arrangement powered by two DC motors, one each for
1
{1}------------------------------------------------
the torso and the leg sections.. Heat is delivered by both halogen heat lamps located inside the jade rollers and carbon epoxy heating panels on the platform beneath them. The panels emit radiant heat and the rollers conductive heat. During use, as the rollers traverse the torso and liefers the device applies light pressure as well as heat to the user.
The motion of the rollers is selectable by the user. different programs as well as a manual, user controlled, mode. There are unfo the programs last forty minutes and concentrate the massage action on different parts of the back, as well as stopping for 20 to 30 seconds at various acupressure points, and the third lasts fifteen minutes and travels continuously up and down the entire back.
The functions of the device, the temperature settings for both the carbon/epoxy emitters and the rollers and the program selection are controlled by a wired remote control with a LED display that indicates the selected program, the temperature settings and current temperatures along with a timer.
The motion of the rollers, the temperature settings and the displays on the remote are controlled by a microprocessor unit employing a microcontroller with 28k of built-in flash ROM as well as timers, EEPROM, a UART to interface with external devices and other elements required for the proper execution of the installed software and the function of the device. The software has been verified by both single step and breakpoint checks in the debugging process and validated by final bench testing on the completed device.
The device measures 27" (68.5 cm) wide by 78" (198 cm) long and has a table height of 20" (51 cm). It weighs 160 lbs. (72 kg) and has a working weight load limit of 300 lbs. (136 kg.). It is wired for 120 volt flass a cycle current and draws a maximum of 480W.
#### Intended Use:
The intended use of the AcQPoint Thermajade Massage Table is to provide the user with muscle relaxation therapy by delivering have as a gentle massage. Additionally the heat elements provide topical heating for temporary relief of minor muscle and joint pain and stiffness; temporary relief of joint pain associated with arthritis; temporary increase in local circulation in the areas affected by the device and relaxation of muscles.
# Technological Characteristics and Substantial Equivalence:
The AcQPoint Thermajade Massage Table and the predicate device both provide heat by both carbon/epoxy heating elements under the roller carriages and halogen heating lamps mounted within the rollers, which heat them. The temperature is user-controlled and limited for
2
{2}------------------------------------------------
safety. In both The AcQPoint Thermajade Massage Table and the predicate device the massage function is delivered by jade rollers that are mounted on a carriage which travels along a curved track corresponding to the curvature of the spine and which is pulled by a cable and pulley mechanism powered by a small electric motor. A microprocessor chip pre-programmed with three different cycles, as well as a manually controlled cvcle, controls the forward and back action of the motion. This is identical to the predicate device. Both devices are controlled by wired handheld remotes with LED displays. Both devices are constructed with all elements mounted on a platform that is surrounded with a dense polyurethane foam mattress that is cut out where the roller mechanism operates. Both have synthetic leather zippered covers.
The AcQPoint Thermajade Massage Table differs from the predicate device only in the particular shape of the jade rollers, that it also provides a second set of rollers powered by an additional motor to massage the leg area, that it has an additional roller attached to the carriage in the torso area to provide massage to the neck area and has a torso section that can be tilted up to an approximate 30 degree angle by means of a small motor should the user so desire.
These distinctions do not negatively impact the safety or effectiveness of the device.
### Performance Data:
Testing was performed to ensure that the device operates in a therapeutic heat range while maintaining safety. These tests show that the tables, after the warm-up period, consistently deliver skin temperatures of between 40° C and 42.5° C for the entire period of use. This is within the range considered therapeutic and below that considered to pose a risk of burn.
There are no performance standards for this type device but, as with the predicate device, the AcQPoint Thermajade Massage Table bears the CE mark. An independent testing laboratory has confirmed conformance to the following standards:
ISO/IEC 60601-1: Medical Electrical Equipment--- Part 1: General Requirements for Safety.
ISO/IEC 60601-1-2: collateral standard: Electromagnetic compatibility for medical devices.
### Risk analysis:
We have conducted a risk analysis with consideration of ISO 14971:2007, application of risk management to medical devices, and relevant FDA guidance documents. We have found no unmitigated risks
{3}------------------------------------------------
that pose any unacceptable hazard and that it is a device with a minor level of concern.
#### Conclusion:
The AcQPoint Thermajade Massage Table Model BL-7600 has the same intended use, with similar functional and performance characteristics and has met similar standards as the predicate device. They are substantially equivalent. Other distinctions do not negatively impact the safety or effectiveness of the device.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the department's name around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized caduceus or a bird-like figure, composed of three angled lines above a wavy base.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MI) 20995-0002
AcQpoint Inc. % Mr. Morris Sheppard 79-825 Highway 111, Suite 101 Loa Quinta, California 92253
Re: K102938
Trade/Device Name: AcQPoint Thermajade Massage Table Regulation Number: 21 CFR 890.5880 Regulation Name: Multi-function physical therapy table Regulatory Class: II Product Code: JFB Dated: September 13. 2011 Received: September 13, 2011
SEP 2 3 2511
Dear Mr. Sheppard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and fisting (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Mr. Morris Sheppard
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable. the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.lda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Pari 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/dgfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours
For
Mark N. Me Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K 102938
Device Name: AcQPoint Thermajade Massage Table
Indications For Use:
The intended use of the AcQPoint Thermajade Massage Table is to provide the user with muscle relaxation therapy by delivering heat and gentle massage. Additionally the heat elements provide topical heating for
* temporary relief of minor muscle and joint pain and stiffness;
* temporary relief of joint pain associated with arthritis
* temporary increase in local circulation in the areas affected by the device
* relaxation of muscles.
Prescription Use _ (Part 21 CFR 801 Subpart D)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
AND/OR
Over-The-Counter Use × (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
---
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K102938
Page 1 of ___ 1_______________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.