K102923 · Molnlycke Health Care · MSH · Nov 19, 2010 · General, Plastic Surgery
Device Facts
Record ID
K102923
Device Name
BARRIER N95 PARTICULATE RESPIRATOR
Applicant
Molnlycke Health Care
Product Code
MSH · General, Plastic Surgery
Decision Date
Nov 19, 2010
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The device is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.
Device Story
Barrier® N95 Particulate Respirators (Models #4272 and #4273) are disposable, non-sterile, flat-folded face masks. The device consists of three layers of non-woven fabric. An integrated aluminum filament nose clamp allows for contouring to the wearer's face, while ultrasonically welded elastic headbands secure the mask. The respirator functions as a physical barrier to reduce the transfer of microorganisms, body fluids, and particulates between the wearer and the environment. It is intended for use by operating room personnel during surgical procedures. The device is NIOSH-certified (TC-84A-4350) for particulate filtration efficiency.
Clinical Evidence
No clinical data was required. Substantial equivalence is supported by bench testing, including fluid resistance (ASTM F1862), flammability (16 CFR 1610), and NIOSH certification (TC 84A-4350) for particulate filtration, bacterial filtration, and differential pressure. Biocompatibility testing confirmed the device is non-cytotoxic, non-irritating, and lacks sensitization potential.
Technological Characteristics
Three-layer non-woven fabric (polypropylene/polyethylene); aluminum nose clamp; polyurethane headbands; ultrasonically welded. Trapezoid-folded design (duckbill when open). NIOSH-certified (TC 84A-4350). Meets ASTM F1862 (fluid resistance) and 16 CFR 1610 (flammability). Non-sterile.
Indications for Use
Indicated for operating room personnel during surgical procedures to protect patients and staff from microorganisms, body fluids, and particulate material.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
Gerson Isolair APR Type N95 Healthcare Particulate Respirator and Surgical Mask Model 2735 (K960778)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/13 description: The image shows the logo for Molnlycke Health Care. The logo consists of a cluster of five black circles on the left side, resembling a molecular structure or a group of cells. To the right of the circles, the text "MÖLNLYCKE" is stacked above "HEALTH CARE" in a simple, sans-serif font. The overall design is clean and professional, conveying a sense of scientific precision and healthcare focus.
-NOV-1-9-2010-
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(c).
| Date Prepared: | November 1, 2010 |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Mölnlycke Health Care US, LLC<br>5550 Peachtree Parkway, Suite 500<br>Norcross, GA 30092 |
| Registration number: | 3004763499 |
| Owner/Operator Number: | 8030877 |
| Official Correspondent: | Caitlin Senter, MS, RAC<br>Regulatory Affairs Specialist<br>Tel: 678-250-7928<br>Fax: 678-245-7746<br>e-mail: caitlin.senter@molnlycke.com |
| Trade/Proprietary Name: | Barrier® N95 Particulate Respirators, Models #4272 and #4273 |
| Common Name: | Particulate Respirator |
| Device Class: | Class II |
| Regulation Number: | 21 CFR 878.4040 |
| Product Code: | MSH |
| Predicate Device Name(s): | Gerson Isolair APR Type N95 Healthcare Particulate Respirator and<br>Surgical Mask Model 2735 (K960778) |
#### Description of Device:
Barrier® N95 Particulate Respirators. Models #4273 are NIOSH certified (TC-84A-4350), trapezoid shaped respirators. Barrier® N95 Particulate Respirators, Models #4273 consists of a flat folded disposable face mask that is a piece of three layered non-woven fabric folded in two. An aluminum filament is enclosed in a binding tape welding the top edge, which forms a conformable nose clamp for the purpose of shaping the mask to the contours of the face. It features elastic head bands, ultrasonically welded, to secure the masks in place for the wearer.
#### Intended Use/Indication for Use:
The devices are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.
#### Technological Characteristics:
The Barrier® N95 Particulate Respirators, Models #4273 are substantially equivalent to the ISOLAIR APR Type N95 (K960778). Both devices have similar indications for use, materials, product design, and method of operation.
> Mölnlycke Health Care 5550 Peachtree Parkway, Suite 500 Norcross, GA 30092 USA
Tel 678 250 7900 Fax 678 250 7984 www.molnlycke.com
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Molnlycke Health Care. The logo consists of five black circles clustered together on the left side of the image. To the right of the circles is the text "MÖLNLYCKE HEALTH CARE" in a simple, sans-serif font. The text is stacked, with "MÖLNLYCKE" on top and "HEALTH CARE" below it.
In the manufacture of these devices, multiple standards were utilized. ... The. Barrier® N95 Particulate Respirators, ... Models #4272 and #4273 characteristics are summarized below as compared to the requirements.
Characteristic
Fluid Resistance
Particulate Filtration Efficiency
Bacterial Filtration Efficiency
Differential Pressure
Flammability
Cytotoxicity
Irritation
Delayed-Type Hypersensitivity
#### Standard
The Barrier® N95 Particulate Respirators, Models #4272 and #4273 meet the requirements of ASTM F1862.
NIOSH Certification Number: TC 84A-4350 (Models #4272 and #4273)
NIOSH Certification Number: TC 84A-4350 (Models #4272 and #4273)
NIOSH Certification Number: TC 84A-4350 (Models #4272 and #4273)
The Barrier® N95 Particulate Respirators, Models #4272 and #4273 meet the requirements of 16 CFR 1610 Standard for Flammability of Clothing Textiles.
The Barrier® N95 Particulate Respirators, Models #4272 and #4273 are not cytotoxic.
The Barrier® N95 Particulate Respirators, Models #4272 and #4273 are not irritating.
The Barrier® N95 Particulate Respirators. Models #4272 and #4273 do not display any potential for sensitization.
> Mölnlycke Health Care 5550 Peachtree Parkway, Suite 500 Norcross. GA 30092 USA
Tel 678 250 7900 Fax 678 250 7984 www.molnlycke.com
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Mölnlycke Health Care. The logo consists of a cluster of five black circles on the left, arranged in a triangular shape. To the right of the circles, the text "MÖLNLYCKE HEALTH CARE" is displayed in a sans-serif font, with "MÖLNLYCKE" on the top line and "HEALTH CARE" on the bottom line.
、 ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------
______________________________________________________________________________________________________________________________________________________________________________ ------
---------The table below presents the similarities and differences of the technological characteristics between the Barrier® N95 Particulate Respirators, Models #4272 and #4273 and the predicate device, Gerson Isolair APR Type N95 Healthcare Particulate Respirator and Surgical Mask Model 2735 (K960778).
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
| Feature | Barrier® N95 Particulate Respirators,<br>Models #4272 and #4273<br>(Proposed Device) | Gerson Isolair APR Type N95 Healthcare Particulate<br>Respirator and Surgical Mask Model 2735<br>(Predicate Device) |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| 510(k) Clearance | TBD | K960778 |
| Manufacturer | Mölnlycke Health Care | LOUIS M. GERSON CO., INC |
| Common Name | Particulate Respirator | Particulate Respirator |
| Classification # | Class II | Class II |
| Classification Name | 21 CFR 878.4040 | 21 CFR 878.4040 |
| Product Code | MSH | MSH |
| Materials | | |
| Outer Material | Polypropylene | Polyester |
| Inner Material | Polyethylene/Polypropylene | Polyester |
| Filter Media | Polypropylene | Polypropylene |
| Nose Clamp | Aluminum | Aluminum/ PVC foam |
| Binding/cover Tapes | Polyester | |
| Head Bands | Polyurethane (blue), ultrasonically welded | Rubber (green), stapled |
| Specifications and<br>Dimensions | Small Mask<br>Length 205 mm ± 5 mm<br>Width 85 ± 5 mm<br>Band Length 190 ± 5mm | Medium Mask<br>Length 240 mm ± 5 mm<br>Width 85 ± 5 mm<br>Band Length 225 ± 5mm |
| | Length 140 mm ± 6 mm<br>Width 95 mm ± 6 mm<br>Height 64 mm ± 3 mm<br>Band Length (top) 318 mm ± 6 mm<br>Band Length (bottom) 260 mm ± 6 mm | |
| Mask Style | Trapezoid mask when folded,<br>Duckbill when open | Cup shaped |
| Design Features | Single use, flat folded, single use, disposable respirator<br>Longer vertical bottom length for more conformability<br>to wearer,<br>aluminum nose clamp to contour to the wearer | |
| | Single use, cup shaped respirator<br>Contoured nosepiece<br>Double stapled headstraps | |
| NIOSH certification | TC 84A-4350 (Models #4272 and #4273) | TC 84A-160 |
| Sterililty | Non-sterile | Non-sterile |
| Fluid Resistance | Pass | Pass |
| Particulate Filtration<br>Efficiency | NIOSH Certification Number: TC 84A-4350<br>(Models #4272 and #4273) | |
| | NIOSH Certification Number: TC 84A-160 | |
| Bacterial Filtration<br>Efficiency | NIOSH Certification Number: TC 84A-4350<br>(Models #4272 and #4273) | |
| | NIOSH Certification Number: TC 84A-160 | |
| Differential Pressure | NIOSH Certification Number: TC 84A-4350<br>(Models #4272 and #4273) | |
| | NIOSH Certification Number: TC 84A-160 | |
| Flammability | Pass | Not Known |
Mölnlycke Health Care 5550 Peachtree Parkway, Suite 500 Norcross. GA 30092 USA
Tel 678 250 7900 Fax 678 250 7984 www.molnlycke.com
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Molnlycke Health Care. On the left side of the logo, there are five black circles clustered together. To the right of the circles, the text "MÖLNLYCKE HEALTH CARE" is written in black, with "MÖLNLYCKE" on the top line and "HEALTH CARE" on the bottom line.
# Performance-Data:- -----------------------------------------------------------------------------------------------------------------------------------------------------------
The performance data are summarized above.
#### Clinical Testing:
No clinical data was required.
#### Conclusion:
Based on the performance testing, it can be concluded that the Barrier® N95 Particulate Respirators, Models #4272 and #4273 are equivalent to the Gerson Isolair APR Type N95 Healthcare Particulate Respirator and Surgical Mask Model 2735 (K960778) predicate with respect to intended use, materials, design, technological characteristics and performance.
> Mölnlycke Health Care 5550 Peachtree Parkway, Suite 500 Norcross. GA 30092 USA
> > .
Tel 678 250 7900 Fax 678 250 7984 www.molnlycke.com
.
{4}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Caitlin Senter Regulatory Affairs Specialist Mölnlycke Health Care US, LLC 5550 Peachtree Parkway, Suite 500 Norcross, Georgia 30092
NOV 19 2010
Re: K102923
Trade/Device Name: Barrier® N95 Particulate Respirators, Models #4273, NIOSH Certification TC-84A-4350 Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: MSH Dated: November 1. 2010 Received: November 2, 2010
Dear Ms. Senter:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
ff your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.goy/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation
Center for Devices and Radiological Health
### Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Molnlycke Health Care. The logo consists of a cluster of five black circles on the left side, arranged in a way that they appear to be grouped together. To the right of the circles, the text "MÖLNLYCKE HEALTH CARE" is written in bold, uppercase letters.
K102923
| | INDICATION FOR USE | NOV 19 2010 |
|--|--------------------|-------------|
|--|--------------------|-------------|
| 510(k) Number (if known): | K102923 |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Barrier® N95 Particulate Respirators, Models #4272 and #4273, NIOSH certification TC-84A-4350 |
| Indication for Use: | The device is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. |
| Prescription Use ______ | And/Or | Over the Counter Use X |
|-------------------------|--------|------------------------|
|-------------------------|--------|------------------------|
(21 CFR Part 801 Subpart D)
And/Or
X (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page of
.
.
Elaine S. Mayhall
(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices
102923
510(k) Number:
Mölnlycke Health Care 5550 Peachtree Parkway, Suite 500 Norcross. GA 30092 USA
Tel 678 250 7900 Fax 678 250 7984 www.molnlycke.com
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.