DISCOVRED

K102821 · Phoenix Dental, Inc. · LFC · Mar 11, 2011 · Dental

Device Facts

Record IDK102821
Device NameDISCOVRED
ApplicantPhoenix Dental, Inc.
Product CodeLFC · Dental
Decision DateMar 11, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.1740
Device ClassClass 2

Indications for Use

DiscovRED™ is indicated for use to detect carious dentine.

Device Story

DiscovRED™ is a red dye agent used by dental professionals in a clinical setting to identify carious dentin. The device is applied as a drop to suspected carious areas, rinsed with water, and air-dispersed. The dye stains carious tissue, which is then removed by the clinician using a low-speed rotary bur. This process is repeated until no further stainable tissue remains. The device does not involve chemical reactions, power-assisted components, or electrical energy. It serves as a visual aid for the clinician to distinguish between healthy and decayed tooth structure, facilitating targeted removal of caries and potentially preserving healthy tooth structure.

Clinical Evidence

No clinical data. Substantial equivalence is based on technological characteristics, intended use, and comparative toxicological analysis of the primary ingredients.

Technological Characteristics

Glycol-based chemical agent containing a red dye colorant. Primary ingredient is polyethylene glycol. Non-powered, manual application via micro-sponge. Shelf-life of two years. No electrical or optical energy source. Not a chemical reaction-based device.

Indications for Use

Indicated for use by dental professionals to detect carious dentin in patients presenting with suspected tooth decay.

Regulatory Classification

Identification

The caries detection device is a device intended to show the existence of decay in a patient's tooth by use of electrical current.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # PHOENIX DENTAL IN "Better Dentistry through Chemistry" 11224 Phoenix Drive Fenton, MI 48430 K102821 ## MAKERS OF SUPER SEAL® Phone: 810.750.2328 or Toll-free: 877.463.9905 Fax: 810.750.7495 www.phoenixdental.com | Submitting Applicant Company | | |--------------------------------------|-------------------------------------------------------------------------------------| | 1) Name | Phoenix Dental, Inc. | | 2) Address | 3452 West Thompson Road<br>Fenton, Michigan, 48430 | | 3) Contact Person | Jeffrey S. Cox, President/CEO<br>Telephone: 877-463-9905<br>Facsimile: 810-750-7495 | | 4) Establishment Registration Number | 1836322 | | 5) Date | September 22, 2010<br>MAR 11 20 | 510(k) SUMMARY | 2. Name of Device | | |------------------------|-------------------------| | 1) Proprietary Name | DiscovRED™ | | 2) Common / Usual Name | Caries Detector | | 3) Classification Name | Caries Detection Device | - 3. Legally marketed device to which the submitter claims equivalence Predicate Device Caries Detector by Kuraray Medical Inc. (K012733) 1) Class = 2) Regulation Number 872.1740 3) Panel Dental 4) Classification Product Number LFC 5) Prescription Use Only Yes - 4. Description of device subject to this premarket notification submission The product, DiscovRED™, is classified into the Caries Detection Device category of CFR 21, Section 872.1740. DiscovRED™, as well as the Predicate Device, Caries Detector, is a red dye agent that when applied to the suspected carious areas of the tooth stains the carious dentine. DiscovRED™ is a caries detector. Both devices have a two year shelf-life. DiscovRED™ is for Prescription Use only and not over-the-counter use. The method of operation and placement of DiscovRED™ is similar to that of the Predicate Device, Caries Detector. A drop of the detector is applied to the carious dentine, rinsed with water and air dispersed. The stained carious tissue is then removed with a low speed rotary bur. The process may be repeated until there are no more stainable tissues in the cavity. DiscovRED™ and the Predicate Device do not involve chemical reactions. DiscovRED™ and the Predicate Device do not use any power-assisted devices such as optical light emitting or electrical energy forms. {2}------------------------------------------------ #### 510(k) Summary (continued) #### 5. Statement of the intended use The intended use of the device, DiscovRED™, is as follows: "DiscovRED™ is indicated for use to detect carious dentine." DiscovRED™ has the same intended clinical use as the Predicate Device, Caries Detector, of detecting carious dentine when a dental patient presents with an area of suspected tooth decay. - 6. Statement of technological characteristics and safety DiscovRED™ has similar technological characteristics as the Predicate Device, Caries Detector. Both devices are composed of glycol based chemistries in combination with a red dye colorant that is used to stain/detect carious dentine of the tooth. The safety of DiscovRED™ is very similar to the commercially available Predicate Device, Caries Detector. Both devices must be handled by a dental professional, i.e. dentist, according to their Instructions for Use and Material Safety Data Sheets. The two devices' main ingredients, polyethylene glycol for DiscovRED™ and propylene glycol for Caries Detector, have similar potential health effects. Repeated or prolonged exposure of either product is not known to aggravate any medical condition. The first aid measures of both devices are similar in that for any skin contact, wash the area with soap and water, and for eye contact, flush with water and seek medical attention. For inhalation, remove to fresh air, and for ingestion, rinse the mouth with water and spit into sink. Toxicological data on the primary ingredients of both DiscovRED™ and Caries Detector are also similar; however, LDso values for oral toxicity indicate that the primary ingredient in Caries Detector, propylene glycol, is more toxic (lower values) than the primary ingredient, polyethylene glycol, in DiscovRED™. Therefore, one may summize that DiscovRED™ is less toxic than the commercially available Predicate Device, Caries Detector. The small amount, less than one drop, used on a micro sponge in a single application of the device is unlikely a sufficient amount to cause harmful affects to a human. The Instructions for Use and MSDS information should be followed. {3}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular arrangement of the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" surrounding a stylized eagle symbol. The eagle is depicted with its wings spread, and its body is formed by three curved lines. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Mr. Jeffrev S. Cox President/ Chief Executive Officer Phoenix Dental. Incorporated 3452 West Thompson Road Fenton, Michigan 48430 - Re: K102821 MAR 1 1 2011 Trade/Device Name: DiscovREDTM Regulation Number: 21 CFR 872.1740 Regulation Name: Caries Detection Device Regulatory Class: II Product Code: LFC Dated: March 4, 2011 Received: March 7, 2011 Dear Mr. Cox: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2- Mr. Cox Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Anthony D. Watson Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health #### Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number: K102821 Device Name: DiscovRED™ Indications for Use: DiscovRED™ is indicated for use to detect carious dentin. Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (Per 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Susan Rump (Division Sign-Off Division of Anesthesiology, General Hospital Infection Control, Dental Devices Page 1 of ____________________________________________________________________________________________________________________________________________________________________ 510(k) Number: Page 2-1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...