GAL-1C BLOOD GLUCOSE MONITORING SYSTEM, GAL-1C TEST STRIPS

K102816 · Apex BioTechnology Corp. · NBW · Sep 15, 2011 · Clinical Chemistry

Device Facts

Record IDK102816
Device NameGAL-1C BLOOD GLUCOSE MONITORING SYSTEM, GAL-1C TEST STRIPS
ApplicantApex BioTechnology Corp.
Product CodeNBW · Clinical Chemistry
Decision DateSep 15, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The GAL-1C Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. Alternative site testing should be performed only during steady-state (when glucose is not changing rapidly). Testing is done outside the body (in vitro diagnostic use). It is indicated for lay use by people with diabetes, as an aid to monitoring levels in Diabetes Mellitus and should only be used by a single patient and it should not be shared. It is not indicated for the diagnosis or screening of diabetes or for neonatal use. The GAL-1C Blood Glucose Test Strips are to be used with the GAL-1C Blood Glucose Meter to quantitatively measure glucose in capillary whole blood taken from fingertips, palm, or forearm. Alternative site testing should be performed only during steady-state (when glucose is not changing rapidly). They are not indicated for the diagnosis or screening of diabetes or for neonatal use. The purpose of the Contrex Plus III Glucose Control Solutions is to validate the performance of the blood glucose monitoring system using a testing solution with a known range of glucose. A control test that falls within the acceptable range indicates the user's technique is appropriate and the test strip and meter are functioning properly.

Device Story

GAL-1C system measures glucose in capillary whole blood via test strips inserted into a handheld meter. User applies blood sample to strip; meter performs electrochemical analysis to quantify glucose concentration. Designed for home use by patients with diabetes; single-patient use only. Includes autocoding feature for strip calibration. Healthcare providers use results to monitor diabetes management. Benefits include convenient, quantitative glucose monitoring for glycemic control. System includes Contrex Plus III control solution to verify meter/strip performance.

Clinical Evidence

Clinical user performance study (N=145) compared patient-obtained capillary blood glucose results (fingertip, palm, forearm) against YSI 2300 reference method. Accuracy met ISO 15197 criteria. Additional bench testing confirmed linearity (16-602 mg/dL), precision (CV 0.7-5.8%), hematocrit interference (30-55%), altitude (up to 10,000 ft), and temperature/humidity stability.

Technological Characteristics

Electrochemical glucose monitoring system. Features: autocoding, 3-button interface, reduced-volume test strip chamber. Software: modified to support autocoding. Disinfection: validated for durability and Hepatitis B inactivation. Connectivity: not specified.

Indications for Use

Indicated for lay use by people with diabetes to monitor blood glucose levels in Diabetes Mellitus. Samples: fresh capillary whole blood from fingertips, forearm, or palm. Contraindications: not for diagnosis or screening of diabetes; not for neonatal use.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K102816 . SEP 1 5 2011 | 510(k) Summary | | |----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter: | Hsue-mei Lee<br>Manager of Quality Assurance Department<br>Apex BioTechnology Corp.<br>No. 7, Li-Hsin Road V, Hsinchu Science Park<br>Hsinchu, 30078<br>CHINA (TAIWAN)<br><br>email: hsue-mei@apexbio.com<br>Phone: 011-886-3-5641952<br>FAX: 011-886-3-5678302 | | Contact Person: | Hsue-mei Lee<br>Manager of Quality Assurance Department<br>Apex BioTechnology Corp.<br>No. 7, Li-Hsin Road V, Hsinchu Science Park<br>Hsinchu, 30078<br>CHINA (TAIWAN)<br><br>email: hsue-mei@apexbio.com<br>Phone: 011-886-3-5641952<br>FAX: 011-886-3-5678302 | | Date Prepared: | September 13, 2011 | | Trade Names: | GAL-1C Blood Glucose Monitoring System<br>GAL-1C Blood Glucose Test Strips<br>Contrex Plus III Glucose Control Solution | | Classification: | Glucose test system, 21 CFR 862.1345, Class II<br>Single (specified) analyte controls (assayed and unassayed), 21 CFR<br>862.1660, Class I | | Product Codes: | CGA, NBW, JJX | | Predicate Device: | GlucoSure STAR Blood Glucose Monitoring System (k073648)<br>Contrex Plus Glucose Control Solution (100747) | | Device Description: | The GAL-1C blood glucose meter and GAL-1C test strips used for<br>testing of blood glucose by self-testers at home with Contrex Plus III<br>Glucose Control Solutions for quality control testing | | Intended Use: | The GAL-1C Blood Glucose Monitoring System is intended for the quantitative<br>measurement of glucose in fresh capillary whole blood samples drawn from the<br>fingertips, forearm, or palm. Alternative site testing should be performed only<br>during steady-state (when glucose is not changing rapidly). Testing is done outside<br>the body (In Vitro diagnostic use). It is indicated for lay use by people with<br>diabetes, as an aid to monitoring levels in Diabetes Mellitus and should only be used<br>by a single patient and it should not be shared. It is not indicated for the diagnosis or<br>screening of diabetes or for neonatal use.<br><br>The GAL-1C Blood Glucose Test Strips are to be used with the GAL-1C Blood<br>Glucose Meter to quantitatively measure glucose in capillary whole blood taken<br>from fingertips, palm, or forearm. Alternative site testing should be performed only<br>during steady-state (when glucose is not changing rapidly). They are not indicated<br>for the diagnosis or screening of diabetes or for neonatal use.<br><br>The purpose of the control solution test is to validate the performance of the Blood<br>Glucose Monitoring System using a testing solution with a known range of glucose.<br>A control test that falls within the acceptable range indicates the user's technique is<br>appropriate and the test strip and meter are functioning properly. | | Comparison of<br>Technological<br>Characteristics: | The GAL-IC meter uses the same test algorithm as the predicate meter. The<br>GAL-1C meter has been modified by changing 4 operating buttons to 3<br>operating buttons. Meter software has been changed to accommodate the<br>new autocoding feature. The test strip holder has been modified to allow<br>automatic detection of the calibration code upon insertion of the test strip.<br>The test strip chemistry has been slightly modified and the blood collection<br>chamber has been reduced in volume. The Contrex Plus III Control Solution<br>includes an added red dye and slightly modified glucose concentrations. | | Non-Clinical<br>Testing: | Testing was conducted as follows: Software verification and validation,<br>software integration, linearity, Lo/Hi detection, drop testing, EMC and<br>Electrical Safety, verification of strip noninterchangeability between new and<br>predicate devices, precision, interferences, minimum sample volume,<br>altitude, hematocrit, humidity and temperature, control solution qualification,<br>environmental conditions testing, test strip shelf life and use life, control<br>solution shelf life and use life, and battery life testing. Disinfection testing<br>was done to show that the system remained accurate after multiple<br>disinfections to simulate a life time of disinfection treatments. Evaluation<br>was done to demonstrate the ability of the selected disinfectant material to<br>inactive Hepatitis B. Results demonstrate substantial equivalence to the<br>predicate device meter, test strips, and control solutions. | | Clinical Testing | An accuracy study was performed with blood testing by healthcare<br>professionals. A User Performance study was conducted with self-testing at<br>finger, palm, and forearm sites. A User Study was conducted to evaluate<br>ease-of-use of the system and ease-of-understanding of the User's Manual.<br>Results demonstrate substantial equivalence to the predicate system. | | Conclusion: | Clinical and non-clinical testing demonstrated that the GAL-1C system<br>performs in a substantially equivalent manner to that of the predicate system.<br>We conclude that the GAL-1C meter and GAL-1C test strips are<br>substantially equivalent to the predicate devices. | {1}------------------------------------------------ ## 510(k) Summary (Continued) {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which features three lines that resemble human figures, and a circle of text around the symbol. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA". The logo is black and white. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Building 66 Silver Spring, MD 20993 Apex BioTechnology Corporation c/o IIsue-mei Lee Manager of Quality Assurance Department No. 7, Li-Hsin Road V, Hsinchu Science Park Hsinchu. 30078 China (Taiwan) SEP 15 2011 Re: k102816 GAL-1C Blood Glucose Monitoring System, GAL-1C Blood Glucose Test Trade/Device Name: Strips, and Contrex Plus III Glucose Control Solutions Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: NBW, CGA, JJX Dated: August 22, 2011 Received: August 23, 2011 Dear Sir/Madam We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You nust comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807). Iabeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. {3}------------------------------------------------ Page 2 - If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Viro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcessorYou/Industry/default.htm. Sincerely yours, signature Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use Statement 510(k) Number (if known): k102816 Device Name: GAL-IC Blood Glucose Monitoring System Indications for Use: GAL-1C Blood Glucose Monitoring System: The GAL-1C Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. Alternative site testing should be performed only during steady-state (when glucose is not changing rapidly). Testing is done outside the body (In 1 Vitro diagnostic use). It is indicated for lay use by people with diabetes, as an aid to monitoring levels in Diabetes Mellitus and should only be used by a single patient and it should not be shared. It is not indicated for the diagnosis or screening of diabetes or for neonatal use. GAL-1C Blood Glucose Test Strips: The GAL-1C Blood Glucose Test Strips are to be used with the GAL-1C Blood Glucose Meter to quantitatively measure glucose in capillary whole blood taken from fingertips, palm, or forearm. Alternative site testing should be performed only during steady-state (when glucose is not changing rapidly). They are not indicated for the diagnosis or screening of diabetes or for neonatal use. Prescription Use Over-The-Counter Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) signature Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) k102816 Page 1 of 2 {5}------------------------------------------------ ## Indications for Use Statement 510(k) Number (if known): k102816 Device Name: Contrex Plus III Glucose Control Solutions Indications for Use: Intended use The purpose of the control solution test is to validate the performance of the Blood Glucose Monitoring System using a testing solution with a known range of glucose. A control test that falls within the acceptable range indicates the user's technique is appropriate and the test strip and meter are functioning properly. Over-The-Counter Use X = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = Prescription Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) k102816 Page 2 of 2
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...