K102808 · GC America, Inc. · LBH · Dec 22, 2010 · Dental
Device Facts
Record ID
K102808
Device Name
MI VARNISH
Applicant
GC America, Inc.
Product Code
LBH · Dental
Decision Date
Dec 22, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3260
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MI Varnish is a fluoride varnish with Recaldent™ (CPP-ACP) that has a desensitizing action when applied to tooth surfaces. The application leaves a film of varnish on tooth and this may allow visual control and verification.
Device Story
MI Varnish is a fluoride varnish containing 5% sodium fluoride and Recaldent (CPP-ACP) for desensitizing tooth surfaces. Delivered in unit-dose packaging; applied to teeth by clinicians using a disposable brush. Forms a visible film on the tooth surface for verification. Acts as a cavity varnish to provide desensitizing action. Used in dental clinical settings.
Clinical Evidence
Bench testing only. Comparative physical testing performed for pH, consistency, and total fluoride content against Duraflor and Vanish Varnish. Results: MI Varnish (pH 6.6, consistency 39mm, fluoride 2.4%) compared to Duraflor (pH 6.8, consistency 40mm, fluoride 2.2%) and Vanish Varnish (pH 6.4, consistency 26mm, fluoride 2.6%).
Technological Characteristics
Fluoride varnish paste containing 5% sodium fluoride and Recaldent (CPP-ACP). Includes hydrogenated rosin (rosin derivative) and diethylene glycol monoethyl ether. Delivered in unit-dose packaging for manual application via disposable brush. Physicochemical properties: pH 6.6, consistency 39mm, total fluoride 2.4 wt%.
Indications for Use
Indicated for patients requiring desensitizing treatment on tooth surfaces. Applied by dental professionals.
Regulatory Classification
Identification
Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.
Special Controls
*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.
{0}------------------------------------------------
Section 6 - 510(k) Summary
ﻨ
Image /page/0/Picture/1 description: The image shows the alphanumeric string "K102808" in a handwritten style. The characters are bold and slightly slanted, with the "K" being the largest and most distinct character. The numbers are connected, giving the impression of a continuous flow.
Image /page/0/Picture/2 description: The image shows the logo for CCI, which is a company that specializes in ammunition. The logo is black and white and features the letters "CCI" in a bold, sans-serif font. There are two apostrophe-like marks on either side of the letters. The registered trademark symbol is on the bottom right.
DEC 2 2 2010
GC AMERICA IN
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
- 1. Submitter Information: GC AMERICA INC.
3737 W. 127ª Street Alsip, IL 60803
Contact Person: Mark Heiss, D.D.S. Phone: (708) 897-4042 Fax: (708) 897-4031
Alternate Contact: Lori Lappo Phone: (708) 897-4036 Fax: (708) 897-4031
Date Prepared: September 20, 2010
- 2. Device Name:
Proprietary Name: MI Varnish Classification Name: Cavity Varnish (21CFR 872.3620) Device Classification: Class II Product Code: гвн
- Predicate Devices: 3.
| Company | Device | K Number |
|------------------------------------------|-----------------------------------------------------|----------|
| A.R. Medical Inc.<br>Pharmascience, Inc. | Duraflor | K961893 |
| 3M ESPE Dental Products | Vanish Varnish, 5% Sodium<br>Fluoride White Varnish | K092141 |
| GC Corporation | MI Paste Plus | K070854 |
- 4. Device description and Intended Use
MI Varnish is a fluoride varnish (5% sodium fluoride with CPP-ACP) which has a desensitizing action when applied to tooth surfaces. The application leaves a film of varnish on tooth and this may allow visual control and verification. MI Vamish is packed in unit dose and can be applied on teeth with a disposable brush.
The applicant device, MI Varnish is substantially equivalent to Duraflor and Vanish Varnish in intended use. Both devices are fluoride varnish as Cavity Varnish, and used for desensitizing action when applied to teeth.
Image /page/0/Picture/18 description: The image contains two logos. The logo on the left is a diamond shape with the text "CERTIFIED TO ISO 13485" inside. The logo on the right is a globe with the text "DTA" in the center and the word "MEMBER" above it.
{1}------------------------------------------------
The chemical composition is similar to the predicate devices (Duraflor and Vanish Varnish) with the exception of Recaldent (CPP-ACP). Recaldent is found in the predicate device MI Paste Plus.
## 5. Components and Mode of Action
The applicant device, MI Varnish is a paste formulation delivered in unit dose package and applied on the tooth surfaces with a disposable brush.
The chemical components of MI Varnish are similar to that of Duraflor. Components of MI Varnish that have been used in approved devices, MI Paste Plus (GC Corporation), Cavit G K875133 (3M ESPE Dental products)/Vanish Varnish, K092141 (3M ESPE Dental products) are listed in the chemical formulation.
The mode of action of the applicant device is substantially equivalent to that of the predicate devices, Duraflor and Vanish Varnish.
## 6. Technological characteristics:
MI Varnish contains two new components that have not been used in dental varnish or predicate devices.
Hydrogenated rosin is a rosin derivative. It is made with hydrogenation treatment of usual rosin (CAS No. 8050-09-7). Rosin (Colophony) and other rosin derivatives are used as adhesive agents in FDA approved dental varnishes. These include Duraflor, Vanish Varnish, Enamel Pro Varnish, K062683 (Premier), and the other fluoride varnishes.
Rosin and hydrogenated rosin are structurally similar. Hydrogenated rosin retains the basic structure of rosin and has extra conjugated double bonds. The difference is reflected in appearance, color and physicochemical properties between rosin and derivatives, but not on toxicological properties.
Diethylene glycol monoethyl ether is a solvent of polyvinyl acetate. The compound is already being used in Japan as medical device in "Caviton EX" for temporary filling material. (The number: 221AABZX00131000 is approved by Health and Welfare Ministry in Japan)
## 7. Summary of Physical tests:
| | pH | Consistency<br>(mm) | Total fluoride<br>(wt %) |
|----------------|-----|---------------------|--------------------------|
| MI Varnish | 6.6 | 39(2) | 2.4(0.1) |
| Duraflor | 6.8 | 40(1) | 2.2(0.1) |
| Vanish Varnish | 6.4 | 26(1) | 2.6(0.1) |
## Summary of Performance Specifications
According to GC Corporation R&D test methods.
As described above, the applicant device, MI Varnish is substantially equivalent to comparative devices, such as Duraflor and Vanish Varnish.
{2}------------------------------------------------
- Description of Safety and Substantial Equivalence: 8.
The applicant device is substantially equivalent to the predicate devices in its intended use.
Most of the chemical components in MI Varnish have been used in the predicate devices. MI Varnish contains two new components, however, one is a rosin derivative and another one is already being used as medical device in Japan. So, we believe that this fact well supports the compatibility of MI Varnish, and the safety of the applicant device is substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its wings and tail feathers. The eagle is positioned within a circular border that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in capital letters.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mark Heiss, DDS Director -Academic and Regulatory Affairs GC America, Incorporated 3737 West 127th Street Alsip, Illinois 60803
DEC 2 2 2010
Re: K102808 Trade/Device Name: MI Varnish Regulation Number: 21 CFR 872.3260 Regulation Name: Cavity Varnish Regulatory Class: II Product Code: LBH Dated: September 27, 2010 Received: September 28, 2010
Dear Dr. Heiss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 – Dr. Heiss
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. - You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
hh for.
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
Indications for Use
K10 2808 510(k) Number (if known): Device Name: MI Varnish
DEC 2 2 2010
Indications for Use:
MI Varnish is a fluoride varnish with Recaldent™ (CPP-ACP) that has a desensitizing action when applied to tooth surfaces. The application leaves a film of varnish on tooth and this may allow visual control and verification.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Quarrer
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K102808
i '
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.