NATRELLE 133 TISSUE EXPANDER WITH SUTURE TABS

K102806 · Allergan Medical · LCJ · Jan 5, 2011 · SU

Device Facts

Record IDK102806
Device NameNATRELLE 133 TISSUE EXPANDER WITH SUTURE TABS
ApplicantAllergan Medical
Product CodeLCJ · SU
Decision DateJan 5, 2011
DecisionSESE
Submission TypeSpecial
Device ClassClass U
AttributesTherapeutic

Indications for Use

The NATRELLE® Tissue Expanders are used in breast reconstruction following mastectomy, treatment of underdeveloped breasts, and treatment of soft tissue deformities. The expanders are intended for temporary subcutaneous implantation and are not intended for use beyond six months.

Device Story

NATRELLE® 133 Tissue Expander is a silicone elastomer device for temporary soft tissue expansion. It consists of an expansion envelope with a BIOCELL® textured surface, three suture tabs, and a MAGNA-SITE® integrated injection site containing a titanium needle guard. The device is implanted subcutaneously by a surgeon. During the expansion phase, sterile saline is injected into the site; the MAGNA-FINDER® Xact external locating device uses rare-earth magnets to identify the injection site location through the skin. The suture tabs facilitate surgical placement. The device is used in clinical settings for breast reconstruction or soft tissue deformity correction. It benefits patients by gradually stretching skin and soft tissue to create space for subsequent reconstructive procedures.

Clinical Evidence

Bench testing only. Biocompatibility testing of the suture tab material was performed per ISO 10993-1:2009. Suture tab bond integrity was validated per ISO 14607:2007(E) and ASTM F1441, meeting all predetermined acceptance criteria.

Technological Characteristics

Silicone elastomer expansion envelope with BIOCELL® textured surface; MAGNA-SITE® injection site with self-sealing port, titanium needle guard, and rare-earth permanent magnets; Mohawk P118 polyester mesh suture tabs. Dimensions vary by style/size. Non-powered, passive implantable device. Sterilization method not specified.

Indications for Use

Indicated for breast reconstruction post-mastectomy, treatment of underdeveloped breasts, and soft tissue deformities. Intended for temporary subcutaneous implantation (max 6 months).

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K102806 page 1/2 # 510(k) SUMMARY JAN - | | | | JAN 5 2011 | |-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|------------| | 510(k) Owner: | Allergan Medical | | | | Contact: | Micah Schloss | | | | Phone: | 805-961-5828 | | | | Fax: | 714-796-9724 | | | | Date Summary Prepared: | September 27, 2010 | | | | Device: | | | | | Trade Name: | NATRELLE® 133 Tissue Expander with Suture Tabs | | | | Common/Classification Name: | Expander, Skin, Inflatable | | | | Classification Regulation: | LCJ (Unclassified) | | | | Predicate Devices: | NATRELLE® Style 133 Series Tissue Expander Matrix (K862203) | | | | Device Description: | Allergan's Tissue Expanders are constructed from silicone elastomer and consist of an expansion envelope with a BIOCELL® textured surface, three suture tabs, a MAGNA-SITE® integrated injection site (incorporating a titanium needle guard), and a stable, reinforced base to enable outward expansion. The three suture tabs are attached to the posterior surface of the shell at approximately the 12 o'clock, 4 o'clock, and 8 o'clock positions of the tissue expander. The Tissue Expanders are available in seven styles and each style consists of six sizes for a total of 42 products, to meet diverse surgical needs. | | | | | The MAGNA-SITE® injection site and MAGNA-FINDER® Xact external locating device contain rare-earth, permanent magnets for an accurate injection system. When the MAGNA-FINDER® Xact external locating device is passed over the surface of the tissue being expanded, its rare-earth, permanent magnet indicates the location of the MAGNA-SITE® injection site. All injection sites contain a self-sealing port and a titanium needle guard to prevent inadvertent puncture through the base of the injection site. | | | | Intended Use: | The NATRELLE® Tissue Expanders are used in breast reconstruction following mastectomy, treatment of underdeveloped breasts, and treatment of soft tissue deformities. The expanders are intended for temporary subcutaneous implantation and are not intended for use beyond six months. | | | {2}------------------------------------------------ K102806 page 2/2 #### Allergan Medical #### Special 510(k): Device Modification NATRELLE® 133 Tissue Expanders with Suture Tabs The predicate device and the modified device utilize the same fundamental Technological Characteristics: technology: - · A silicone expansion envelope with a BIOCELL® textured surface, which expands with sequential injections of sterile saline - · A silicone injection site, with self-sealing port, a magnetic locating system and a titanium needle guard In accordance with ISO 10993-1:2009 and Allergan's internal standards and Biocompatibility procedures, a biocompatibility review of Mohawk P118 polyester mesh was Data: performed. Based on the test results, the material is approved and considered qualified for use in the suture tabs of NATRELLE® Tissue Expanders. All other materials used in NATRELLE® Tissue Expanders are identical to those used in the predicate device. Suture tab bond integrity was tested in accordance with ISO 14607:2007(E), Performance ASTM F1441 and Allergan's internal standards and procedure, and Testing: predetermined acceptance criteria were met. The device passed all preclinical testing, indicating that no new questions of Conclusions: safety and effectiveness are introduced with the proposed modification. NATRELLE® 133 Tissue Expander with Suture Tabs is substantially equivalent to Allergan's currently marketed NATRELLE® Style 133 Series Tissue Expander Matrix (K862203). {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of a human figure. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002 3 - 11 the state of the state of the states of the states of Allergan Medical % Mr. Micah Schloss Associate Regulatory Analyst 71 S. Los Carneros Road Goleta, California 93117-5506 Re: K102806 Trade/Device Name: NATRELLE® Tissue Expanders Regulation Number: 21 CFR 878.3600 Regulation Name: Tissue expander Regulatory Class: II Product Code: LCJ Dated: December 3, 2010 Received: December 8, 2010 Dear Mr. Schloss: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {4}------------------------------------------------ Page 2 - Mr. Micah Schloss CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electionic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1118809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, pease note the regulation entitled, "Misbranding by reference to premarket notification" (21 CIFR Preu 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll the (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. Arij. Vs. Nh fr. Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Allergan Medical ### Special 510(k): Device Modification NATRELLE® 133 Tissue Expanders with Suture Tabs ## Indications for Use 510(k) Number (if known): K102806 Device Name: NATRELLE® Tissue Expanders Indications for Use: The NATRELLE® Tissue Expanders are used in breast reconstruction following mastectomy, treatment of underdeveloped breasts, and treatment of soft tissue deformities. The expanders are intended for temporary subcutaneous implantation and are not intended for use beyond six months. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE) | | Daniel Krone for MXM | |-----------------------------------------------------------|----------------------| | (Division Sign-Off) | | | Division of Surgical, Orthopedic, and Restorative Devices | | | 510(k) Number | K102806 |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...