ADVIA 2120 WITH AUTOSLIDE SYSTEM, AND ADVIA 2120I

K102644 · Siemens Healthcare Diagnostics · GKZ · Nov 23, 2011 · Hematology

Device Facts

Record IDK102644
Device NameADVIA 2120 WITH AUTOSLIDE SYSTEM, AND ADVIA 2120I
ApplicantSiemens Healthcare Diagnostics
Product CodeGKZ · Hematology
Decision DateNov 23, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5220
Device ClassClass 2

Indications for Use

The ADVIA 2120 and ADVIA 2120i with autoslide are quantitative, automated hematology analyzers that provide: - A complete blood count (CBC) consisting of WBC, RBC, Hgb, CN-Free Hgb, Calculated Hgb, MCV, Hct, MCH, MCHC, CHCM, RDW, HDW, CH, Plt, MPV. - A leukocyte differential count consisting of Neut (%/#), Lymph (%/#), Mono (%/#), Eos (%/#), Baso (%/#), LUC (%/#). - A reticulocyte analysis consisting of Retic (%/#), MCVg, MCVr, CHCMg, CHCMr, CHg, CHr. - A nucleated red blood cell count consisting of NRBC (%/#). - Enumeration of the total nucleated cell (TNC) count and RBC count for pleural, peritoneal, and peritoneal dialysis (PD) specimens. Note: Above measurands are determined (in whole blood, pleural, peritoneal, or peritoneal dialysis specimens with K2 and/or K3 EDTA anti-coagulants). - Quantitative determination of blood cells in Cerebrospinal Fluid (CSF) consisting of WBC, RBC, Neut (%/#), Lymph (%/#), Mono (%/#), MN (%/#), PMN (%/#). In addition, the system provides the added capability to automatically prepare and stain high quality blood smears on a microscope slide.

Device Story

ADVIA 2120/2120i are integrated hematology analyzers; system aspirates, dilutes, and analyzes whole blood, CSF, and body fluid samples; includes auto-sampler for tube processing and auto-slide module for wedge smear preparation/staining. Device replaces legacy 386ex CPU board with ARM9 CPU board. System uses analytical module to perform quantitative cell counts; computer workstation manages data, user interface, and reporting. Healthcare providers use results for CBC, differential, and morphology assessment to aid clinical decision-making. Benefits include automated, high-throughput hematology analysis and reflexive slide preparation.

Clinical Evidence

Bench testing only. Performance evaluated across three configurations (legacy hardware/software, legacy hardware/new software, new hardware/new software) using 219 whole blood, 20 body fluid, 80 NRBC, and 52 CSF samples. Regression analysis demonstrated equivalence in analytical performance (r > 0.98 for most parameters). Precision (within-run) and linearity met specifications for all modes. Carryover was <1% for all channels.

Technological Characteristics

Automated hematology analyzer; analytical module with auto-sampler and slide stainer. Sensing via flow cytometry/optical analysis. ARM9 CPU board running Nucleus OS. Connectivity via 100Base-TX Ethernet. Power: 100-240 Vac, 50/60Hz. Dimensions: 141cm x 86cm x 68cm. Weight: 193 kg.

Indications for Use

Indicated for quantitative, automated hematology analysis of whole blood, pleural, peritoneal, peritoneal dialysis, and cerebrospinal fluid (CSF) specimens in clinical settings. Provides CBC, leukocyte differential, reticulocyte analysis, NRBC count, and body fluid cell counts. For prescription use only.

Regulatory Classification

Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k102644 B. Purpose for Submission: Replacement of the Analytical Module (AM) CPU board utilized in the ADVIA 2120/2120i Hematology Analyzers C. Manufacturer and Instrument Name: Siemens Healthcare Diagnostics Inc.; ADVIA® 2120 Hematology Auto analyzer and 2120i Hematology Auto analyzer with Auto slide D. Type of Test or Tests Performed: Quantitative test for WBC, RBC, Hgb, CN-Free Hgb, Calculated Hgb, MCV, Hct, MCH, MCHC, CHCM, RDW, HDW, Plt, MPV, Neut (%/#), Lymph (%/#), Mono (%/#), Eos (%/#), Baso (%/#), LUC (%/#), NRBC (%/#), Reticulocyte [Retic (%/#), MCVg, MCVr, CHCMg, CHCMr, CHg, CHr], CSF [WBC, RBC, Neut (%/#), Lymph (%/#), Mono (%/#), MN (%/#), PMN (%/#)], and Body Fluid [TNC, RBC] parameters. E. System Descriptions: 1. Device Description: The ADVIA 2120/210i Hematology Systems with Auto Slide are an integrated option of hematology analyzers with complete blood cell count, leukocyte differential cell count, reticulocyte analysis capability, nucleated red blood cell count, quantitative determination of blood cells in Cerebrospinal Fluid (CSF), enumeration of the total nucleated cell (TNC) count and RBC count for pleural, peritoneal, and peritoneal dialysis (PD) specimens and a slide stainer designed to provide reflexive slide making/staining without user intervention based upon pre-selected, user-definable criteria. The ADVIA 2120/210i Hematology systems with Auto slide consists of the following: an analytical module that aspirates, dilutes, and analyzes whole blood samples; an auto sampler that automatically mixes, identifies, and presents the samples for processing; a computer workstation that controls the instrument, provides primary user interface with the instrument and manages the data produced by the instrument; a printer that optionally generates reports based on the instrument results and an auto slide module that prepares a wedge smear from a drop of blood, places it on a microscope slide and stains the slide in accordance with Wright, Wright-Giemsa and May-Grunwald Giemsa Staining techniques. 2. Principles of Operation: Refer to individual 510(k) clearances below: k930148 (H*3 Reticulocyte measurands: Retic, MCVg, MCVr, CHCMg, CHCMr, CHg, CHr) k954954 (Reticulocyte measurands: Retic, MCVg, MCVr, CHCMg, CHCMr, CHg, CHr) k971998 (ADVIA 120 measurands except NRBC, body fluids and CSF) k003796 (ADVIA 120 CSF measurands: CSF WBC, Neut, Lymph, Mono, MN, PMN) {2} k012904 (ADVIA 120 CN-free Hgb measurand) k022668 (ADVIA 120 Calculated Hgb measurand) k022331 (ADVIA 120 revised measurands: CSF WBC, Neut, Lymph, Mono MN, PMN) k042251 (ADVIA 2120 measurands except NRBC and body fluids) k051693 (ADVIA 2120 with Auto slide, NRBC, all measurands except body fluids) k090346 (ADVIA 2120/2120i Body fluids Application measurands: RBC, TNC) 3. Modes of Operation: Refer to individual 510(k) clearances in item 2 above. 4. Specimen Identification: Refer to individual 510(k) clearances in item 2 above. 5. Specimen Sampling and Handling: Refer to individual 510(k) clearances in item 2 above. 6. Calibration: Refer to individual 510(k) clearances in item 2 above. 7. Quality Control: Refer to individual 510(k) clearances in item 2 above. 8. Software: FDA has reviewed applicant’s Hazard Analysis and Software Development processes for this line of product types: Yes ☐ X ☐ or No ☐ F. Regulatory Information: 1. Regulation section: 21 § CFR 864.5220, Automated differential cell counter 2. Classification: Class II 3. Product code: GKZ, Counter, differential cell 4. Panel: Hematology (81) G. Intended Use: 1. Indication(s) for Use: The ADVIA 2120 and ADVIA 2120i with autoslide are quantitative, automated hematology analyzers that provide: - A complete blood count (CBC) consisting of WBC, RBC, Hgb, CN-Free Hgb, Calculated Hgb, MCV, Hct, MCH, MCHC, CHCM, RDW, HDW, CH, Plt, MPV. - A leukocyte differential count consisting of Neut (%/#), Lymph (%/#), Mono (%/#), Eos (%/#), Baso (%/#), LUC (%/#). - A reticulocyte analysis consisting of Retic (%/#), MCVg, MCVr, CHCMg, CHCMr, CHg, CHr. - A nucleated red blood cell count consisting of NRBC (%/#). - Enumeration of the total nucleated cell (TNC) count and RBC count for pleural, peritoneal, and peritoneal dialysis (PD) specimens. Note: Above measurands are determined (in whole blood, pleural, peritoneal, 2 {3} or peritoneal dialysis specimens with K2 and/or K3 EDTA anti-coagulants). - Quantitative determination of blood cells in Cerebrospinal Fluid (CSF) consisting of WBC, RBC, Neut $(\%/\#)$ , Lymph $(\%/\#)$ , Mono $(\%/\#)$ , MN $(\%/\#)$ , PMN $(\%/\#)$ . In addition, the system provides the added capability to automatically prepare and stain high quality blood smears on a microscope slide. 2. Special Conditions for Use Statement(s): For prescription use only. # H. Substantial Equivalence Information: 1. Predicate Device Name(s) and 510(k) numbers: ADVIA 2120 Hematology Auto analyzer with current CPU board; k042251 2. Comparison with Predicate Device: | Similarities | | | | | --- | --- | --- | --- | | Item | ADVIA 2120 and 2120i with current 386ex CPU | | ADVIA 2120/2120i with ARM9 CPU | | Parameters | CBC Results | WBC, RBC, HGB, HCT, MCV, MCH, MCHC, CHCM, RDW, HDW, CH, PLT | Same | | | Differential Results | NEUT, LYMPH, MONO, EOS, BASO, LUC, NRBC (% and absolute) | | | | Platelet Results | PLT, MPV | | | | Reticulocyte Results | %RETIC, #RETIC, MCVr, CHCMr, CHr, MCVg, MCVr, CHCMg, CHg | | | | CSF Results | CSF RBC, CSF WBC, CSF MN, CSF PMN, CSF NEUT, CSF LYMPH, CSF MONO | | | | BF Results | TNC, RBC | | | Morphology Results | WBC: | Left Shift, Atypical Lymph, Blasts, Immature Granulocytes, Myeloperoxidase Deficiency | Same | | | RBC and PLT: | NRBC, ANISO, MICRO, MACRO, HC VAR, HYPO, HYPER, RBC Fragments, RBC Ghosts, Platelet Clumps, Large Platelets | | | User Interface personal computer | Intel based processorWindows 2000Floppy driveNetwork cardModem | | Same | | User Interface external peripherals | KeyboardMouseHand-held barcode scannerPrinter | | Same | {4} | Similarities | | | | --- | --- | --- | | Item | ADVIA 2120 and 2120i with current 386ex CPU | ADVIA 2120/2120i with ARM9 CPU | | Electrical Power | Voltage selectable for single-phase: 100 Vac (6 Amps) - 240 Vac (3 Amps) Frequency: 50/60Hz | Same | | Temperature | Operating: 18°C to 32°C Storage:-29°C to 60°C | Same | | Relative Humidity | Operating: 15%-80% (noncondensing) | Same | | Sample Mode Volumes | Automatic Closed-Tube: 175μL Manual Closed-Tube: 175μL Manual Open-Tube: 175μL | Same | | Test Selectivity/Throughput | CBC 120 samples/hr CBC/Diff 120 samples/hr CBC/Diff/Retic 74 samples/hr CBC/Retic 74 samples/hr Retic 74 samples/hr With Autoslide slide making enabled: CBC 108 samples/hr CBC/Diff 108 samples/hr | Same | | Sample Capacity | 150 samples 15 racks of 10 tubes | Same | | Barcode reader | Reads up to 14 digits Automatic label code discrimination Compatible barcode types: 1. Codabar 2. Interleaved 2 of 5 3. Code 39 4. Code 128 5. EAN and JAN (8 and 13) | Same | | Data Management | • TDC version 9 or higher • Database storage capacity of 10,000 records, including graphics • Review and edit capability ○ User-defined windows ○ User-defined reports ○ User-defined ranges based on age and sex for Normal, Rerun, Panic, and Delta Check criteria • Bi-directional and host query communication protocols • Quality control ○ 3D bar graph ○ Levey-Jennings plot ○ SDI graph ○ Table format • Remote QC • ILQC programs • Patient moving average • User assistance ○ Context sensitive help ○ Operator’s guide ○ Procedure wizards ○ Problem solving diagnostics ○ Remote diagnostics | Same | 4 {5} | Similarities | | | | --- | --- | --- | | Item | ADVIA 2120 and 2120i with current 386ex CPU | ADVIA 2120/2120i with ARM9 CPU | | Reagents | • CBC TIMEPAC ○ Baso ○ HGB ○ RBC/PLT ○ Defoamer • CN-Free CBC TIMEPAC ○ Baso ○ CN-Free HGB ○ RBC/PLT ○ Defoamer • DIFF TIMEPAC ○ Perox 1 ○ Perox 2 ○ Perox 3 ○ Perox Sheath • autoRetic • EZ KLEEN • Sheath/Rinse • CSF | Same | | Calibrators | • ADVIA OPTIpoint • ADVIA SETpoint | Same | | Controls | • ADVIA TESTpoint Low • ADVIA TESTpoint Normal • ADVIA TESTpoint High • ADVIA TESTpoint Retic Low • ADVIA TESTpoint Retic High • ADVIA TESTpoint 3-in-1 Abnormal1 • ADVIA TESTpoint 3-in-1 Normal • ADVIA TESTpoint 3-in-1 Abnormal2 | Same | | Differences | | | | --- | --- | --- | | Item | ADVIA 2120 and 2120i with current 386ex CPU | ADVIA 2120/2120i with ARM9 CPU | | Real-time Control | Intel 386ex CPU running Nucleus OS | ARM9 CPU running Nucleus OS | | User Interface software | ADVIA 2120/2120i user interface software (v5.8) | ADVIA 2120/2120i user interface software (v6.0) | | Communication Interface | BNC Ethernet cable (10Base-2) | RJ-45 Ethernet Cable (100Base-TX) | I. Special Control/Guidance Document Referenced (if applicable): CLSI EP9-A2, Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline-Second Edition CLSI EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline CLSI H26-A2, Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Proposed Standard-Second Edition J. Performance Characteristics: 1. Analytical Performance: {6} # a. Accuracy: Testing was performed in three hardware/software configurations as follows: | Configuration 1 | Configuration 2 | Configuration 3 | | --- | --- | --- | | System software V5.8 | System software V6.0 | System software V6.0 | | 386ex CPU board (P/N: 067-1452-01) | | ARM 9 CPU board, Cycle 2 (P/N: 067-B794-01) | | CPU Cover/Shield (P/N: 067-1448-01) | | CPU Cover/Shield (P/N: 067-1627-01) | | BNC Cable (P/N: 067-B391-02) | | RJ45 Ethernet Cable (P/N: 664-1516-01) | Configuration 1 is the current hardware/software configuration. Configuration 2 is the current hardware configuration, but uses a software version that can support both hardware configurations. Configuration 3 has both the new hardware and new software configuration. Testing was performed on 219 whole blood samples, 20 body fluid samples, 80 NRBC samples, and 52 CSF samples. All samples were assayed on an ADVIA 2120i in all three configurations. Samples were run duplicate, but only the first replicate was used for the analysis. The results of the regression analysis comparing all three configurations show equivalence in analytical performance, meeting bias specifications set by the sponsor. Configuration 1 versus Configuration 2 | CBC/Diff Parameter | r | Slope | Intercept | Syx | Average X | Average Y | Observed Bias | Bias Specs | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | WBC | 0.999 | 1.03 | -0.27 | 0.5 | 11.45 | 11.54 | 0.09 | ±0.15 | | RBC | 0.998 | 1.00 | 0.01 | 0.05 | 3.28 | 3.29 | 0.01 | ±0.05 | | Hgb | 0.997 | 0.98 | 0.1 | 0.1 | 9.2 | 9.2 | 0.0 | ±0.1 | | MCV | 0.998 | 0.99 | 0.9 | 0.6 | 91.1 | 91.0 | -0.1 | ±1.0 | | CHCM | 0.995 | 0.99 | 0.4 | 0.2 | 32.8 | 32.9 | 0.1 | ±0.5 | | RDW | 0.999 | 1.00 | 0.0 | 0.1 | 16.9 | 16.9 | 0.0 | ±1.0 | | HDW | 0.999 | 0.99 | 0.04 | 0.03 | 3.23 | 3.25 | 0.02 | ±0.2 | | Plt | 0.998 | 0.98 | 3 | 14 | 236 | 234 | -2 | ±6 | | MPV | 0.890 | 0.84 | 1.8 | 0.8 | 10.5 | 10.7 | 0.2 | ±0.5 | | %Neut | 0.995 | 0.99 | 0.8 | 2.5 | 61.4 | 61.3 | -0.1 | ±0.5 | | %Lymph | 0.991 | 0.98 | 0.7 | 2.6 | 22.1 | 22.3 | 0.2 | ±0.5 | | %Mono | 0.989 | 0.98 | 0.1 | 1.1 | 8.0 | 7.9 | -0.1 | ±1.5 | | %Eos | 0.944 | 0.99 | 0.1 | 0.7 | 1.9 | 1.9 | 0.0 | ±0.2 | | %Baso | 0.969 | 1.00 | 0.0 | 0.4 | 0.7 | 0.7 | 0.0 | ±0.2 | | %LUC | 0.985 | 0.98 | 0.0 | 1.7 | 6.2 | 6.0 | -0.2 | ±0.5 | | %Retic | 0.992 | 0.99 | 0.02 | 0.31 | 2.2 | 2.2 | 0.0 | ±0.5 | | CHr | 0.989 | 0.98 | 0.4 | 0.6 | 32.7 | 32.5 | -0.2 | ±0.5 | | %NRBC | 0.819 | 0.75 | 1.1 | 2.0 | 5.2 | 5.0 | -0.2 | ±0.5 | | | | | | | | | | | | Body Fluid Parameter | r | Slope | Intercept | Syx | Average X | Average Y | Observed Bias | Bias Specs | | TNC | 0.998 | 1.03 | -2.7 | 25.4 | 299 | 304 | 5 | ±10 | | RBC | 0.99997 | 1.01 | 0.1 | 0.8 | 29 | 29 | 0 | ±10 | | | | | | | | | | | | CSF Parameter | r | Slope | Intercept | Syx | Average X | Average Y | Observed Bias | Bias Specs | | TNC | 0.999 | 1.00 | 0.0 | 0.0 | 1.0 | 1.0 | 0 | ±0.1 | | RBC | 0.99997 | 0.99 | 0.0 | 0.0 | 0.0 | 0.0 | 0 | 0.0 | {7} | CBC/Diff Parameter | r | Slope | Intercept | Syx | Average X | Average Y | Observed Bias | Bias Specs | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | WBC | 0.995 | 1.06 | -9 | 33 | 245 | 251 | 6 | ±15 | | RBC | 0.999 | 1.01 | -3 | 23 | 355 | 357 | 2 | ±15 | | %MN | 0.992 | 1.00 | 0.6 | 2.3 | 52 | 53 | 1 | n/a | | %PMN | 0.992 | 1.00 | -0.7 | 2.3 | 48 | 47 | -1 | n/a | | %Neut | 0.992 | 1.00 | -0.6 | 2.2 | 47 | 47 | 0 | n/a | | %Lymph | 0.988 | 1.06 | -1.3 | 2.5 | 38 | 39 | 1 | n/a | | %Mono | 0.826 | 0.71 | 4.0 | 1.9 | 14 | 14 | 0 | n/a | Configuration 1 versus Configuration 3 | CBC/Diff Parameter | r | Slope | Intercept | Syx | Average X | Average Y | Observed Bias | Bias Specs | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | WBC | 0.999 | 1.03 | -0.24 | 0.6 | 11.45 | 11.56 | 0.11 | ±0.15 | | RBC | 0.998 | 1.00 | -0.01 | 0.04 | 3.28 | 3.27 | -0.01 | ±0.05 | | Hgb | 0.996 | 0.99 | 0.0 | 0.2 | 9.2 | 9.2 | 0.0 | ±0.1 | | MCV | 0.996 | 0.99 | 0.7 | 0.8 | 91.1 | 90.9 | -0.2 | ±1.0 | | CHCM | 0.994 | 0.98 | 0.7 | 0.2 | 32.8 | 32.9 | 0.1 | ±0.5 | | RDW | 0.999 | 1.00 | 0.0 | 0.2 | 16.9 | 16.9 | 0.0 | ±1.0 | | HDW | 0.998 | 1.00 | 0.01 | 0.04 | 3.23 | 3.26 | 0.03 | ±0.2 | | Plt | 0.997 | 0.99 | 0 | 18 | 234 | 234 | 0 | ±6 | | MPV | 0.906 | 0.83 | 2.0 | 0.7 | 10.5 | 10.8 | 0.3 | ±0.5 | | %Neut | 0.996 | 0.99 | 0.6 | 2.3 | 61.4 | 61.5 | 0.1 | ±0.5 | | %Lymph | 0.991 | 1.00 | 0.1 | 2.7 | 22.1 | 22.2 | 0.1 | ±0.5 | | %Mono | 0.987 | 0.98 | 0.0 | 1.2 | 8.0 | 7.8 | -0.2 | ±1.5 | | %Eos | 0.962 | 0.97 | 0.0 | 0.6 | 1.9 | 1.8 | -0.1 | ±0.2 | | %Baso | 0.966 | 0.98 | 0.0 | 0.4 | 0.7 | 0.7 | 0.0 | ±0.2 | | %LUC | 0.986 | 0.97 | 0.2 | 1.6 | 6.2 | 6.1 | -0.1 | ±0.5 | | %Retic | 0.991 | 1.02 | 0.03 | 0.34 | 2.21 | 2.28 | 0.1 | ±0.5 | | CHr | 0.992 | 0.98 | 0.4 | 0.5 | 32.7 | 32.6 | -0.1 | ±0.5 | | %NRBC | 0.814 | 0.78 | 1.0 | 2.1 | 5.2 | 5.0 | -0.2 | ±0.5 | | | | | | | | | | | | Body Fluid Parameter | r | Slope | Intercept | Syx | Average X | Average Y | Observed Bias | Bias Specs | | TNC | 0.996 | 1.03 | -16 | 34 | 299 | 292 | -7 | ±10 | | RBC | 0.9998 | 1.00 | 0.0 | 2.0 | 29 | 29 | 0 | ±10 | | | | | | | | | | | | CSF Parameter | r | Slope | Intercept | Syx | Average X | Average Y | Observed Bias | Bias Specs | | WBC | 0.996 | 1.02 | -4 | 29 | 245 | 246 | 1 | ±15 | | RBC | 0.999 | 0.99 | 3 | 17 | 355 | 354 | -1 | ±15 | | %MN | 0.993 | 1.01 | 0.3 | 2.2 | 52 | 53 | 1 | n/a | | %PMN | 0.993 | 1.01 | -0.8 | 2.2 | 48 | 47 | -1 | n/a | | %Neut | 0.993 | 1.01 | -1.0 | 2.1 | 47 | 47 | 0 | n/a | | %Lymph | 0.988 | 0.99 | 0.9 | 2.4 | 38 | 39 | 1 | n/a | | %Mono | 0.878 | 0.85 | 2.0 | 1.8 | 14 | 14 | 0 | n/a | b. Precision/Reproducibility: {8} Whole blood samples, simulated body fluid and CSF samples were assayed in 20 consecutive replicates on an ADVIA 2120i in configurations 2 and 3 to verify that the two test configurations meet the within run precision specifications. The observed results pass the precision specifications if either the SD or the CV% specification is met. Whole Blood Precision | | | Configuration 2 | | | Configuration 3 | | | Specifications | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Parameter | n | Mean | SD | % CV | Mean | SD | % CV | SD | % CV | | WBC | 20 | 4.32 | 0.1 | 2.36 | 4.36 | 0.1 | 2.24 | 0.2 | 2.66 | | RBC | 20 | 4.35 | 0.03 | 0.7 | 5.04 | 0.02 | 0.4 | 0.06 | 1.2 | | Hgb | 20 | 12.8 | 0.10 | 0.76 | 15.1 | 0.09 | 0.58 | 0.14 | 0.93 | | MCV | 20 | 90.1 | 0.2 | 0.19 | 88.9 | 0.2 | 0.25 | 0.7 | 0.78 | | CHCM | 20 | 33.5 | 0.04 | 0.1 | 34.7 | 0.06 | 0.2 | 0.25 | 0.8 | | RDW | 20 | 13.0 | 0.07 | 0.52 | 13.3 | 0.06 | 0.43 | 0.25 | 1.92 | | HDW | 20 | 2.52 | 0.012 | 0.46 | 2.50 | 0.009 | 0.38 | 0.1 | 3.57 | | Plt | 20 | 325 | 8.0 | 2.45 | 205 | 6.0 | 2.91 | 8.8 | 2.93 | | %Neut | 20 | 53.3 | 0.7 | 1.29 | 68.0 | 0.5 | 0.69 | 1.4 | 2.15 | | %Lymph | 20 | 26.9 | 0.9 | 3.2 | 15.1 | 0.6 | 4.1 | 1.1 | 4.4 | | %Mono | 20 | 8.5 | 0.6 | 6.6 | 9.9 | 0.5 | 5.1 | 0.9 | 15 | | %Eos | 20 | 7.6 | 0.4 | 5.4 | 5.0 | 0.4 | 7.5 | 0.5 | 25 | | %Baso | 20 | 0.8 | 0.1 | 16.2 | 0.7 | 0.1 | 10.2 | 0.5 | 50 | | %LUC | 20 | 2.9 | 0.3 | 10.8 | 1.2 | 0.2 | 17.2 | 0.5 | 25 | | %Retic | 20 | 1.2 | 0.1 | 7.1 | 0.9 | 0.1 | 6.2 | 0.4 | 20 | Body Fluid Precision | Autosampler | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | Configuration 2 | | | Configuration 3 | | | Specifications | | | Parameter | n | Mean | SD | % CV | Mean | SD | % CV | SD | % CV | | TNC | 20 | 521 | 22.4 | 4.3 | 523 | 25.6 | 4.9 | n/a | ≤15 | | RBC | 20 | 53 | 1.8 | 3.4 | 52 | 2.0 | 3.7 | n/a | ≤10 | | | | | | | | | | | | | Manual Closed Tube Sampler | | | | | | | | | | | | | Configuration 2 | | | Configuration 3 | | | Specifications | | | Parameter | n | Mean | SD | % CV | Mean | SD | % CV | SD | % CV | | TNC | 20 | 517 | 18.8 | 3.6 | 516 | 22.2 | 4.3 | n/a | ≤15 | | RBC | 20 | 52 | 1.6 | 3.0 | 53 | 1.8 | 3.3 | n/a | ≤10 | {9} 9 | Manual Open Tube Sampler | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | Configuration 2 | | | Configuration 3 | | | Specifications | | | Parameter | n | Mean | SD | % CV | Mean | SD | % CV | SD | % CV | | TNC | 20 | 505 | 23.5 | 4.7 | 460 | 33.9 | 7.4 | n/a | ≤15 | | RBC | 20 | 54 | 2.0 | 3.8 | 50 | 1.6 | 3.1 | n/a | ≤10 | CSF Precision | | | Configuration 2 | | | Configuration 3 | | | Specifications | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Parameter | n | Mean | SD | % CV | Mean | SD | % CV | SD | % CV | | WBC | 20 | 88 | 3.8 | 4.3 | 100 | 3.3 | 3.3 | ≤15 | ≤15 | | RBC | 20 | 111 | 6.5 | 5.9 | 102 | 4.5 | 4.4 | ≤15 | ≤15 | | %MN | 20 | 93 | 5.1 | 5.5 | 94 | 4.2 | 4.5 | ≤20 | ≤20 | | %PMN | 20 | 91 | 4.4 | 4.8 | 105 | 5.9 | 5.6 | ≤20 | ≤20 | c. Linearity: Linearity was tested using commercially prepared linearity controls. Each level of the control was assayed 4 times in configurations 2 and 3. Linearity was tested in whole blood, body fluid, and CSF modes. The table below summarizes linearity results. Over the ranges tested, none of the results exceeded the maximum deviation specified. | Parameter | Analytical Range | Configuration 2 Range tested | Configuration 2 Range tested | Specifications | | --- | --- | --- | --- | --- | | WBC (10³cells/μL) | 0.02 – 400 | 0.00 – 448.60 | 0.00 – 405.52 | Maximum deviation: ±0.5 x 10³cells/μL or 5.0% (whichever is greater) | | RBC (10⁶cells/μL) | 0.0 – 7.0 | 0.00 – 7.67 | 0.00 – 7.67 | Maximum deviation: ±0.1 x 10⁶ cells/μL | | Hgb (g/dL) | 0.0 – 22.5 | 0.0 – 25.0 | 0.0 – 24.5 | Maximum deviation: ±0.2g/dL or ±2.0% (whichever is greater) | | PLT (10³cells/μL) | 5 – 3500 | 0 – 5196 | 0 – 3675 | Maximum deviation: ±5 or ±5.0% (whichever is greater) | | %Retic (%) | 0.2 – 24.5 | 0.8 – 26.4 | 0.7 – 26.3 | Maximum deviation: ±5.0% | | CSF WBC Low (cells/μL) | 0 – 50 | 0 – 52 | 0 – 52 | Maximum Deviation: ±5 cells/μL | | CSF RBC Low (cells/μL) | 0 – 50 | 0 – 46 | 0 – 51 | Maximum Deviation: ±5 cells/μL | | CSF RBC High (cells/μL) | 50 – 1500 | 0 – 1509 | 0 – 1508 | Maximum Deviation: ±10% | | Body Fluid TNC Low (cells/μL) | 0.02 – 400 | 1 – 542 | 0 – 618 | Maximum deviation: ±10 cells/μL or ±10.0% (whichever is greater) | | Body Fluid RBC (10³cells/μL) | 0.1 – 6.76 | 0 – 7190 | 0 - 7105 | Maximum deviation: ±10x10³cells/μL or ±10.0% (whichever is greater) | {10} d. Carryover: Carryover was tested for the WBC, RBC, Hgb, Plt, and retic channels of the system. A carryover set consisted of assaying a high pool followed by 3 aspirations of phosphate buffered saline. The carryover procedure consisted of performing 10 carryover sets. Carryover was tested in configurations 2 and 3, and within each configuration, carryover was tested with the Autosampler (AS), the manual closed tube sampler (MCTS), and the manual opened tube sampler (MOTS). Carryover in the reticulocyte channel was calculated using the RTC-RBC count (i.e. the RBC count derived from the reticulocyte channel). The observed results for the 10 carryover sets met specifications for all parameters, having less than $1\%$ carryover in both configurations. | Parameter / Configuration | Average % Carryover | | | | --- | --- | --- | --- | | | ACTS | MCTS | MOTS | | WBC – Configuration 2 | 0.18 | 0.33 | 0.21 | | WBC – Configuration 3 | 0.19 | 0.26 | 0.17 | | RBC – Configuration 2 | 0.14 | 0.14 | 0.14 | | RBC – Configuration 3 | 0.14 | 0.14 | 0.40 | | Hgb – Configuration 2 | 0.00 | 0.00 | 0.00 | | Hgb – Configuration 3 | 0.00 | 0.00 | 0.00 | | PLT – Configuration 2 | 0.23 | 0.25 | 0.22 | | PLT – Configuration 3 | 0.82 | 0.83 | 0.83 | | RTC-RBC – Configuration 2 | 0.14 | 0.15 | 0.21 | | RTC-RBC – Configuration 3 | 0.14 | 0.14 | 0.24 | e. Interfering Substances: Refer to related $510(\mathrm{k})$ clearances in item 2 above. f. Background Counts: Refer to related $510(\mathrm{k})$ clearances in item 2 above. 2. Other Supportive Instrument Performance Data Not Covered Above: Not applicable K. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. L. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...