Vertex cold-curing denture base materials are indicated for: 1. Manufacture of full and partial dentures 2. Repair of full and partial dentures 3. Rebasing of full and partial dentures 4. Relining of full and partial dentures
Device Story
Vertex cold-curing denture base materials (Self Curing, Castavaria, Castapress) are self-polymerizing acrylic resins composed of polymethyl methacrylate powder and methyl methacrylate liquid. Used by dental professionals in clinical or laboratory settings to fabricate, repair, rebase, or reline full and partial dentures. Vertex Self Curing polymerizes in 10 minutes using a pressure vessel; Castavaria is a pourable, multifunctional material with minimized shrinkage and extended working time; Castapress is a pouring/casting material featuring a unique accelerator system for color stability. The materials are shaded to simulate gum tissue using cadmium-free pigments. The final cured acrylic provides the structural base for dental prostheses, restoring function and aesthetics for patients.
Clinical Evidence
Bench testing only. Performance data provided includes flexural strength, flexural modulus, water absorption, water solubility, and residual monomer content. Results demonstrate compliance with ISO 1567, ISO 20795, ISO 179-1, ISO 7405, and ASTM F 895-84 standards.
Technological Characteristics
Polymethyl methacrylate powder and methyl methacrylate liquid. Cadmium-free pigments. Self-polymerizing (cold-curing) acrylic resin. Standards: ISO 1567, ISO 20795, ISO 179-1, ISO 7405, ASTM F 895-84. No software or electronic components.
Indications for Use
Indicated for the manufacture, repair, rebasing, and relining of full and partial dentures in patients requiring dental prosthetics.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
{0}------------------------------------------------
K102640v
Submitter: Vertex Dental B.V. Vertex cold-curing denture base material Premarket Notification: Traditional 510(k) MAR 2 5 2011
.
ー
| 510(k) Summary |
|----------------|
| Submitter Name:<br>Submitter Address: | Vertex Dental B.V.<br>Johan van Oldenbarneveltlaan 62,<br>3705 HJ Zeist<br>The Netherlands |
|--------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Phone Number:<br>Fax Number: | 31 30 69 767 49<br>31 30 69 551 88 |
| Contact Person:<br>Contact Address: | William Greenrose<br>220 River Rd.<br>Claremont, NH 03743 |
| Phone Number:<br>Fax Number: | 603-369-3550<br>603-369-3562 |
| Date Prepared: | August 9, 2010 |
| Device Trade Name: | Vertex Self Curing, Vertex Castavaria, Vertex Castapress |
| Common Name | Denture base material |
| Classification Name,<br>Number &<br>Product Code: | resin, denture, relining, repairing, rebasing<br>872.3760<br>EBI |
| Predicate Devices: | Probase Cold (K913655), Major Repair (K082153) |
| Device Description<br>and Statement of<br>Intended Use | Device Description: Vertex cold-curing denture base materials are self curing acrylic denture base materials composed of polymethyl methacrylate powder and a liquid consisting of methyl methacrylate and other ingredients. The polymer is shaded to simulate the color of gum tissue using pigments that are approved for alimentary or similar use and are Cadmium free. The devices covered in this submission are the Vertex Self Curing, Vertex Castavaria, and the Vertex Castapress denture base materials.<br>VertexTM Self Curing is a self-curing acrylic denture base material intended for the repair and relining of full and partial dentures, made by heat-curing acrylics. This acrylic can be polymerized in 10 minutes using a pressure vessel.<br>VertexTM Castavaria is a multifunctional self polymerizing denture base material intended as a pouring and as a repair acrylic. The advantages |
{1}------------------------------------------------
of this acrylic are: minimized shrinkage, color stability, stable polymerization cycle and the acrylic is pourable for a long period of time. In addition Vertex™ Castavaria can be worked and modeled over a relatively long period of time.
Vertex™ Castapress is a self polymerizing pouring/casting type denture base material also suitable for repair, relining, rebasing and extensions of partial dentures. The colour stability of the material is excellent because of the use of an unique accelerator system
The Vertex Cold-Curing denture base materials are substantially equivalent to predicate denture base devices presently on the USA market and safety and effectiveness are well documented in dental literature.
## Intended Use:
Vertex cold-curing denture base materials are indicated for:
- 1. Manufacture of full and partial dentures
- 2. Repair of full and partial dentures
- Rebasing of full and partial dentures 3.
- Relining of full and partial dentures ধ :
Summary of Technological Characteristics All of the components found in Vertex™ cold-curing denture base materials have been used in legally marketed devices and were found safe for dental use.
0
{2}------------------------------------------------
| Feature | Vertex Cold-Curing Denture<br>base material | Probase Cold | Major Repair |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>Number | Not yet assigned | K913655 | K082153 |
| Manufacturer | Vertex Dental B.V. | Ivoclar | Major Prodotti Dentari<br>S.p.a |
| Classification<br># & Product<br>Code | 872.3760<br>EBI | 872.3760<br>EBI | 872.3760<br>EBI |
| Indications<br>for Use | Vertex cold-curing denture<br>base materials are indicated<br>for:<br>1. Manufacture of full<br>and partial dentures<br>2. Repair of full and<br>partial dentures<br>3. Rebasing of full and<br>partial dentures<br>4. Relining of full and<br>partial dentures | The cold-curing ProBase<br>Cold is suitable for both the<br>pouring and the packing<br>technique.<br>Indication<br>- Partial dentures<br>- Combination dentures<br>- Relining<br>- Repairs<br>- Complete dentures | Major.Repair is a cold-<br>curing polymer for dental<br>prosthesis. Poly-<br>methylmethacrylate<br>based. For reparing and<br>rebasing dentures.<br>Powder and liquid.<br>It is used to:<br>- repair and rebase<br>prosthesis<br>- temporary prosthesis |
| Physical<br>Properties | flexural strength:<br>Vertex Self Curing: 68 MPa<br>Vertex Castavaria: 79 MPa<br>Vertex Castapress: 75 MPa | flexural strength: 63.44 ±<br>4.24 MPa | flexural strength: 66.4<br>MPa |
| | flexural modulus:<br>Vertex Self Curing: 2028 MPa<br>Vertex Castavaria: 2316 MPa<br>Vertex Castapress: 2293 MPa | flexural modulus: 1832.00<br>± 88.87 MPa | flexural modulus: 2217<br>MPa |
| | water absorption:<br>Vertex Self Curing: 20.3 µg/mm³<br>Vertex Castavaria: 23.2 µg/mm³<br>Vertex Castapress: 22.1 µg/mm³ | | water absorption: 21.2<br>µg/mm³ |
| | Water solubility:<br>Vertex Self Curing:1.8 µg / mm³<br>Vertex Castavaria: 1.8 µg / mm³<br>Vertex Castapress:1.19 ± 0.18<br>µg / mm³ | | Water solubility: 1.4 µg /<br>mm³ |
| | Residual monomer:<br>Vertex Self Curing: 3.76 ± 0.15%<br>Vertex Castavaria: 3.91 ± 0.05%<br>Vertex Castapress: 3.22 ± 0.06% | Residual monomer:<br><4.5% | Residual monomer:<br>4.0% |
| Standards of<br>Conformity | ISO 1567<br>ISO 20795<br>ISO 179-1<br>ISO 7405<br>ASTM F 895-84 | ISO 1567 | ISO 1567 |
## Summary of Technical Characteristics
{3}------------------------------------------------
| Submitter:<br>Vertex Dental B.V. | Vertex cold-curing denture base material<br>Premarket Notification: Traditional 510(k) |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Substantial<br>Equivalence | The Vertex cold-curing denture base materials are substantially<br>equivalent to the Probase Cold (K913655) and Major Repair (K082153),<br>with respect to mode of action and intended use. It is substantially<br>equivalent to the Probase Cold (K913655) and Major Repair (K082153)<br>denture base materials, with respect to material of composition. The<br>submitted devices pose the same types of questions about safety or<br>effectiveness as the existing device. |
| Conclusion | The information discussed above demonstrates that the Vertex cold-<br>curing denture base materials are substantially equivalent to the<br>predicate devices. |
| Declarations | • This summary includes only information that is also covered in the<br>body of the 510(k).<br>• This summary does not contain any puffery or unsubstantiated<br>labeling claims.<br>• This summary does not contain any raw data, i.e., contains only<br>summary data.<br>• This summary does not contain any trade secret or confidential<br>commercial information.<br>• This summary does not contain any patient identification information. |
・
.
.
:
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Vertex Dental B.V. C/O Mr. William F. Greenrose Qserve America 220 River Road Claremont, New Hampshire 03743
MAR 2 5 2011
Re: K102640 Trade/Device Name: Vertex Self Curing Vertex Castavaria
Vertex Castapress Regulation Number: 21 CFR 872.3760
Regulation Name: Denture Relining, Repairing, or Rebasing Resin Regulatory Class: II Product Code: EBI Dated: March 22, 2011 Received: March 23, 2011
Dear Mr. Greenrose:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2- Mr. Greenrose
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D.m
Anthony D. Watson, B.S., M.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
.102640
510(k) Number (if known):
Device Name:
Vertex Self Curing Vertex Castavaria Vertex Castapress
Indications For Use:
Vertex cold-curing denture base materials are indicated for:
- Manufacture of full and partial dentures 1.
- Repair of full and partial dentures 2.
- ડે. Rebasing of full and partial dentures
- 4. Relining of full and partial dentures
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Puan
(Division Sign-Off)
Divisitation Sign-Off)
Divisites of Sentrol, Dental Devices Infection Cor
510(k) Number:
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.