← Product Code [JWH](/productcode/JWH) · K102580

# VANGUARD REMOVABLE MOLDED POLY TIBIA (K102580)

_Biomet Manufacturing Corp · JWH · Mar 3, 2011 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K102580

## Device Facts

- **Applicant:** Biomet Manufacturing Corp
- **Product Code:** [JWH](/productcode/JWH.md)
- **Decision Date:** Mar 3, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Vanguard® Removable Molded Poly Tibia is intended for: 1. Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis traumatic arthritis where one or more compartments are involved; 2. Failure of a previous joint replacement procedure 3. Correction of varus, valgus, or post traumatic deformity: 4. Correction or revision of unsuccessful osteotomy or arthrodesis. The device is a sinqle use implant intended for use with bone cement.

## Device Story

Vanguard® Removable Molded Poly Tibia is a knee prosthesis component replacing the tibial articulating surface. System consists of three parts: Co-Cr-Mo tibial plate, UHMWPE tibial bearing, and Ti-6AL-4V insert. Device is implanted by orthopedic surgeons in a clinical/OR setting during total knee arthroplasty. It functions as a semi-constrained cemented prosthesis to restore joint function and alleviate pain in patients with degenerative or traumatic knee conditions. The device is a single-use implant. It is used in conjunction with previously cleared Vanguard® Knee Systems. Clinical benefit is the restoration of knee joint mobility and reduction of pain.

## Clinical Evidence

No clinical testing was necessary for a determination of substantial equivalence. Evidence consists of bench-level mechanical fatigue testing, including CoCr I-Beam Tray 80/20 fatigue, Removable Molded Tibia PS post fatigue, intercondylar polyethylene thickness analysis, and retention testing.

## Technological Characteristics

Materials: Co-Cr-Mo (tibial plate), UHMWPE (tibial bearing), Ti-6AL-4V (insert). Design: Semi-constrained, cemented, fixed cruciate stem. Sizing: Includes 59mm, 87mm, and 91mm options. Connectivity: None (implantable). Sterilization: Not specified.

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- Vanguard Removable Molded Poly Tibia ([K060525](/device/K060525.md))
- MCK [Maxim] Knee System ([K915132](/device/K915132.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K102580
p1/2.

510(k) Summary

| Preparation Date:    | January 3, 2010                                                                                             |
|----------------------|-------------------------------------------------------------------------------------------------------------|
| Applicant/Sponsor:   | Biomet Manufacturing Corp.                                                                                  |
| Contact Person:      | Tamara West                                                                                                 |
| Proprietary Name:    | Vanguard® Removable Molded Poly Tibia                                                                       |
| Common Name:         | Knee Prosthesis                                                                                             |
| Classification Name: | Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis (21CFR §888.3560) |

MAR - 3 2011

## Legally Marketed Devices To Which Substantial Equivalence Is Claimed:

| K060525 | Vanguard Removable Molded Poly Tibia | Biomet Manufacturing Corp. |
|---------|--------------------------------------|----------------------------|
| K915132 | MCK [Maxim] Knee System              | Biomet Manufacturing Corp. |

## Device Description:

The Vanguard® Removable Molded Poly Tibia (also referred to as the Vanguard® Mono-Lock) is intended to replace the tibial articulating surface in a joint replacement and works in conjunction with previously cleared Vanguard® Knee Systems. There are three parts to the Vanguard® Removable Cruciate Molded Polyethylene Tibia component: the tibial plate (Co-Cr-Mo), tibial bearing (UHMWPE) and an insert (Ti-6AL-4V).

The profile of the tibial plate is identical to that of the predicates. The stem geometry of the Vanguard® Removable Molded Poly Tibia is non modular and does not include the conical core geometry to house the female taper as the predicate K915132.

The articulation surface is identical to the surface cleared in K060525.

Intended Use: The Vanguard® Removable Molded Poly Tibia is intended for: 1. Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis traumatic arthritis where one or more compartments are involved;

- 2. Failure of a previous joint replacement procedure
- 3. Correction of varus, valgus, or post traumatic deformity:
- 4. Correction or revision of unsuccessful osteotomy or arthrodesis.

The device is a sinqle use implant intended for use with bone cement.

{1}------------------------------------------------

K102580
p.2/2

### Summary of Technologies:

The technological characteristics (material, design, and sizing) of the Vanguard® Removable Molded Poly Tibia are identical to the predicate devices. The only change made to the Vanquard® Removable Molded Poly Tibia from the predicate device K060525 is to the stem design; the new device has a fixed cruciate stem whereas the predicate has a fixed I-beam stem. Additional sizes to the tibial plate and bearing include 59mm, 87mm and 91mm, which are identical to the predicate devices in K915132. The material, locking mechanism, and bearing articulation remain identical to the predicate devices. Therefore, Biomet believes the Vanguard® Removable Molded Poly Tibia is substantially equivalent to the previously cleared devices.

#### Non-Clinical Testing:

Mechanical testing was previously performed to determine substantial equivalence and the following test reports were provided in the predicate 510(k): CoCr I-Beam Tray 80/20 Fatique Test Report, Removable Molded Tibia PS Post Fatigue Strength, 10mm PS Component - Intercondylar Polyethylene Thickness Analysis, Vanguard Removable Retention Test. The CoCr Cruciate Plate 80/20 Fatigue Test Report is also provided to further indicate the devices are functional within its intended use.

## Clinical Testing:

No clinical testing was necessary for a determination of substantial equivalence.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS) in the USA. The logo features a stylized image of an eagle or bird with outstretched wings, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird, indicating the department's name and country.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Biomet Manufacturing Corp. % Ms. Tamara J. West Senior Regulatory Affairs Specialist 56 East Bell Drive P.O. Box 587 Warsaw, Indiana 465801

MAR - 3 2011

Re: K102580

Trade/Device Name: Vanguard® Removable Molded Poly Tibia

Regulation Number: 21 CFR 888.3560

Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis

- Regulatory Class: II Product Code: JWH Dated: January 3, 2011 Received: January 7, 2011
Dear Ms. West:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

Page 2 - Ms. Tamara J. West

comply with all the Act's requirements, including, but not limited to: registration and listing (21) CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known):_K102580

Device Name: The Vanquard® Removable Molded Poly Tibia

Indications For Use:

The Vanquard® Cruciate Molded Poly Tibia is intended for:

1. Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis traumatic arthritis where one or more compartments are involved;

2. Failure of a previous joint replacement procedure

3. Correction of varus, valgus, or post traumatic deformity;

4. Correction or revision of unsuccessful osteotomy or arthrodesis.

The device is a single use implant intended for use with bone cement.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

for M. Melkerson

(Division Sign-Off) (Division Sign-Off)
Division of Surgical, Orthopedic,
Division of Surgices Divisious Surges and Restorative Devices

Page 1 of 1

510(k) Number _

---

**Source:** [https://fda.innolitics.com/device/K102580](https://fda.innolitics.com/device/K102580)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
