K102579 · Biological Signal Processing, Ltd. · MWI · Dec 2, 2010 · Cardiovascular
Device Facts
Record ID
K102579
Device Name
HYPERQ AD-100 SYSTEM
Applicant
Biological Signal Processing, Ltd.
Product Code
MWI · Cardiovascular
Decision Date
Dec 2, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2300
Device Class
Class 2
Indications for Use
ECG is intended to disclose either normal condition or patterns of arrhythmia, myocardial ischemia, rate abnormalities, or features of prognostic value in the following cases: Patients with suspected cardiac abnormalities Populations of patients at an age or period in which a routine baseline evaluation of ECG characteristics is desired. Stress Testing Angina pectoris (chest pain) is a clinical syndrome resulting from myocardial ischemia, indicative of reduced blood supply to the cardiac muscle. The electrocardiogram may establish the diagnosis of ischemic heart disease if characteristic changes are present. Stress testing is the most widely used method to decide whether this chest pain is related to myocardial ischemia, and thus to coronary artery disease. In stress testing, the contractile capability of the heart muscle is monitored via ECG during patient exercise. Patients exercise by bicycle, treadmill, or other means, while the ECG is monitored continuously. Exercise loads are determined by predefined protocols. The ECG signals as well as the HF-QRS signals are recorded for the resting, exercise, and recovery phase portions of the exercise protocol. The changes in both ECG waveforms are compared to the resting ECG,records. In the HyperQ™ stress test, changes in the high frequency of the mid QRS complex, calculated as root, meansquare (RMS) values, are compared to the resting values. Most of the commercial stress test systems control the bicycle or treadmill automatically according to the requirements of the chosen protocol, although this is not essential. ST segment monitoring is intended as an aid in the evaluation of myocardial ischemia in patients with known or suspected coronary artery disease. The ST segment algorithm has been tested for accuracy of the ST segment data, and a database is used as a tool for performance testing. The significance of the ST segment changes must be determined by a physician. HyperQ™ The HyperQ™ Software is intended to be used as an aid to stress ECG test by means of analysis of high frequency components present within the central portion of the QRS complex. The significance of the HF-QRS changes must be determined by a physician.
Device Story
HyperQ AD-100 System is a compact monitor for measuring, processing, storing, and displaying ECG data. It functions as an accessory to existing ECG stress systems. Input: standard 12-lead ECG signals (1000 samples/sec). Operation: device analyzes high-frequency components of the QRS complex; calculates root-mean-square (RMS) values of the mid-QRS complex; compares exercise/recovery phase values to resting baseline. Output: processed HF-QRS data displayed for physician review. Used in clinical settings during exercise stress testing (treadmill/bicycle). Physician interprets output to aid in diagnosing myocardial ischemia and coronary artery disease. Benefits: provides additional prognostic information beyond standard ST-segment analysis.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, including electrical safety (IEC 60601-1, IEC 60601-2-25, IEC 60601-2-27) and EMC testing. ST segment algorithm accuracy was validated using a database as a tool for performance testing.
Technological Characteristics
12-lead ECG monitor; 1000 samples/sec; 16-bit A/D resolution; 0.05-300 Hz frequency range; >100M Ohm input impedance; >100 dB CMRR. Connectivity: accessory to external stress test systems. Power: 5V, 200mA. Dimensions: 17 x 11 x 2.5 cm. Software includes ST segment analysis and HF-QRS analysis algorithms.
Indications for Use
Indicated for patients with suspected cardiac abnormalities or those requiring routine baseline ECG evaluation. Used as an aid in stress ECG testing to evaluate myocardial ischemia in patients with known or suspected coronary artery disease. Significance of findings must be determined by a physician.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
K102579 pl/3
# 510(k) Summary of Safety & Effectiveness
.:'
DEC = 2 2011
# Pursuant to CFR 807.92, the following 510(k) Summary is provided:
| 1. (a) | Submitter<br>Address: | George J. Hattub<br>MedicSense, USA<br>291 Hillside Avenue<br>Somerset, MA 02726<br>www.medicsense.com |
|------------------------------------|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. (b)<br>Manufacturer<br>Address: | | Biological Signal Processing Ltd.<br>22a Raul Wallenberg Street<br>Tel-Aviv 69719, Israel |
| | Mfg. Phone: | Tel.: 972-3-647-4840 |
| | Contact Person: | Eran Toledo |
| | Date: | August 31, 2010 |
| 2. | Device &<br>Classification<br>Name: | Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)<br>device as a class II device (product code MWI, 21 CFR 870.2300)<br>HyperQ™ AD-100 System |
| 3. | Predicate Device: | HyperQ™ System K082564 |
| 4. | Description: | The HyperQ™ AD-100 System is a compact monitor for measuring,<br>processing, storing, and displaying information derived from an<br>electrocardiogram (ECG). The device analyzes and records the high<br>frequency components of the QRS complex of standard ECG. |
| 5. | Intended Use: | ECG |
| | | ECG is intended to disclose either normal condition or patterns of<br>arrhythmia, myocardial ischemia, rate abnormalities, or features of<br>prognostic value in the following cases:<br>Patients with suspected cardiac abnormalities<br>Populations of patients at an age or period in which a routine baseline<br>evaluation of ECG characteristics is desired. |
| | | Stress Testing |
| | | Angina pectoris (chest pain) is a clinical syndrome resulting from myocardial<br>ischemia, indicative of reduced blood supply to the cardiac muscle. The<br>electrocardiogram may establish the diagnosis of ischemic heart disease if<br>characteristic changes are present.<br>Stress testing is the most widely used method to decide whether this chest |
.
: ,
{1}------------------------------------------------
K102579
p2/3
pain is related to myocardial ischemia, and thus to coronary artery disease. In stress testing, the contractile capability of the heart muscle is monitored via ECG during patient exercise. Patients exercise by bicycle, treadmill, or other means, while the ECG is monitored continuously. Exercise loads are determined by predefined protocols. The ECG signals as well as the HF-QRS signals are recorded for the resting, exercise, and recovery phase portions of the exercise protocol. The changes in both ECG waveforms are compared to the resting ECG,records. In the HyperQ™ stress test, changes in the high frequency of the mid QRS complex, calculated as root, meansquare (RMS) values, are compared to the resting values. Most of the commercial stress test systems control the bicycle or treadmill automatically according to the requirements of the chosen protocol, although this is not essential.
ST segment monitoring is intended as an aid in the evaluation of myocardial ischemia in patients with known or suspected coronary artery disease. The ST segment algorithm has been tested for accuracy of the ST segment data, and a database is used as a tool for performance testing. The significance of the ST segment changes must be determined by a physician.
### HyperQ TM
The HyperQ™ Software is intended to be used as an aid to stress ECG test by means of analysis of high frequency components present within the central portion of the QRS complex.
The significance of the HF-QRS changes must be determined by a physician.____________________________________________________________________________________________________________________________________________________________________ Carlos Concession Company Concession Company Concession Company Concession Company Controller Company Controller Company Controller Company Controller Company Controller Comp
Comparison of Technological Characteristics:
6.
With respect to technology and intended use, the Modified HyperQ™ System (HyperQ™ AD-100 System) is substantially equivalent to its predicate device which is the HyperQ™ System. The primary difference is that the modified device has the ability to be an accessory (add-on) to a legally marketed ECG Stress System to provide HyperQ analysis from the acquired ECG. Based upon the validation results, BSP believes this change does not raise additional safety of efficacy concerns. In addition, the system has passed IEC Electrical Safety Testing as well as EMC.
The following predicate comparison table delinerates the specific similarities and differences:
{2}------------------------------------------------
K102579
P3/3
| Characteristic | Predicate Device (System) | Modified Device (System) |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| Number of<br>Electrodes | 10 | same |
| Number of Leads | 12 | same |
| ECG<br>Samples/second | 1000 | same |
| A/D Bits | 16 (0.15 µVLSB) | same |
| Defibrillation<br>Protection | Built In | same (utilizes a<br>defibrillation proof<br>patient cable) |
| Input Signal<br>dynamic range | 10 mV | same |
| Simultaneously<br>12L | Yes, with<br>sample/hold<br>circuitry to assure<br>zero delay<br>between leads | same |
| CMMR | > 100 dB | same |
| Input Impedance | > 100M Ohm | same |
| DC max input | ±330 mV | same |
| Frequency Range<br>(-3db) | 0.05-300 Hz | same |
| Low pass filter<br>(software) | 35 Hz | same |
| Line pass filter<br>(software) | 50/60 Hz | same |
| Safety Standard | IEC 60601-1, EN<br>60601-2, IEC<br>60601-2-25, IEC<br>60601-2-27, EC11<br>ST Segment<br>Analysis<br>EC 38-1999 | Same |
| Operating<br>Temperature | 0°C to 50°C | same |
| Treadmill Control | yes | no |
| On Line ECG | yes | no |
| HyperQ calculated<br>from on line ECG | yes | yes |
| Power | 5 V | same |
| Current | 80 mA | 200 mA |
| Dimensions | 15 x 12 x 2 cm | 17 x 11 x 2.5 cm |
| Weight | 200 grams | same |
| Storing<br>Temperature | -20°C to 60°C | same |
| Humidity | 0.85% | same |
·
:
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three lines representing its wings and body. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Biological Signal Processing, Ltd. c/o Mr. George J. Hattub Senior Staff Consultant MedicSense, USA 291 Hillside Avenue Somerset, MA 02726
DEC - 2 2010
Re: K102579
> Trade/Device Name: Hyper-QTM AD-100 System Regulatory Number: 21 CFR 870.2300 Regulation Name: Patient Physiological Monitor (without arrhythmia detection or alarms) Regulatory Class: II (two) Product Code: 74 MWI Dated: October 26, 2010 Received: November 2, 2010
Dear Mr. Hattub:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 – Mr. George J. Hattub
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K102579
## Indications for Use
### 510(k) Number (if known):
DEC - 2 2010
### Device Name: HyperQ™System
Indications For Use: ECG is intended to disclose either normal condition or patterns of arrhythmia, myocardial ischemia, rate abnormalities, or features of prognostic value in the following cases:
· Patients with suspected cardiac abnormalities
• Populations of patients at an age or period in which a routine baseline evaluation of ECG characteristics is desired.
### Stress Testing
Angina pectoris (chest pain) is a clinical syndrome resulting from myocardial ischemia, indicative of reduced blood supply to the cardiac muscle. The electrocardiogram may establish the diagnosis of ischemic heart disease if characteristic changes are present.
Stress testing is the most widely used method to decide whether this chest pain is related to myocardial ischemia, and thus to coronary artery disease. In stress testing, the contractile capability of the heart muscle is monitored via ECG during patient exercise. Patients exercise by bioycle, treadmill, or other means, while the ECG is monitored continuously. Exercise body of treading, on one may of The ECG signals as well as the HF-QRS signals are recorded for the resting, exercise, andreasonals phase portions of the exercise protocol. The changes in both ECG waveforms are compared to the resting ECG records. In the HyperQ™ stress test, changes in the high frequency of the mid QRS complex, calculated as root-mean-square (RMS) values, are compared to the resting values.
Most of the commercial stress test systems control the bicycle or treadmill automatically according to the requirements of the chosen protocol, although this is not essential.
ST segment monitoring is intended as an aid in the evaluation of myocardial ischemia in patients with known or suspected coronary artery disease. The ST segment algorithm has been tested for accuracy of the ST segment data, and a database is used as a tool for performance testing.
The significance of the ST segment changes must be determined by a physician.
### HyperQ™
The HyperQ™ Software is intended to be used as an aid to the ECG stress test by means of analysis of high frequency components present within the central portion of the QRS complex.
The significance of the HF-QRS changes must be determined by a physician.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
1. J.M.
Page 1 of 1
(Division Sign-Off)
Division of Cardiovascular Devices