← Product Code [HGX](/productcode/HGX) · K102516

# LUCINA MODEL MELODI (K102516)

_Genadyne Biotechnologies, Inc. · HGX · Aug 11, 2011 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K102516

## Device Facts

- **Applicant:** Genadyne Biotechnologies, Inc.
- **Product Code:** [HGX](/productcode/HGX.md)
- **Decision Date:** Aug 11, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5160
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

The Lucina-Melodi Powered Breast Pump is intended to express and collect milk from the breasts of lactating women.

## Device Story

Lucina-Melodi is a portable, battery-powered breast pump for single-user, dual-breast expression. Device utilizes a diaphragm motor to generate vacuum suction (50-250 mmHg) controlled by an internal PCB and software. Features a color LCD screen and 6-button interface for user-selected settings. Operates in a 2-phase cycle: Phase 1 (higher speed) and Phase 2 (lower speed). Powered by rechargeable Li-Ion battery or AC adapter. Used by lactating women in home or clinical settings to collect milk into bottles via breast shields and tubing. Output is expressed milk; device provides consistent vacuum pressure and cycle rates to facilitate collection. Benefits include portability, rechargeable operation, and adjustable suction/speed settings for user comfort.

## Clinical Evidence

Bench testing only. Evaluated vacuum pressure consistency and cycles per minute (CPM) over time in both single and double pumping modes. Results demonstrated stable, consistent performance equivalent to the predicate device.

## Technological Characteristics

Powered breast pump; diaphragm motor; rechargeable Li-Ion battery; 19V, 30W power; color LCD interface; 6-button control; non-sterile; dual-pumping capability. Software-controlled 2-phase suction cycle. Dimensions: 6 x 4 x 2.36 inches; weight: 1.5 lbs.

## Regulatory Identification

A powered breast pump in an electrically powered suction device used to express milk from the breast.

## Predicate Devices

- Medela Pump In Style Breast Pump ([K950750](/device/K950750.md))
- Medela Pump In Style Advance Breast Pump ([K031614](/device/K031614.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K102516 page lots

## Traditional 510k Summary

AUG 1 1 2011

### General Information

- 1. Applicant:
Genadyne Biotechnologies Inc. 65 Watermill Lane, Great Neck, NY 11021 (t) 516.487.8787 (f) 516.487.7878 www.genadyne.com

2. Contact Person:

Mr. Chien-Ming GOH (Andrew) Vice President Genadyne Biotechnologies Inc. 65 Watermill Lane, Great Neck, NY 11021 Tel: 516-487-8787 Fax: 516-487-7878 Andrew@genadyne.com

### 3. Trade/Proprietary Name Including Model Number of Device:

Lucina-Melodi Powered Breast Pump

### 4. Common Name or Classification Name (21 CFR Part 807.87) of Device:

Powered Breast Pump (21 CFR 884.5160, Product Code HGX)

#### 5. Class in which Device has been placed:

Class II

#### 6. Reason for Premarket Notification:

Introduction of a new device.

## 7. Identification of Legally Marketed Device Which We Claim Substantial Equivalence (Predicate Device):

Medela Pump In Style Breast Pump (K950750) Medela Pump In Style Advance Breast Pump (K031614)

{1}------------------------------------------------

K102516 page 2 of 5

### 8. Description of the Device

The Lucina-Melodi Powered Breast Pump is an advanced breast pump powered by a smart software. It has a 2 phase suction cycle. It is portable and is battery powered, with a rechargeable battery build in the system. It weighs less than 1kg. It is intended for single user usage, and can be use on both breasts at the same time. The device has a build in color LCD screen, and easy to use buttons and graphic user interface.

### 9. Intended use of the Device

The Lucina-Melodi Powered Breast Pump is intended to express and collect milk from the breasts of lactating women.

### 10. Description of Device

The device can be run on battery power and also while plugged in to the AC adapter. Internally, it gets its suction from a diaphragm motor. The PCB board will control the speed and the suction of the motor to provide optimal suction or based on the settings that the user provided. The device has a smart 2 phase software running which provide 2 different speeds to the user. Phase 1 runs at a higher speed than phase 2. The device has a large color LCD screen, with 6 easy to use buttons.

{2}------------------------------------------------

K102516 page 3 of 5

# Comparative Information

# 11. Comparison to Predicate Device

# Table of Comparison to Legally Marketed Device:

| Comparative Information |                                           |                          |
|-------------------------|-------------------------------------------|--------------------------|
|                         | Predicate Device                          | New Device               |
| Company                 | Medela, Inc.                              | Genadyne Biotechnologies |
| Device Name             | Pump in Style Advance                     | Lucina-Melodi            |
| 510 (K) Number          | K031614                                   |                          |
| Technical Data          |                                           |                          |
| Vacuum Range            | 50-250 MMHG                               | 50-250 MMHG              |
| Power Requirements      | 60 Hz                                     | 19 V, 30W                |
| Battery Type            | Non-Rechargeable Ni-MH                    | Rechargeable Li-Ion      |
| Dimensions              | 11 5/8 x 7 7/8 x 11 $3\frac{1}{4}$ inches | 6 x 4 x 2.36 inches      |
| Weight                  | 9 Lbs                                     | 1.5 Lbs                  |
| Dual Pumping            | Yes                                       | Yes                      |
| Accessories             |                                           |                          |
| Filter                  | No                                        | Yes                      |
| Breast Shields          | Yes                                       | Yes                      |
| Storage Bag             | Yes                                       | Yes                      |
| Baby Bottles            | Yes                                       | Yes                      |
| Valve flanges           | Yes                                       | Yes                      |
| Tubing                  | Yes                                       | Yes                      |
| Splitter                | No                                        | Yes                      |
| Sterile                 | Non Sterile                               | Non Sterile              |

{3}------------------------------------------------

| Indications for Use |                                                                                                                                                                                                                                                           |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                     | The Pump In Style Advanced Breast pump is a powered breast pump to be used by lactating women to express and collect milk from their breasts. Intended use of the powered breast pump is to express and collect milk from the breasts of lactating women. |
| Testing             |                                                                                                                                                                                                                                                           |
|                     | NA IEC 60601-1-2                                                                                                                                                                                                                                          |
|                     | NA FCC part 15 Class B                                                                                                                                                                                                                                    |
|                     | NA EN 55011                                                                                                                                                                                                                                               |
|                     | NA IEC 61000-4-2                                                                                                                                                                                                                                          |
|                     | NA IEC 61000-4-3                                                                                                                                                                                                                                          |
|                     |                                                                                                                                                                                                                                                           |

## Discussion of Similarities and Differences Device Similarities

### Indication for use

The Lucina-Melodi Powered Breast Pump and the Medela Pump In Style Advanced Breast Pump are both intended to express and collect milk from the breasts of lactating women. The indication for use is identical.

### Technological characteristics

Functional designs are the same, where by the system services as a suction device which is connected to a collection bottles where by breasts shields are use to place onto the breasts to collect milk. All devices are sold non-sterile. Devices are for dual pumping of breasts, and have the same accessories to collect milk.

### Basic Product Function

The Lucina-Melodi and the predicate device have the same product function of generating a vacuum to provide general use suction and collection of milk into a bottle.

## Device Differences

In comparison to the predicate devices, the Lucina-Melodi has several differences which do not affect the device safety and effectiveness of the intended use. These differences are described in further detail below.

## Lucina-Melodi

The Lucina-Melodi is lighter in weight and smaller in dimension. It also have a rechargeable battery so that users could use it without it being plugged onto the wall all the time. In all other aspects, the Lucina-Melodi Powered Breast Pump and the predicate device are substantially equivalent.

{4}------------------------------------------------

K10z516 page. 5ofs

## 12. Conclusions:

I believe that the Lucina-Melodi Powered Breast Pump is substantially equivalent to the predicate device. Results from the bench test, proves that the data collected is equally as good as the predicate.

The pressure at different levels was consistent and stable, in both single and double pumping usage. In another pressure vs time test, the peak pressure was consistent throughout the whole test at several pressure levels. Lastly, in the cycles per minute vs time test, the CPM were consistent over time.

## 13. Determination of Substantial Equivalence

Bench testing were done on the Lucina-Melodi Powered Breast Pump.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three lines forming its body and wings. The symbol is enclosed within a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is written around the border of the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring. MD 20993-0002

Mr. Chien-Ming Goh Vice President Genadyne Biotechnologies Inc. 65 Watermill Lane GREAT NECK NY 11021

AUG 1 1 2011

Re: K102516

Trade/Device Name: Lucina-Melodi Powered Breast Pump Regulation Number: 21 CFR& 884.5160 Regulation Name: Powered breast pump Regulatory Class: II Product Code: HGX Dated: July 26, 2011 Received: July 27, 2011

Dear Mr. Goh :

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related

{6}------------------------------------------------

### Page 2

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Hubert Hennen MD

Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

### Indications for Use

510(k) Number (if known):

K102516

Device Name: Lucina-Melodi Powered Breast Pump

Indications For Use:

The Lucina-Melodi Powered Breast Pump is intended to express and collect milk from the breasts of lactating women.

Prescription Use (Per 21 CFR 801 Subpart D)

OR

Over-The Counter Use X (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Tomi M. Whay

on Sidn-Off ion of Reproductive, Gastro-Renal, and plogical Devices 510(k) Number KID2516

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/device/K102516](https://fda.innolitics.com/device/K102516)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
