QUADROX-I NEONATAL MICROPOROUS MEMBRANE OXYGENATOR WITH INTEGRATED ARTERIAL FILTER WITH SOFTLINE/BIOLINE COATING
K102464 · Maquet Cardiopulmonary, AG · DTZ · Mar 4, 2011 · Cardiovascular
Device Facts
Record ID
K102464
Device Name
QUADROX-I NEONATAL MICROPOROUS MEMBRANE OXYGENATOR WITH INTEGRATED ARTERIAL FILTER WITH SOFTLINE/BIOLINE COATING
Applicant
Maquet Cardiopulmonary, AG
Product Code
DTZ · Cardiovascular
Decision Date
Mar 4, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4350
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The membrane oxygenator QUADROX-i Neonatal is intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. It is designed for a blood flow rate of 0.2 - 1.5 l/min and is intended for the treatment of pediatric (neonate and infant) patients. Within the specified flow rate range, the device oxygenates the blood, removes carbon dioxide and regulates the blood temperature. The QUADROX-i Neonatal (HMO 11000) model with integrated arterial filter also filters air bubbles and particles larger than 33 um. The utilization period of this device is restricted to six hours. Responsibility for deciding whether to use an oxygenator rests solely with the attending physician.
Device Story
Blood-gas exchanger with integrated heat exchanger and optional arterial filter; designed for neonate/infant patients requiring cardiopulmonary bypass. Input: venous blood from patient; Output: oxygenated, CO2-depleted, temperature-regulated blood returned to patient. Integrated arterial filter (33 micron) removes air bubbles/particles. Used in clinical cardiac surgery settings; operated by perfusionists/physicians. Device provides gas exchange surface (0.38 m²) and heat exchange (0.07 m²). SOFTLINE or BIOLINE coatings applied to blood-contacting surfaces to improve biocompatibility. Single-use, sterile, non-pyrogenic; max 6-hour duration. Benefits include small priming volume (38-40 ml) suitable for low-blood-volume pediatric patients.
Clinical Evidence
Bench testing only. Compliance with ISO 10993-1 (biocompatibility), ISO 7199 (blood gas exchangers), and ISO 15675 (arterial line blood filters) demonstrated safety and performance. No clinical data presented.
Technological Characteristics
Microporous membrane oxygenator; integrated heat exchanger; optional 33-micron woven arterial filter. Blood-contacting surfaces feature SOFTLINE or BIOLINE coatings. Priming volume 38-40 ml. Gas exchange surface 0.38 m²; heat exchange surface 0.07 m². Sterile, single-use. Complies with ISO 7199 and ISO 15675.
Indications for Use
Indicated for pediatric (neonate and infant) patients undergoing cardiopulmonary bypass during cardiac surgery requiring extracorporeal circulation at blood flow rates of 0.2-1.5 l/min.
Regulatory Classification
Identification
A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”
QUADROX-i Adult microporous membrane oxygenator with and without integrated arterial filter with SOFTLINE coating (K082117)
QUADROX-i Adult Microporous Membrane Oxygenator with and without Arterial Filter with Bioline Coating (K090511)
Reference Devices
QUADROX-iD Pedatric Diffusion Membrane Oxyenator with BIOLINE Coating (K100278)
Medos Hilte Infant Oxygenator, model 1000, 800 LT (K090449)
Submission Summary (Full Text)
{0}------------------------------------------------
#### MAR - 4 2011
## 510 (K) Summary [as required by 21 CFR 807.92(c) ]
March 3, 2011
K102464
GETINGE GR
Submitter:
Maguet Cardiopulmonary AG Hechinger Strasse 38 72145 Hirrlingen Germany
Contact Person:
Katrin Schwenkglenks Phone: +49 7478 921-151 +49 7478 921-400 Fax: E-mail: katrin schwenkglenks@maquet-cp.com
Date Prepared:
Device Trade Name:
QUADROX-i Neonatal Microporous Membrane Oxygenator with and without integrated Arterial Filter with SOFTLINE Coating and with BIOLINE Coating
Common/Usual name:
Oxygenator with integrated heat exchanger and optional integrated arterial filter
Classification names:
Oxygenator, cardiopulmonary bypass Heat Exchanger, cardiopulmonary bypass Filter, blood, cardiopulmonary bypass, arterial line
#### Predicate Devices:
Capiox RX 05 Baby Hollow Fiber Oxygenator, Terumo Cardiovascular Systems Corp., cleared under K022115,
Capiox Arterial Filter, CX*AF02, Terumo Cardiovascular Systems Corp., cleared under K943917,
QUADROX-i Adult microporous membrane oxygenator with and without integrated arterial filter with SOFTLINE coating, MAQUET Cardiopulmonary AG, cleared under K082-117,
QUADROX-i Adult Microporous Membrane Oxygenator with and without Arterial Filter with Bioline Coating, MAQUET Cardiopulmonary AG, K090511.
{1}------------------------------------------------
QUADROX-iD Pedatric Diffusion Membrane Oxyenator with BIOLINE Coating, MAQUET Cardiopulmonary AG, K100278
Medos Hilte Infant Oxygenator, model 1000, 800 LT, Gish Biomedical Inc. K 090449
#### Device Description:
The QUADROX-i Neonatal Oxygenator is a blood-gas exchanger with integrated heat exchanger and optional integrated arterial filter, intended for the treatment of neonate and infant patients. The Oxygenator is designed for a blood flow range from 0.2 to 1.5 Ilmin. Since these patients have a very small blood volume of their own, the small priming volume of the oxygenator is an important parameter. The priming volume is 38 ml for the version without integrated arterial filter. The version with integrated arterial filter has a priming volume of 40 ml.
The effective gas exchange surface is 0.38 m²; the one for the heat exchanger is 0.07 m². The integrated arterial filter is composed of a planar area from a woven 33 micron filter. The filter surface is 20 cm2.
The blood contacting surfaces are coated with SOFTLINE coating or with BIOLINE Coating.
The QUADROX-i Neonatal oxygenator is a sterile and non-pyrogenic device, for single use only and is not to be re-sterilized by the user.
#### Indications for Use:
The membrane oxygenator QUADROX-i Neonatal is intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. It is designed for a blood flow rate of 0.2 - 1.5 l/min and is intended for the treatment of pediatric (neonate and infant) patients. Within the specified flow rate range, the device oxygenates the blood, removes carbon dioxide and requiates the blood temperature.
The QUADROX-i Neonatal (HMO 11000) model with integrated arterial filter also filters air bubbles and particles larger than 33 um.
The utilization period of this device is restricted to six hours.
Responsibility for deciding whether to use an oxygenator rests solely with the attending physician.
{2}------------------------------------------------
# MAQUET GETINGE GROUP
#### Statement of Technical Comparison:
The QUADROX-i Neonatal Microporous Membrane Oxygenator ાં ક comparable to the QUADROX-i Adult Microporous Membrane Oxygenator regarding the design principles, biocompatibility and sterility process. Both products come with and without an Integrated Arterial Filter with Softline Coating. The Softline Coating and the BIOLINE Coating are the same coatings contained in the QUADROX-i Adult Oxygenator.
#### Non-clinical Testing:
The QUADROX-i Neonatal Microporous Membrane Oxygenator with and without Integrated Arterial Filter with Softline Coating and with BIOLINE Coating have been tested or evaluated for compliance to ISO 10993-1 Biologic Evaluation of Medical Devices, ISO 7199 "Cardiovascular implants and artificial organs - blood gas exchangers (oxygenators) as well as the requirements of ISO 15675 "Cardiovascular implants and artificial organs -Cardiopulmonary Bypass - Arterial line blood filters". The products met these requirements.
### Determination of Substantial Equivalence
Testing and evaluation on safety and effectiveness was conducted to demonstrate that the QUADROX-1 Neonatal Microporous Membrane Oxygenator with and without Integrated Arterial Filter with SOFTLINE Coating and BIOLINE Coating is substantially equivalent to the Capiox RX 05 Baby Hollow Fiber Oxygenator from Terumo as well as to the Capiox Arterial Filter. CX*AF02 from Terumo.
The following areas have been tested or evaluated:
- Indications for Use
- Integrity
- Performance
- Biocompatibility
- Sterility
### Conclusion
The data given demonstrate that the QUADROX-i Neonatal Microporous Membrane Oxygenator with and without Integrated Arterial Filter with Softline Coating and with BIOLINE Coating is substantially equivalent to the named predicate devices which currently hold market clearance.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is a stylized image of an eagle or bird-like figure with three curved lines representing its body and wings. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Maquet Cardiopulmonary AG c/o Ms. Katrin Schwenkglenks Hechinger Strasse 38 D-72145 Hirrlingen, Germany
MAR - 4 2011
Re: K102464
Trade/Device Name: QUADROX-i Neonatal Microporous Membrane Oxygenator with SOFTLINE COATING (HMO 10000), QUADROX-i Neonatal Microporous Membrane Oxygenator with integrated Arterial Filter with SOFTLINE COATING (HMO 11000), QUADROX-i Neonatal Microporous Membrane Oxygenator with BIOLINE COATING (BE-HMO 10000, BEQ-HMO 10000), and QUADROX-i Neonatal Microporous Membrane Oxygenator with integrated Arterial Filter with BIOLINE COATING (BEQ-HMO 11000, BE-HMO 11000) Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary bypass oxygenator Regulatory Class: II
Regulatory Class. II
Product Code: DTT, DTR
Product Code: DTZ, DTR, DTM
Dated: February 21, 2011
Received: February 24, 2011
Dear Ms. Schwenkglenks:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 = Ms. Katrin Schwenkglenks
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
ina R. Vahner
E
DE
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): __ K | 0246
## Device Name: QUADROX-i Neonatal Microporous Membrane Oxygenator with and without integrated Arterial Filter
Indications for Use:
The membrane oxygenator QUADROX-i Neonatal is intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. It is designed for a blood flow rate of 0.2 - 1.5 //min and is intended for the treatment of pediatric (neonate and infant) patients. Within the specified flow rate range, the device oxygenates the blood, removes carbon dioxide and regulates the blood temperature.
The QUADROX-i Neonatal (HMO 11000) model with integrated arterial filter also filters air bubbles and particles larger than 33 um.
The utilization period of this device is restricted to six hours.
Responsibility for deciding whether to use an oxygenator rests solely with the attending physician.
Prescription Use × AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
R.v.lumer
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number_K102 46 4
Page 1 of 1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.