I-NEB AAD SYSTEM WITH TIM; I-NEB INSIGHT AAD SYSTEM
K102454 · Respironics Respiratory Drug Delivery (Uk) , Ltd. · CAF · Oct 25, 2011 · Anesthesiology
Device Facts
Record ID
K102454
Device Name
I-NEB AAD SYSTEM WITH TIM; I-NEB INSIGHT AAD SYSTEM
Applicant
Respironics Respiratory Drug Delivery (Uk) , Ltd.
Product Code
CAF · Anesthesiology
Decision Date
Oct 25, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The I-neb AAD system with TIM is a nebulizer system designed to aerosolize liquid medication for inhalation by the patient in the home care, nursing home, sub-acute institution, or hospital environment. The I-neb Insight AAD System is monitoring software that provides feedback to the patient recording treatment events, including treatment times and compliance data which is also available to the clinician. It is an accessory to an accessory for use with the I-neb AAD system. The I-neb AAD system with TIM is intended for patients 3 years and older who can coordinate breathing.
Device Story
I-neb AAD System with TIM is a vibrating mesh nebulizer for aerosolizing liquid medication. It utilizes breath-triggered technology, sensing patient inhalation via pressure signals to synchronize drug delivery. The system includes a restrictive flap valve in the mouthpiece to guide controlled breathing, with tactile vibration feedback to indicate target inhalation times. The I-neb Insight AAD System is an accessory software component that records treatment events, times, and compliance data, providing feedback to patients and clinicians. Used in home, nursing home, or hospital environments, the device helps patients manage medication delivery. Clinicians review recorded compliance data to monitor treatment adherence. The system uses RF discs to provide pre-set parameters for nebulizer performance, ensuring consistent dosing. The device benefits patients by improving delivery efficiency through breath-actuation and providing objective compliance tracking.
Clinical Evidence
No clinical data. Substantial equivalence supported by comparative bench testing. Particle characterization (MMAD, GSD, respirable fraction) performed using 3 drugs at 15 lpm and 30 lpm. Comparative dosing studies measured gravimetric dose, filter dose, and treatment time, demonstrating equivalence to predicates Omron U22 and Salter 8900.
Technological Characteristics
Vibrating mesh nebulizer; breath-actuated via pressure signal; restrictive flap valve for flow control; tactile vibration feedback; RF disc for parameter setting; software-driven monitoring; materials identical to K042991; intended for home/clinical use.
Indications for Use
Indicated for patients >3 years old capable of coordinating breathing, requiring nebulized liquid medication (excluding pentamidine) in home, nursing home, sub-acute, or hospital settings.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
K102454
## OCT 2 5 2011
### 510(k) Summary Page 1 of 5 26-Sep-1 l
| Respironics Respiratory Drug Delivery (UK) Limited | |
|----------------------------------------------------|-----------------------------------------------------|
| Chichester Business Park | |
| City Fields Way, Tangmere | Tel - +44 800 1300 840 |
| Chichester PO20 2FT UK | Fax - +44 800 1300 841 |
| Official Contact: | Clare Fripp, Quality and Regulatory Affairs Manager |
| Proprietary or Trade Name: | I-neb AAD Systems with TIM and |
| | I-neb Insight AAD System |
| Common/Usual Name: | Nebulizer (direct patient interface) |
| Classification Name: | Nebulizer |
| | CAF - 868.5630 |
| Predicate Devices: | K062263 – Omron U-22 |
| | K870027 - Salter 8900 nebulzier |
| | K072019 - Activaero - AKITA2 APIXNEB |
| | K042991 - Profile - I-neb AAD |
| | K052491 - Profile - I-neb Insight |
### Device Description:
The I-neb AAD System is an ultrasonic (vibrating mesh) nebulizer system designed to aerosolize liquid medication (except pentamidine) for inhalation by the patient in the home care, nursing home, sub-acute institution, or hospital environment.
The I-neb Insight AAD System is an accessory for use with the I-neb AAD system and controls monitoring software that provides feedback to the patient recording treatment events, including treatment times and compliance data which is also available to the clinician.
The I-neb AAD system with TIM incorporates a modification of the predicate I-neb AAD system with (tidal breathing mode (TBM)), K042991. While the I-neb Insight incorporates updated software to accommodate the added TIM feature, it is a modification of the predicate I-neb Insight AAD System, K052941.
### Indications for Use:
The I-neb AAD system with TIM is a nebulizer system designed to aerosolize liquid medication for inhalation by the patient in the home care, nursing home, sub-acute institution, or hospital environment.
The I-neb Insight AAD System is monitoring software that provides feedback to the patient recording treatment events, including treatment times and compliance data which is also available to the clinician. It is an accessory to an accessory for use with the I-neb AAD system.
### Patient Population:
The I-neb AAD system with TIM is intended for patients > 3 years and older who can coordinate breathing.
{1}------------------------------------------------
### 510(k) Summary Page 2 of 5 26-Sep-1 l
### Environment of Use:
Home care, nursing home, sub-acute institution, or hospital environment.
### Contraindications:
None
### Performance Testing:
I erformed comparative bench testing to demonstrate that the I-neb AAD System with TIM is we performed comparist remotion of System (K042991) in TBM mode, and Omron U22 Courtain'to the production in the comparative testing demonstrates that the proposed device is substantially equivalent to the predicate devices.
Previous testing of the I-neb AAD system (K042991) included IEC 60601 and electrical safety, I revious testing of the Photo in the vironmental testing. As the proposed device is a modification Dires, Entil and Moonanteer with TBM (K042991) and 1-neb AAD with Insight (K052491) and no or the predicate i noo in he basic hardware was performed, thus repeating these tests was not required to demonstrate safety.
Particle characterization via Cascade Impactor was performed with 3 drugs. This testing was performed at flow rates of 15 lpm and 30 lpm to simulate the intended patient population.
Comparative dosing was performed. This testing demonstrated that the TBM and the TIM modes Comparative dosing was perfectived. This and Salter 8900 (K870027) were equivalent.
### Substantial Equivalence:
The I-neb AAD System with TIM is viewed as substantially equivalent to the predicate devices because:
### Indications -
As a general purpose nebulizer, identical to predicate – Omron U-22 (K062263), Salter 8900 (K870027) and Activaero AKITA2 APIXNEB (K072019)
### Technology -
Identical vibrating mesh nebulizer technology to predicates – K042991 – I-neb AAD and K062263 - Omron U-22
Identical breath triggered nebulization technology to predicate - K042991 - I-neb AAD System and K072019 Activaero AKITA2 APIXNEB
### Materials -
The materials in the gas and fluid pathway are identical to predicate device - K042991 -- I-neb AAD System.
{2}------------------------------------------------
### 510(k) Summary Page 3 of 5 26-Sep-11
### Environment of Use -
Environment of Ose –
Identical to predicate – K062263 – Omron U-22, K072019 - Activaero AKITA² APIXNEB and K042991 - I-neb AAD.
### Patient Population -
Faction I o predicates -K062263 - Omron U-22, K042991 - I-neb AAD System and K072019 Activaero AKITA2 APIXNEB.
The I-neb Insight AAD System is viewed as substantially equivalent to the predicate devices because:
Indications -Identical to predicate - K052491 - I-neb Insight AAD System
### Technology -
Identical technology to predicate - K052491 - I-neb Insight AAD system
### Summary of Specific Particle Characterization and Dose testing
- Comparison of particle characterization testing included the evaluation of .
- MMAD, GSD, Respirable Fraction (%) the predicates and the proposed device o were found to be substantially equivalent.
- Comparative Dose for the I-neb AAD system in TBM vs. TIM mode and to the . predicates Omron U22 (K062263) and Salter 8900 (K870027) was performed.
- Parameters measured and compared included gravimetric dose, Filter dose and O Treatment time
- Results the predicates and the proposed device were substantially equivalent. O
{3}------------------------------------------------
| Attribute | Omron<br>U22 nebulizer<br>K062263 | Salter<br>8900<br>K870027 | Activaero<br>AKITA2 APIXNEB<br>K072019 | Profile<br>I-neb AAD<br>K042991 | Proposed<br>I-neb AAD with TIM |
|---------------------------------------------------------------------------------------|-----------------------------------|---------------------------|---------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | General purpose use | General purpose use | General purpose use | Specific to a drug | General Purpose use |
| Patient population | Adult and pediatric | Adult and Pediatric | 3 and older who can<br>coordinate breathing | 2 and older who can<br>coordinate breathing | 3 and older who can coordinate<br>breathing |
| Nebulizer technology | Vibrating Mesh | Jet / venturi | Vibrating Mesh | Vibrating Mesh | Vibrating Mesh |
| Accessory mouthpiece | Yes | Yes | N/A | Yes | Yes |
| Software driven | Yes | No | Yes | Yes | Yes |
| Mode of Operation | Continuous | Continuous | Breathe activated | Breathe activated | Breathe activated |
| Synchronized delivery of<br>nebulized drug | No | No | Yes | Yes<br>Inhalation only | Yes<br>Inhalation only |
| Drug delivery on demand | No | No | Yes | Yes | Yes |
| Drug delivery triggered by | N/A | N/A | Patient inhalation<br>Pressure signal | Patient inhalation<br>Pressure signal | Patient inhalation<br>Pressure signal |
| Drug delivery pulse range<br>(seconds) | N/A | N/A | 1 to 7 sec | 0.5 to 5 sec | Up to 7 sec |
| Delivers medication | Continuously | Continuously | Up to the last 1 second of<br>the inhalation cycle | Between 50 to 80% of<br>the inhalation cycle | Up to the last 1 second of the<br>inhalation cycle |
| Method of guiding patient for<br>Controlled inhalation | None | None | Auxiliary flow is<br>provided from a<br>compressor set at<br>15 Lpm | Use of a flap valve in the<br>mouthpiece in which the<br>patient entrains room air<br>which mixes with the<br>nebulized medication.<br>Approximate flow rate is<br>not controlled in TBM<br>mode | Use of a restrictive flap valve in the<br>mouthpiece to control breathing.<br>Vibration felt by the patient to<br>indicate target time. The patient<br>entrained room air which mixes<br>with the nebulized medication.<br>Approximate flow rate is 15 Lpm<br>for TIM mode |
| Method of providing pre-set<br>parameters for nebulizer<br>performance | None | None | Yes<br>Smart Card | Yes<br>RF disc | Yes<br>RF disc |
| Accessory for monitoring and<br>recording patient data for review<br>by the clinician | N/A | N/A | N/A | Yes | Yes |
1.
{4}------------------------------------------------
## 510(k) Summary
# I-neb Insight with TIM
ーーーー -
।
| Results | Profile | Proposed |
|-------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | I-neb Insight<br>(K052491) | I-neb AAD with TIM |
| Indications for Use<br>(all have except pentamidine) | Will be used with patients for whom<br>doctors have prescribed medication<br>for nebulization | Will be used with patients for whom<br>doctors have prescribed medication for<br>nebulization |
| Accessory to I-neb AAD system with<br>the indications of monitoring and<br>providing patient feedback | The I-neb Insight is monitoring<br>software that provides feedback to the<br>patient recording treatment events,<br>including treatment times and<br>compliance data which is also<br>available to the clinician | The I-neb Insight is monitoring<br>software that provides feedback to the<br>patient recording treatment events,<br>including treatment times and<br>compliance data which is also available<br>to the clinician |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of two parts: the department's seal and the department's symbol. The seal is a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. The symbol is a stylized representation of a human figure, with three wavy lines representing the body and head.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Respironics Respiratory Drug Delivery (UK) Limited C/O Mr. Paul Dryden President ProMedic, Incorporated 24301 Woodsage Drive · Bonita Springs, Florida 34134
Re: K102454
Trade/Device Name: I-neb AAD Systems with TIM and I-neb Insight AAD System Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: October 6, 2011
Received: October 7, 2011
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
OCT 2 5 2011
{6}------------------------------------------------
### Page 2 – Mr. Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### Indications for Use Statement
### Page 1 of 1
510(k) Number:
K102454
Device Name:
I-neb AAD system with TIM and I-neb Insight AAD system
Indications for Use:
The I-neb AAD system with TIM is a nebulizer system designed to aerosolize liquid medication for inhalation by the patient in the home care, nursing home, sub-acute institution, or hospital environment.
The I-neb Insight AAD System is monitoring software that provides feedback to the patient recording treatment events, including treatment times and compliance data which is also available to the clinician. It is an accessory to an accessory for use with the I-neb AAD system.
The I-neb AAD system with TIM is intended for patients 3 years and older who can coordinate breathing.
Prescription Use XX (Part 21 CFR 801 Subpart D) or
Over-the-counter use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schultheis
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: 4102454
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.