← Product Code [HEB](/productcode/HEB) · K102445

# ROSTAM SCENTED AND UNSCENTED BIODEGRADABLE PLASTIC APPLICATOR TAMPONS (K102445)

_Rostam , Ltd. · HEB · Nov 10, 2010 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K102445

## Device Facts

- **Applicant:** Rostam , Ltd.
- **Product Code:** [HEB](/productcode/HEB.md)
- **Decision Date:** Nov 10, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5470
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

The Rostan Scented and Unscented Biodegradable Plastic Applicator Tampons are inserted into the vagina and used to absorb menstrual or other vaginal discharge.

## Device Story

Rostam menstrual tampons consist of absorbent rayon/cotton core, polymeric overwrap, cotton/polyester thread, and cotton cord, housed in a biodegradable plastic applicator. Device is used by patients for self-administration to absorb menstrual or vaginal fluid. Available in four absorbencies: light, regular, super, and super plus. The primary innovation is the biodegradable material of the applicator. Device functions as a standard menstrual tampon; no electronic or mechanical processing involved.

## Clinical Evidence

Bench testing only. Biocompatibility testing included cytotoxicity, rabbit vaginal irritation assays (saline and cottonseed oil), and rabbit skin irritation assays (saline and cottonseed oil). Absorbency was verified using standard Syngina testing. No clinical trials were conducted.

## Technological Characteristics

Absorbent core: rayon and/or cotton; overwrap: polymeric; thread: cotton/polyester; cord: cotton. Applicator: biodegradable plastic. Absorbency levels: light, regular, super, super plus. Non-electronic, non-software device.

## Regulatory Identification

An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.

## Predicate Devices

- Rostam Scented and unscented Tampons ([K042773](/device/K042773.md))
- Rostam Scented and unscented Tampons ([K062001](/device/K062001.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## KI02445

Robert J. Staab Ph.D. President

(860) 434 5872 (860) 434 5892 ph/fx NOV 1 0 2010

# RTA

## Regulatory and Technical Associates 30 Neck Road Old Lyme CT 06371 Official Correspondent for Rostam

Prepared 11/10/10

### VII

#### 510K SUMMAKY

Device Name: Rostam Scented and unscented plastic biodegradable applicator Tampois (Various Trade Tampons Sold Under Private Labels As Plastic Applicapors)

Legally marketed device: These Tampors are substantially cquivalent to logally marketed Roslam Scented and unscented Tampons with plastic applicators. They are Class II devices, 21 CFR 884.5460 and 21 CFR 884.5470. The prodicate tampons have been marketed under K 042773 and K062001.

Regulation Name: (Unscented menstrual tampon & Scented menstrual tampon), regulation number (21 CFR 884.5470 & 21 CFR 884.5460), and procode (HEB & HIL)

Device description: Rostam plastic scented and unscented biodegradable applicator tampons are being submitted to obtain clearance for a new material applicator. They are menstrual tampons used to absorb menstrual fluid. These Tanıpons will be provided with 4 absorbencies, light, regular, super and super plus.

These Tampons are made from rayon and/or cotton (as the absorbent material) polymeric overwrap, cotton/polyester sewing thread and cotton cord.

#### Intended Use:

The materials used in these tampons are similar to those used in the predicate device, a that with the marked tampon and the Intended Use: These tampous are menstrual tampons prodicate designed and used to absorb mensitual fluid is the same as in the predicate dovice a legally marketed tampon.

## Technological Characteristics.

The subject and predicate device have different technological characteristics, as a result of the new material upplicator. The new applicators are biodegradable.

#### PERFORMANCE TESTING

K

{1}------------------------------------------------

Biocompatibility testing and sufety evaluations of tampon components were historically carried out. The results of these tests demonstrate that these Tamports we coulisabled in tenns of safety and effectiveness to legally marketed tampons. Stardard Syngina testing confirmed the absorbency of these Tampons. In addition to the review of existing toxicological data in the public literature, the following tests were conducted and are relevant to the safety of Rostam plastic biodegradable applicator tampons.

Cytotoxicity, rabbit vaginal irritation assays with saline and cottonseed oil, rabbit skin irritation ussays with saline and cottonseed oil were carried out to substantiate artis the biodegradable applicator.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its wings, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white and appears to be a scanned image.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Rostam Ltd. c/o Robert J. Staab, Ph.D. Official Correspondent RTA Regulatory and Technical Associates, Inc. 30 Neck Road OLD LYME CT 06371

NOV 1 0 .2010

Re: K102445

> Trade Name: Rostam Ltd Scented and Unscented Biodegradable Plastic Applicator Tampons

Regulation Number: 21 CFR 8884.5470 Regulation Name: Unscented menstrual tampon Regulatory Class: II Product Code: HEB Dated: August 23, 2010 Received: August 26, 2010

Dear Dr. Staab:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA) . You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing. of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related

{3}------------------------------------------------

#### Page 2

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21-CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please . go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Hubert Lemon MD

Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive. Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health.

#### Enclosure

{4}------------------------------------------------

# Indications for Use Form

Indications for Use

S10(k) Number (if known): _K102445

Device Name:Rostain Ltd Scented and Unscented Biodegradable Plastic Applicator Tampons

Indications for Use:

The Rostan Scented and Unscented Biodegradable Plastic Applicator Tampons are inserted into the vagina and used to absorb menstrual or other vaginal discharge.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use X (21 CFR 801 Subpart C)

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

ision Sian-C Gastm

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/device/K102445](https://fda.innolitics.com/device/K102445)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).