← Product Code [NHA](/productcode/NHA) · K102382

# SFI-BAR IMPLANT ADAPTER STRAUMANN, SFI-BAR IMPLANT ADAPTER FOR NEOSS IMPLANT (K102382)

_Cendres & Metaux SA · NHA · Dec 3, 2010 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K102382

## Device Facts

- **Applicant:** Cendres & Metaux SA
- **Product Code:** [NHA](/productcode/NHA.md)
- **Decision Date:** Dec 3, 2010
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 872.3630
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

The SFI-Bar® is intended to be used with the implant manufacturer's (Table 1) implant to provide support for fixation of overdentures.

## Device Story

SFI-Bar® is a dental abutment system providing connection between endosseous dental implants and removable overdentures. Device consists of an implant adapter (abutment) and a stress-free bar. Adapter is screwed into compatible dental implants (Thommen, Neoss, Straumann); bar provides retentive system for overdenture fixation. Used in clinical settings by dental professionals. Functions to stabilize overdentures, distribute shear forces, and compensate for resilience. Benefits include secure denture fixation and primary splinting of implants. Non-sterile, supplied in dental blisters.

## Clinical Evidence

Bench testing only. No clinical data. Testing included torque tests, application testing, and functional testing to evaluate performance characteristics across compatible implant platforms. Acceptance criteria were met, demonstrating performance equivalent to the predicate device.

## Technological Characteristics

Material: Wrought Titanium-6Aluminium-4Vanadium ELI Alloy (ASTM F 136, UNS R 56401). Design: Machined, screw-retained abutment and bar system. Connectivity: Mechanical interface with specific dental implant platforms. Manufacturing: CNC machining with validated cleaning process. Non-sterile.

## Regulatory Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

## Predicate Devices

- SFI-Bar® ([K083876](/device/K083876.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a logo with the text "CENDRES + MÉTAUX". To the left of the text is a stylized letter "C". The text is stacked, with "CENDRES +" on top and "MÉTAUX" on the bottom. The logo is black and white.

K102382

21.10.2010

# DEC - 3 2010

# 510(k) Summary

#### Applicant's Name and Address

Submitter:

Contact Person:

October 21, 2010

Tanja Bongni

Phone: Fax:

Cendres+Métaux SA

Rue de Boujean 122

2501 Biel/Bienne, Switzerland

Consultant Regulatory Affairs

+41 58 360 20 00

+41 58 360 20 10

Date of Submission:

## Name of the Device

Trade Name: Common Name: Classification Name: Regulation Number:

SFI-Bar® Abutment, Dental, Endosseous implants Endosseous Dental Implant Abutment 21 CFR 872.3630

Legally Marketed Device to which Equivalence is Claimed (Predicate Device) K083876 Predicate Device(s):

Description of the Device

Device Description:

The SFI-Bar® provides the connection between compatible dental implant systems for the fixation of removable overdentures. The SFI-Bar® consists of an implant adapter (abutment) and a stress-free bar for the fixation of removable overdentures. The implant adapter is screwed into the dental implant.

The implant adapter (abutment) fit the Thommen SPI® Element Platform Ø 4.0 mm / the Neoss ProActive Implant Ø 3.5 / 4.0 / 4.5 / 5.0 / 5.5 mm and the Straumann dental implants / ITI Dental Implant System® Standard Ø 4.1 and Ø 4.8 mm / Standard Plus Ø 4.1 mm and Ø 4.8 mm / Tapered Effect Ø 4.1 and Ø 4.8 mm and Regular Neck (RN) Ø 4.8 mm.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a logo with the text "CENDRES+" on the top line and "MÉTAUX" on the bottom line. To the left of the text is a stylized letter "M". The text and the letter are all in black. The logo appears to be for a company that deals with ashes and metals.

### Intended Use of the Device:

The SFI-Bar® is intended to be used with the implant manufacturer's . (Table 1) implant to provide support for fixation of overdentures.

Table 1 Compatible Commercial Implant Manufacturers

| Implant Company                       | Implant System                                                                                                                                                                                                                                                               | Implant Platform Diameter                                                                                                                                                                                                                                                    | Attribute             | Candidates                                                                                     | Predicate Device                                                                                                              |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Thommen Medical                       | SPI® Element Platform                                                                                                                                                                                                                                                        | 4.0 mm                                                                                                                                                                                                                                                                       |                       | SFI-Bar® Implant adapter<br>Straumann®<br>and<br>SFI-Bar® Implant adapter for<br>Neoss Implant | SFI-Bar® Implant adapter SPI®<br>Element PF Ø 4.0<br><br>Already cleared in combination<br>with the SFI-Bar®<br><br>(K083876) |
| Neoss                                 | Neoss ProActive Implant                                                                                                                                                                                                                                                      | 3.5 / 4.0 / 4.5 / 5.0 / 5.5 mm                                                                                                                                                                                                                                               | Design / construction | Machined,<br>screw-retained                                                                    | Machined,<br>screw-retained                                                                                                   |
| Institut Straumann                    | ITI Dental Implant System®                                                                                                                                                                                                                                                   | Standard 4.1 and 4.8 mm /<br>Standard Plus 4.1 and 4.8 mm /<br>Tapered Effect 4.1 and 4.8 mm /<br>Regular Neck (RN) 4.8 mm                                                                                                                                                   | Anatomical Site       | Oral Cavity                                                                                    | Oral Cavity                                                                                                                   |
| Platform compatibility                | ITI Dental Implant System®:<br>Standard Ø 4.1 mm and Ø 4.8 mm/<br>Standard Plus Ø 4.1 mm and Ø 4.8 mm /<br>Tapered Effect Ø 4.1 mm and Ø 4.8 mm /<br>Regular Neck (RN) Ø 4.8 mm<br><br>Neoss ProActive Implant:<br>Ø 3.5 / 4.0 / 4.5 / 5.0 / 5.5 mm                          | Thommen Implant System:<br>SPI® Element Platform Ø 4.0 mm                                                                                                                                                                                                                    |                       |                                                                                                |                                                                                                                               |
| Device Material                       | Wrought Titanium-6Aluminium-<br>4Vanadium ELI Alloy for<br>Surgical Implant applications                                                                                                                                                                                     | Wrought Titanium-6Aluminium-<br>4Vanadium ELI Alloy for<br>Surgical Implant applications                                                                                                                                                                                     |                       |                                                                                                |                                                                                                                               |
| Indications for Use                   | The SFI-Bar® is intended to be<br>used with the implant<br>manufacturer's (Table 1)<br>implant to provide support for<br>fixation of overdentures:                                                                                                                           | The SFI-Bar® is intended to be<br>used with the SPI Element<br>Platform 4.0 mm implant to<br>provide support for fixation of<br>overdentures.                                                                                                                                |                       |                                                                                                |                                                                                                                               |
| Operating principle /<br>Basic Design | Impression taking: Optional,<br>preassembled (plug-in<br>connection).<br><br>Abutment implant connection:<br>Screw fixation.<br><br>Connecting principle to<br>overdenture: Retentive system.<br><br>Bar fixation on implant:<br>Screwed.<br><br>Function: Stabilization and | Impression taking: Optional,<br>preassembled (plug-in<br>connection).<br><br>Abutment implant connection:<br>Screw fixation.<br><br>Connecting principle to<br>overdenture: Retentive system.<br><br>Bar fixation on implant:<br>Screwed.<br><br>Function: Stabilization and |                       |                                                                                                |                                                                                                                               |

### Summary Technological Characteristics:

Comparison /Compatibility Substantially Equivalence: The proposed implant adapters are substantially equivalent to the currently marketed predicate devices. The intended use, basic design, fundamental operating principles and manufacturing procedures are the same as the predicate device.

The material of the implant adapters conform to ASTM F 136, Wrought Titanium-6Aluminium-4Vanadium ELI Alloy for Surgical Implant applications (UNS R 56401). The parts for the SFI-Bar® System are manufactured from wires.

The proposed implant adapters are substantially equivalent to the currently marketed predicate devices. The intended use, basic design, fundamental operating principles and manufacturing procedures are the same as the predicate device.

To ensure compatibility the following process was carried out: The implant adapters are developed.and manufactured in close cooperation with the implant companies (see Table 1 , column "Implant Company").

There are Quality Agreements between Cendres + Métaux and the implant companies in place. Those agreements handle among other things the Design Control, Change Control, Complaint Handling and Post Market Surveillance.

Table 2 summarizes the substantial equivalence comparison to the predicate device:

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows a logo with the letters 'CM' on the left and the words 'CENDRES' and 'MÉTAUX' stacked on the right. The 'C' and 'M' are large and bold, with the 'M' slightly overlapping the 'C'. The word 'CENDRES' is above 'MÉTAUX', and there is a plus sign to the right of 'CENDRES'. The text is in a simple, sans-serif font.

・・ :

.

....

.

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

. . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

-

.

Table 2 Substantial Equivalence Comparison to Predicate Devices

:

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a logo for a company called "CENDRES + MÉTAUX". The logo consists of a stylized letter "C" on the left, followed by the company name in a sans-serif font. The word "CENDRES" is on top of the word "MÉTAUX", and there is a plus sign after the word "CENDRES".

## Special 510(k) - SFI-Bar® 510(k) Summary

| Attribute                             | Candidates                                                                                                         | Predicate Device                                                                                                   |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
|                                       | primary splinting of implants.                                                                                     | primary splinting of implants.                                                                                     |
|                                       | Countering forces that would<br>dislodge the denture,<br>distribution of shear forces,<br>resilience compensation. | Countering forces that would<br>dislodge the denture,<br>distribution of shear forces,<br>resilience compensation. |
|                                       | Cleaning procedures for patient:<br>Common procedure for oral<br>hygiene.                                          | Cleaning procedures for patient:<br>Common procedure for oral<br>hygiene.                                          |
|                                       | Patient handling:<br>Common cleaning and insertion                                                                 | Patient handling:<br>Common cleaning and insertion                                                                 |
|                                       | of denture.                                                                                                        | of denture.                                                                                                        |
| Shelf life                            | 95% after 10 years                                                                                                 | 95% after 10 years                                                                                                 |
| Packaging, materials and<br>processes | Produced on process orientated<br>CNC machines. The last step is                                                   | Produced on process orientated<br>CNC machines. The last step is                                                   |
|                                       | a validated cleaning process<br>(same processes).                                                                  | a validated cleaning process<br>(same processes).                                                                  |
|                                       | Packaging: Dentalblister,<br>non-sterile.                                                                          | Packaging: Dentalblister,<br>non-sterile.                                                                          |

Performance Data:

Summary of Testing to Demonstrate Safety and Effectiveness / Conclusion:

Torque tests, application testing and functional testing have been conducted to evaluate the performance characteristics of the additional SFI-Bar®. The test methods used were the same as in the predicate device. Testing has shown that the SFI-Bar® is equivalent in performance characteristics to the predicate SFI-Bar®. The acceptance criteria were met.

Non-clinical test data was used to support the substantially equivalence claim. Clinical testing was not necessary. Non-clinical testing consisted of analysis of platforms to identify worst-case test samples. Fatigue testing was not done as the basic design is the same than the predicate device. The evaluation was based on FDA guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments." Torque tests, application and functional tests have been carried out.

The summary of technological characteristics as well as the torque test, application and functional testing indicate that the device is safe and effective for its intended use and performs as well or better than the predicate devices.

{4}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol with three curved lines, which is the official symbol of the department.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Tanja Bongni Consultant Regulatory Affairs Cendres & Metaux SA Rue De Boujean 122 Biel/Bienne Switzerland 2501

DEC - 3 2010

Re: K102382

Trade/Device Name: SFI-Bar® Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: October 21, 2010 Received: November 5, 2010

Dear Mr. Bongni:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

## Page 2- Mr. Bongni

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

h for

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices

Office of Device Evaluation

Center for Devices and

Radiological Health

Enclosure

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a logo with the letters 'CM' on the left and the words 'CENDRES' and 'MÉTAUX' on the right. The 'CM' is stylized with the 'C' wrapping around the 'M'. The word 'CENDRES' is above the word 'MÉTAUX'. The logo is black and white.

·

20.09.2010

## Indications for Use Statement

510(k) Number:

K102382

Device Name:

SFI-Bar®

Indications for Use:

The SFI-Bar® is intended to be used with the implant manufacturer's (Table 1) implant to provide support for fixation of overdentures.

DEC

ന്ന
 1

2010

612387

Table 1 Compatible Commercial Implant Manufacturers

| Implant Company    | Implant System                | Implant Platform<br>Diameter                                                                                                           |
|--------------------|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Thommen Medical    | SPI® Element<br>Platform .    | 4.0 mm                                                                                                                                 |
| Neoss              | Neoss ProActive<br>Implant    | 3.5 / 4.0 / 4.5 / 5.0 /<br>5.5 mm                                                                                                      |
| Institut Straumann | ITI Dental Implant<br>System® | Standard 4.1 and 4.8<br>mm /<br>Standard Plus 4.1 and<br>4.8 mm /<br>Tapered Effect 4.1 and<br>4.8 mm /<br>Regular Neck (RN) 4.8<br>mm |

PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Üff) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

Prescription Use X OR 510(k) NumbeQver-The-Counter

---

**Source:** [https://fda.innolitics.com/device/K102382](https://fda.innolitics.com/device/K102382)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
