RADIAL JAW 4 PULMONARY STANDARD CAPACITY BIOPSY FORCEPS; RADIAL JAW 4 PULMONARY STANDARD CAPACITY BIOPSY FORCEPS WITH NE

K102336 · Boston Scientific Corporation · BWH · Jan 24, 2011 · Ear, Nose, Throat

Device Facts

Record IDK102336
Device NameRADIAL JAW 4 PULMONARY STANDARD CAPACITY BIOPSY FORCEPS; RADIAL JAW 4 PULMONARY STANDARD CAPACITY BIOPSY FORCEPS WITH NE
ApplicantBoston Scientific Corporation
Product CodeBWH · Ear, Nose, Throat
Decision DateJan 24, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.4680
Device ClassClass 2

Indications for Use

These single-use biopsy forceps are specifically designed to collect tissue endoscopically for histologic evaluation. These forceps should not be used for any purpose other than their intended function.

Device Story

Radial Jaw 4 Pulmonary Biopsy Forceps are sterile, single-use instruments for endoscopic tissue acquisition. Device consists of handle, dual pull wires, and distal jaws. User operates device by sliding spool on handle to actuate jaws; dual pull wire design enables pivoting for tangential tissue approach. Procedure involves opening jaws, pressing against biopsy site, closing jaws, and withdrawing. Used in clinical settings by physicians during bronchoscopy. Provides tissue samples for histologic evaluation, aiding in diagnosis. Available in large and standard capacity versions, with or without needle, compatible with 2.0mm or 2.8mm endoscope working channels.

Clinical Evidence

Bench testing only. Evaluations included jaw volume, opening, alignment, engagement, passability, device throw, pivoting head, and integrity under load. Biocompatibility testing performed per EN ISO 10993-1:2009 (cytotoxicity, sensitization, intracutaneous reactivity). Sterilization validated per EN ISO 11137-1:2006 and 11137-2:2006 (VDmax25).

Technological Characteristics

Single-use biopsy forceps; dual pull-wire actuation; 100cm working length; compatible with 2.0mm or 2.8mm endoscope channels. E-Beam sterilization (SAL 10^-6). Materials similar to predicate RJ3 device.

Indications for Use

Indicated for endoscopic tissue collection for histologic examination in patients undergoing bronchoscopy.

Regulatory Classification

Identification

A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K102336 SECTION 6 510(k) SUMMARY 1. Submitter Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752 Telephone: 508-683-4359 Fax: 508-683-5939 Contact: Ashley Pyle Sr. Regulatory Affairs Specialist Date Prepared: January 5, 2011 ### 2. Proposed Device: Trade Name: Radial Jaw™ 4 Pulmonary Biopsy Forceps Classification Name: Bronchoscope (flexible or rigid) and accessories Regulation Number: 874.4680 Product Code: BWH Classification: Class II #### 3. Predicate Device: Boston Scientific Radial Jaw"11 Pulmonary Biopsy Forceps (K895415) Boston Scientific Radial Jawan 3 Pulmonary Biopsy Forceps (K895415) #### 4. Proposed Device Description: The Radial Jaw"15 4 Pulmonary Biopsy Forceps (RJ4 Pulmonary) are sterile, single-use (disposable) devices. The devices are packaged in a tyvek pouch, which is packaged into a carton box. The Radial Jawns 4 Pulmonary Biopsy Forceps are available in two jaw sizes: RJ4 Pulmonary Large Capacity is compatible with a 2.8mm or larger working channel endoscope and the RJ4 Pulmonary Standard Capacity is compatible with a 2.0mm or larger working channel endoscope. The RJ4 Pulmonary Large Capacity is only available without a needle, The RJ4 Pulmonary Standard Capacity is available with or without a needle. Both the RJ4 Pulmonary Large Capacity and Standard Capacity devices have a 100cm working length. To operate the device, the user slides the spool back and forth over the handle body to open and close the jaws. The spool simultaneously actuates the dual pull wires, each of which run the length of the device and terminate with a connection to the jav. The dual pull wire design allows the javs to pivot, thus enabling tissue acquisition with a tangential approach if desired. Using the RJ4 Pulmonary Biopsy Forceps the user can obtain a tissue sample by opening the jaws, pressing the jaws against the biopsy site, closing the jaws, and pulling the jaws away from the biopsy site. ## 5. Intended Use/ Indications for Use: These single-use biopsy forceps are specifically designed to collect tissue endoscopically for histologic evaluation. These forceps should not be used for any purpose other than their intended function. JAN 2 4 2011 {1}------------------------------------------------ # 6. Technological Characteristics: The technological characteristics of the proposed Radial Jaw"™ 4 Pulmonary Biopsy Forceps are similar to the predicate Radial Jaw-Thi Pulmonary Biopsy Forceps (K895415) and the Radial Jawan 3 Pulmonary Biopsy Forceps (K895415). The proposed device has the same intended use and requires the same technique for obtaining biopsy samples as the predicate devices. The RJ4 Pulmonary Standard Capacity Biopsy Forceps working length and endoscope working channel compatibility specifications are identical to the RJ3 Pulmonary Standard Capacity Biopsy Forceps. The RJ4 Pulmonary Large Capacity Biopsy Forceps working length and endoscope working channel compatibility specifications are identical to the RJ3 Pulmonary Large Capacity Biopsy Forceps. The materials of the proposed RJ4 Pulmonary device are similar to the predicate RJ3 Pulmonary Biopsy Forceps. #### 7. Performance Data: In-vitro testing has been performed on the finished Radial Jaw™ 4 Pulmonary Biopsy Forceps device and the proposed device met the required specifications for the completed tests. The proposed RJ4 Pulmonary Biopsy Forceps were evaluated in accordance with EN ISO 10993-1: 2009. The following biocompatibility tests were performed on the biopsy forceps: Cytotoxicity, Sensitization, and Intracutaneous Reactivity. The RJ4 Pulmonary Biopsy Forceps are E-Beam sterilized. The RJ4 Pulmonary Biopsy Forceps were adopted into the sterilization validation of the master product RJ4 Large Capacity. The sterilization was performed in accordance with EN ISO 11137-1:2006 and EN ISO 11137-2:2006. The validation method was VDmax25. The sterility assurance level is 10°. The RJ4 Pulmonary Biopsy Forceps are not labeled as non-pyrogenic. The following bench tests were conducted on the RJ4 Pulmonary Biopsy Forceps: Jav Volume, Jaw Opening, Jaw Alignment and Engagement, Multi-Loop and Retroflex, Passability Force, Device Throw, Pivoting Head, Forceps Length for the Working Channel Portion of the Device, Forceps Operation after High Load, Forceps Integrity During Extreme Loading, Separation of Coil from Clevis, Separation of Sheath from Handle, Split Handle Design-Handle Separation, Split Handle Design- Spool Separation, Device Robustness, Endoscope Compatibility, Smooth Edges, and Distal End Protector. {2}------------------------------------------------ # 8. Conclusion: All biocompatibility tests conducted on the RJ4 Pulmonary Biopsy Forceps passed. Therefore, the RJ4 Pulmonary Biopsy Forceps are considered biocompatible for their intended use. The RJ4 Pulmonary Biopsy Forceps met the requirements of the sterilization validation. Therefore, the RJ4 Pulmonary Biopsy Forceps are validated for E-Beam sterilization. All device bench testing results were acceptable. The data demonstrate that the RJ4 Pulmonary Biopsy Forceps sufficiently met the design specifications and are suitable for the intended use. Boston Scientific Corporation has demonstrated that the proposed Radial Jawns 4 Pulmonary Biopsy Forceps are substantially equivalent to Boston Scientific Corporation's currently marketed Radial Jaw M Pulmonary Biopsy Forceps (K895415) and Radial Jaw™ 3 Pulmonary Biopsy Forceps (K895415), {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circle with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is an image of an eagle with its wings spread. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002 Boston Scientific Corporation c/o Ms. Ashley Pyle Sr. Regulatory Affairs Specialist 100 Boston Scientific Way Marlborough, MA 01752 JAN 2 4 201 Re: K102336 Trade/Device Name: Radial Jaw™ 4 Pulmonary Biopsy Forceps Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (flexible or rigid) and accessories Regulatory Class: Class II Product Code: BWH Dated: January 5, 2011 Received: January 6, 2011 We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Insting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Ms. Ashley Pyle Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. Rutherford Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # INDICATIONS FOR USE STATEMENT Indications for Use: 510(k) Number (if known): To Be Determined K102336 Device Name: Radial Jaw 4 Pulmonary Biopsy Forceps Indications for Use: The Radial Jaw 4 Pulmonary Biopsy Forceps are specifically designed to collect tissue endoscopically for histologic examination. These instruments should not be used for any purpose other than their intended use. Prescription Use X (Part 21 CFR 801 Part D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Susan Rudy CRNP (Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices 510(k) Number K102336 000014
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...