Retrospective clinical serum specimens from a US reference laboratory
Retrospective clinical serum specimens were used to perform method comparison studies against a predicate device (EIA) and to evaluate sensitivity/specificity against clinical gold standards (culture/India ink).
Serum specimens submitted to a US reference laboratory for cryptococcal antigen testing; Sample Size: 197; Number of Sites: 1 (US reference laboratory)
Cryptococcal Antigen EIA
Percent agreement positive and negative
Indications for Use
The CrAg Lateral Flow Assay is an immunochromatographic test system for the qualitative or semi-quantitative detection of capsular polysaccharide antigens of Cryptococcus species complex (Cryptococcus neoformans and Cryptococcus gattii) in serum.
Device Story
The CrAg Lateral Flow Assay is a dipstick sandwich immunochromatographic test for detecting cryptococcal antigen in serum. The device uses specimen wicking to transport diluted serum across a membrane containing gold-conjugated anti-cryptococcal monoclonal antibodies. If antigen is present, it forms a sandwich complex with immobilized anti-cryptococcal antibodies at the Test Line (T), creating a visible line. A control line (C) confirms proper flow using gold-conjugated goat IgG and immobilized bovine anti-goat IgG. The assay is performed in a laboratory by trained personnel using test tubes or microtiter plates. Results are interpreted visually by the healthcare provider; a positive result shows two lines, while a negative result shows one. The test aids in the diagnosis of cryptococcosis, particularly in immunocompromised populations such as AIDS patients, organ transplant recipients, and those on steroid therapy. It provides a faster, simpler alternative to traditional latex agglutination methods by eliminating the need for serum pre-treatment (e.g., pronase).
Clinical Evidence
Performance evaluated via method comparison against gold standard (culture/India Ink) and EIA. Sensitivity 100% (96.0-100% CI) and specificity 100% (97.0-100% CI) observed against culture/India Ink. Comparison with EIA showed 100% positive agreement and 93% negative agreement. Analytical sensitivity (cut-off) is 1.25 ng/ml. Cross-reactivity testing performed on various fungal and non-fungal pathologies; interference testing performed on icteric, hemolyzed, and lipemic samples. High-dose hook effect identified at concentrations >200 ug/ml.
Technological Characteristics
Dipstick sandwich immunochromatographic assay. Components: lateral flow test strips with gold-conjugated anti-cryptococcal monoclonal antibodies, immobilized anti-cryptococcal monoclonal antibodies (test line), and immobilized bovine anti-goat IgG (control line). Specimen diluent included. Manual, non-instrumented. Storage at 20-25°C. No electronic components or software algorithms.
Indications for Use
Indicated for qualitative or semi-quantitative detection of Cryptococcus neoformans and Cryptococcus gattii capsular polysaccharide antigens in serum to aid in the diagnosis of cryptococcosis. For prescription-use in laboratory settings.
Regulatory Classification
Identification
Cryptococcus neoformans serological reagents are devices that consist of antigens used in serological tests to identify antibodies to Cryptococcus neoformans in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) and are used to identify Cryptococcus neoformans directly from clinical specimens or from cultured isolates derived from clinical specimens. The identification aids in the diagnosis of cryptococcosis and provides epidemiological information on this type of disease. Cryptococcosis infections are found most often as chronic meningitis (inflammation of brain membranes) and, if not treated, are usually fatal.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.
Predicate Devices
Latex-Cryptococcus Antigen Detection System (K791382)
Submission Summary (Full Text)
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K102286
B. Purpose for Submission:
To obtain substantial equivalence for an original 510(k) for a device which detects Cryptococcal Antigen.
C. Measurand:
Capsular polysaccharide antigens of *Cryptococcus* species complex (*Cryptococcus neoformans* and *Cryptococcus gattii*)
D. Type of Test:
Qualitative and semi-quantitative dipstick sandwich lateral flow immunochromatographic assay
E. Applicant:
Immuno-Mycologics, Inc.
F. Proprietary and Established Names:
CrAg Lateral Flow Assay
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| GMD | II | 866.3165 | 83-Microbiology |
H. Intended Use:
1. Intended use:
The CrAg Lateral Flow Assay is an immunochromatographic test system for the qualitative or semi-quantitative detection of capsular polysaccharide antigens of *Cryptococcus* species complex (*Cryptococcus neoformans* and *Cryptococcus gattii*) in serum.
{2}
2. Indication for use:
The CrAg Lateral Flow Assay is an immunochromatographic test system for the qualitative or semi-quantitative detection of capsular polysaccharide antigens of *Cryptococcus* species complex (*Cryptococcus neoformans* and *Cryptococcus gattii*) in serum.
The CrAg Lateral Flow Assay is a prescription use laboratory assay which can aid in the diagnosis of cryptococcosis
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Not applicable
I. Device Description:
The CrAg Lateral Flow Assay is a dipstick sandwich immunochromatographic assay which detects cryptococcal antigen in serum. The assay consists of CrAg Lateral Flow test strips which have a gold-conjugated antibody and a gold-conjugated, anti-cryptococcal antibody deposited onto a sample membrane and anti-Crypto antibody and control-line capture antibody striped onto a membrane. Also in the kit is a specimen diluent.
J. Substantial Equivalence Information:
1. Predicate Device names
Latex-*Cryptococcus* Antigen Detection System
2. Predicate K number:
K791382
{3}
Comparison with predicate:
Table 1: Comparison Between New Device and Predicate Device
| SIMILARITIES | | |
| --- | --- | --- |
| Feature | Cryptococcal Antigen Lateral Flow Assay | Latex-Cryptococcus Antigen Detection System (K791382) |
| Intended Use | | |
| Intended Use | Test for the qualitative or semi-quantitative detection of capsular polysaccharide antigens of Cryptococcus in serum | Test for the qualitative or semi-quantitative detection of capsular polysaccharide antigens of Cryptococcus in serum and CSF |
| Indication For Use | Prescription-use laboratory assay, which can aid in the diagnosis of cryptococcosis | Prescription-use laboratory assay, which can aid in the diagnosis of cryptococcosis |
| Device Description | | |
| Sample Matrix | Serum | Serum and CSF |
| Instruments | None | None |
| Detection Antibody | Anti-cryptococcal monoclonal antibody | Anti-cryptococcal antibody |
| | | |
| DIFFERENCES | | |
| Feature | Cryptococcal Antigen Lateral Flow Assay | Latex-Cryptococcus Antigen Detection System (K791382) |
| Intended Use | | |
| Intended Use | No difference | No difference |
| Indication For Use | No difference | No differences |
| Device Description | | |
| Assay Principle | Lateral flow assay | Latex agglutination |
| Assay Components | Specimen diluent, lateral flow strips, built-in control, gold conjugated antibodies | Positive control, negative control, latex cards, latex conjugated anti-cryptococcal polyclonal antibodies |
| Serum Pre-Treatment | Serum dilution | Pronase with pronase inhibitor |
| CSF Pre-Treatment | none | Boil |
| Detection Antibodies | Gold-conjugated anti-cryptococcal monoclonal antibodies | Latex-conjugated anti-cryptococcal polyclonal antibodies |
| Storage Requirements | 20-25°C | 4°C ± 2° |
| | | |
# K. Standard/Guidance Document Referenced (if applicable):
Not Applicable
{4}
L. Test Principle:
The CrAg Lateral Flow Assay is a dipstick sandwich immunochromatographic assay, which detects cryptococcal antigen in serum. For the qualitative procedure, specimens are diluted 1:2 in 1x Specimen Diluent and analyzed. For the semi-quantitative procedure, specimens are diluted 1:5 in 1x Specimen Diluent followed by 1:2 serial dilutions. All dilutions are then analyzed as in the qualitative procedure. Specimens are placed into an appropriate reservoir, such as a test tube or microtiter plate, and the lateral flow device is then placed into the reservoir allowing the specimen to come into contact with the test membrane. The test uses specimen wicking to capture gold-conjugated, anti-cryptococcal monoclonal antibodies and gold-conjugated control antibodies that are deposited onto a membrane. If cryptococcal antigen is present in the specimen, it binds to the gold-conjugated, anti-cryptococcal antibodies. The gold-labeled antibody-antigen complex will continue to wick up the membrane where it will interact with the Test Line (T). The Test Line is immobilized anti-cryptococcal monoclonal antibodies. If the specimen contains cryptococcal antigen, a sandwich is created with the gold-labeled antibodies and the immobilized antibodies, causing a visible line to develop at the test line site (T). If proper flow occurs and the reagents are reactive at the time of use, the wicking of any specimen, positive or negative, will cause the gold-conjugated control goat IgG antibody to move to the Control Line (C) which is immobilized bovine anti-goat IgG antibody. The immobilized anti-goat antibody will bind to the gold-conjugated goat IgG Control antibody and will cause a visible line to develop (C). A positive test result will create two lines, while a negative test result will create one line (Figure 1). If the control line fails to develop a line, then the test is not valid.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Serum:
Repeatability and reproducibility with serum specimens were determined by spiking a serum specimen pool that was negative by the IMMY Latex-Cryptococcus Antigen Detection System with cryptococcal antigen at four concentrations: Negative, high negative (C₅), low positive (near C₉₅), and medium positive. The samples were analyzed on the CrAg Lateral Flow Assay in triplicate on five different days, at three different sites with a total of five different operators, on one lot, according to EP5-A2. One site was internal (Site 1) and two were a South African reference laboratory (Site 2) and a US hospital laboratory (Site 3). For repeatability, percent positive and percent negative detected were calculated for each site (Table 2). For reproducibility, overall percent positive and percent
{5}
negative detected were calculated by combining the data from all three sites (last two rows of Table 2).
| SERUM PANEL | Site 1 % Pos | Site 2 % Pos | Site 3 % Pos | Overall % Pos |
| --- | --- | --- | --- | --- |
| Negative | 0% (0/30) | 0% (0/30) | 0% (0/15) | 0% (0/75) |
| High Negative | 7% (2/30) | 0% (0/30) | 0% (0/15) | 3% (2/75) |
| Low Positive | 100% (30/30) | 100% (30/30) | 100% (15/15) | 100% (75/75) |
| Moderate Positive | 100% (30/30) | 100% (30/30) | 100% (15/15) | 100% (75/75) |
As expected, the high negative tested positive nearly 5% of the time and 3% of the time in serum. All other samples performed 100% as expected across all sites, all runs, and all operators.
## b. Linearity/assay reportable range:
Not Applicable
## c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The CrAg Lateral Flow Assay has a three-month expiration date when stored at room temperature. Real-time stability testing is on-going. Initial accelerated stability was determined by testing three lots of kits stored at 25°C and 37°C for three months. Buffer was spiked with cryptococcal antigen at four concentrations: 1.5, 3.0, 6.0 and 9.0ng/ml. Non-spiked buffer was used as a negative control. Each sample was tested in triplicate by one operator. Each spiked sample replicate tested positive and each negative control tested negative on all lots of the CrAg Lateral Flow Assay, stored at both 25°C and 37°C. This initial data indicate that the CrAg Lateral Flow Assay is stable at room temperature for three months. Stability studies are on-going to demonstrate a one-year expiration dating.
## d. Detection limit:
### Detection Limit/Analytical Cut-off
Analytical sensitivity for the CrAg Lateral Flow Assay was estimated by running varying concentrations of cryptococcal antigen diluted in LF Specimen Diluent. Test results determined that the analytical cut-off is 1.25ng/ml (Table 3).
{6}
Table 3. Analytical Cut-Off
| Sample Concentration (ng/ml) | No. Positive | No. Tested | % Positive |
| --- | --- | --- | --- |
| 0.50 | 0 | 24 | 0% |
| 0.75 | 0 | 24 | 0% |
| 1.00 | 4 | 24 | 17% |
| 1.25 | 12 | 24 | 50% |
| 1.50 | 21 | 24 | 88% |
| 1.75 | 24 | 24 | 100% |
| 2.00 | 24 | 24 | 100% |
| 2.50 | 24 | 24 | 100% |
| 3.00 | 24 | 24 | 100% |
| 3.50 | 24 | 24 | 100% |
| 4.00 | 24 | 24 | 100% |
Analytical specificity:
Table 4. Analytical Specificity
| Pathology | No. of Specimens | No. of Replicates Per Specimen | Total Tests | Total Number of Positives | % Positive (No Pos/Total) |
| --- | --- | --- | --- | --- | --- |
| Non-Fungal Pathologies | | | | | |
| HAMA (Human anti-Mouse antibody) Positive | 5 | 3 | 15 | 0 | 0% |
| Syphilis | 10 | 3 | 30 | 0 | 0% |
| Rubella | 5 | 3 | 15 | 0 | 0% |
| Mycoplasma | 10 | 3 | 30 | 0 | 0% |
| Toxoplasmosis | 7 | 3 | 21 | 0 | 0% |
| CMV Infection | 10 | 3 | 30 | 0 | 0% |
| Rheumatoid factor* | 10 | 3 | 30 | 0 | 0% |
| Total | 57 | 3 | 171 | 0 | 0% |
| Fungal Pathologies | | | | | |
| Penicilliosis | 5 | 3 | 15 | 0 | 0% |
| Sporotrichosis | 6 | 3 | 18 | 0 | 0% |
{7}
| Blastomycosis | 10 | 3 | 30 | 0 | 0% |
| --- | --- | --- | --- | --- | --- |
| Coccidioidomycosis | 10 | 3 | 30 | 0 | 0% |
| Histoplasmosis | 10 | 3 | 30 | 0 | 0% |
| Candidiasis | 10 | 3 | 30 | 0 | 0% |
| Aspergillosis ** | 10 | 3 | 30 | 3 | 10% |
| Aspergillus terreus
Culture Filtrate | 3 | 3 | 9 | 0 | 0% |
| Aspergillus niger
Culture Filtrate | 3 | 3 | 9 | 0 | 0% |
| Aspergillus flavus
Culture Filtrate | 3 | 3 | 9 | 0 | 0% |
| Aspergillus fumigatus
Culture Filtrate | 3 | 3 | 9 | 0 | 0% |
| Paracoccidioides brasiliensis
Culture Filtrate | 3 | 3 | 9 | 6 | 67% |
| Total | 76 | 3 | 228 | 9 | 3.9% |
* Rheumatoid factor concentrations tested ranged from 112IU/ml to 6479IU/mls.
** The three positives were the result of three replicates of the same specimen, which was positive.
In addition to the cross-reactivity study, interference testing was also performed on five icteric, five hemolyzed, and five lipemic serum specimens. Each specimen was spiked with cryptococcal antigen at three times the C95 concentration. All specimens were then tested at IMMY, on one lot of CrAg Lateral Flow assay in triplicate: spiked and unspiked. Percent positivity was determined for each condition. All of the unspiked specimens had negative results on the CrAg Lateral Flow Assay. All spiked specimens were positive, thus, these types of serum specimens do not interfere with the CrAg Lateral Flow Assay. However, it is possible that hemolyzed samples could lead to false negatives due to the high background color on the strip. As such, a statement is included in the Limitations of the Procedure section of the package insert:
"Hemolyzed serum samples could lead to false negative results due to the high background color on the strip."
Due to specimen availability, the following conditions were not tested in the CrAg Lateral Flow Assay: Candida dubliniensis, Candida tropicalis, Candida parapsilosis, Candida krusei, Candida glabrata, Cladosporium trichoides, Neisseria meningitidis, Salmonella typhi, Pneumocystis carinii, Trichosporon beigelii,
{8}
Zygomycetes, ANA+, HAV, HCV, Staph, and Strep.
# f. Assay cut-off:
High dose hook effect concentrations with serum specimens were determined by spiking negative serum that were negative by the IMMY Latex-Cryptococcus Antigen Detection System and CrAg Lateral Flow Assay, with cryptococcal antigen at various concentrations between 20 and $500\mathrm{ug / ml}$ . Each concentration was tested in triplicate at IMMY on one lot of CrAg Lateral Flow Assay, according to the package insert. It was determined that serum specimens with a cryptococcal antigen concentration higher than $200\mathrm{ug / ml}$ can produce a high dose hook effect and therefore may produce a false negative result.
# 2. Comparison studies:
# a. Method comparison with predicate device:
The CrAg Lateral Flow Assay was compared to the gold standard diagnoses of cryptococcosis (culture and/or India Ink) to evaluate the sensitivity and specificity of the assay. These studies contained a mix of both prospective and retrospective specimens. A summary table of the data collected is included below.
| Serum | | Culture/India Ink | |
| --- | --- | --- | --- |
| | | Positive | Negative |
| CrAg LFA Assay | Positive | 91 | 0 |
| | Negative | 0 | 123 |
| Serum | Calculated | 95% CI |
| --- | --- | --- |
| Sensitivity | 100% | 96.0% - 100% |
| Specificity | 100% | 97.0% - 100% |
# b. Matrix comparison:
# Serum Specimens
A panel of 197 serum specimens that were submitted to a US reference laboratory for cryptococcal antigen testing, using a Cryptococcal Antigen EIA, were collected and stored frozen until method comparison studies were performed. Of the 197 specimens, 96 were positive and 101 were negative by the EIA. All specimens were analyzed, according to EP12-A2, concurrently in the CrAg Lateral Flow Assay and in the IMMY Latex-Cryptococcus Antigen Detection System to ensure the data was not affected by the freeze-
{9}
thaw cycle. Each specimen was tested at IMMY in duplicate in both tests according to each test's package insert. The data is presented in a 2x2 contingency table (Table 8). The percent agreement positive, percent agreement negative, and 95% confidence interval are also presented (Table 9).
Table 8. Serum 2x2 Contingency Table
| | EIA (+) | EIA (-) |
| --- | --- | --- |
| CrAg Lat Flow (+) | 96 | 7 |
| CrAg Lat Flow (-) | 0 | 94 |
Table 9. Serum Statistical Analysis
| | | 95% CI |
| --- | --- | --- |
| % Agreement Positive | 100% | 96-100% |
| % Agreement Negative | 93% | 86-97% |
Semi-Quantitative Method Comparison
In addition, 62 of these specimens were tested using the semi-quantitative titration procedure in both the CrAg Lateral Flow Assay and the IMMY Late Cryptococcal Antigen Detection System (REF CR1003). Linear regression analysis of the data yielded an R² value of 0.890
3. Clinical studies:
Not Applicable
4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range:
Not Applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.