MERIT BIPOLAR COAGULATION PROBE

K102273 · Merit Medical Systems, Inc. · KNS · Sep 7, 2010 · Gastroenterology, Urology

Device Facts

Record IDK102273
Device NameMERIT BIPOLAR COAGULATION PROBE
ApplicantMerit Medical Systems, Inc.
Product CodeKNS · Gastroenterology, Urology
Decision DateSep 7, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.4300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Bipolar Coagulation Probe is intended to be passed through an endoscope's working channel to provide hemostasis throughout the gastrointestinal tract.

Device Story

Bipolar Coagulation Probe; endoscopic accessory for GI tract hemostasis. Device consists of proximal plug for electrosurgical generator connection and luer flush hub for irrigation; Teflon-coated wire encased in tubing; distal tip features spiral-configured metal electrodes on rounded ceramic base. Operated by physician via endoscope working channel. Device delivers electrical energy to tissue to achieve hemostasis; irrigation hub allows fluid delivery to site. Benefits patient by providing controlled coagulation during endoscopic procedures.

Clinical Evidence

Bench testing only. Verification testing included tensile, flow, leak, high frequency leakage, high frequency dielectric strength, mains frequency dielectric strength, wiring continuity, kink resistance, and tip separation. All results met predetermined acceptance criteria.

Technological Characteristics

Materials: Teflon-coated wire, metal electrodes, ceramic tip. Energy: Electrosurgical (bipolar). Form factor: Flexible probe for endoscopic working channel. Connectivity: Connects to electrosurgical generator and irrigation source. No software or electronic processing components.

Indications for Use

Indicated for providing hemostasis throughout the gastrointestinal tract via an endoscope's working channel. For prescription use only.

Regulatory Classification

Identification

An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ VA AERITAEDICA K102273 py 1 of 2 Merit Medical Systems, Inc. 1,600 Viest Merit Parkway Scuth Iopdan. Utah 54095 PHONE: 201-253-1600 FAX 801-250 14:33 :www.pepl.com ## 5.0 510(k) Summary | General<br>Provisions | Submitter Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Primary Contact:<br>Date of Preparation: | Merit Medical Systems, Inc.<br>1600 West Merit Parkway<br>South Jordan, UT 84095<br>(801) 208-4187<br>(801) 253-6905<br>Dan W. Lindsay<br>August 9, 2010 | SEP 07 2010 | |-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|-------------| | Subject<br>Device | Trade Name:<br>Common/Usual Name:<br>Classification Name: | To be assigned<br>Bipolar Coagulation Probe<br>Unit, electrosurgical, endoscopic (with or without<br>accessories) | | | Predicate<br>Device | Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer: | Merit Bipolar Coagulation Probe<br>Unit, electrosurgical, endoscopic (with or without<br>accessories)<br>K912129<br>Merit Medical Systems, Inc. | | | Classification | Class II<br>21 CFR §876.4300<br>Gastroenterology/Urology | | | | Intended Use | The Bipolar Coagulation Probe is intended to be passed through an<br>endoscope's working channel to provide hemostasis throughout the<br>gastrointestinal tract. | | | | Device<br>Description | The proximal end of the Merit Bipolar Coagulation Probe consists of a plug<br>which connects to an electrosurgical generator and a luer flush hub which<br>connects to an irrigation source. The plug comes in one of two<br>configurations: a single or dual bipolar plug. The tubing from the plug<br>encases a Teflon coated wire which is united with the flushing lumen from<br>the luer in a Y connection. A single tube with the coated wire and flushing<br>lumen leads from the Y connection to the distal probe tip. The distal end<br>(tip) consists of uniformly spaced metal electrodes configured in a spiral<br>fashion on the rounded ceramic tip. | | | {2}------------------------------------------------ | pg 2 of 2 | | |------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Technological Characteristics | The subject device has the same technological characteristics as the predicate device. Minor design changes were made to improve aesthetics, ergonomics and manufacturability of the device. | | Performance Testing | The verification testing required by the risk analysis was completed and included the following tests: tensile, flow, leak, high frequency leakage, high frequency dielectric strength, mains frequency dielectric strength, wiring continuity, kink resistance and tip separation. Testing results met the predetermined acceptance criteria. | | Testing Conclusion | The results of the testing performed demonstrate the subject device is as safe, as effective, and performs as well as or better than the predicate device. | | Summary of Substantial Equivalence | Based on the indications for use, design, and safety and performance testing, the subject Merit Bipolar Coagulation Probe meets the requirements that are considered essential for its intended use and is substantially equivalent to the predicate device, the Merit Bipolar Coagulation Probe manufactured by Merit Medical Systems, Inc. | 、バスト : . . . . . . . . バ K 102273 Pet 2 of 2 . {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the three branches of government. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002 Mr. Dan W. Lindsay Regulatory Affairs Specialist II Merit Medical Systems, Inc. 1600 West Merit Parkway SOUTH JORDAN UT 84095 SEP 0 7 2010 Re: K102273 Trade/Device Name: Merit Bipolar Coagulation Probe Regulation Number: 21 CFR8 876,4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: II Product Code: KNS Dated: August 10, 2010 Received: August 11, 2010 Dear Mr. Lindsay: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related {4}------------------------------------------------ adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Hubert H Humphrey Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive. Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health. Enclosure {5}------------------------------------------------ K102273 Merit Bipolar Coagulation Probe Premarket Notification 510(k) Merit Medical Systems, Inc. SEP 0 7 2010 ## Indications for Use Statement 4.0 510(k) Number (if known): Device Name: Indications for Use: The Bipolar Coagulation Probe is intended to be passed through an endoscope's working channel to provide hemostasis throughout the gastrointestinal tract. K102273 Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Hul ition Sign Off (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...