K102248 · Innovasis, Inc. · NKB · Oct 7, 2010 · Orthopedic
Device Facts
Record ID
K102248
Device Name
EXCELLA II SPINAL SYSTEM MODEL EC2-00
Applicant
Innovasis, Inc.
Product Code
NKB · Orthopedic
Decision Date
Oct 7, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3070
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Innovasis® Excella® Spinal System is intended for use in the non-cervical area of the spine. The INNOVASIS® Excella® Spinal System, when used for pedicle screw fixation is intended only for patients: a) Having severe spondylolisthesis (Grade 3 & 4) at the L5-S1 joint; b) Who are receiving fusion using autogenous bone graft only; c) Who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and d) Who are having the device removed after the development of a solid fusion mass. The INNOVASIS® Excella® Spinal System, when used as a pedicle screw system in skeletally mature patients, is intended to provide immobilization of spinal segments, as an adjunct to fusion, in treatment of the following acute and chronic deformities of the thoracic, lumbar, and sacral spine: i. Degenerative spondylolisthesis with objective evidence of neurologic impairment; ii. Fracture; iii. Dislocation; iv. Scoliosis; v. Kyphosis; vi. Spinal tumor; and vii. Previous failed fusion (pseudoarthrosis). The INNOVASIS® Excella® Spinal System, when used for anterolateral non-pedicle fixation, is intended for the following indications: i. Degenerative disc disease (defined as back pain of discogenic origin with degenerative disc confirmed by patient history and radiographic studies); ii. Spinal stenosis; iii. Spondylolisthesis; iv. Spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis); v. Pseudoarthrosis; vi. Tumor; vii. Trauma (i.e. fracture or dislocation); and viii. Previous failed fusion. The INNOVASIS® Excella® Spinal System, when used for posterior non-pedicle screw fixation to the noncervical spine, is intended for the following indications: i. Degenerative disc disease (defined as back pain of discogenic origin with degenerative disc confirmed by patient history and radiographic studies); ii. Spinal stenosis; iii. Spondylolisthesis; iv. Spinal deformities (i.e. scoliosis, kyphosis and/or lordosis); v. Pseudoarthrosis; vi. Tumor; vii. Trauma (i.e. fracture or dislocation); viii. Previous failed fusion.
Device Story
Excella II Spinal System; polyaxial pedicle screw system for non-cervical spine stabilization. Components: 6AL-4V titanium alloy cannulated screws (4.75mm-8.5mm diameter), titanium rods, locking caps with buttress threads, and cross connectors. Features self-tapping, double-lead screw threads; reduction heads available for scoliosis correction. Used in OR by orthopedic surgeons; implants provide mechanical immobilization of spinal segments to facilitate fusion. System components locked via threaded caps to ensure construct stability. Benefits: provides rigid fixation for spinal deformities/trauma; reduction heads facilitate rod placement in complex anatomy. Non-sterile; requires sterilization by user.
Clinical Evidence
Bench testing only. Performance verified per ASTM F543 (bone screw testing), ASTM F1717 (vertebrectomy model), and ASTM F1798 (interconnection mechanisms). Testing included static compression bend, static torsion, dynamic compression bend, axial pullout strength, and torque to failure. No clinical data presented.
Indicated for skeletally mature patients requiring non-cervical spinal immobilization/stabilization as an adjunct to fusion. Treats degenerative spondylolisthesis, fractures, dislocations, scoliosis, kyphosis, spinal tumors, pseudoarthrosis, degenerative disc disease, and spinal stenosis. Contraindicated in morbid obesity, mental illness, alcoholism/drug abuse, pregnancy, metal sensitivity, severe osteopenia, active infection, or patients unable to follow post-operative instructions.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
{1}------------------------------------------------
# 1.0 510(k) Summary Prepared October 5, 2010 Company: Innovasis, Inc. 614 E. 3900 South Salt Lake City, UT 84107 Marshall C. McCarty Contact: OCT 7 2010 Phone: (801) 261-2236 mmccarty@innovasis.com Trade Name: Excella® Spinal System Excella-M Excella-P Excella II™ Common Name: Pedicle Screw Spinal System Classification: Regulation No.: 21CFR 888.3050, 3060. Class II Regulation No .: 21CFR 888.3070, Class III Product Code: NKB, KWP, KWQ, MNI, MNH Review Panel: Orthopedic Applicable Standards: ASTM F543-07 Standard Specification and Test . Methods for Metallic Medical Bone Screws. ASTM F1717-09 Standard Test Methods for Spinal . Implant Constructs in a Vertebrectomy Model ASTM F1798-97(Reapproved 2008) Standard Guide . for Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants ASTM F2193-02(Reapproved 2007) Standard Specifications and Test Methods for Components Used in the Surgical Fixation of the Spinal Skeletal System Substantially Excella® Spinal System Equivalent Device: K071921 Device Description: The Excella IIM Spinal System consists of 6AL-4V Titanium alloy implants meant to be used in a system. The single use devices are offered in a variety of different
Pg 1 of 5
{2}------------------------------------------------
K1022-18
lenaths ranging from 25mm to 60mm as well as the following diameters, 4.75mm, 5.5mm, 6.5mm and 7.5mm, 8.5mm-and are sold non-sterile. The Excella II implants are cannulated and feature a multi-axial joint for free motion of the screw head to allow the physician greater flexibility when placing the screws, including use of a minimally invasive surgical procedure. The bone thread portion of the screws features a self-tapping, double lead with an aggressive thread.
The Excella II system utilizes 5.5mm Titanium rods in various lengths from 35mm to 460mm (straight) or 35mm to 85mm (bent). The rods are placed in the head of the screws and the system is locked in place with Titanium locking caps. The locking caps design includes a threaded locking screw attached to an alignment cap. Each locking cap includes a buttress thread design for axial strength and to prevent screw loosening. The alignment cap is attached to the bottom of the locking screw and rotates about the radial axis. This cap is intended to aid the alignment of the locking cap during insertion. The alignment cap also improves the rotational stiffness at the rod to locking cap interface.
The Excella II system also has available alignment caps with extended length head (reduction head) that can be broken off after installation. The available lengths go from 5 inches (12.7mm) to 4.72 inches (120mm). Primarily used in scoliosis correction, the reduction head allows easier use of longer rods that may be difficult to place in the standard head due to the anatomy of the patient and the curvature of the spine needing correction. A retaining ring is provided to prevent splay of the reduction head while tightening down the rod. Once the locking screw has been tightened in place, the ring is removed and the added length on the alignment cap can be broken off at a tab, resulting in a final profile and assembly that is identical to a standard screw and locking cap assembly.
The cross connectors come in various adjustable lengths and allow for variation in perpendicular placement on two rods. Each cross connector features a clamping mechanism to grip the rods and stabilize the relation between the rods.
(Non-clinical)--Performance testing per ASTM F543, Performance Data: F1717 and F1798 for Static Compression Bend, Static Torsion and Dynamic Compression Bend, Axial Pullout Strength, Torque to Failure, Axial, Flexion and Torsional Grip indicates that the Excella II Spinal System is capable of performing in accordance with its intended use.
Pg 2 of 5
{3}------------------------------------------------
K102248
| Materials: | The implants are machined from Medical Grade Titanium<br>(6AL-4V) per ASTM F136-02a. |
|------------|--------------------------------------------------------------------------------------|
|------------|--------------------------------------------------------------------------------------|
The intended use is the same as the predicate Excella® Intended Use: Spinal System (K071921). The Innovasis Excella II™ Spinal System is intended for use in the non-cervical area of the spine.
Indications for Use: The intended use is the same as the predicate Excella® Spinal System (K071921).
> The Excella Pedicle Screw System, when used for pedicle screw fixation is intended only for patients:
- a) Having severe spondylolisthesis (Grade 3 & 4) at the L5-S1 joint;
- b) Who are receiving fusion using autogenous bone graft only;
- c) Who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and
- d) Who are having the device removed after the development of a solid fusion mass.
The Excella Pedicle Screw System, when used as a pedicle screw system in skeletally mature patients, is intended to provide immobilization and stabilization of spinal segments, as an adiunct to fusion, in treatment of the following acute and chronic deformities of the thoracic, lumbar, and sacral spine:
- a) Degenerative spondvlolisthesis with objective evidence of neurologic impairment;
- b) Fracture:
- c) Dislocation;
- d) Scoliosis;
- e) Kyphosis;
- Spinal tumor; and ﮯ
- g) Previous failed fusion (pseudoarthrosis).
The Excella Pedicle Screw System, when used for anterolateral non-pedicle fixation, is intended for the following indications:
- a) Degenerative disc disease (as defined as back pain of discogenic origin with degenerative disc confirmed by patient history and radiographic studies);
- b) Spinal stenosis;
- c) Spondylolisthesis;
- d) Spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis);
- e) Pseudoarthrosis;
- c Tumor:
- g) Trauma (i.e. fracture or dislocation); and
{4}------------------------------------------------
K10aa48?
- h) Previous failed fusion.
The Excella Pedicle Screw System, when used for posterior non-pedicle screw fixation to the non-cervical spine, is intended for the following indications:
- a) Degenerative disc disease (defined as back pain of discogenic origin with degenerative disc confirmed by patient history and radiographic studies);
- b) Spinal stenosis;
- Spondylolisthesis; c)
- d) Spinal deformities (i.e. scoliosis, kyphosis and/or lordosis);
- e) Pseudoarthrosis;
- Tumor: ()
- g) Trauma (i.e. fracture or dislocation);
- h) Previous failed fusion.
### Contraindications (Exclusions for Use): .
- a.) Morbid obesity
- b.) Mental illness
- c.) Alcoholism or drug abuse
- d.) Pregnancy
- e.) Metal sensitivity / allergies
- f.) Severe osteopenia
- g.) Active infection near the surgical site
- h.) Patients unwilling or unable to follow post-operative care instructions
## Basis for Substantial
Equivalence:
The Excella II Spinal System has been subjected to risk analysis, engineering analysis and testing to recognized standards and has been shown to be substantially equivalent to the predicate device, K071921, Excella Spinal System.
Design configurations are substantially equivalent.
Applied mechanical loads are substantially equivalent.
Product sizes and shapes are substantially equivalent.
Materials used are equivalent.
Biocompatibility requirements are equivalent.
Manufacturing and processing methods are substantially equivalent.
Training of Physicians and Representatives is substantially equivalent.
Indications for Use are equivalent.
Shelf life is equivalent.
{5}------------------------------------------------
K102248
# Summary of Safety
**And Effectiveness:**
And Effectiveness: The Innovasis Excella II Spinal System is shown to be safe and effective for use as a polyaxial pedicle screw system and in the indications associated with device product codes KWP, MNH, MNI and NKB.
> Safety: The following have been performed to ensure the safety of the product:
> Use of biocompatible materials that have a proven history of acceptable performance.
Material selection with certifications to ASTM F136.
Additional biocompatibility testing verification (LAL, MEM analysis)
Mechanical testing performed to applicable ASTM standards.
Effectiveness: The following have been performed to ensure the effectiveness of the product:
Dimension verification (inspection) to the design.
Process verification and validation.
Product verification and validation.
Physician evaluation
Calibrated inspection systems.
5 of 5
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle or bird with outstretched wings, depicted in a simple, flowing design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
OCT 7 2010
Innovasis, Inc. % Mr. Marshall C. McCarty Manager QA/RA 614 East 3900 South Salt Lake City, Utah 84107
Re: K102248
Trade/Device Name: Excella II Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNI, MNH, KWP, KWQ Dated: September 10, 2010 Received: September 13, 2010
Dear Mr. McCarty:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{7}------------------------------------------------
Page 2- Mr. Marshall C. McCarty
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Mark N. Melkers Director Division of Surgical. Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use Statement
510(k) Number:
Device Name: Excella® Spinal System
### Indications for Use:
The Innovasis® Excella® Spinal System is intended for use in the non-cervical area of the spine.
The INNOVASIS® Excella® Spinal System, when used for pedicle screw fixation is intended only for patients: a) Having severe spondylolisthesis (Grade 3 & 4) at the L5-S1 joint; b) Who are receiving fusion using autogenous bone graft only; c) Who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and d) Who are having the device removed after the development of a solid fusion mass.
The INNOVASIS® Excella® Spinal System, when used as a pedicle screw system in skeletally mature patients, is intended to provide immobilization of spinal segments, as an adjunct to fusion, in treatment of the following acute and chronic deformities of the thoracic, lumbar, and sacral spine:
- i. Degenerative spondylolisthesis with objective
- evidence of neurologic impairment;
- · Fracture;
- Dislocation; iii.
- iv. Scoliosis;
- V. Kyphosis;
- Spinal tumor; and vi.
- vii. Previous failed fusion (pseudoarthrosis).
Prescription Use (21 CFR 801 Subpart D) OR
Over-The-Counter-Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K102248
Page 1 of 2
{9}------------------------------------------------
The INNOVASIS® Excella® Spinal System, when used for anterolateral non-pedicle fixation, is intended for the following indications:
- i. Degenerative disc disease*
- ii . Spinal stenosis;
- iii. Spondylolisthesis;
Pseudoarthrosis V.
- vi. Tumor
- vii. Trauma (i.e. fracture or dislocation); and
- iv. Spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis);
- viii. Previous failed fusion.
*(defined as back pain of discogenic origin with degenerative disc confirmed by patient history and radiographic studies).
The INNOVASIS® Excella® Spinal System, when used for posterior non-pedicle screw fixation to the noncervical spine, is intended for the following indications:
- i. Degenerative disc disease;
- ii. Spinal stenosis;
- iii. Spondylolisthesis;
- V. Pseudoarthrosis
- Tumor: vi.
- vii. Trauma (i.e. fracture or dislocation);
- iv. Spinal deformities (i.e. scoliosis, kyphosis and/or lordosis); viii. Previous failed fusion. *(defined as back pain of disogenic origin with degenerative disc confirmed by patient history and radiographic studies).
Prescription Use (21 CFR 801 Subpart D) OR Over-The-Counter-Use (21 CFR 807 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
KIO 2248 510(k) Number.
Page 2 of 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.