← Product Code [CAW](/productcode/CAW) · K102245

# CARE ZY5BA OXYGEN CONCENTRATOR (K102245)

_Foshan Care Medical Technology Co., Ltd. · CAW · Nov 9, 2010 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K102245

## Device Facts

- **Applicant:** Foshan Care Medical Technology Co., Ltd.
- **Product Code:** [CAW](/productcode/CAW.md)
- **Decision Date:** Nov 9, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5440
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

The CARE ZY 5BA Oxygen Concentrator is intended for use as an oxygen concentrator to provide supplemental low flow oxygen therapy in the home, nursing homes, patient care facilities, etc. The CARE ZY5BA is available by prescription only under the supervision of a physician, and is not intended to support or sustain life.

## Device Story

AC-powered oxygen concentrator; draws ambient room air via compressor; utilizes molecular sieve adsorbent to selectively retain nitrogen; releases nitrogen via pressure venting; cycle controlled by motorized valve; delivers high-purity oxygen via standard tubing, nasal cannula, or mask; optional bubble humidifier. Used in home or clinical settings; operated by patient or caregiver. Front panel provides status indicators (power, oxygen purity, pressure, hour meter) and flow control. Device provides supplemental oxygen to assist patients; does not sustain life.

## Clinical Evidence

Bench testing only. Validated against design specifications and standards including ASTM F1464, ISO 8359, IEC60601-1, IEC60601-1-2, IEC60601-1-4, and ISO14971. Hazard analysis performed for system and software.

## Technological Characteristics

AC-powered portable oxygen generator; molecular sieve adsorbent technology; motorized valve control; pressure relief valve. Includes status indicators, flow meter, and monitor display for pressure/purity. Complies with ASTM F1464, ISO 8359, IEC60601-1, IEC60601-1-2, IEC60601-1-4, and ISO14971.

## Regulatory Identification

A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).

## Predicate Devices

- A&J-POCA01 Oxygen Concentrator ([K071608](/device/K071608.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.

The assigned 510(k) number is:

#### Submitter:

KIDSSAR

Foshan Care Medical Technology Co., Ltd

#### Address:

No.1 HuaBao South Road, Foshan, Guangdong, China

### Contact Person:

Polo Chen R&D Manager Email: polochen888@hotmail.com Tel: 0086-757-88023253 Fax: 0086-757-83828996

#### Applicant:

Foshan Care Medical Technology Co., Ltd

## Address:

No.1 HuaBao South Road, Foshan, Guangdong, China

# Date of the summary prepared:

June 15.2010

## Name of the device:

Trade/Proprietary Name: CARE ZY5BA Oxygen Concentrator Common Name: Oxygen Concentrator

#### Classification:

Class II as per 21 CFR 868.5440, Portable Oxygen Generator

#### Legally Marketed Predicate Device:

A&J-POCA01 Oxygen Concentrator K071608

# Intended Use:

1

#### Chapter 4

{1}------------------------------------------------

The CARE ZY 5BA Oxygen Concentrator is intended for use as an oxygen concentrator to provide supplemental low flow oxygen therapy in the home, nursing homes, patient care facilities, etc. The CARE ZY5BA is available by prescription only under the supervision of a physician, and is not intended to support or sustain life.

## Description:

The CARE ZY5BA is an AC power electrically operated device. The unit separates oxygen from room air (ambient air), which allows high-purity supplemental oxygen to be delivered through the oxygen outlet, although the concentrator filters the oxygen in a room. It will not affect the normal amount of oxygen in user's room. Air is drawn into the device with a compressor and exposed to molecular sieve adsorbent that selectively retains nitrogen and other components until they are released when the pressure is vented to the atmosphere. This cycle is controlled by a motorized valve and protected from over pressurization by the compressor's pressure relief valve.

Oxygen provided by the CARE ZYSBA Oxygen Concentrator is delivered to the user by means of standard oxygen supply tubing and a standard nasal cannula or oxygen mask. A standard bubble humidifier may be used, if physician has prescribed an oxygen humidifier as part of therapy.

The front panel of the CARE ZY5BA contains the controls and indicators. These include the status lights (included power light, normal oxygen light, low oxygen light and service required light), standard power switch, flow meter and the flow meter knob, a circuit breaker which could reset the device after electrical overload shutdown, an oxygen outlet which oxygen is dispersed through, a monitor display which indicates the condition of system status (included pressure status, oxygen purity status and electric hour meter, etc). The user could operate the device conveniently according the instructions.

### Technological Characteristics:

Technologies utilized by the CARE ZY5BA Oxygen Concentrator bring forth no new questions of safety and effectiveness. These technologies are also currently being used in the identified predicate device.

Bench performance testing has demonstrated that the CARE ZY5BA Oxygen

{2}------------------------------------------------

Concentrator is substantially equivalent to the predicate device.

# Testing:

Laboratory testing was conducted to validate and verify that the CARE ZYSBA Oxygen Concentrator met all design specifications and was substantially equivalent to the predicate device. This testing consisted of all environmental testing identified in the FDA's "Reviewer Guidance for Nebulizers, Metered Dose Inhalers, Spacers and Actuators (October 1, 1993)". Additional testing was performed to demonstrate compliance with the standards of ASTM F1464 and ISO 8359. Finally, a hazard analysis of the system and its software was performed and testing was conducted to validate the systems overall operation. The CARE ZY5BA Oxygen Concentrator has also been tested to assure compliance to the requirements of various published standards, including IEC60601-1, IEC60601-1-2, IEC60601-1-4, and ISO14971.

# Conclusion:

The conclusions drawn from the testing of the CARE ZY5BA Oxygen Concentrator demonstrates that the device is substantially equivalent to the predicate device.

{3}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Foshan Care Medical Technology Company, Limited C/O Mr. Marc M. Mouser Responsible Third Party Official Underwriters Laboratories, Incorporated 2600 NW Lake Road Camas, Washington 98607-9526

NOV - 9 2010

Re: K102245

Trade/Device Name: CARE ZY5BA Oxygen Concentrator Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: II Product Code: CAW Dated: October 28, 2010 Received: November 4, 2010

Dear Mr. Mouser:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however. that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

### Page 2- Mr. Mouser

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Susan Pease

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

NOV - 9 2010

510(k) Number (if known): 长/ロフ24/5 Device Name: CARE ZY5BA Oxygen Concentrator Indications for Use:

The CARE ZY5BA Oxygen Concentrator is intended for use as an oxygen concentrator to provide supplemental low flow oxygen therapy in the home, nursing homes, patient care facilities, etc. The CARE ZY5BA Oxygen Concentrator is available by prescription only under the supervision of a physician, and is not intended to support or sustain life.

Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

L. Schult
Page 1 of 1

(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K102245

---

**Source:** [https://fda.innolitics.com/device/K102245](https://fda.innolitics.com/device/K102245)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
