The Elecsys PreciControl Multimarker is used for quality control of specified Elecsys immunoassays on the Elecsys and cobas e immunoassay analyzers. For prescription use only.
Device Story
Elecsys® PreciControl Multimarker is a lyophilized quality control product consisting of analytes (C-Peptide, Insulin, ACTH, hGH) in an equine serum matrix. The device is used by laboratory personnel on Elecsys and cobas e immunoassay analyzers to monitor the performance of specified immunoassays. The product is reconstituted with distilled or deionized water before use. By comparing measured control values against expected ranges, clinicians and laboratory staff verify the accuracy and precision of the immunoassay system, ensuring reliable patient test results. The device serves as an essential component of laboratory quality assurance programs.
Clinical Evidence
No clinical data. Bench testing only. Stability studies confirmed recovery within 90-110% of target concentrations under various storage conditions (reconstituted, frozen, and on-analyzer). Value assignment validated via duplicate testing on multiple analyzer platforms with acceptance criteria for precision and inter-analyzer variability.
Technological Characteristics
Lyophilized equine serum matrix; two concentration levels. Analyte-specific values assigned via master calibrators traceable to NIBCS or WHO reference materials. Compatible with Roche Elecsys and cobas e immunoassay analyzers. Stability: 2-8°C (unopened), 2-8°C (72 hours reconstituted), -20°C (31 days reconstituted), 20-25°C (5 hours on-analyzer).
Indications for Use
Indicated for quality control of specified Elecsys immunoassays on Elecsys and cobas e immunoassay analyzers in a clinical laboratory setting.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
{0}------------------------------------------------
Confidential
| Submitter<br>Name, Address,<br>Contact | Roche Diagnostics<br>9115 Hague Road<br>Indianapolis, IN 46250<br><br>Contact Person: Sarah Baumann<br>Phone: 317-521-3952<br>Fax: 317-521-2324<br>Email: sarah.baumann@roche.com<br><br>Secondary Contact: Stephanie Greeman<br>Phone: 317-521-2458<br>Fax: 317-521-2324<br>Email: stephanie.greeman@roche.com<br><br>Date Prepared: July 29, 2010 | AUG 27 2010 | |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|--------------------------------------|
| Device name | Proprietary name: Elecsys® PreciControl Multimarker<br>Common name: PreciControl Multimarker<br>Classification: Multi-Analyte Controls, All Kinds (assayed and unassayed) | | |
| Device<br>description | Elecsys® PreciControl Multimarker is a lyophilized product consisting of<br>analytes in an equine serum matrix. During manufacture, the analytes are<br>spiked into the matrix at the desired concentration levels. | | |
| Predicate<br>device | The Elecsys® PreciControl Multimarker is substantially equivalent to other<br>products in commercial distribution intended for similar use. We claim<br>equivalency to the currently marketed Elecsys® MultiAnalyte (K033937). | | |
| Intended use | Elecsys® PreciControl Multimarker is used for quality control of specified<br>Elecsys immunoassays on the Elecsys and cobas e immunoassay analyzers. | | |
| | Continued on next page | | |
| Device<br>Comparison—<br>Similarities | The table below presents the similarities between the Elecsys® PreciControl<br>Multimarker and the predicate device, Elecsys® PreciControl MultiAnalyte<br>(K033937). | | |
| | Characteristic | Predicate Device<br>Elecsys® PreciControl<br>MultiAnalyte<br>(K033937) | Elecsys® PreciControl<br>Multimarker |
| | Analyzer system | Elecsys and cobas e | Same |
In accordance with 21 CFR 807.92, Roche Diagnostics hereby submits Introduction official notification as required by Section 510(k) of the Federal Food, Drug and Cosmetics Act of our intention to market the device described in this Premarket Notification [510(k)].
. . . .
510(k) Summary— Elecsys® PreciControl Multimarker
K 102157
510(k) Summary
Elecsys® PreciControl Multimarker
{1}------------------------------------------------
Roche Diagnostics
Elecsys® PreciControl Multimarker 510(k) Summary
## 510(k) Summary— Elecsys® PreciControl Multimarker, Continued
Analyte
Format
Matrix
Levels
concentration
immunoassay analyzers
· C-Peptide (ng/mL)
Level 1 = 2
Level 2 = 10
Lyophilized
Two
Equine serum Same Same
Same
Same
Continued on next page
{2}------------------------------------------------
K102157
Roche Diagnostics
Elecsys® PreciControl Multimarker
510(k) Summary
# 510(k) Summary— Elecsys® PreciControl Multimarker, Continued
| Device<br>Comparison—<br>Differences | The table below presents the differences between the Elecsys® PreciControl<br>Multimarker and the predicate device, Elecsys® PreciControl MultiAnalyte<br>(K033937). | | |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Characteristic | Predicate Device<br>Elecsys® PreciControl<br>MultiAnalyte<br>(K033937) | Elecsys® PreciControl<br>Multimarker |
| | Intended use | PreciControl MultiAnalyte<br>is used for quality control<br>of the Elecsys C-Peptide<br>and Elecsys Insulin<br>immunoassays on the<br>Elecsys immunoassay<br>systems: | Elecsys PreciControl<br>Multimarker is used for<br>quality control of specified<br>Elecsys immunoassays on<br>the Elecsys and cobas e<br>immunoassay analyzers. |
| | Analyte<br>concentration | • Insulin (µU/mL)<br>Level 1 = 20<br>Level 2 = 100 | • Insulin (µU/mL)<br>Level 1 = 25<br>Level 2 = 80<br>• ACTH (pg/mL)<br>Level 1 = 50<br>Level 2 = 1000<br>• hGH (ng/mL)<br>Level 1 = 1.00<br>Level 2 = 10.0 |
| | Handling | Dissolve carefully the<br>contents of one bottle by<br>adding exactly 2.0 mL of<br>distilled water and allow to<br>stand closed for 15 minutes<br>to reconstitute. Mix<br>carefully, avoiding the<br>formation of foam. | Dissolve carefully the<br>contents of one bottle by<br>adding exactly 2.0 mL of<br>distilled or deionized water<br>and allow to stand closed for<br>30 minutes to reconstitute.<br>Mix carefully, avoiding the<br>formation of foam. |
Continued on next page
{3}------------------------------------------------
K102157
### 510(k) Summary— Elecsys® PreciControl Multimarker, Continued
The table below presents the differences between the Elecsys® PreciControl Device Multimarker and the predicate device, Elecsys® PreciControl MultiAnalyte Comparison-Differences (K033937). (continued)
| Characteristic | Predicate Device<br>Elecsys® PreciControl<br>MultiAnalyte<br>(K033937) | Elecsys® PreciControl<br>Multimarker |
|----------------|------------------------------------------------------------------------|-----------------------------------------------------------------|
| Stability | Unopened :<br>Store at 2-8°C until expiration<br>date | Unopened :<br>Store at 2-8°C until expiration<br>date |
| | Reconstituted:<br>• on the analyzer at 20-25°C:<br>up to 3 hrs | Reconstituted:<br>• on the analyzer at 20-25 °C:<br>up to 5 hrs |
| | • at -20°C: 3 months (freeze<br>only once) | • at -20°C: 31 days (freeze<br>only once) |
| | | • or at 2-8°C for 72 hours |
| | After Thawing:<br>use only once. | |
The Elecsys PreciControl Multimarker was evaluated for value assignment, Performance Characteristics stability, and duration of reconstitution. C-The data demonstrate that the performance of the Elecsys® PreciControl Conclusion Multimarker is substantially equivalent to that of the predicate device, Elecsys® PreciControl MultiAnalyte (K033937).
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/2 description: The image shows a logo with a stylized eagle or bird-like figure with outstretched wings. The logo is associated with the Department of Health, as indicated by the text surrounding the emblem. The overall design is simple and monochromatic, with a focus on the symbolic representation of the organization.
Roche Professional Diagnostics Ms. Sarah Baumann 9115 Hague Road P.O. Box 50410 Indianapolis, IN 46250-0416
Food & Drug Administration 10903 New Hampshire Avenue Building 66 Silver Spring, MD 20993
AUG 2 7 2010
Re: k102157
Trade/Device Name: Elecsys PreciControl Multimarker Regulation Number: 21 CFR 862.1660 Regulation Name: Quality Control Material (Assayed and Unassayed) Regulatory Class: Class I, Reserved Product Code: JJY Dated: July 29, 2010 Received: July 30, 2010
Dear Ms. Baumann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{5}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
CA
Courtney C. Harper, Ph.D. Directør Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K102157
# Indications for Use Form
510(k) Number (if known): _ K 102157 · '
Device Name: Elecsys PreciControl Multimarker
Indications for Use: Elecsys PreciControl Multimarker is used for quality control of specified Elecsys immunoassays on the Elecsys and cobas e immunoassay analyzers.
Over-The-Counter Use Prescription Use × Prescription Use
(Part 21 CER 801 Subpart D) AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Noug Rheinheine
Division S an-Off Office of In Vitro Diagnostic Device Evaluation and Safety
长102157 510(k)
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.