The ODFS® Pace is intended to provide ankle dorsiflexion in individuals who have a dropped foot as a consequence of upper motor neuron injury. By detecting the swing phase of gait through a foot switch signal, appropriate electrical stimulation of the leg and ankle muscles may improve gait by flexing the foot of persons who have lost or impaired function. There may be additional benefits from FES such as muscle re-education, prevention/retardation of disuse atrophy, increased or maintained range of joint motion and increase in local blood flow. The ODFS® Pace is a medical device and should only be used under medical supervision for the treatment of dropped foot following an upper motor neuron injury.
Device Story
ODFS Pace is a body-worn, single-channel external functional neuromuscular stimulator (FES). Input: gait signal from a wired footswitch worn in the shoe. Operation: device detects swing phase of gait; triggers electrical pulses to leg/ankle muscles via self-adhesive skin surface electrodes (typically common peroneal nerve). Output: muscle contraction causing ankle dorsiflexion. Optional mode: powered muscle stimulator for exercise (footswitch disabled). Used by patients with drop foot from upper motor neuron injury; clinician-supervised setup. Benefits: improved gait, muscle re-education, reduced disuse atrophy, maintained joint range of motion, increased local blood flow.
Clinical Evidence
Bench testing only. Verification of functional and output specifications (stimulator controls, settings, indicators, stimulus output/timing, footswitch operation, power consumption). Compliance with ISO 60601-1:2006 (safety), ISO 60601-1-2:2007 (EMC), and ISO 60601-2-10:2001 (nerve/muscle stimulators).
Technological Characteristics
Battery-powered external FES/NMES. Waveform: Biphasic (symmetrical/asymmetrical). Pulsewidth: 0-360μs. Frequency: 20-60Hz. Dimensions: small, body-worn. Connectivity: wired footswitch. Electrodes: 5cm x 5cm self-adhesive. Standards: ISO 60601-1, ISO 60601-1-2, ISO 60601-2-10.
Indications for Use
Indicated for individuals with dropped foot due to upper motor neuron injury. Used to provide ankle dorsiflexion to improve gait via FES. Optional use as a powered muscle stimulator for muscle re-education, prevention of disuse atrophy, maintenance of joint range of motion, and increased local blood flow. Contraindicated for use without medical supervision.
Regulatory Classification
Identification
An external functional neuromuscular stimulator is an electrical stimulator that uses external electrodes for stimulating muscles in the leg and ankle of partially paralyzed patients (e.g., after stroke) to provide flexion of the foot and thus improve the patient's gait.
{0}------------------------------------------------
K10211.5
Image /page/0/Picture/1 description: The image shows a logo with the letters "OML" in a bold, sans-serif font. The letters are white and are set against a dark, textured background. The logo is contained within a rounded rectangle with a thick border. The overall appearance of the logo is somewhat distressed or vintage due to the texture and shading.
LEADING REHABILITATION
THROUGH TECHNOLOGY
# Odstock Medical Limited
National Clinical FES Centre MAR 3 0 2011 Salisbury District Hospital Salisbury Wilitshire SP2 8BJ വന
Tel +44 (0) 1722 439541 Fax +44 (0) 1722 425263 steven.crook@odstockmedical.com www.odstockmedical.com
# 510(k) Summary For the
# ODFS® Pace
# [Section 5 of the Submission]
5.1 Sponsor/Owner
Odstock Medical Limited Laing Building Salisbury District Hospital Salisbury Wiltshire SP2 8BJ United Kingdom
Odstock Medical Ltd is a Registered Establishment, number: 3005344585
| Contact person: | Steven Crook, Quality and Regulatory Manager |
|-----------------|----------------------------------------------|
| Telephone: | 0044 1722 439541 |
| Fax: | 0044 1722 425263 |
| Email: | steven.crook@odstockmedical.com |
21 July 2010, with revisions 11&14 March 2011 Date prepared
#### US Agent
| <b>Boston Brace International</b><br>20 Ledin Drive<br>Avon, MA 02322 USA | Phone: 800-262-2235<br>Fax: 800-634-5048 |
|---------------------------------------------------------------------------|------------------------------------------|
|---------------------------------------------------------------------------|------------------------------------------|
Barry Bull (Chairman) • Malcom Cassells (Director) • Philip Casson (CEO) • Ian Swain (Director)
Odstock Medical Limited (Company No: 5532620) is part of Salisbury NHS Foundation Trust
{1}------------------------------------------------
#### 510(k) Summary for the ODFS® PACE Dropped Foot Stimulator
## 5.2 DEVICE NAME
| Trade/Proprietary Name: | ODFS®-Pace |
|-------------------------|----------------------------------------------------------------------------------------|
| Common/Usual Name: | External Functional Neuromuscular Stimulator (FES) |
| Classification Name: | External Functional Neuromuscular Stimulator (FES)<br>Powered Muscle Stimulator (NMES) |
| Classification No. | 21 CFR 882.5810 |
| Classification Code: | Class II |
| Product Codes: | GZI and IPF |
#### 5.3 Predicate Devices
| Predicate Device | K number | Comment | |
|----------------------|----------|---------------------------------------|--|
| EMPI Focus Model 795 | K951951 | EMPI | |
| ODFS III V6.2 | K020991 | Odstock Medical Ltd (transferred from | |
| | | Salisbury District Hospital) | |
#### 5.4. Device Description
The ODFS® Pace is an external functional neuromuscular stimulator. It consists of a small, body worn, single channel stimulator which is triggered by a wired footswitch worn in the shoe. Electrical pulses are generated by the device in order to stimulate muscles in the lower limb. Triggering with a footswitch permits the stimulation to be varied according to the gait of the user. Stimulation produces contraction of the appropriate muscles to cause dorsiflexion of the ankle in individuals who have impaired or absent ability to pick up their foot during walking as a result of a neurological injury. This partial paralysis affecting the foot is commonly known as 'drop foot' or 'dropped foot'.
The electrical stimulation is delivered via self adhesive skin surface electrodes. The electrodes may be flexibly placed according to the optimum response of the patient but typically are located over the common peroneal nerve as it passes near the head of the fibula.
For individuals presenting with weakness and/or spasticity a muscle stimulation exercise program can be delivered by the ODFS® Pace. In this case the footswitch trigger is not used and the ODFS® Pace functions as a powered muscle stimulator. A clinician would determine the need for this application which can enabled from a setup option in the ODFS® Pace. (Default is to have the exercise disabled)
#### 5.5 Intended Use
The ODFS® Pace is intended to provide ankle dorsiflexion in individuals who have a dropped foot as a consequence of upper motor neuron injury. By detecting the swing phase of gait through a foot switch signal, appropriate electrical stimulation of the leg and ankle muscles may improve gait by flexing the foot of persons who have lost or impaired function. There may be additional benefits from FES such as muscle re-education,
Odstock Medical Ltd 510(k) Summary, v4, ODFS® PACE
Page 2 of 4, Section 5
{2}------------------------------------------------
prevention/retardation of disuse atrophy, increased or maintained range of joint motion and increase in local blood flow.
The ODFS® Pace is a medical device and should only be used under medical supervision for the treatment of dropped foot following an upper motor neuron injury.
#### 5.6 Technological Characteristics and Substantial Equivalence
The technological characteristics of the ODFS® Pace and predicate devices are tabulated below. All the devices are battery powered and deliver electrical stimulation pulses of similar waveform, pulsewidth and frequency. The ODFS® Pace is a development of the ODFS III v6.2 and therefore has identical indications for use as an FES device. These indications correspond to the FES functions of the EMPI device. The exercise function of the ODFS® Pace corresponds to and is substantially equivalent to the NMS/NMES indications of the EMPI device.
| Feature | ODFS® Pace | ODFS III v6.2 | EMPI Focus 795 |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Waveform | Biphasic (symmetrical<br>and balanced<br>asymmetrical) | Biphasic (symmetrical<br>and balanced<br>asymmetrical) | Biphasic (symmetrical<br>and balanced<br>asymmetrical) |
| Pulsewidth | $0μs - 360μs (±10%)$<br>according to patient<br>selection (3.6 μs steps) | $7μs - 365μs (±10%)$<br>according to patient<br>selection | $300μs$ |
| Frequency | Default is 40Hz.<br>20 - 60 Hz in 5Hz steps<br>available | 40Hz ± 10% | 25, 30, 35, 45,50,80Hz |
| Indications for use | The ODFS® Pace is<br>intended to provide<br>ankle dorsiflexion in<br>individuals who have a<br>dropped foot as a<br>consequence of upper<br>motor neuron injury. By<br>detecting the swing<br>phase of gait through a<br>foot switch signal,<br>appropriate electrical<br>stimulation of the leg<br>and ankle muscles may<br>improve gait by flexing<br>the foot of persons who<br>have lost or impaired<br>function. There may be<br>additional benefits from<br>FES such as muscle re-<br>education,<br>prevention/retardation of<br>disuse atrophy,<br>increased or maintained<br>range of joint motion<br>and increase in local | To provide ankle<br>dorsiflexion in<br>individuals who have a<br>dropped foot as a<br>consequence of upper<br>motor neuron injury. By<br>detecting the swing<br>phase of gait through a<br>foot switch signal,<br>appropriate electrical<br>stimulation of the leg<br>and ankle muscles may<br>improve gait by flexing<br>the foot of persons who<br>have lost or impaired<br>function. There may be<br>additional benefits from<br>FES such as muscle re-<br>education,<br>prevention/retardation of<br>disuse atrophy,<br>increased or maintained<br>range of joint motion<br>and increase in local<br>blood flow. | TENS device<br>Symptomatic relief and<br>management of chronic<br>intractable pain.<br>Adjunctive treatment for<br>post-surgical and post<br>trauma acute pain<br>As a NMS/NMES<br>device: relaxation of<br>muscle spasm;<br>prevention or retardation<br>of disuse atrophy;<br>increasing local blood<br>circulation; muscle<br>reduction; immediate<br>post surgical stimulation<br>of the calf muscles to<br>prevent venous<br>thrombosis; maintaining<br>or increasing range of<br>motion.<br>As a FES device:<br>Stimulating muscles in |
# 5.6 Technological Characteristics and Substantial Equivalence - Table
Odstock Medical Ltd
{3}------------------------------------------------
# 510(k) Summary for the ODFS® PACE Dropped Foot Stimulator
| | blood flow.<br>As a NMS/NMES<br>device: relaxation of<br>muscle spasm;<br>prevention or retardation<br>of disuse atrophy;<br>increasing local blood<br>circulation; muscle re-<br>education; maintaining<br>or increasing range of<br>motion. | | the leg and ankle of<br>partially paralysed<br>patients to provide<br>flexion of the foot and<br>thus improve the<br>patients gait. |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Accessories | Heel switch | Heel switch | Heel switch, hand<br>switch |
| Electrode size & shape | 5cm x 5cm | Scm x 5cm | 5.1cm4 |
# 5.7 Performance testing
The ODFS® Pace system has been subjected to testing to verify that the device meets its functional and output specifications. Production units pass functional testing of stimulator controls, settings and indicators, stimulus output and timing, footswitch operation, and power consumption.
Parameters measured under the conditions specified in FDA Guidance Document for Powered Muscle Stimulator 510(k)s are included in the ODFS® Pace 510(k) submission.
In addition to functional and operational performance testing the ODFS® Pace passed the applicable requirements of the following standards:
- ISO 60601-1-1:2006 Medical Electrical Equipment General requirements for . basic safety and essential performance.
- I ISO 60601-1-2:2007 Medical Electrical Equipment – General requirements for basic safety and essential performance. Collateral standard: Electromagnetic compatibility - Requirements and tests
- . ISO 60601-2-10:2001 Medical electrical equipment. Particular requirements for the safety of nerve and muscle stimulators
Based on the information provided above the ODFS® Pace is substantially equivalent to legally marketed predicate devices.
| | --X-- | |
|--|-------|--|
|--|-------|--|
| Version | Comment | Author | |
|---------|------------------------------------------------------------------------------|-------------------|--|
| 3->4 | Amend indications for use and add regulation number for classification name. | 1 S.Crook 14/3/11 | |
| | | | |
| | | | |
Odstock Medical Ltd 510(k) Summary, v4, ODFS® PACE
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized caduceus, a symbol often associated with medicine and healthcare. The symbol consists of three curved lines that converge at the bottom.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Odstock Medical, Ltd. c/o Dr. Steven Crook Quality and Regulatory Manager Salisbury District Hopital Salisbury, Wilitshire United Kingdom, SP2 8BJ
MAR 3 0 2011
Re: K102115
Trade/Device Name: ODFS® Pace Regulation Number: 21 CFR 882.5810 Regulation Name: External Functional Neuromuscular Stimulator Regulatory Class: Class II Product Code: GZI Dated: January 21, 2011 Received: January 26, 2011
Dear Dr. Crook:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
signature
Malvina B. Eydelman, M.D. Director Division of Ophthalmic. Neurological. and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
K102115 510(k) Number (if known)
Device Name: ODFS® Pace
Indications for use: The ODFS® Pace is intended to provide ankle dorsiflexion in individuals who have a dropped foot as a consequence of upper motor neuron injury. By detecting the swing phase of gait through a foot switch signal, appropriate electrical stimulation of the leg and ankle muscles may improve gait by flexing the foot of persons who have lost or impaired function. There may be additional benefits from FES such as muscle re-education, prevention/retardation of disuse atrophy, increased or maintained range of joint motion and increase in local blood flow.
The ODFS® Pace is a medical device and should only be used under medical supervision for the treatment of dropped foot following an upper motor neuron injury.
Prescription use X (Part CFR 801 Subpart D) and/or
Over-The-Counter Use (CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
of CDRH, Office of Device Evaluation, (ODE)
(Division Sign-O Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K102115
Odstock Medical Ltd Indications statement v4, ODFS® PACE
[section 4]
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.