ENDOSKELETON TO

K102067 · Titan Spine, LLC · MAX · Nov 5, 2010 · Orthopedic

Device Facts

Record IDK102067
Device NameENDOSKELETON TO
ApplicantTitan Spine, LLC
Product CodeMAX · Orthopedic
Decision DateNov 5, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Endoskeleton TO® Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. The device must be used with supplemental fixation. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). It is indicated to be used with autograft bone.

Device Story

Endoskeleton TO is an intervertebral body fusion device designed to provide mechanical stability in the lumbar disc space to facilitate biologic fusion. It is implanted by a surgeon during spinal fusion procedures. The device is manufactured in various sizes to accommodate patient anatomy and pathology. It functions as a structural spacer to maintain disc height and support the vertebral column while bone graft (autograft) facilitates fusion. It requires supplemental fixation to ensure stability. The device is intended for use in patients with degenerative disc disease who have failed conservative treatment.

Clinical Evidence

Bench testing only. Testing included static and dynamic axial compression, static and dynamic compression shear, and static torsion per ASTM F2077-03. Subsidence testing performed per ASTM F2267-04. Expulsion testing conducted per recognized protocol. No clinical data provided.

Technological Characteristics

Constructed from medical grade titanium alloy (Ti6Al4V-ELI). Device is an intervertebral body fusion cage available in various sizes. Mechanical testing performed per ASTM F2077-03 and ASTM F2267-04.

Indications for Use

Indicated for skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1, including those with up to Grade I spondylolisthesis. Requires 6 months of prior non-operative treatment. Must be used with supplemental fixation and autograft bone.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K102067 # 510(k) SUMMARY Titan Spine's Endoskeleton TO NOV-5 2010 | Date: | May 28, 2010 | | |-----------------|--------------------------------------------------------------|-----------------------------------------------------------------------------------------------| | Contact: | Kevin Gemas<br>President<br>866-822-7800 | Titan Spine, LLC<br>Mequon Research Center<br>6140 Executive Dr., Suite A<br>Mequon, WI 53092 | | Trade Name: | Endoskeleton TO® Interbody Fusion Device | | | Product Class: | Class II | | | Classification: | 21 CFR §888.3080 Orthosis, intervertebral body fusion device | | | Product Codes: | MAX | | | Panel Code: | 87 | | Name of Device and Name/Address of Sponsor Endoskeleton TO Titan Spine, LLC Mequon Research Center 6140 W. Executive Drive, Suite A Mequon, WI 53092 ## Common or Usual Name Intervertebral body fusion device #### Predicate Devices The Endoskeleton TO was shown to be substantially equivalent to legally marketed predicate devices. The predicate devices include the Ray Threaded Fusion Cage (P950019), the Lumbar I/F Cage (P960025), the Endoskeleton TT Interbody Fusion Device (K083714) and the BAK Proximity (P950002-S3). # Intended Use / Indications for Use The Endoskeleton TO® Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. The device must be used with supplemental fixation. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). It is indicated to be used with autograft bone. 1 of 2 {1}------------------------------------------------ # Technological Characteristics The Endoskeleton TO is comprised of a variety of implant sizes to accommodate various patient's anatomy and pathology, and associated instrumentation. All implantable components are manufactured from medical grade titanium alloy (Ti6Al4V-ELI). # Performance Data Static and dynamic axial compression, static and dynamic compression shear, and static torsion were completed following ASTM F2077-03. Subsidence was tested following ASTM F2267-04. Expulsion testing was conducted following a recognized protocol to allow comparison evaluation of intervertebral body fusion device assemblies, and characterize their resistance to expulsion. The above pre-clinical testing performed on the Endoskeleton TO Interbody Fusion Device indicated that the Endoskeleton TO Interbody Fusion Device is substantially equivalent to the predicate devices and is adequate for the intended use. # Summary: The Endoskeleton TO Interbody Fusion Device and predicate devices have the same intended use, to provide mechanical stability in the lumbar disc space to facilitate biologic fusion. The indications for use of the Endoskeleton TO Interbody Fusion Device contain no new language that is not already included in at least one of the predicate devices. Moreover, the device is very similar in its size to the predicate devices. The materials used are also the same as in some predicate devices. There are no significant differences in technological characteristics compared to the predicates, and the minor differences that do exist do not raise any new types of safety or efficacy issues. Furthermore, bench testing demonstrates that these differences do not adversely impact device performance, as discussed below. # 2 of 2 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services, USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" arranged around the perimeter. Inside the circle is a stylized emblem resembling an abstract eagle or bird-like figure, composed of three curved lines that suggest wings or feathers. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Titan Spine, LLC % Silver Pine Consulting, LLC Rich Jansen, Pharm.D. 13540 Guild Avenue Apple Valley, Minnesota 55124 NOV - 5 2010 Re: K102067 Trade/Device Name: Endoskeleton® TO Interbody Fusion Device Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: July 22, 2010 Received: August 11, 2010 Dear Dr. Jansen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Or te and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual revistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your davine can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements not hot hact {3}------------------------------------------------ Page 2 - Rich Jansen, Pharm.D. or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. Mark A. Millan Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use Statement $$\mathtt{510}(\mathtt{k})$$ $$\mathtt{Number (if $$\mathtt{k}\mathtt{noown}$$ ): $$\mathtt{510}\mathtt{?00}\mathtt{?}$$ $$ NOV - 5 2010 Device Name: Endoskeleton TO Indications for Use: The Endoskeleton TO® Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. The device must be used with supplemental fixation. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). It is indicated to be used with autograft bone. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Surgical, Orthopedic. and Restorative Devices K102067 510(k) Number_
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...