ELECSYS DIGOXIN CALCHECK 5

K102044 · Roche Diagnostics · JJX · Aug 31, 2010 · Clinical Chemistry

Device Facts

Record IDK102044
Device NameELECSYS DIGOXIN CALCHECK 5
ApplicantRoche Diagnostics
Product CodeJJX · Clinical Chemistry
Decision DateAug 31, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Intended Use

The Elecsys Digoxin CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys Digoxin reagent on the indicated Elecsys and cobas e immunoassay analyzers. For prescription use only.

Device Story

Elecsys Digoxin CalCheck 5 is an in vitro diagnostic control set containing 5 levels of Digoxin in a bovine serum matrix; used for calibration verification and assay range verification on Roche immunoassay analyzers. Device provides 5 concentration levels (0 to 5 ng/mL) to verify instrument performance across the assay range. Healthcare providers use the output to confirm that the analyzer's calibration remains valid and that the assay is performing within established specifications. This ensures the accuracy of patient Digoxin test results, which is critical for therapeutic drug monitoring and preventing toxicity.

Clinical Evidence

No clinical data. Bench testing only. Performance was established through value assignment using a master calibration curve on the Elecsys 2010 analyzer, verified across multiple MODULAR ANALYTICS E170 / cobas e 601 analyzers. Stability studies confirmed 18-month shelf life at 2-8°C and 6-hour open-vial stability at 20-25°C.

Technological Characteristics

Liquid product; bovine serum matrix; five concentration levels; stored at 2-8°C; opened stability 6 hours at 20-25°C. Analyte: Digoxin. Used on Elecsys and cobas e immunoassay analyzers.

Indications for Use

Indicated for use as an assayed control for calibration verification and assay range verification of the Elecsys Digoxin reagent on Roche Elecsys 2010 / cobas e 411 and MODULAR ANALYTICS E170 / cobas e 601 immunoassay analyzers.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary K102044 | Introduction | According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. | | |-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------| | Submitter<br>name, address, contact | Roche Diagnostics<br>9115 Hague Road, P.O. Box 50416<br>Indianapolis, IN 46250-0416<br>317-521-3501 | AUG 3 1 2010 | | | Contact Person: Kelly Colleen O'Maine Adams<br>Phone: 317-521-3577<br>Fax: 317-521-2324<br>Email: colleen.adams@roche.com | | | | Secondary Contact: Stephanie Greeman<br>Phone: 317-521-2458<br>Fax: 317-521-2324<br>Email: stephanie.greeman@roche.com | | | | Date Prepared: July 20, 2010 | | | Device Name | Proprietary name: Elecsys Digoxin CalCheck 5<br>Common name: Digoxin CalCheck 5<br>Classification name: Single (specified) analyte controls (assayed and unassayed) | | | Predicate device | The Elecsys Digoxin CalCheck 5 is substantially equivalent to other products in commercial distribution intended for similar use. We claim equivalency to the currently marketed Elecsys Digoxin CalCheck (K973973). | | | Device Description | The Elecsys Digoxin CalCheck 5 is a liquid product consisting of Digoxin in a buffer/protein (bovine serum) matrix. During manufacture, the analyte is spiked into the matrix at the desired concentration levels. | | | Intended use | The Elecsys Digoxin CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys Digoxin reagent on the indicated Elecsys and cobas e immunoassay analyzers. | | Continued on next page {1}------------------------------------------------ ## 510(k) Summary, Continued #### The table below compares Elecsys Digoxin CalCheck 5 with the predicate Comparison Table device, Elecsys Digoxin CalCheck (K973973). | Characteristic | Elecsys Digoxin CalCheck 5<br>(Candidate Device) | Elecsys Digoxin CalCheck<br>(K973973) | |----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | The Elecsys Digoxin Calcheck 5 is an<br>assayed control for use in calibration<br>verification and for use in the<br>verification of the assay range<br>established by the Elecsys Digoxin<br>reagent on the indicated Elecsys and<br>cobas e immunoassay analyzers. | For use in the verification of the<br>calibration established by the Elecsys<br>Digoxin reagent on the indicated<br>Elecsys and cobas e immunoassay<br>analyzers. | | Analyte | Digoxin | Digoxin | | Levels | Five | Three | | Format | Liquid | Liquid | | Handling | Mix gently by inversion to ensure<br>homogeneity. | Mix gently by inversion to ensure<br>homogeneity. | | Stability | Unopened:<br>• Store at 2-8°C until expiration date<br>Opened:<br>• 20-25°C: 6 hours | Unopened:<br>• Store at 2-8°C until expiration date<br>Opened:<br>• 15-25°C: 5 hours | | Matrix | Bovine serum matrix | Bovine serum matrix | Performance Characteristics The Elecsys Digoxin CalCheck 5 was evaluated for value assignment and stability. · のお気になる。 {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human form or a caduceus, rendered in black. The symbol is composed of three curved lines that converge at the bottom, creating a sense of movement or flow. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002 Roche Professional Diagnostics Ms. Kelly Colleen O'Maine Adams 9115 Hague Road Indianapolis, IN 46250-0416 Re: k102044 AUG 3 1 2010 Trade/Device Name: Elecsys Digoxin CalCheck 5 Regulation Number: 21 CFR 862.1660 Regulation Name: Quality Control Material (Assayed and Unassayed) Regulatory Class: Class I, reserved Product Code: JJX Dated: 20 July, 2010 Received: 21 July, 2010 Dear Ms. Adams: We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section 310(x) premained is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Americal Coloral Food. Days commerce proc to May 20, 1970, the enaoutines with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval applicians of the Act . The and Cosment Act (Act) market the device, subject to the general controls provisions of the Act. The You may, merelore, market the device, becares for annual registration, listing of general controls provisions of the Act include requirements for annual registration and general controls provisions of the fiel labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such If your device is classified (see above) mis one stice can be found in Title 21, additional controls. Existing major regulations asonaly of the may publish further Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further Couc of Pedoral Regaration your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issualies of a subscribes with other requirements of the Act that FDA has made a determination that your device areasies. You must that FDA has made a delemination inal your devices by other Federal agencies. You must or any Federal statules and regulations and including, but not limited to: registration and listing (2 l earing of comply with all the Act S requirements, medical device reporting (reporting of CFR Part 807); labeling (21 CFR Parts (11 am 800); and anonvifiesturing practice CFR Part 807); labeling (21 CFR 803); and good manufacturing practice medical device-related adverse evelis) (21 CER OS), and giove and requirements as set form in the quality systems (10) rour Section 510(k) premarket will allow you to begin marketing your device as described in your Section 510 a legally ma will allow you to begin marketing your device of your device to a legally marketed notification. The FDA finding of substantial equivalence of thus parmits your device to notitication. The FDA Inding of substantial equivalise of this, permits your device to proceed to the market. {3}------------------------------------------------ #### Page 2 If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, AC Courtney C. Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indication for Use 510(k) Number (if known): K1020 Device Name: Elecsys Digoxin CalCheck 5 Indication For Use: The Elecsys Digoxin CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys Digoxin reagent on the indicated Elecsys and cobas e immunoassay analyzers. Prescription Use _ X (21 CFR Part 801 Subpart D) And/Or Over the Counter Use (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Signature Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) k102044 Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...