← Product Code [GEH](/productcode/GEH) · K102006

# CRYOMEGA (K102006)

_Stc Consulting, LLC · GEH · Oct 6, 2010 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K102006

## Device Facts

- **Applicant:** Stc Consulting, LLC
- **Product Code:** [GEH](/productcode/GEH.md)
- **Decision Date:** Oct 6, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4350
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The CryOmega is a disposable device intended for the surgical destruction of target tissue by applying cryogenic gases at extreme low temperatures. The list below shows examples of the type of lesions that can be treated. -Genital Lesions -Molluscum Contagiosum -Seborrheic Keratoses -Skin Tags -Verruca Plantaris -Verruca Vulgaris -Verruca Plana -Actinic Keratosis -Lentigo

## Device Story

CryOmega is a self-contained, disposable cryosurgical device; utilizes 16g nitrous oxide (N2O) cartridge; delivers continuous stream of liquid N2O at -89°C to target tissue via topical spray; operated by physician or licensed practitioner; manual on/off actuator lever controls spray; device discarded after gas depletion; no maintenance or servicing required; intended for destruction of benign skin lesions; provides safe, controlled tissue destruction; eliminates risks associated with cartridge replacement or high-pressure gas handling.

## Clinical Evidence

No clinical data provided. Substantial equivalence supported by bench testing and literature verification of N2O efficacy for tissue destruction.

## Technological Characteristics

Disposable cryosurgical unit; 16g N2O cartridge; plastic housing; butadiene-rubber O-rings; manual actuator lever; N2O delivery at -89°C; complies with ASTM F882-84 (96) for Cryosurgical Medical Instruments; non-electronic; no software.

## Regulatory Identification

(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.

## Predicate Devices

- Cryoprobe-C ([K024009](/device/K024009.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K102006

# CryOmega

# PREMARKET NOTIFICATION SUMMARY

#### I. Device Name:

Proprietary Name - CryOmega Usual Name - Cryosurgical unit and accessories Classification Name - Class II, GEH

QCT 6 2010

#### II. Establishment Registration Number:

This 510(k) is submitted by STC Consulting LLC. The product will be manufactured by CryoConcepts LLP, a company owned by STC Consulting LLC.

### III. Device Classification:

The CryOmcga is a Class II device according to Cryosurgical unit and accessories (21 CFR 878.4350) and has been assigned a product code of GEH.

#### IV. Performance Standards:

To date, no performance standards have been finalized which affect the CryOmega device.

. .

### V. Proposed Labels, Labeling, Advertisements and Enginecring Drawings:

This document contains the proposed Instruction Manual for the CryOmega device

#### VI. Statement of Similarities and Differences to Legally Marketed Devices:

A summary of the substantial equivalence ("SE") analysis for the CryOmega versus the Cryoprobe-C is provided in the following table.

| 510k Predicate<br>Characteristic                     | Cryoprobe-C<br>Predicate<br>K024009                                                                | CryOmega-<br>Substantial<br>Equivalence                                                                                                                |
|------------------------------------------------------|----------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cryoprobe-C<br>K024009<br>-Intended Use<br>Statement | Intended use: To<br>destroy tissue<br>during surgical<br>procedures by<br>applying extreme<br>cold | The CryOmega is a<br>disposable device<br>intended for the<br>surgical<br>destruction of<br>target tissue by<br>applying cryogenic<br>gases at extreme |
|                                                      |                                                                                                    |                                                                                                                                                        |
|                                                      |                                                                                                    | low temperatures.                                                                                                                                      |
|                                                      |                                                                                                    | The list below                                                                                                                                         |
|                                                      |                                                                                                    | shows examples of                                                                                                                                      |
|                                                      |                                                                                                    | the type of                                                                                                                                            |
|                                                      |                                                                                                    | lesions that can                                                                                                                                       |
|                                                      |                                                                                                    | be treated.                                                                                                                                            |
|                                                      |                                                                                                    |                                                                                                                                                        |
|                                                      |                                                                                                    | -Genital Lesions                                                                                                                                       |
|                                                      |                                                                                                    | -Molluscum                                                                                                                                             |
|                                                      |                                                                                                    | Contagiosum                                                                                                                                            |
|                                                      |                                                                                                    | -Seborrheic                                                                                                                                            |
|                                                      |                                                                                                    | Keratoses                                                                                                                                              |
|                                                      |                                                                                                    | -Skin Tags                                                                                                                                             |
|                                                      |                                                                                                    | -Verruca Plantaris                                                                                                                                     |
|                                                      |                                                                                                    | -Verruca Vulgaris                                                                                                                                      |
|                                                      |                                                                                                    | -Verruca Plana                                                                                                                                         |
|                                                      |                                                                                                    | -Actinic Keratosis                                                                                                                                     |
|                                                      |                                                                                                    | -Lentigo                                                                                                                                               |
|                                                      |                                                                                                    |                                                                                                                                                        |
|                                                      |                                                                                                    |                                                                                                                                                        |
| Cryoprobe-C                                          | Nitrous Oxide, N20                                                                                 | Nitrous Oxide, N20                                                                                                                                     |
| K024009                                              | at 50 bar pressure.                                                                                | at 50 bar                                                                                                                                              |
| -Cryogen                                             | 8g or 16g                                                                                          | pressure. 16g                                                                                                                                          |
| Characteristics                                      | cartridges                                                                                         | cartridge                                                                                                                                              |
|                                                      |                                                                                                    |                                                                                                                                                        |
| Cryoprobe-C                                          | Housing: Aluminum                                                                                  | Housing: Plastic                                                                                                                                       |
| K024009                                              | Micro-Applicator:                                                                                  | Filter: Plastic                                                                                                                                        |
| -Materials                                           | unknown                                                                                            | O-rings:                                                                                                                                               |
|                                                      | Lock Cap: Plastic                                                                                  | Butadiene-rubber                                                                                                                                       |
|                                                      | Filter: Unknown                                                                                    | Cartridge: Metal                                                                                                                                       |
|                                                      | O-rings: Unknown                                                                                   |                                                                                                                                                        |
|                                                      | Cartridge: Metal                                                                                   |                                                                                                                                                        |
| Cryoprobe-C                                          | Apply Spray                                                                                        | Apply Spray                                                                                                                                            |
| K024009                                              | Topically                                                                                          | Topically                                                                                                                                              |
| -Mode of Use                                         |                                                                                                    |                                                                                                                                                        |
|                                                      |                                                                                                    |                                                                                                                                                        |
| Cryoprobe-C                                          | N20 qas is                                                                                         | N20 gas is                                                                                                                                             |
| K024009                                              | delivered to the                                                                                   | delivered to the                                                                                                                                       |
| -Mechanism of                                        | treatment site at -                                                                                | treatment site at                                                                                                                                      |
| action                                               | 89C to effect                                                                                      | -89C to effect                                                                                                                                         |
|                                                      | cellular                                                                                           | cellular                                                                                                                                               |
|                                                      | destruction                                                                                        | destruction                                                                                                                                            |
| Cryoprobe-C                                          | <50°C                                                                                              | <50°C                                                                                                                                                  |
| K024009                                              |                                                                                                    |                                                                                                                                                        |
| -Storage                                             |                                                                                                    |                                                                                                                                                        |
| Conditions                                           |                                                                                                    |                                                                                                                                                        |
|                                                      | Complies with ASTM:                                                                                | Complies with                                                                                                                                          |
| Cryoprobe-C                                          | F882-84 (96) for                                                                                   | ( 96)<br>ASTM: F882-84                                                                                                                                 |
| K024009                                              |                                                                                                    |                                                                                                                                                        |
| -Safety                                              | Cryosurgical<br>Medical Instruments                                                                | for Cryosurgical<br>Medical<br>Instruments                                                                                                             |
| Cryoprobe-C<br>K024009<br>-Gas Cartridge<br>Safety   | Cartridge can expel<br>unused gas under<br>pressure during<br>cartridge<br>replacement             | Unit is discarded<br>after liquefied<br>gas is emptied.                                                                                                |
| Cryoprobe-C<br>K024009<br>-Treatment<br>Procedure    | Suggests Freezing<br>of target tissue by<br>spray                                                  | Same                                                                                                                                                   |
| Cryoprobe-C<br>K024009<br>-Operation                 | Spray begins when<br>gas cartridge is<br>engaged. Continuous<br>spray unless capped                | Gas dispensed<br>using actuator<br>lever. Spray<br>controlled by<br>on/off actuator                                                                    |
| Cryoprobe-C<br>K024009<br>-Disposal                  | Main Unit is<br>reusable with<br>replaceable gas<br>cartridges                                     | Whole Unit is<br>disposable after<br>liquefied gas is<br>emptied from the<br>cartridge.                                                                |
| Cryoprobe-C<br>K024009<br>-Defined Operators         | Physician or<br>Licensed<br>Practitioner                                                           | Same                                                                                                                                                   |
| Cryoprobe-C<br>K024009<br>-Service/Repair            | Return to<br>Manufacturer                                                                          | Disposable once<br>liquefied gas is<br>emptied. No<br>Servicing                                                                                        |

P. i/4

. * . * *

{1}------------------------------------------------

\$\rho_{2/4}\$

{2}------------------------------------------------

It is believed that the data presented in the Testing Section of this submission is sufficient to demonstrate substantial equivalence between the predicate and CryOmega Device.

#### Intended Use

The CryOmega is a disposable device intended for the surgical destruction of target tissue by applying cryogenic gases at extreme low temperatures. The list below shows examples of the type of lesions that can be treated.

3/6

- -Genital Lesions -Molluscum Contagiosum -Seborrheic Keratoses -Skin Tags -Verruca Plantaris -Verruca Vulgaris -Verruca Plana

{3}------------------------------------------------

-Actinic Keratosis -Lentigo

#### Technological Characteristics

The CryOmega device is designed to dispense a continuous stream of liquid nitrous oxide when actuated. The device contains a 16g cartridge of nitrous oxide that is dispensed once the device is activated. Physicians or medical professionals can then dispense the gas for use in procedures requiring the destruction of tissue using the extreme cold of nitrous oxide (-89°C). Unlike the predicate device, the CryOmega is designed to be self contained and disposable after all of the liquefied gas has been dispensed. It is believed that this feature increases the safety of the device since operators cannot replace or access the high pressure gas cartridge. Additionally because the device is disposable there is no servicing or maintenance of the CryOmega device.

### CONCLUSION

In summary, the CryOmega's path through the 510(k) flowchart ` is as follows:

- The CryOmega has the same intended use as the predicate 피 cryogenic device that is intended for the destruction of tissue using nitrous oxide.
- The CryOmega has the same technological characteristics as . the Cryoprobe-C Device which sprays liquid nitrous oxide gas.
- Both the CryOmega and Predicate using valves to dispense . nitrous oxide gas. Therefore technological differences between the CryOmega and its predicate devices do not raise new questions of safety or efficacy.
- The 510(k) notice includes data and literature to verify that the 다 nitrous oxide in the CryOmega device can safely and effectively cause destruction when used as instructed.

Therefore, STC Consulting believes that FDA can find the CryOmega to be substantially equivalent to the Cryoprobe-C device.

'FDA Guidance-Format for Traditional & Abbreviated 510k's 2005

P4/4

{4}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized lines.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

STC Consulting LLC % R. Sam Niedbala, Ph.D. 4093 Maulfair Drive Allentown, Pennsylvania 18103

OCT 6 2010

Re: K102006

Trade/Device Name: CryOmega™ Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: Class II Product Code: GEH Dated: September 03, 2010 Received: September 13, 2010

Dear Dr. Niedbala:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you: however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{5}------------------------------------------------

Page 2 - R. Sam Niedbala, Ph.D.

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Peter Der Dik

Mark N. Melkers Director Division of Surgical. Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# Indications for Use

510(k) Number (if known): K102006

Device Name: CryOmega ™

Indications For Use:

The CryOmega is a disposable device intended for the surgical destruction of target tissue by applying cryogenic gases at extreme low temperatures. The list below shows examples of the type of lesions that can be treated.

-Genital Lesions -Molluscum Contagiosum -Seborrheic Keratoses -Skin Tags -Verruca Plantaris -Verruca Vulgaris -Verruca Plana -Actinic Keratosis -Lentigo

AND/OR Prescription Use Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE QNTINUE ON ANOTHER - Kis PAGE IF (Division Sign-Off) NEEDED) Division of Surgical. Orthopede, and Restorative Devices 2006

Concurrence of CDRH, Office of Device

---

**Source:** [https://fda.innolitics.com/device/K102006](https://fda.innolitics.com/device/K102006)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
