The Global UNITE humeral stems, suture collars, epiphyseal components and humeral heads are intended for total shoulder or hemi-shoulder replacement.
Device Story
Global UNITE Shoulder System is a modular orthopedic implant for total or hemi-shoulder arthroplasty. System features two-piece humeral stem design allowing conversion from primary fracture prosthesis to reverse prosthesis without removing well-fixed distal stem. Components include humeral stems, suture collars, epiphyseal components, and humeral heads. Used in surgical settings by orthopedic surgeons. Device provides structural replacement for damaged or diseased shoulder joints; modularity allows intraoperative flexibility for fracture management or revision scenarios. Benefits include restoration of joint function and pain relief. System integrates with existing DePuy glenoids for total shoulder replacement and Delta Xtend components for reverse conversion.
Clinical Evidence
Bench testing only. Testing included collar taper axial/torque disassembly strength, collar taper fatigue, humeral head taper axial disassembly strength, and fatigue testing for both anatomic and reverse configurations per ASTM F2009-00 and ASTM 1378-05.
Technological Characteristics
Modular two-piece humeral stem system. Materials and porous coating are similar to predicate devices. Features include suture holes on stem and collar for tuberosity reattachment, male taper on epiphyseal component, and modular epiphyseal design. Testing performed per ASTM F2009-00 (taper disassembly) and ASTM 1378-05 (fatigue).
Indications for Use
Indicated for patients with severely painful/disabled joints due to osteoarthritis, traumatic arthritis, or rheumatoid arthritis; proximal humerus fracture-dislocations with severe comminution or vascular compromise; or difficult clinical problems like failed primary component revision. Hemi-shoulder replacement indicated for ununited humeral head fractures, avascular necrosis, or deformity/limited motion. Conversion to reverse shoulder indicated for grossly deficient rotator cuff with severe arthropathy or failed joint replacement, provided the patient has a functional deltoid muscle and anatomically suitable joint.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
{0}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the logo is a stylized image of a caduceus, a symbol of medicine and healing, with a staff entwined by two snakes and topped by wings.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 2, 2016
Depuy (Ireland) Ms. Rhonda Myer Senior Regulatory Affairs Associate Loghbeg Ringaskiddy Co. Cork, IRELAND
Re: K101996 Trade/Device Name: DePuy Global UNITE Shoulder System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, MBF, KWS, HSD Dated: November 23, 2010 Received: November 24, 2010
Dear Ms. Myer:
This letter corrects our substantially equivalent letter of December 15, 2010.
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviewed your be determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use sured in the encreations of the enactment date of the Medical Device Amendments of to commore of they 20, 1970, are excerdance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Oosmetic I so ( roo ) 100 ) 100 the general controls provisions of the Act. The I ou may, morelors, manns of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be may or subject to additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{1}------------------------------------------------
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Lori A. Wiggins -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Section 4: Indications for Use Statement
# 510 (k) Number (if known): K101996
# Device Name: DePuy Global UNITE Shoulder System
# Indications for Use:
The Global UNITE humeral stems, suture collars, epiphyseal components and humeral heads are intended for cemented or uncemented total shoulder or hemi-shoulder replacement in treatment of the following:
- 1. A severely painful and/or disabled joint resulting from osteoarthritis, traumatic arthritis or rheumatoid arthritis
- 2. Fracture-dislocations of the proximal humerus where the articular surface is severely comminuted, separated from its blood supply or where the surgeon's experience indicates that alternative methods of treatment are unsatisfactory
- 3. Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. revision of a failed primary component)
Hemi-shoulder replacement is also indicated for:
- 1. Ununited humeral head fractures
- 2. Avascular necrosis of the humeral head
- 3. Deformity and/or limited motion
When used in a total shoulder replacement, the Global UNITE implants are to be used with DePuy glenoids. The glenoids are for cemented use only. ",
When well-fixed, the Global UNITE humeral stems, in conjunction with existing Delta Xtend epiphyseal components, are also indicated for conversion to a reverse, in treatment of a grossly deficient rotator cuff joint with severe arthropathy or a previously failed joint replacement with a grossly deficient rotator cuff joint. The patient's joint must be anatomically and structurally suited to receive the selected implant(s), and a functional deltoid muscle is necessary. The Delta Xtend metaglene component is HA-coated and is intended for uncemented use with the addition of screws for fixation. The Delta Xtend epiphyseal components are HA-coated and are intended for uncemented use.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use. (21 CFR 807 Subpart C)
. :
K101996
for M. Melkerson
(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices
510(k) Number K
{3}------------------------------------------------
K101996
p 1/3
:・・・
| | Section 5: 510 (k) Summary<br>(As required by 21 CFR 807.92 and 21 CFR 807.93) |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | DEC 1 5 2010 |
| NAME OF SPONSOR: | DePuy (Ireland)<br>Loughbeg<br>Ringaskiddy<br>Co. Cork Ireland<br>Establishment Registration Number: 9616671 |
| 510(K) CONTACT: | Rhonda Myer<br>Senior Regulatory Affairs Associate<br>Telephone: (574) 371-4927<br>Facsimile: (574) 371-4987<br>Electronic Mail: Rmyer7@its.jnj.com |
| DATE PREPARED: | December 15, 2010 |
| PROPRIETARY NAME: | DePuy Global UNITE Shoulder System |
| COMMON NAME: | Total and Hemi Shoulder Arthroplasty Prosthesis |
| CLASSIFICATION AND<br>REGULATION: | Class II per 21 CFR 888.3660: Shoulder joint<br>metal/polymer semi-constrained cemented<br>prosthesis (KWS) |
| | Class II per 21 CFR 888.3690: Shoulder joint<br>humeral (hemi-shoulder) metallic uncemented<br>prosthesis (HSD) |
| | Class II per 21 CFR 888.3670: Shoulder joint<br>metal/polymer/metal non-constrained or semi-<br>constrained porous-coated uncemented prosthesis<br>(MBF) |
| DEVICE PRODUCT CODE AND<br>DESCRIPTION: | KWS: Prosthesis, Shoulder, Semi-Constrained,<br>Metal/Polymer Cemented<br>HSD: Prosthesis, Shoulder, Hemi-, humeral,<br>Metallic Uncemented<br>MBF: Prosthesis. Shoulder, Semi-constrained,<br>Metal/Polymer, Uncemented |
| PREDICATE DEVICES | DePuy Global AP, K063652 (stem) and<br>K060874 (head)<br>DePuy Delta Xtend, K071379 |
{4}------------------------------------------------
……………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………
์
# DEVICE DESCRIPTION:
The Global UNITE Shoulder System includes a two-piece stem design that provides the option to convert a primary fracture prosthesis to a reverse prosthesis, a suture collar, an epiphyseal component, and a humeral head. In the case of failed tuberosity healing, the surgeon will be able to remove the Global UNITE epiphyseal component and replace it with a Delta Xtend Reverse epiphyseal component without removing the well-fixed distal stem.
# INDICATIONS AND INTENDED USE:
# Indications:
The Global UNITE humeral stems, suture collars, epiphyseal components and humeral heads are intended for cemented or uncemented total shoulder or hemi-shoulder replacement in treatment of the following:
- 1. A severely painful and/or disabled joint resulting from osteoarthritis, traumatic arthritis or rheumatoid arthritis
- 2. Fracture-dislocations of the proximal humerus where the articular surface is severely comminuted, separated from its blood supply or where the surgeon's experience indicates that alternative methods of treatment are unsatisfactory
- 3. Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. revision of a failed primary component)
Hemi-shoulder replacement is also indicated for:
- 1. Ununited humeral head fractures
- 2. Avascular necrosis of the humeral head
- 3. Deformity and/or limited motion
When used in a total shoulder replacement, the Global UNITE implants are to be used with DePuy glenoids. The glenoids are for cemented use only.
When well-fixed, the Global UNITE humeral stems, in conjunction with existing Delta Xtend epiphyseal components, are also indicated for conversion to a reverse, in treatment of a grossly deficient roator cuff joint with severe arthropathy or a previously failed joint replacement with a grossly deficient rotator cuff joint. The patient's joint must be anatomically and structurally suited to receive the selected implant(s), and a functional deltoid muscle is necessary. The Delta Xtend metaglene component is HA-coated is intended for uncemented use with the addition of screws for fixation. The Delta Xtend epiphyseal components are HA-coated and are intended for uncemented use.
## Intended Use:
The Global UNITE humeral stems, suture collars, epiphyseal components and humeral heads are intended for total shoulder or hemi-shoulder replacement.
# BASIS OF SUBSTANTIAL EQUIVALENCE:
Based on the similarities in intended use, indications for use, materials, and design, DePuy believes the subject Global UNITE Shoulder System is substantially equivalent to the previously beneves the subject Global AP and Delta Xtend Shoulder Systems, cleared in K063652, K060874 and K071379. :
# Technological Characteristics
{5}------------------------------------------------
K01996
p 3/3
The technological characteristics of the DePuy Global UNITE Shoulder System are similar to the predicate devices, including geometry, design, size range, material, and porous coating. The differences include the following: The predicate Global AP only has suture holes on the stem, and the subject Global UNITE System has suture holes on the stem and also a suture collar for tuberosity reattachment. The predicate Global AP stem has a female taper with linking components that attach to the head while the subject Global UNITE has a male taper on the epinhyseal component that attaches to the head. The predicate Global AP does not have a modular epiphyseal component and the subject Global UNITE does have a modular epiphyseal component. The predicate Global AP heads have more material on the underside of the head than the Global UNITE heads. The predicate Delta Xtend does not have sizes 6 and 8 while the subject Global UNITE does offer sizes 6 and 8.
#### Performance Data
The Global UNITE Shoulder System was subjected to the following pre-clinical testing: Collar taper axial disassembly strength (ASTM F2009-00), Collar taper torque disassembly strength, Collar taper fatigue testing and axial disassembly strength, Humeral head taper axial disassembly strength (ASTM F2009-00), Anatomic configuration fatigue testing (ASTM 1378-05, and Reverse configuration fatigue testing (ASTM 1378-05).
## Conclusion
DePuy believes the DePuy Global UNITE Shoulder System is substantially equivalent to the predicate devices, Global AP and Delta Xtend, as confirmed through device testing and production activities. No new issues of safety and efficacy have been raised with the Global UNITE Shoulder System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.