TAPERED SCREW-VENT T IMPLANT

K101977 · Zimmer Dental, Inc. · DZE · Sep 14, 2010 · Dental

Device Facts

Record IDK101977
Device NameTAPERED SCREW-VENT T IMPLANT
ApplicantZimmer Dental, Inc.
Product CodeDZE · Dental
Decision DateSep 14, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3640
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Tapered Screw-Vent® T Implants are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load.

Device Story

Tapered Screw-Vent® T Implant is an endosseous dental implant composed of titanium alloy. Designed for implantation in maxilla or mandible to replace missing teeth. Features tapered body with triple-lead threads and surface treatment to facilitate osseointegration. Modified from predicate to include MTX surface texture on implant top and small grooves on implant collar. Used by dental professionals in clinical settings. Implant provides structural foundation for prosthetic teeth; supports immediate or delayed loading based on primary stability and occlusal load assessment. Benefits patient by restoring dental function and aesthetics.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Endosseous dental implant; material: titanium alloy; design: tapered body with triple-lead threads; surface: MTX treated; diameters: 3.7mm, 4.1mm, 4.7mm, 6.0mm; connection: implant/abutment interface.

Indications for Use

Indicated for patients requiring replacement of one or more missing teeth in the maxilla or mandible. Suitable for immediate loading given good primary stability and appropriate occlusal load, or for loading after conventional healing.

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a logo for Zimmer Dental. The logo consists of a circle with a stylized letter Z inside it. Below the circle, the word "zimmer" is written in lowercase letters, with the word "dental" written in smaller letters underneath and underlined. Zimmer Dental 1900 Aston Avenue Carlsbad, CA 92008 760.929.4300 (ph) 760.431.7811 (fax) 510k No .: Page No.: ## Traditional 510(k PRE-MARKET NOTIFICATION 510(k) # 510(k) SUMMARY (21CFR807.92(a)) - Submitter's Information: 1. - Zimmer Dental Inc. Name: SFP 1 4 2018 A5-1 Address: 1900 Aston Ave. Carlsbad, CA 92008 760-929-4300 Phone: Contact: Jeremy Markovich Date Prepared: July 12, 2010 - 2. Device Name: Tapered Screw-Vent® T Implant Device Classification Name: Endosseous Dental Implant - Predicate Device(s): Zimmer Dental Tapered Screw-Vent® Implant System 3. NobelReplace Tapered Conical Connection - Device Description: 4. The Tapered Screw-Vent® T Implant is an endosseous dental implant. The implant is composed of titanium alloy. The implant section is designed for ease of implantation and with greater surface area for osseointegration. The implant section surface is treated to facilitate osseointegration. In addition, the implant section is tapered with triplelead threads. - 5. Indications for Use: The Tapered Screw-Vent® T Implants are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing {2}------------------------------------------------ 510(k) No. Page No. _____________________________________________________________________________________________________________________________________________________________________ A5-2 teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. #### . Device Comparison: The Tapered Screw-Vent® T Implant is the same as the predicate Tapered Screw-Vent® Implant in the implant/abutment connection, implant body design, materials, and manufacturing. This device has been modified to add the MTX texture to the top of the implant and add small grooves on the implant collar similar to the predicate NobelReplace Tapered implant. The new device will feature MTX surface equivalent to existing Zimmer Dental implants. The new implant will be offered in 3.7mm, 4.1mm, 4.7mm and 6.0mm diameters. {3}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The caduceus is a common symbol associated with healthcare and medicine. ### Public Health Service Food and Drug Administration 10903 New Hamoshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Mr. Jeremy Markovich Regulatory Affairs Zimmer Dental Incorporated 1900 Aston Avenue Carlsbad, California 92008-7308 ## SEP 1 4 2010 Re: K101977 Trade/Device Name: Tapered Screw-Vent® T Implant Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE Dated: September 3, 2010 Received: September 7, 2010 Dear Mr. Markovich: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register. {4}------------------------------------------------ ### Page 2- Mr. Markovich Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. 1h for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health ### Enclosure {5}------------------------------------------------ K101977 Image /page/4/Picture/1 description: The image shows the logo for Zimmer Dental. The logo consists of a stylized letter "Z" inside a circle, with the word "zimmer" written in a bold, sans-serif font below it. Underneath "zimmer" is the word "dental" in a smaller, less bold font. The logo is black and white. Indications for Use 510(k) Number (if known): Device Name: Tapered Screw-Vent® T Implant Indications For Use: The Tapered Screw-Vent® T Implants are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. × Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Susan Runne (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K101927
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...