OSTEOBRIDGE IKA INTRAMEDULLARY KNEE ARTHRODESIS ROD FIXATION SYSTEM

K101939 · Merete Medical GmbH · HSB · Oct 7, 2010 · Orthopedic

Device Facts

Record IDK101939
Device NameOSTEOBRIDGE IKA INTRAMEDULLARY KNEE ARTHRODESIS ROD FIXATION SYSTEM
ApplicantMerete Medical GmbH
Product CodeHSB · Orthopedic
Decision DateOct 7, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3020
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Intramedullary knee arthrodesis

Device Story

Modular intramedullary rod system for knee arthrodesis; consists of rod segments, 10° angled attachment shell, and interlocking screws. Components implanted into femoral and tibial medullary canals; segments joined via attachment shell and screws to provide rigid fixation. Used in surgical settings by orthopedic surgeons. Provides structural stability for bone fusion; bone graft recommended over attachment shell to promote callus formation. Fixation achieved with interlocking screws, optionally with bone cement. Enables adjustment of extension/flexion and valgus/varus positioning via modular design.

Clinical Evidence

Bench testing only. Mechanical and functional testing performed, including static torsion tests and four-point/three-point bending fatigue tests on the rod system and locking screws. Testing represented worst-case scenarios; all tests passed.

Technological Characteristics

Modular intramedullary rod system; materials: Ti-6Al-4V (ISO 5832-3). Components: nails (10-18mm diameter), 10° angled attachment shell (50mm), 5mm diameter interlocking screws. Energy source: none (mechanical). Sterilization: Gamma radiation. Connectivity: N/A.

Indications for Use

Indicated for patients requiring knee arthrodesis due to irretrievably failed total knee arthroplasty, limb salvage, oncology surgery, or conditions with insufficient soft or bony tissue for support.

Regulatory Classification

Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ ## 2. 510(k) Summary of Safety and Effectiveness Information | Date Prepared: | June 14, 2010 | |-------------------------------|---------------------------------------------------------------------------------------------------------------------------------| | Date Revised: | October 6, 2010 | | Submitted by: | Merete Medical GmbH<br>Alt Lankwitz 102,<br>12247 Berlin Germany | | | OCT 7 2010 | | Contact Person: | Donna Coleman<br>Merete Medical, Inc.<br>49 Purchase Street<br>Rye, New York 10580<br>Phone: 914 967 1532<br>Fax: 914 967 1542 | | FDA Registration Number: | 3002949614 | | Proprietary Name: | OsteoBridge® IKA Intramedullary Knee Arthrodesis<br>Rod Fixation System | | Classification Name: | 21 CFR 888.3020 - Intramedullary fixation rod | | Product Code: | HSB – Rod, Fixation, Intramedullary and Accessories | | Proposed Regulatory Class: | Class II | | Sterilizer (Gamma Radiation): | BBF Sterilisationsservice GmbH<br>Willy-Rüsch-Str. 4-10<br>71394 Kernen-Rommelshausen<br>Certificates are enclosed in Annex III | #### Predicate Device: | OsteoBridge® IDSF Fixation Rod System (K051965) | |---------------------------------------------------| | Femorotibial Medullary Nail Zimmer Inc. (K853250) | | Knee Fusion Nail, Smith & Nephew, Inc. (K050938) | #### Device Description: OsteoBridge® Intramedullary Knee Arthrodesis Rod Fixation System (IKA) is a series of modular intramedullary rod segments that may be used as either proximal or distal segments. The segments are designed to be attached together to form a complete intramedullary rod using a semicircular hollow angled attachment shell that is clamped together with multiple screws to create a firm fixation of the bone. We strongly recommend covering the attachment shell with bone graft to enhance callus formation. all components are manufactured from Ti-6Al-4V conforming to ISO 5832-3. The intramedullary rods can be fixed with interlocking screws without or with bone cement. {2}------------------------------------------------ #### The device consists of: Nails with collar: Dia.: 10, 12, 14 and 16 mm in Lengths: 110, 130 and 150 mm Nails without collar: Dia.: 14, 16 and 18 mm in Lengths: 130, 150 and 200 mm Attachment Shells: 10° Angled: 50 mm length Interlocking Screws: Dia. 5 mm in lengths: from 20 mm to 56 mm in two millimeter intervals #### Intended use: Intramedullary knee arthrodesis Indications include: - 1. Irretrievably failed total knee arthroplasty - 2. Limb salvage - 3. Oncology surgery - 4. Any other condition where there is little soft tissue or bony tissue available for support and arthrodesis is the treatment of choice The intramedullary rods can be fixed with interlocking screws without or with bone cement. #### Summary of Technologies: #### Similarities The OsteoBridge® IKA has the same functional characteristics (modular internal rod fixation system) as the predicate device (K051965 / K853250 / K050938) and is made of the same titanium alloy material. The OsteoBridge® IKA has the same intended use as the predicate device (K050938). #### Differences In comparison to the above mentioned predicate device (K051965), the OsteoBridge® IKA System is designed for fusion of the knee joint. Both devices are anchored in the diaphysis of the bone. The performance is therefore comparable. In comparison to the predicate device (K050938) the OsteoBridge® IKA is a modular system with two nails. In comparison to the predicate device (K853250) the two nails of the OsteoBridge® IKA must be joined using the attachment shell. The OsteoBridge® IKA provides an angled attachment shell to allow the adjustment of extension/flexion and valgus/varus positioning. The nails are modified to meet the needs of the damage following failed total knee arthroplasty. #### Mechanical Testing: In order to demonstrate that the OsteoBridge® IKA has the mechanical properties necessary to perform its intended use, and that the device performs as well as or better than the predicate device, Merete has conducted mechanical and functional testing. This includes: - Static Torsion Test . - Four-Point Bending Fatigue Test - Three-Point Bending Fatigue Test of the IDSF Locking Screws (used also for IKA System) The tests represent the worst case scenario for the strength of the system. All tests were passed. The "worst case" was tested. Page 5 of 64 {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three horizontal lines that curve and converge, resembling a stylized human figure. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Merete Medical GmbH % Mr. Emmanuel Anapliotis President & CEO Alt Lankwitz 102 12247 Berlin, Germany OLT - 7 2010 Re: K101939 Trade/Device Name: OsteoBridge IKA Intramedullary Knee Arthrodesis Rod Fixation System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: II Product Code: HSB Dated: July 7, 2010 Received: July 12, 2010 Dear Mr. Anapliotis: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {4}------------------------------------------------ ### Page 2 - Mr. Emmanuel Anapliotis device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.h1m for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. for Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## 1. Indications for Use of OsteoBridge® IKA # Indications for Use #### 510(k) Number (if known): K101939 ## Device Name: OsteoBridge® IKA Intramedullary Knee Arthrodesis Rod Fixation System Intended Use: Intramedullary knee arthrodesis Indications include: - 1. Irretrievably failed total knee arthroplasty - 2. Limb salvage - 3. Oncology surgery - 4. Any other condition where there is little soft tissue or bony tissue available for support and arthrodesis is the treatment of choice The intramedullary rods can be fixed with interlocking screws without or with bone cement. | Prescription Use | <span></span> | |-----------------------------|---------------| | (Part 21 CFR 801 Subpart D) | <span></span> | AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) | | Concurrence of CDRH, Office of Device Evaluation (ODE) | |--|--------------------------------------------------------| |--|--------------------------------------------------------| (Division Sign-Off) (Division Sign-Sical, Orthopedic, Division or and Restorative Devices 510(k) Number K101939 Page 1 of 1 Merete Medical GmbH Submission Date: June 2010 Revision Date: October 2010 Page 3 of 64
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...