← Product Code [FLL](/productcode/FLL) · K101912

# THE INFRARED FOREHEAD THERMOMETER, MODEL FS-300&301 (K101912)

_Hubdic Co., Ltd. · FLL · Nov 10, 2010 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K101912

## Device Facts

- **Applicant:** Hubdic Co., Ltd.
- **Product Code:** [FLL](/productcode/FLL.md)
- **Decision Date:** Nov 10, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2910
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Pediatric

## Indications for Use

The Infrared Forehead Thermometer, FS-300&301 is intended for intermittent measurement of human body temperature in people of all ages.

## Device Story

The Infrared Forehead Thermometer (FS-300&301) is a handheld electronic device used to measure human body temperature, room temperature, and relative humidity. It utilizes an infrared thermopile sensor to detect natural thermal radiation from the forehead and temple area. During operation, the device scans the forehead while the measurement button is pressed; it performs 5,000 measurements per second to identify and display the highest temperature reading on an LCD screen. The device is intended for use by individuals of all ages. Healthcare providers or users view the output on the integrated display to assess body condition and environmental factors. The device provides a non-invasive method for temperature monitoring.

## Clinical Evidence

Bench testing only. The device was tested to meet safety requirements of IEC 60601-1 (General requirements for safety) and IEC 60601-1-2 (Electromagnetic compatibility). No clinical data provided.

## Technological Characteristics

Infrared thermopile sensor; measures thermal radiation; 5,000 samples per second sampling rate; LCD display; includes room temperature and relative humidity sensors; handheld form factor; battery-powered; complies with IEC 60601-1 and IEC 60601-1-2.

## Regulatory Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

## Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; 
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and 
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

## Predicate Devices

- Sensor touch ([K011291](/device/K011291.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### 510(k) Summary న.

#### [as required by 807.92(c)]

# A .- Applicant :-

-Company name: HuBDIC Co., Ltd

-Address: 195-42, Anyang 7-dong, Manan-gu, Anyang-si, Gyeonggi-do, Korea (zip. 430-815)

> TEL : +82-31-441-8637

► FAX : +82-31-442-4994

► http://www.hubdic.com

-Contact person: Peter Chung 412-687-3976

-Date: June 28th, 2010

## B. Proprietary and Established Names:

Trade Name: The Infrared Forehead Thermometer, FS-300&301 Common Name: Clinical Electronic Thermometer Requiation Name: Clinical Electronic Thermometer Regulation Number:880.2910

NOV 1 0 2010

Regulatory Classification: II Product Code: FLL

C. Predicate device: Sensor touch (K011291)

#### D. Device Description

The Infrared Forehead Thermometer, Modelfype FS-300&301 is an electronic thermometer using an infrared i sensor (thermopile) to measure forehead temperature, then get a reading and display it on the LCD.

Its operation is based on measuring the natural thermal radiation emanating from the forehead and the adjacent surfaces.

This product is an Infrared Forehead Thermometer with measuring room temperature and relative humidity.

#### E. Intended use

The Infrared Forehead Thermometer, FS-3008301 is internittent measurement of human body temperature in people of all ages.

## F. Technological Characteristics:

To increase the accuracy, body temperature is measured 5,000 times in one second instantly and displays highest temperature on the LCD.

Purpose of measuring body temperature is to check body condition by measuring correct temperature in important part of body and both measured room temperature and relative humidity are to show environmental conditions.

In case of Forehead measurement, this product searches the highest temperature from the forehead to the temple. The temperature of the artery between the temple that emits thermal infrared from the inner shin to the exterior body is scanned continuously while the measurement button is pressed down

#### G. Performance (Safety and Effectiveness Information)

The Infrared Forehead Thermometer, FS-3008.301 has been manufactured and tested to meet the safety requirements of IEC. The Infrared Forehead Thermometer, FS-300&301 complies with IEC 60601-1 Medical electrical equipment - Part 1: General requirements for safety and IEC 60601-1-2:2001. Medical electrical equipment - Part 1: General requirements for safety - Collateral standard: Electromagnetic compatibility -Requirements and tests.

### I. Conclusion:

The performance tests demonstrated that The Infrared Forehead Thermometer. FS-3008301 is as safe, as effective and performs in a substantially equivalent manner to the predicate device

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features an abstract image of an eagle with three wavy lines extending from its body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Hubdic Company, Limited C/O Mr. Peter Chung 300 Atwood Pittsburgh, Pennsylvania 15213

NOV 1 0 2010

Re: K101912

Trade/Device Name: The Infrared Forehead Thermometer, FS-300&301 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: October 8. 2010 Received: October 20, 2010

Dear Mr. Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

## Page 2- Mr. Chung

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/Offices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Susan Parmer

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

| 510(k) Number (if known): | K101912                                       |
|---------------------------|-----------------------------------------------|
| Device Name:              | The Infrared Forehead Thermometer, FS-300&301 |

Indications For Use:  
The Infrared Forehead Thermometer, FS-300&301 is intended for intermittent measurement of human body temperature in people of all ages.

| Prescription Use X<br>(Part 21 CFR 801 Subpart D) | AND/OR | Over-The-Counter Use<br>(21 CFR 807 Subpart C) |
|---------------------------------------------------|--------|------------------------------------------------|
|---------------------------------------------------|--------|------------------------------------------------|

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| (Division Sign-Off)                                                               | <div> <img alt="Signature" src="signature.png"/> </div> |
|-----------------------------------------------------------------------------------|---------------------------------------------------------|
|                                                                                   | 11/4/2010                                               |
| Division of Anesthesiology, General Hospital<br>Infection Control, Dental Devices | Page 1 of 1                                             |
| 510(k) Number:                                                                    | K101912                                                 |

NOV 10 2010
1 ?

-

---

**Source:** [https://fda.innolitics.com/device/K101912](https://fda.innolitics.com/device/K101912)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
