K101719 · Dalton Instrument Corp. · IOR · Aug 25, 2010 · Physical Medicine
Device Facts
Record ID
K101719
Device Name
TACAMO MANUAL WHEELCHAIR
Applicant
Dalton Instrument Corp.
Product Code
IOR · Physical Medicine
Decision Date
Aug 25, 2010
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 890.3850
Device Class
Class 1
Intended Use
The intended use of the TACAMO Manual Wheelchair is to provide mobility to persons limited to a sitting position.
Device Story
TACAMO is a rigid, non-folding, manually operated mechanical wheelchair. Input: user or attendant physical force for propulsion and steering. Operation: frame constructed of 1.25" and 1" OD steel tubing; 24" rear wheels with hand rims; 8" front casters. Features a manually operated 'Tilt in Space' mechanism for patient stability, comfort, and positioning; activated by rear release levers that disengage gear rack segments from notched actuator rods via flexible cables. Provides 30 degrees of tilt and 30 degrees of recline. Used in clinics, nursing homes, or home settings. Output: physical mobility and postural adjustment. Benefits: provides independent or assisted mobility and facilitates patient care tasks (e.g., feeding) through tilt/recline functionality.
Clinical Evidence
Bench testing only. Device performance verified against ANSI/RESNA WC/Vol.1-1998 (Requirements and Test Methods for Wheelchairs) and DIS ISO 7176 (1992) Part 8 (Static, Impact, and Fatigue Strengths).
Technological Characteristics
Rigid, non-folding manual wheelchair. Materials: 1.25" and 1" OD steel tubing. Mechanical actuation: manual 'Tilt in Space' via release levers, flexible cables, gear rack segments, and notched actuator rods. Dimensions: 24" rear wheels, 8" front casters. Standards: ANSI/RESNA WC/Vol.1-1998, DIS ISO 7176 Part 8.
Indications for Use
Indicated for persons limited to a sitting position requiring mobility assistance. Suitable for use by individuals or as an attendant-propelled transport device in clinical or home environments.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Predicate Devices
Invacare Manual Wheelchair Model Action AT II (K984447)
Sunrise Medical/Quickie Design Model "Quickie TSR" Manual Wheelchair (K952641)
Related Devices
K984447 — MODEL ACTION AT II MANUAL WHEELCHAIR · Invacare Corp. · Jan 8, 1999
K032133 — BRODA MODEL 587 MANUAL WHEELCHAIR · Broda Ent., Inc. · Jul 21, 2003
K181090 — Invacare® Solara® 3G and Solara® 3G Spree Manual Wheelchairs · Invacare Corporation · Nov 14, 2018
K151893 — P.R.O. CG Manual Wheelchair · Freedom Designs, Inc. · Oct 22, 2015
K221435 — SUPERTILT PLUS (STP) · Power Plus Mobility · Oct 28, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
May 13, 2010
# - TH DMC
Image /page/0/Picture/2 description: The image shows the logo for Dalt Instruments. The logo features the letters 'DI' in a stylized font, with an oval shape surrounding the 'I'. Above the logo, the text 'K101719' is written in a handwritten style. The overall image appears to be a scan or photocopy, with some imperfections and variations in contrast.
JUN. 1.8 2010
### Received
# 510(k) SUMMARY DALTON INSTRUMENT CORPORATION'S Model TACAMO Manual Wheelchair
AUG 2 5 2010
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared.
Dalton Instrument Corp. 3788 Arapaho Road Addison, TX 75001 Tel: 469-522-1200 Fax: 469-522-1202
Contact Person: President, Mei Lein Date Prepared: May 13, 2010
# Name of Device and Name/Address of Sponsor
Model TACAMO Manual Wheelchair
Dalton Instrument Corp. 3788 Arapaho Road Addison, TX 75001 Tel: 469-522-1200 Fax: 469-522-1202
Common or Usual Name: Manual Wheelchair
Classification Name: Wheelchair, Mechanical 89IOR
#### Predicate Devices
The Dalton Instrument Corp. Model TACAMO substantially equivalent to the Invacare Manual Wheelchair Model Action AT II (K984447, 1/8/1999) and Sunrise Medical/Quickie Design Model "Quickie TSR" Manual Wheelchair (K952641, 6/9/1995)
Intended Use: the intended use of the Model TACAMO Manual Wheelchair is to provide mobility to persons limited to a sitting position.
{1}------------------------------------------------
## Technological Characteristics and Substantial Equivalence
#### A. Device Description
The DALTON INSTRUMENT CORPORATION Model TACAMO is a manually operated, self propelled, manual, mechanical wheelchair. The primary design function and use is to provide mobility to persons who may be restricted to a sitting position. This product may also be used as an attendant propelled patient transport device in a health care environment such as a hospital, nursing home or long term care facilities.
The product has standard manual wheelchair features. It consists of the frame, 24" rear wheels with hand rims for propelling the chair, and 8'' front casters for steering and turning.
The wheelchair frame is constructed from both 1.25" and 1" outside diameter(OD) round, mechanical, steel tubing. The side frames are of welded construction and are secured to the remaining frame members using screws and bolts. This device is a rigid, "non folding" type wheel chair that incorporates a solid seating surface. TACAMO also comes with wheelchair cushion and seating system.
TACAMO also includes a "Tilt in Space" feature which allows the seat and back of the wheelchair to be tilted. This feature is used for those patients who require a tilt feature for stability, comfort or head control. It also serves as an attendant aid in those situations where a patient need to be tilted to be fed, or attended to in some situations.
The "Tilt in Space" feature is manually operated and is activated by two release levers located at the rear of the wheelchair. Adjusting the tilt is achieved by squeezing the two release lever handles inward simultaneously. When the levers are squeezed, flexible cables attached to each release lever disengage gear rack segments from independent actuator rods which are notched to match the teeth of the gear segments. Once the desired tilt angle has been obtained, the handles are released and the chair will remain at the angle chosen. The chair has a tilt angle range of 30 degrees and 30 degrees of reclining angle.
### B. Substantial Equivalence
The DALTON INSTRUMENT CORPORATION Model TACAMO substantially equivalent to numerous manual, mechanical wheelchairs currently in market place,
{2}------------------------------------------------
for which FDA has granted marketing clearance through the 510(k) premarket notification process. More specifically the DALTON INSTRUMENT CORPORATION Model TACAMO is substantially equivalent to the Sunrise Medical/Quickie Design Model "Quickie TSR" Manual Wheelchair(K952641 6/9/95) and the Invacare Model AT II Manual Wheelchair(K984447 1/8/1999)
Each of the products is a manual operated, attendant or user propelled, manual mechanical wheelchair with the same intended function and use which is to provide mobility to persons that may be limited to a seated position. All products consist basically of a mechanical frame to support the wheelchair, 24" rear wheels for propelling the wheelchair, and 8" casters for turning and steering. Additionally, each of these wheelchairs incorporates a manually operated "tilt in space" feature for patient comfort and positioning.
#### Performance Data
The DALTON INSTRUMENT CORPORATION wheelchair TACAMO meets the applicable performance requirements specified in:
- . Rehabilitation Engineering Society of North America (RESNA) Standard ANSI/RESNA WC/Vol.1-1998 "Requirements and Test Methods for Wheelchairs(including Scooters)"
- DIS ISO 7176 (1992) Part 8: Requirements and Test Methods for Static, ● Impact and Fatigue Strengths
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with three stripes forming its body and wings. The eagle is facing right. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Dalton Instrument Corporation % Mei Lein President 3788 Arapaho Road Addison, Texas 75001
AUG 2 5 2010
Re: K101719
Trade/Device Name: TACAMO Manual Wheelchair Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical wheelchair Regulatory Class: Class I Product Code: IOR Dated: August 9, 2010 Received: August 16, 2010
Dear Mei Lein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
#### Page 2 - Mei Lein
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Marki McMullerson
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
K101719
510(K) Number (if known): K101719
Device Name: TACAMO Manual Wheelchair
Indications for Use: The intended use of the TACAMO Manual Wheelchair is to provide mobility to persons limited to a sitting position.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _X (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
.
Page 1 of
: Sim-Off sion of Surgical. Orthopedic, i Kestorative Devices
510(k) Number K101719
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.