← Product Code [GEI](/productcode/GEI) · K101711

# CERULEAU ELECTROSURGICAL PROBE (K101711)

_Nuortho Surgical, Inc. · GEI · Jul 1, 2010 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K101711

## Device Facts

- **Applicant:** Nuortho Surgical, Inc.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Jul 1, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

NuOrtho CERULEAU™ Probes are intended for use as surgical instruments. The NuOrtho CERULEAU™ Probes are indicated for resection, coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures. These procedures include the following indications: [List of specific joint procedures provided in document].

## Device Story

Bipolar electrosurgical probe; directs radiofrequency (RF) energy from external generator to target tissue during arthroscopic/orthopedic surgery. Features protected electrode geometry inhibiting direct electrode-to-tissue contact; housing acts as mechanical implement adjunct to energy delivery. Delivers non-ablation, low-level RF energy. Used by surgeons in OR setting. Output allows for precise resection, coagulation, and hemostasis; reduces collateral tissue damage compared to traditional ablation probes. Benefits patient via minimized thermal injury to surrounding healthy tissue.

## Clinical Evidence

No clinical data. Bench testing and in vitro studies on excised human tissue provided. Study REP-100-1220 demonstrated statistically significant reduction in collateral tissue damage compared to predicate devices using histological evidence.

## Technological Characteristics

Bipolar electrosurgical probe; radiofrequency energy source. Features protected electrode geometry. Compliant with IEC 60601-1 (safety) and IEC 60601-2-2 (high-frequency surgical equipment). Biocompatibility per ISO 10993-5, -10, -11. Sterilization via radiation (ISO 11137-1). Packaging per ISO 11607-1/2 and ASTM F1980.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- ArthroCare Arthrowands ([K011083](/device/K011083.md))
- Stryker SERFAS ([K991960](/device/K991960.md))
- Mitek VAPR S90 ([K002422](/device/K002422.md))
- ConMed UltrAblator ([K993885](/device/K993885.md))
- ConMed Bicap Superconductor™ ([K012018](/device/K012018.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K 1 01711

Page 1 of (4)

NuOrtho Surgical, Inc.

ip

CERULEAU™ Electrosurgical Probe Traditional 510(k)

## 5. 510(K) SUMMARY

| CERULEAU™ Electrosurgical Probe - 510(k) Summary |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |             |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Submitter:                                       | NuOrtho Surgical, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |             |
| Contact:                                         | Roy Morgan, PE, PMP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |             |
| Date Summary<br>Prepared:                        | June 30th 2010                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | JUL -1 2010 |
| Device Trade Name:                               | CERULEAU™ Electrosurgical Probe                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |             |
| Common Name:                                     | Electrosurgical Probe                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |             |
| Classification Name:                             | Electrosurgical cutting and coagulation device and<br>accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |             |
| Classification Code:                             | 21 CFR §878.4400 Product Code: GEI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |             |
| Equivalent Device(s):                            | ArthroCare Arthrowands (K011083) Stryker SERFAS (K991960) Mitek VAPR S90 (K002422) ConMed UltrAblator (K993885) ConMed Bicap Superconductor™ (K012018)                                                                                                                                                                                                                                                                                                                                                                                                                                              |             |
| Device Description:                              | The CERULEAU™ Electrosurgical Probe is a bipolar<br>device designed to direct radiofrequency energy from an<br>electrosurgical generator into target tissue during<br>arthroscopic and orthopedic surgical procedures.<br>Ceruleau delivers non-ablation, low-level amounts of<br>radiofrequency energy to the treatment site. The<br>radiofrequency energy is delivered via a protected<br>electrode geometry which inhibits direct electrode-to-<br>tissue contact. Design of the electrode's protective<br>housing can function as a mechanical implement for use<br>adjunct to energy delivery. |             |
| Intended and<br>Indications for Use:             | NuOrtho CERULEAU™ Probes are intended for use as<br>surgical instruments.<br>The NuOrtho CERULEAU™ Probes are indicated for<br>resection, coagulation of soft tissue and hemostasis of<br>blood vessels in arthroscopic and orthopedic procedures.<br>These procedures include the following indications:                                                                                                                                                                                                                                                                                           |             |

{1}------------------------------------------------

K1017"

Page 2 of 4

.

CERULEAU™ Electrosurgical Probe

| Arthroscopic and Orthopedic Procedures | Joint Specific or All Joints<br>(ankle, elbow, hip, knee, shoulder, and wrist) |
|----------------------------------------|--------------------------------------------------------------------------------|
| Resection and Excision                 |                                                                                |
| Acetabular Labrum                      | Hip                                                                            |
| Articular Labrum                       | All Joints                                                                     |
| Capsule                                | All Joints                                                                     |
| Capsular Release                       | Knee                                                                           |
| Cartilage Flaps                        | Knee                                                                           |
| Cysts                                  | All Joints                                                                     |
| Discoid Meniscus                       | Knee                                                                           |
| Frozen Shoulder Release                | Shoulder                                                                       |
| Glenoid Labrum                         | Shoulder                                                                       |
| Labral Tear                            | Shoulder                                                                       |
| Lateral Release                        | Knee                                                                           |
| Ligament                               | All Joints                                                                     |
| Loose Bodies                           | All Joints                                                                     |
| Meniscal Cystectomy                    | Knee                                                                           |
| Meniscectomy                           | Knee                                                                           |
| Plica Removal                          | All Joints                                                                     |
| Scar Tissue                            | All Joints                                                                     |
| Soft Tissue                            | All Joints                                                                     |
| Synovial Membrane                      | All Joints                                                                     |
| Tendon                                 | All Joints                                                                     |
| Triangular Fibrocartilage (TFCC)       | Wrist                                                                          |
| Villusectomy                           | Knee                                                                           |
| Debridement                            |                                                                                |
| ACL/PCL                                | Knee                                                                           |
| Acromioplasty                          | Shoulder                                                                       |
| Articular Cartilage                    | All Joints                                                                     |
| Bursectomy                             | All Joints                                                                     |
| Chondroplasty                          | All Joints                                                                     |
| Fascia                                 | All Joints                                                                     |
| Ligament                               | All Joints                                                                     |
| Notchplasty                            | Knee                                                                           |
| Scar Tissue                            | All Joints                                                                     |
| Soft Tissue                            | All Joints                                                                     |
| Subacromial Decompression              | Shoulder                                                                       |
| Synovectomy                            | All Joints                                                                     |
| Tendon                                 | All Joints                                                                     |
| Coagulation                            |                                                                                |
| ACL/PCL                                | Knee                                                                           |
| Articular Cartilage                    | All Joints                                                                     |
| Carpal Ligaments                       | Wrist                                                                          |
| Glenohumeral Capsule                   | Shoulder                                                                       |
| Ligament                               | All Joints                                                                     |
| Medial Retinaculum                     | Knee                                                                           |
| Rotator Cuff                           | Shoulder                                                                       |
| Tendon                                 | All Joints                                                                     |
| Wrist Tendons                          | Wrist                                                                          |

{2}------------------------------------------------

K 1017"

Page 3 of (4)

CERULEAU™ Electrosurgical Probe Traditional 510(k)

Non-clinical Performance Data:

Performance testing per standardized methods and NSI test protocols including bench and in vitro human data was conducted and provides support that the CERULEAU™ Electrosurgical Probe is substantially equivalent to currently marketed predicate devices. Non-clinical performance test protocols and results are included in sections 14 through 18 of this 510(k) submission. Test protocols and reports of results demonstrate that. in consideration of its intended use, the design, labeling, packaging and sterilization of the CERULEAU™ Electrosurgical Probe is compliant with the following standards:

- ANSI/AAMI ISO 11137-1:2006 (Cor 1:2007) , . Sterilization of health care products - Radiation - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices
- . ISO 11607-1:2006-04-15 Packaging for terminally sterilized medical devices -- Part 1: Requirements for materials, sterile barrier
- . ISO 11607-2: 1st Edition-2006-04-15; Packaging for terminally sterilized medical devices -- Part 2: Validation requirements for forming, sealing and assembly processes, including Annex B (informative) listing of standardized test methods and procedures
- ASTM F1980-07; Shelf-life and accelerated aging . techniques for standard evaluation of packaging performance
- . ASTM D4169 (2009):Standard Test Method for Testing of Shipping Containers and Systems
- . ISO 10993-5:2009, Biological evaluation of medical devices -- Part 5: Tests for In Vitro Cytotoxicity .
- ISO 10993-10:2002 (A1:2006). Biological evaluation of . medical devices - Part 10: Tests for irritation and delayed-type hypersensitivity.
- t ISO 10993-11:2006, Biological evaluation of medical devices -- Part 11: Tests for systemic toxicity.
- IEC 60601-1; Medical Electrical Equipment - Part 1: General Requirements for Safety and Essential Performance.
- IEC 60601-2-2- 2009. Medical electrical equipment -. Part 2-2: Particular requirements for the safety of high frequency surgical equipment

Section 18, (Bench testing) includes a summary of a comparison study (REP-100-1210) that was conducted to evaluate the effectiveness of the CERULEAU™ Probe performance to other legally available probes intended for similar usage in arthroscopic and orthopedic surgery. The full

{3}------------------------------------------------

CERULEAU™ Electrosurgical Probe Traditional 510(k)

report of REP-100-1210 entitled "Evaluation of Ceruleau™ Performance on Human tissue -Establishment of Clinically Relevant Functional Substantial Equivalence" is provided in Appendix 12.

Section 18 also includes a summary of REP-100-1220 entitled, "Comparative_Evaluation of Ceruleau™ Chondroplasty Performed on Excised Human Articular Cartilage -

Assessment of Clinical Efficacy and Extent of Collateral Damage", which was conducted invitro on excised human tissue and is provided in Appendix 12.

The REP-100-1220 provides clear histological evidence to substantiate that, as compared to predicates, the design and use of the CERULEAU™ Probe consistently and significantly reduces incidence of damage to collateral and subject tissues. The study proves that, due to its design, the CERULEAU™ Probe consistently performs to reduce incidence of damage to collateral and subject tissues, while the designs of the predicate devices are inherently prone to cause damage to collateral and subject tissues. Additionally, Section 21 includes a summary of scientific publications that support the need for the technological features and clinical utility of the CERULEAU Electrosurgical Probe.

Clinical Performance Data:

Clinical data was not necessary to support that the CERULEAU™ Electrosurgical Probe is substantially equivalent to currently marketed predicate devices.

{4}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JUL - 1 2010

NuOrtho Surgical, Inc. % Regulatory Technology Services, LLC Mr. Mark Job 1394 25th Street, NW Buffalo, Minnesota 55313

DEPARTMENT OF HEALTH & HUMAN SERVICES

Re: K101711

Trade/Device Name: Ceruleau™ Electrosurgical Probe Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI, HRX Dated: June 17, 2010 Received: June 18, 2010

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act 

{5}------------------------------------------------

Page 2 - Mr. Mark Job

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours

Mark A. Milliman

Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Page 1 of 2

CERULEAU™ Electrosurgical Probe Traditional 510(k)

## 4. INDICATIONS FOR USE STATEMENT

510(k) Number if known: N/A

## Device Name: CERULEAU™Electrosurgical Probe

The NuOrtho CERULEAU™ Probes are indicated for resection, coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures.

These procedures include the following indications:

| Arthroscopic and Orthopedic Procedures | Joint Specific or All Joints<br>(ankle, elbow, hip, knee, shoulder, and wrist) |
|----------------------------------------|--------------------------------------------------------------------------------|
| Resection and Excision                 |                                                                                |
| Acetabular Labrum                      | Hip                                                                            |
| Articular Labrum                       | All Joints                                                                     |
| Capsule                                | All Joints                                                                     |
| Capsular Release                       | Knee                                                                           |
| Cartilage Flaps                        | Knee                                                                           |
| Cysts                                  | All Joints                                                                     |
| Discoid Meniscus                       | Knee                                                                           |
| Frozen Shoulder Release                | Shoulder                                                                       |
| Glenoid Labrum                         | Shoulder                                                                       |
| Labral Tear                            | Shoulder                                                                       |
| Lateral Release                        | Knee                                                                           |
| Ligament                               | All Joints                                                                     |
| Loose Bodies                           | All Joints                                                                     |
| Meniscal Cystectomy                    | Knee                                                                           |
| Meniscectomy                           | Knee                                                                           |
| Plica Removal                          | All Joints                                                                     |
| Scar Tissue                            | All Joints                                                                     |
| Soft Tissue                            | All Joints                                                                     |
| Synovial Membrane                      | All Joints                                                                     |
| Tendon                                 | All Joints                                                                     |
| Triangular Fibrocartilage (TFCC)       | Wrist                                                                          |
| Villusectomy                           | Knee                                                                           |
| Debridement                            |                                                                                |
| ACL/PCL                                | Knee                                                                           |
| Acromioplasty                          | Shoulder                                                                       |
| Articular Cartilage                    | All Joints                                                                     |
| Bursectomy                             | All Joints                                                                     |
| Chondroplasty                          | All Joints                                                                     |
| Facia                                  | All Joints                                                                     |
| Ligament                               | All Joints                                                                     |
| Notchplasty                            | Knee                                                                           |
| Scar Tissue                            | All Joints                                                                     |
| Soft Tissue                            | All Joints                                                                     |
| Subacromial Decompression              | Shoulder                                                                       |
| Synovectomy                            | All Joints                                                                     |
| Tendon                                 | All Joints                                                                     |
| Coagulation                            |                                                                                |
| ACL/PCL                                | Knee                                                                           |

K1017
Number
(k)

ision Sign-Off)
Restoralive Devices

Page 11

{7}------------------------------------------------

K101711

Page 2 of (2)

NuOrtho Surgical, Inc.

## CERULEAU™ Electrosurgical Probe Traditional 510(k)

| Coagulation (continued) |            |
|-------------------------|------------|
| Articular Cartilage     | All Joints |
| Carpal Ligaments        | Wrist      |
| Glenohumeral Capsule    | Shoulder   |
| Ligament                | All Joints |
| Medial Retinaculum      | Knee       |
| Rotator Cuff            | Shoulder   |
| Tendon                  | All Joints |
| Wrist Tendons           | Wrist      |

Prescription Use __X (Per 21 CFR 801 Subpart D)

AND/OR

Over-the-Counter Use (Per 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Signature

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K101711

---

**Source:** [https://fda.innolitics.com/device/K101711](https://fda.innolitics.com/device/K101711)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
