← Product Code [LLZ](/productcode/LLZ) · K101707

# FULACCESS MODEL 1.0 (K101707)

_Fulcrum Medical, Inc. · LLZ · Oct 29, 2010 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K101707

## Device Facts

- **Applicant:** Fulcrum Medical, Inc.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Oct 29, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

fulAccess is a standalone software product that provides basic image processing and communication functionality for use in radiology, oncology, and other clinical specialties. With fulAccess the user can review images from different modalities including CT, MR, and PET, perform basic functions including zoom, pan, measure, sample pixel intensity, and make annotations. Users may also create and save a set of files that include a copy of the data being viewed, any annotations, and an instance of the viewer executable. These files can be stored locally or on portable media (e.g. CD and DVD). fulAccess is a Windows application and will be used on recommended hardware that is provided by the company or the end user. fulAccess is a stand-alone software package that assists users in the display, analysis, comparison and communication of medical imaging data from various sources. It allows the display, analysis, annotation, and communication of medical images. The intended users of fulAccess are trained medical professionals including physicians, nurses, physicists and other medical technologists. fulAccess is not intended for use in mammography.

## Device Story

Standalone Windows software for radiology/oncology; inputs include medical images (CT, MR, PET); provides tools for review, zoom, pan, measurement, pixel intensity sampling, and annotation; allows saving data with viewer executable for local/portable media distribution; operated by trained medical professionals in clinical settings; assists in image display and analysis to support clinical decision-making; benefits include improved accessibility and communication of imaging data.

## Clinical Evidence

Bench testing only. Verification testing included unit, system, and integration tests covering DICOM import, image control, and data export. Validation testing performed by a board-certified clinician in a simulated clinical environment using representative data types. All tests passed with no anomalies.

## Technological Characteristics

Standalone Windows-based software application. Functions include DICOM image import, display, manipulation (zoom, pan, measure), and annotation. No specific hardware requirements beyond standard PC specifications. No specialized energy source or materials.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- MIM 4.1 ([K071964](/device/K071964.md))
- Velocity AI ([K081076](/device/K081076.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K101707

# Section 5

# 510(k) Summary or 510(k) Statement

Premarket Notification [510(k)] Summary OCT 2 9 2010

(Provided in conformance with 21 CFR 807.92)

### Submitter: 1.

Company Name and Address: Fulcrum Medical, Inc 17209 Chesterfield Airport Rd. Chesterfield, MO 63005

Ryan S. Brame, Ph.D. Contact Person: President Contact Title: 800-515-9132 Contact Phone: 888-788-5946 Contact Fax:

Date of Summary:

May 28, 2010

### Device Name and Classification: 2.

| Trade Name:            | fulAccess                                               |
|------------------------|---------------------------------------------------------|
| Common and Usual Name: | Medical Imaging Software                                |
| Classification Name:   | Image Processing System, Radiology<br>(21 CFR 892.2050) |
| Regulatory Class:      | Class II                                                |
| Product Code:          | LLZ                                                     |

### Predicate Device(s): 3.

| K071964 | MIM 4.1     | MIMVista Corporation            |
|---------|-------------|---------------------------------|
| K081076 | Velocity AI | Velocity Medical Solutions, LLC |

### Description of Device: 4.

fulAccess is a standalone software product that provides basic image runcessor is a communication functionality for use in radiology, oncology, processing and solecialties. With fulAccess the user can review images and other entired on only ing CT, MR, and PET, perform basic functions including zoom, pan, measure, sample pixel intensity, and make annotations. meraamig 200m, passe and save a set of files that include a copy of the data being viewed, any annotations, and an instance of the viewer executable. being viewed, any annellatorally or on portable media (e.g. CD and DVD).

fulAccess is a Windows application and will be used on recommended hardware that is provided by the company or the end user.

Image /page/0/Picture/18 description: The image shows a black and white drawing of a stylized mountain range. The mountains are depicted with sharp, angular lines, creating a geometric and somewhat abstract representation. The interior of the mountain on the left is filled with a stippled pattern, adding texture and visual interest to the design.

Pi of 2

{1}------------------------------------------------

# fulcrummedical

# 5. Indication for Use:

fulAccess is a stand-alone software package that assists users in the display, analysis, comparison and communication of medical imaging data from various sources. It allows the display, analysis, annotation, and communication of medical images. The intended users of fulAccess are trained medical professionals including physicians, nurses, physicists and other medical technologists. fulAccess is not intended for use in mammography.

## Comparison with Predicate Devices: 6.

fulAccess is substantially equivalent to the identified predicate devices. The fulAccess product is similar in characteristics, materials, and features, and has similar technological features, intended use and indications for use as the predicates, and does not pose any new issues of safety and effectiveness.

A detailed comparison can be found in Section 12 of this submittal.

### Non-Clinical Performance Summary 7.

Fulcrum Medical has verified and validated that the fulAccess software meets its functional specifications and performance requirements. Verification testing was accomplished using a combination of unit, system and integration tests. Verification testing was organized with reference to discrete units of software functionality (e.g., DICOM import, image control, structure control, data export/saving, etc.) and the results of hazard analyses. All verification tests were passed and no anomalies remained at the conclusion of testing. Validation testing was performed by a board certified clinician who organized the testing according to the intended clinical use (i.e. importing data, viewing data, manipulating data, saving data, etc.). Validation testing was performed in a simulated clinical environment using a variety of data types and combinations that were judged by the clinician to be representative of the types of data the device will encounter in clinical use. All validation tests were passed and no anomalies remained at the conclusion of testing.

#### 8. Conclusions

In summary, Fulcrum Medical, Inc. is of the opinion that fulAccess does not introduce any new potential safety risks, is as effective, and performs as well mat devices currently on the market, and thus concludes that the fulAccess software is substantially equivalent to the predicate devices.

K101767 7. 2 of 2

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Mr. Scott Brame President Fulcrum Medical, Inc 17209 Chesterfield Airport Rd. CHESTERFIELD MO 63005

# UCI 2 9 20:0

Re: K101707

> Trade Name: fulAccess Regulation Number: 21 CFR § 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: September 21, 2010 Received: October 1, 2010

Dear Mr. Brame:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{3}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

David G. Brown, Ph.D. Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Section 4

Indications for Use Statement

510(k) Number (if known) _K101707_

# Device Name: fulAccess

fulAccess is a standalone software product that provides basic image processing and communication functionality for use in radiology, oncology, and other clinical specialties. With fulAccess the user can review images from different modalities including CT, MR, and PET, perform basic functions including zoom, pan, measure, sample pixel intensity, and make annotations. Users may also create and save a set of files that include a copy of the data being viewed, any annotations, and an instance of the viewer executable. These files can be stored locally or on portable media (e.g. CD and DVD).

fulAccess is a Windows application and will be used on recommended hardware that is provided by the company or the end user.

fulAccess is a stand-alone software package that assists users in the display, analysis, comparison and communication of medical imaging data from various sources. It allows the display, analysis, annotation, and communication of medical images. The intended users of fulAccess are trained medical professionals including physicians, nurses, physicists and other medical technologists. fulAccess is not intended for use in mammography.

(DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use     | X |
|----------------------|---|
| (Per 21 CFR 801.109) |   |

OR

| Over-The-Counter Use |  |
|----------------------|--|
|----------------------|--|

(Division Sign-Off)

Division of Radiological Devices

Office of In Vitro Diagnostic Device Evaluation and Safety

| §10K | K101707 |
|------|---------|
|------|---------|

K101707

P. 1 of 1

OCT 2 9 2010

KI 01707

---

**Source:** [https://fda.innolitics.com/device/K101707](https://fda.innolitics.com/device/K101707)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
