← Product Code [MWI](/productcode/MWI) · K101674

# WVSM (WIRELESS VITAL SIGNS MONITOR), MODEL 5.0 (K101674)

_Athena Gtx · MWI · Jul 21, 2010 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K101674

## Device Facts

- **Applicant:** Athena Gtx
- **Product Code:** [MWI](/productcode/MWI.md)
- **Decision Date:** Jul 21, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2300
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Wireless Vital Signs Monitor (WVSM) is intended to be used as an adult patient monitor. It is indicated as a single or multi-parameter vital signs monitor for ECG, noninvasive blood pressure (NIBP) and SpO2. It may be used in the following locations: Hospitals, healthcare facilities, emergency medical applications, during transport, and other healthcare applications. The monitor uses wireless communications to transmit vital signs data to a handheld or PC computer. The monitor is intended to be used by trained healthcare providers.

## Device Story

WVSM is a small, lightweight, rugged, portable patient monitor; measures ECG, NIBP, and SpO2. Device displays vital signs locally; transmits data via WiFi 802.11b/g to handheld devices or PCs. Used by trained healthcare providers in hospitals, clinics, emergency settings, and transport. Provides real-time physiological data to clinicians; facilitates patient monitoring and clinical decision-making during care or transport; improves portability of vital signs monitoring.

## Clinical Evidence

Bench testing only. Compliance with recognized national and international standards for safety and performance verified.

## Technological Characteristics

Small, lightweight, rugged portable monitor. Measures ECG, NIBP, SpO2. Connectivity via WiFi 802.11b/g to PC or PDA. Class II device.

## Regulatory Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

## Predicate Devices

- Welch Ally Propaq LT Vital Signs Monitor ([K033378](/device/K033378.md))
- Nonin Medical, Inc. Model # 7500 Digital Pulse Oximeter ([K071285](/device/K071285.md))
- HomMed Genesis Patient Monitor System ([K040799](/device/K040799.md))
- Welch Allyn Acuity Central Monitoring Station ([K052160](/device/K052160.md), [K022453](/device/K022453.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# Section 5 – Summary of Safety and Effectiveness

## 510(k) Summary of Safety and Effectiveness

| Submission Date:                         | June 11th, 2010                                                                                                                                                                                                                                                                         | JUL 21 2010 |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| 510(k) Submitter/Holder:                 | Athena GTX<br>3620 SW 61st Street, Suite 395<br>Des Moines, lowa 50321<br>Ph.: 515.288.3360 Fax.: 515.288.3394                                                                                                                                                                          |             |
| Company Contact:                         | Sean Mahoney (Chief Technical Officer)<br>Office Phone Number: 515.288.3360<br>Email: smahoney@athenagtx.com                                                                                                                                                                            |             |
| Trade Name:                              | WVSM Wireless Vital Signs Monitor                                                                                                                                                                                                                                                       |             |
| Common Name:                             | Cardiac Monitor                                                                                                                                                                                                                                                                         |             |
| Classification Name:                     | Cardiac Monitor (Including cardiotachometer and rate alarm)<br>(Refer to 21 CFR 870.2300)<br>NIBP Measurement System (Refer to 21 CFR 870.1130<br>Oximeter (Refer to 21 CFR 870.2700)<br>Radiofrequency Physiological Signal Transmitter and<br>Receiver (Refer to 21 CFR 870.2910)     |             |
| Classification Regulation:               | Class II                                                                                                                                                                                                                                                                                |             |
| Basis for Submission:                    | New device design                                                                                                                                                                                                                                                                       |             |
| Legally Marketed<br>(Predicate) Devices: | Welch Ally Propaq LT Vital Signs Monitor, (K033378)<br>Nonin Medical, Inc. Model # 7500 Digital Pulse Oximeter<br>(K071285)<br>HomMed Genesis Patient Monitor System (K040799)<br>Welch Allyn Acuity Central Monitoring Station<br>(K052160, K022453)                                   |             |
| Device Description:                      | The Athena GTX (WVSM) Wireless Vital Signs Monitor is a<br>small, lightweight, rugged, and highly portable patient<br>monitor designed to measure SpO2, NIBP and ECG. Vital<br>signs are displayed directly on the device, and may be<br>transmitted via WiFi 802.11b/g radio frequency |             |

5 - I

{1}------------------------------------------------

# Section 5 – Summary of Safety and Effectiveness

|                               | communication to a Personal Computer (PC) or Personal<br>Digital Assistant (PDA).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate Device<br>Overview: | The WVSM is designed for the same application and<br>intended use as the combination of listed predicate devices.<br>The WVSM is capable of the same ECG, heart rate, systolic<br>and diastolic blood pressure, functional oxygen saturation,<br>and pulse rate measurements as have been provided by the<br>combination of predicate devices referenced above.                                                                                                                                                                                                                             |
| Intended Use:                 | The Wireless Vital Signs Monitor (WVSM) is intended to be<br>used as an adult patient monitor. It is indicated as a single<br>or multi-parameter vital signs monitor for ECG, noninvasive<br>blood pressure (NIBP) and SpO2. It may be used in the<br>following locations: Hospitals, healthcare facilities,<br>emergency medical applications, during transport, and other<br>healthcare applications. The monitor uses wireless<br>communications to transmit vital signs data to a handheld or<br>PC computer.<br>The monitor is intended to be used by trained healthcare<br>providers. |
| Summary of<br>Testing:        | Testing On the WVSM 5.0 has been completed to verify<br>compliance with recognized national and international<br>standards for safety and performance for medical devices,<br>and particular requirements applicable to this device                                                                                                                                                                                                                                                                                                                                                         |
| Conclusion:                   | Based on the results for all safety and compliance testing<br>performed, it is the opinion of Athena GTX the WVSM<br>Wireless Vital Signs Monitor is safe and effective, and is<br>substantially equivalent to the above listed predicate<br>devices.                                                                                                                                                                                                                                                                                                                                       |

5 - 2

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird-like figure, with flowing lines representing its wings and body.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

Mr. Sean Mahoney Chief Technical Officer Athena GTX 3620 SW 61st Street, Suite 395 Des Moines, Iowa 50321

JUL 21 2010

Re: K101674

Device Name: Athena GTX Wireless Vital Signs Monitor (WVSM) Regulation Number: 21 CFR 870.2300 Regulation Name: Patient Physiological Monitor (Without Arrhythmia Detection or Alarms) Regulatory Class: Class II (Two) Product Codes: MWI, DQA, DXN, DRG Dated: June 14, 2010 Received: June 15, 2010

Dear Mr. Mahoney:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Footd, Orug and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act . The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Sean Mahoney

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

To: Bram D. Zuckerman, M.D.

Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### Indications for Use Statement

510(k) Number (if known): K101674

Device Name: Wireless Vital Signs Monitor

Indications for Use:

The Wireless Vital Signs Monitor (WVSM) is intended to be used as an adult patient monitor. It is indicated as a single or multi-parameter vital signs monitor far ECG, noninvasive blood pressure (NIBP) and SpO2. It may be used in the following locations: Hospitals, healthcare facilities, emergency medical applications, during transport, and other healthcare applications. The monitor uses wireless communications to transmit vital signs data to a handheld or PC computer. The monitor is intended to be used by trained healthcare providers.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

W.W

Concurrence of CDRH, Office of Device Evaluation (ODE)

at

| (Division Sign-Off) Division of Cardiovascular Devices |         |
|--------------------------------------------------------|---------|
| 510(k) Number                                          | K101674 |

[ 4 - 1 ]

1 of 1

---

**Source:** [https://fda.innolitics.com/device/K101674](https://fda.innolitics.com/device/K101674)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
