← Product Code [NTE](/productcode/NTE) · K101651

# ANGIOGUARD XP EMBOLI CAPTURE GUIDEWIRE, ANGIOGUARD RX EMBOLI CAPTURE GUIDEWIRE (K101651)

_Cordis Corp. · NTE · Jul 9, 2010 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K101651

## Device Facts

- **Applicant:** Cordis Corp.
- **Product Code:** [NTE](/productcode/NTE.md)
- **Decision Date:** Jul 9, 2010
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The ANGIOGUARD™ XP Emboli Capture Guidewire is indicated for use as a guidewire and embolic protection system to contain and remove embolic material (thrombus/debris) while performing carotid artery angioplasty and stenting procedures in carotid arteries. The diameter of the artery at the site of filter basket placement should be from 3mm to 7.5 mm.

## Device Story

Device is an interventional guidewire with an integrated distal emboli filter basket; used during carotid artery angioplasty and stenting. Guidewire delivered via OTW or RX deployment sheath; filter basket deploys in umbrella-like fashion to provide passive hemo-filtration and capture thrombus/debris. At procedure end, filter is collapsed and retrieved. Used in clinical settings by physicians. Benefits patient by containing and removing embolic material during stenting, reducing risk of distal embolization.

## Clinical Evidence

Bench testing only. Verification activities included deployment and capture testing, visual inspections, bioburden, endotoxin, EtO residual testing, and biocompatibility testing per ISO 10993-1. Results demonstrated that design outputs meet device inputs.

## Technological Characteristics

Guidewire with integrated distal filter basket; umbrella-like deployment mechanism. Materials include radiopaque markers attached via modified adhesive. Sterilization via EtO. Dimensions accommodate 3mm to 7.5mm artery diameters. Mechanical operation; no software or electronics.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- ANGIOGUARD™ XP Emboli Capture Guidewire ([K062531](/device/K062531.md))
- ANGIOGUARD™ RX Emboli Capture Guidewire ([K062531](/device/K062531.md))
- Boston Scientific FilterWire EX Embolic Protection System ([K023691](/device/K023691.md))
- PercuSurge, Inc. Percusurge Guardwire Temporary Occlusion and Aspiration System ([K013913](/device/K013913.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KU1651

Special 510(k) Notification

Cordis Corporation ANGIOGUARD XP & RX Emboli Capture Guidewires

## 510(k) Summary

| Applicant            | Cordis Corporation, a Johnson & Johnson Company<br>430 Rt. 22 East<br>Bridgewater, New Jersey 08807<br>Telephone: 908-541-4888<br>Fax: 908-541-4559                                                                                                                                                                                               |  | JUL - 9 2010 |  |  |  |  |  |  |  |  |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--------------|--|--|--|--|--|--|--|--|
| Contact<br>Person    | Joan Martin<br>Manager<br>Regulatory Affairs                                                                                                                                                                                                                                                                                                      |  |              |  |  |  |  |  |  |  |  |
| Date                 | June 10, 2009                                                                                                                                                                                                                                                                                                                                     |  |              |  |  |  |  |  |  |  |  |
| Device Name          | Trade Name: ANGIOGUARD™ XP Emboli Capture Guidewire<br>ANGIOGUARD™ RX Emboli Capture Guidewire   Common or Usual<br>Name: Common Name: Embolic Protection Guidewire<br>Cardiovascular Percutaneous Catheter per 21 CFR<br>870.1250 and Catheter Guidewire per 21 CFR<br>870.1330   Classification: Class II   Classification Panel Cardiovascular |  |              |  |  |  |  |  |  |  |  |
| Predicate<br>Devices | The device is substantially equivalent to Cordis' current, legally-marketed<br>ANGIOGUARD™ XP Emboli Capture Guidewire and ANGIOGUARD™                                                                                                                                                                                                            |  |              |  |  |  |  |  |  |  |  |

RX Emboli Capture Guidewire which were found to be substantially equivalent to Boston Scientific's FilterWire EX Embolic Protection System (K023691) and PercuSurge, Inc.'s Percusurge Guardwire Temporary Occlusion and Aspiration System (K013913) via 510(k) K062531 on September 22, 2006.

{1}------------------------------------------------

Cordis Corporation ANGIOGUARD XP & RX Emboli Capture Guidewires Special 510(k) Notification

Both subject and predicate ANGIOGUARD XP and RX devices consist of a Device guidewire with integrated emboli filter basket at the distal end. The devices Description function as an interventional guidewire and distal protection device during delivery and placement of stents and interventional devices in carotid procedures. The guidewire is delivered via an OTW (over-the-wire) or RX (rapid-exchange) deployment sheath and is captured via an OTW or RX capture sheath. ANGIOGUARD devices have a fitter basket at the distal end that is deployed prior to the stenting procedure. When deployed, the filter basket opens in an umbrella-like fashion, allowing passive hemo-filtration with subsequent emboli capture. At the end of the procedure, the filter is collapsed and retrieved.

The subject and predicate ANGIOGUARD XP and ANGIOGUARD RX Intended Use Emboli Capture Guidewire devicves are indicated for use as a guidewire and embolic protection system to contain and remove embolic material (thrombus/debris) while performing carotid artery angioplasty and stenting procedures in carotid arteries. The diameter of the artery at the site of filter basket placement should be from 3mm to 7.5 mm.

## Summary of Substantial Equivalence

With the exception of a change to the adhesive material used to adhere radiopaque markers to the filter basket struts, the subject and predicate devices share the same:

- Intended use .
- Indication for use
- Design & Dimensions .
- Size Range .
- Materials (with exception of adhesive material change described herein) .
- Fundamental technology and operating principle .
- Manufacturing site and methods .
- Sterilization site, method, parameters, and sterility assurance level .
- Packaging .
- Labeling / Instructions for use .
- . Shelf Life

subject ANGIOGUARD XP Emboli Capture Guidewire and The ANGIOGUARD RX Emboli Capture Guidewire devices are substantially equivalent to the predicate ANGIOGUARD XP Emboli Capture Guidewire and ANGIOGUARD RX Emboli Capture Guidewire devices.

{2}------------------------------------------------

Cordis Corporation ANGIOGUARD XP & RX Emboli Capture Guidewires Special 510(k) Notification

Design verification activities (Deployment and Capture Testing, Visual Summary of Inspections, Bioburden, Endotoxin, EtO residual testing and biocompatibility & Conclusions testing per ISO 10993-1) were conducted to confirm that the modified devices Drawn from meet the same requirements as the current, legally marketed predicate Design devices. Verification Activities

Results demonstrated that design outputs continue to meet device inputs. Hence, the subject ANGIOGUARD XP and RX Emboli Capture Guidewire devices are substantially equivalent to the current, legally marketed ANGIOGUARD XP and RX Emboli Capture Guidewire predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of a bird, possibly an eagle, with its wings spread. The bird is stylized with thick, curved lines, giving it a modern and symbolic appearance.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

JUL - 9 2010

Cordis Corporation c/o Joan Martin Manager, Regulatory Affairs 430 Route 22 East Bridgewater, NJ 08807

Re: K101651

Trade/Device Name: Angioguard XP and RX Emboli Capture Guidewires Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: NTE Dated: June 10, 2010 Received: June 11, 2010

Dear Ms. Martin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Joan Martin

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

R. Vilmer

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Special 510(k) Notification

Cordis Corporation ANGIOGUARD XP & RX Emboli Capture Guidewires

## Indication for Use

K10165 510(k) Number: ___

Device Name: ANGIOGUARD™ XP Emboli Capture Guidewire System

Indications For Use: The ANGIOGUARD™ XP Emboli Capture Guidewire is indicated for use as a guidewire and embolic protection system to contain and remove embolic material (thrombus/debris) while performing carotid artery angioplasty and stenting procedures in carotid arteries. The diameter of the artery at the site of filter basket placement should be from 3mm to 7.5 mm.

Prescription Use: X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use: (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Duna D. Valuer

(Division Sign-Off) Division of Cardiovascular Devices

510(k) Number_K101651

{6}------------------------------------------------

Special 510(k) Notification

Cordis Corporation ANGIOGUARD XP & RX Emboli Capture Guidewires

## Indication for Use

510(k) Number: _______________________________________________________________________________________________________________________________________________________________

Device Name: ANGIOGUARD™ RX Emboli Capture Guidewire System

Indications For Use: The ANGIOGUARD™ RX Emboli Capture Guidewire is indicated for use as a guidewire and embolic protection system to contain and remove embolic material (thrombus/debris) while performing carotid artery angioplasty and stenting procedures in carotid arteries. The diameter of the artery at the site of filter basket placement should be from 3mm to 7.5 mm.

Prescription Use: X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use: (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

brmea D. McAlusen

Division Sian-Off) Division of Čardiovascular Devices

510(k) Number

---

**Source:** [https://fda.innolitics.com/device/K101651](https://fda.innolitics.com/device/K101651)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
