K101599 · Revascular Therapeutics, Inc. · PDU · Jan 18, 2011 · Cardiovascular
Device Facts
Record ID
K101599
Device Name
RVT CTO DEVICE
Applicant
Revascular Therapeutics, Inc.
Product Code
PDU · Cardiovascular
Decision Date
Jan 18, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The RVT CTO Device is indicated to facilitate the intra-luminal placement of conventional guidewires beyond peripheral artery chronic total occlusions. The device is contraindicated for use in carotid arteries. The RVT Extension Wire is designed to extend the RVT CTO Device so that a catheter can be exchanged for another catheter.
Device Story
RVT CTO Device is a sterile, disposable, steerable 0.018" guidewire with a diamond-coated active tip; used with a battery-operated control unit. Operates in two modes: 'Active Mode' uses a DC motor to generate rotational energy (approx. 13,000 RPM) transferred via a Nitinol driveshaft to the active tip to penetrate occlusions; 'Passive Mode' allows manual steering like a standard guidewire. Used by physicians in clinical settings to facilitate intra-luminal placement of conventional guidewires beyond peripheral artery chronic total occlusions (CTOs). Includes a shaping tool and extension wire for catheter exchange. Provides mechanical recanalization of occluded vessels, enabling subsequent interventional procedures. Benefits patients by providing a crossing mechanism when conventional techniques fail, potentially avoiding bypass surgery.
Clinical Evidence
Prospective, multi-center, single-arm study (Re-OPEN) of 85 patients with infrainguinal CTOs. Primary safety endpoint: 0% clinical perforation rate at 30 days. Primary efficacy endpoint (device advancement): 76.5% (65/85). Primary performance endpoint (technical success): 80% (68/85). Study demonstrated safety and effectiveness in facilitating CTO crossing where conventional guidewires failed, with no unanticipated adverse events.
Technological Characteristics
Materials: Nitinol, gold-plated stainless steel, platinum tungsten, hydrophilic coating, tin, silver, nickel, natural diamonds. Energy: Mechanical rotational energy (13,000 RPM). Dimensions: 0.018" OD, 165 cm length. Connectivity: Standalone battery-operated control unit. Sterilization: Ethylene Oxide. Design: Guidewire-based with internal driveshaft and stationary outer shaft.
Indications for Use
Indicated for patients with peripheral artery disease (PAD) presenting with infrainguinal chronic total occlusions (100% occlusion, no antegrade flow) resistant to conventional guidewire crossing. Contraindicated for use in carotid arteries.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Chronic Total Occlusion Device
# 510(k) Summary
JAN 1 8 2011
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
## A. Name, Address, Phone and Fax Number of Applicant
ReVascular Therapeutics, Inc. 766 Palomar Avenue Sunnyvale, CA 94085 Phone: (408) 617-0770 Fax: (408) 481-9045
## B. Contact Person
.
Nancy Lincé, RAC Regulatory Affairs Consultant (650) 759-6186
- C. Date Prepared June 4, 2010
## D. Device Name
| Trade Name: | RVT CTO Device |
|----------------------|-----------------------------------------|
| Common Name: | Guidewire |
| Classification Name: | 21 CFR §870.1330; Wire, Guide, Catheter |
| Product Code: | DQX |
## E. Device Description
The RVT CTO Device is a sterile, disposable, steerable 0.018" OD Guidewire with an Active Tip for penetrating chronic total occlusions and is used with a sterile, disposable, battery-operated Control Unit during a single patient procedure.
The RVT CTO System consists of four (4) components: the RVT CTO Device, Control Unit, Shaping Tool, and the RVT Extension Wire.
- 1. RVT CTO Device (Guidewire Assembly, Torquer, Motor Housing with integrated Connector Cable) - The RVT CTO Device consists of a distal 0.018" Guidewire Assembly and a Motor Housing with Connector Cable. The RVT CTO Device is 165 cm in working length and is hydrophilically coated. The RVT CTO Device has a shapeable Distal Tip and the cone-shaped portion of the Active Tip is diamond coated. The distal 3 cm is radiopaque to facilitate visualization under fluoroscopy.
{1}------------------------------------------------
- Materials used in construction with patient contact: Nitinol. Gold Plated a. Stainless Steel, Platinum Tungsten, Hydrophilic Coating, Tin, Silver, Nickel, Natural Diamonds
- 2. Control Unit - Pressing the Activation Button allows current to flow from the battery in the Control Unit to the DC Motor within the Motor Housing. Mechanical energy (via rotational torque) is generated and transferred from the DC Motor to the Nitinol Driveshaft located within the stationary hollow Outer Shaft. The Driveshaft turns the Active Tip at the very distal end of the Guidewire Assembly at approximately 13,000 revolutions per minute (RPM) under "no load" conditions. The hollow Outer Shaft remains stationary while the Active Tip rotates independently. In the Active Mode, the RVT Guidewire creates a pathway through the lesion via mechanical rotation. In the "Passive Mode" the operator steers the Guidewire Assembly to seek a path of least resistance through the lesion consistent with standard quidewires.
- a. Materials used in construction with patient contact: none
- 3. Shaping Tool used to shape the tip of the RVT CTO Device, if desired.
- a. Materials used in construction with patient contact: none
- 4. RVT Extension Wire Attachment of the RVT Extension Wire to the RVT CTO Device creates an extended guidewire that can be used to exchange a catheter without removing the RVT CTO Device from the artery.
- a. Materials used in construction with patient contact: none
### F. Intended Use
The RVT CTO Device is indicated to facilitate the intra-luminal placement of conventional quidewires beyond peripheral artery chronic total occlusions. The device is contraindicated for use in carotid arteries. The RVT CTO Device is only intended for use with the RVT Control Unit.
The RVT Extension Wire is designed to extend the RVT CTO Device so that a catheter can be exchanged for another catheter.
### G. Predicate Device Comparison
The RVT CTO Device is substantially equivalent to the FlowCardia CROSSER (K072776, K091119) and Baylis Medical RF Tunneler (K051670). The devices have similar components with respect to materials, technological characteristics, intended use, and procedural steps as presented in the following table.
{2}------------------------------------------------
# ReVascular Therapeutics
·
#
| FEATURE | FlowCardia<br>CROSSER System | Baylis<br>PowerWire RF<br>Guidewire<br>(formerly RF<br>Tunneler Wire) | ReVascular Therapeutics<br>RVT CTO Device |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>Reference<br>Number | K072776 and K091119 | K051670 | K101599 |
| Indications for<br>Use | The CROSSER System is<br>indicated to facilitate the<br>intra-luminal placement of<br>conventional guidewires<br>beyond peripheral artery<br>chronic total occlusions. The<br>device is contraindicated for<br>use in carotid arteries. | The RF Tunneler<br>Wire is intended to<br>create a channel in<br>totally occluded<br>peripheral vessels<br>3mm or greater. | The RVT CTO Device is<br>indicated to facilitate the<br>intra-luminal placement of<br>conventional guidewires<br>beyond peripheral artery<br>chronic total occlusions.<br>The device is<br>contraindicated for use in<br>carotid arteries. |
| Design Features | The FlowCardia System<br>consists of reusable<br>electronics (Generator and<br>Transducer) and the single<br>use CROSSER Catheter.<br>The CROSSER Catheter is<br>connected to the electronics<br>by attaching the proximal<br>hub of the CROSSER to the<br>distal end of the Transducer.<br>The Generator converts AC<br>line power into high<br>frequency current. This<br>current is then delivered to<br>piezoelectric crystals<br>contained within the<br>transducer resulting in<br>crystal expansion and<br>contraction. The Transducer<br>horn amplifies the rapid<br>crystal expansion and<br>contraction propagating high<br>frequency mechanical<br>vibration down a Nitinol core<br>wire to the tip of the<br>CROSSER Catheter. This<br>vibrational energy provides<br>mechanical impact which<br>aids in the recanalization of | The RF Tunneler<br>Wire consists of a<br>core wire<br>surrounded with a<br>polymer insulation.<br>The wire connects<br>to a RF generator<br>at the proximal end<br>via a connector<br>cable, and has an<br>active tip at the<br>distal end to<br>delivery RF energy. | The RVT CTO Device has<br>a Nitinol Outer Shaft and<br>an internal Driveshaft<br>connected to an external<br>distal Active Tip. Active<br>rotational energy is<br>delivered to the distal<br>Active Tip via by the<br>internal Nitinol Driveshaft.<br>The proximal end of the<br>internal Nitinol Driveshaft is<br>connected to an external<br>Motor Housing and a<br>Control Unit. |
| FEATURE | FlowCardia<br>CROSSER System | Baylis PowerWire RF<br>Guidewire<br>(formerly RF<br>Tunnelor Wire) | Revascular Therapeutics<br>RVT-CTO Device |
| Main<br>Components | • Core Wire<br>• Catheter<br>• Transducer<br>• Reusable AC Powered<br>Generator<br>• Footswitch | • Nitinol Core<br>w/Insulating<br>Polymer Coating<br>• Proximal<br>Connector Cable<br>• Patient<br>Grounding Pad<br>• Reusable RF<br>Generator and Foot<br>Switch | • Driveshaft with Active<br>Tip<br>• Outer Shaft Assembly<br>• Motor Housing<br>• Control Unit<br>• Shaping Tool<br>• Extension Wire |
| Energy Type | Mechanical Energy | Radio Frequency<br>Energy | Mechanical Energy |
| Power Level | 20,000 cycles per second | 10 - 25 Watts | 13,000 revolutions per<br>minute |
| Mechanism of<br>Action | High frequency current<br>converted into vibrational<br>energy delivered at the distal<br>tip | RF energy<br>delivered at the<br>distal tip | Rotational energy<br>delivered at the distal tip |
| Shaft Size(s)<br>(OD) | 0.043" and 0.061" | 0.035" | 0.018" |
| Tip Diameter<br>(OD) | 0.043" and 0.061" | 0.028" | 0.017" |
| Tip Attributes | Steerable | Steerable, straight,<br>and angled | Steerable, shapeable |
| Working<br>Length(s) | 146 cm and 125 cm | 250 cm | 165 cm |
| Radiopacity | Yes | Yes | Yes |
| Guidewire<br>Exchange<br>Possible | Yes | Yes | Yes |
| Sterilization<br>Method | Gamma | Ethylene Oxide | Ethylene Oxide |
| Pyrogen Free | Yes | Yes | Yes |
| Single Use Only | Yes | Yes | Yes |
| Meets<br>applicable IEC<br>60601-1 Testing | Yes | Yes | Yes |
·
{3}------------------------------------------------
## ReVascular Therapeutics
## Traditional 510(k) Notification Chronic Total Occlusion Device
The subject and predicate systems provide "active" energy to power the distal Tip of the device through the occlusion. The ReVascular Therapeutics (RVT) CTO Device and its predicates are designed to cross Chronic Total Occlusions (CTO's) in the periphery by providing a means of energy at the distal tip. All the devices are similar in that they convert an external electrical energy source to energy at the device tip. The ReVascular Therapeutics CTO Device and the FlowCardia CTO Device (CROSSER) are designed to cross CTO's in the periphery by providing a means of
Appendix B, Page 4 of 7
{4}------------------------------------------------
mechanical energy at the distal tip and the Baylis PowerWire RF Guidewire System provides radio frequency energy to power the guidewire through the occlusion. Like the PowerWire System, the RVT System uses a guidewire-based platform, while the Crosser is catheter-based. All devices focus energy at the active tip to create channels in the target occlusion. Both the RVT and Baylis devices may be used as standard guidewire systems when used in their "passive" modes.
## H. Summary of Performance Data
The RVT CTO Device performance characteristics were evaluated in the following invitro bench and in-vivo animal studies:
- Dimensional Verification .
- . Torqueability
- ◆ Torque Strength
- . Active Tip Rotation
- . Tip Stiffness
- Catheter Compatibility
- Flexibility/Trackability ●
- Device Advancement .
- Friction: Static and Dynamic ●
- t Tip Temperature
- Tensile Testing .
- Electrical Safety and ● Electromagnetic Testing
- · Tip Shaping
- · Coating Adhesion Integrity
- · Support Catheter Removal and Advancement Over Extension
- · Coil Integrity
- · Packaging Testing
- · Shipping Testing
- · Sterility Testing
- · Shelf Life Testing
- · Evaluation of safety and performance of the RVT CTO Device in the porcine model
- · Biocompatibility
Results of the non-clinical testing demonstrate that the materials chosen, the manufacturing process, and design of the RVT CTO Device meet the established specifications necessary for consistent performance during its intended use. In addition, the testing demonstrates that the RVT CTO Device does not raise new questions of safety or effectiveness when compared to the predicate devices.
### l. Summary of Clinical Data
The RVT CTO Device was the subject of an investigational study, the Re-OPEN Study, a prospective, multi-center, single-arm (non-randomized) study conducted to evaluate the device in the treatment of infrainguinal target vessel occlusions classified angiographically as absolute (100% occlusion with no flow), where no antegrade filling beyond the occlusion was visible.
The purpose of the study was to demonstrate that the RVT CTO Device can safely and effectively facilitate the crossing of chronic total infrainguinal artery occlusions
Appendix B, Page 5 of 7 .
{5}------------------------------------------------
following demonstration of resistance to crossing with conventional guidewire techniques or prior failed treatment attempt. In addition, this study was intended to evaluate the impact of successfully crossing a CTO and allowing subsequent treatment using interventional methods rather than bypass surgery or medical management alone.
A total of 85 patients were enrolled across 3 medical centers in Europe between January and December 2009. The population in this study was composed of patients who presented with Peripheral Artery Disease who met all eligibility criteria and exhibited any of the following conditions: ABI of less than 1.0, abnormal duplex ultrasound, claudication, leg numbness or weakness, leg rest pain, leg skin color changes or non-healing ulcers of the legs or feet, or previously documented lower extremity artery occlusions. Each patient must have had a prior failed attempt or a concurrent and reasonable attempt (at least 5 minutes) during the procedure to demonstrate resistance to conventional guidewire crossing. The occlusions treated under this protocol were all confirmed to be 100% CTOs that had been in existence for 30 days or greater.
The Primary Safety Endpoint was defined as freedom from clinical perforation (any perforation requiring treatment) of the index lesion to 30 days. There were no (0.0%) clinical perforations reported at time of intervention with use of the RVT CTO Device and only 1 (1.2%) patient experienced a clinical perforation as a result of a cutting balloon and was successfully treated with stent placement. No further clinical sequelae were reported. This event was considered by the Clinical Events Committee (CEC) to be unlikely related to the RVT CTO Device with a definite relation to the procedure and a probable relation to a preexisting condition.
All SAEs (Serious Adverse Events) were adjudicated by an independent CEC. In addition, the CEC reviewed all AEs in aggregate to ensure that there were no trends or AEs occurring in greater severily or degree of incidence than anticipated for this patient population. There were no AEs reported that were considered unanticipated in nature, severity or degree of incidence.
The Primary Clinical Efficacy Endpoint, defined as advancement of the RVT CTO Device through the CTO and subsequent distal guide wire positioning was achieved in 65 out of 85 cases (76.5%). This allowed subsequent positioning of conventional guide wire(s) distal to the vessel occlusion,
The Primary Performance (Technical Success) Endpoint, defined as facilitation of crossing the CTO with any guide wire was achieved in 68 of 85 patients (80%).
{6}------------------------------------------------
In conclusion, the ReOPEN Study provides strong evidence that the RVT CTO Device is safe and effective at facilitating crossing of Chronic Total Occlusions (CTOs) in patients with Peripheral Vascular Disease (PAD) where conventional guidewires have failed. The results demonstrated that the RVT CTO device was able to successfully facilitate crossing in the majority of cases where a conventional guidewire was unsuccessful and use of the RVT CTO did not increase the rate of perforations beyond the rate seen with use of a conventional quidewire.
In addition to the non-clinical testing and descriptive comparisons, the data from the ReOPEN study provide further evidence that use of the RVT CTO Device does not raise new questions of safety or effectiveness when compared to the predicate devices.
#### J. Summary of Data
Non-clinical testing of the subject device has shown that the RVT CTO Device meets the functional, performance, and design requirements derived from a variety of design inputs; including specifications resulting from examination of the predicate devices, external standards and clinician input. In addition to the non-clinical testing and descriptive comparisons, the data from the ReOPEN study provide further evidence that use of the RVT CTO device does not raise new questions of safety or effectiveness when compared to the predicate devices, as demonstrated by a comparison of the investigational study results reported for the FlowCardio predicate and the subject device. The studies were similar in design, both were prospective clinical studies of 85 patients designed to assess the safety and efficacy of the devices in the treatment of infrainguinal target vessel occlusions classified angiographically as absolute (100% occlusion with no flow), where no antegrade filling beyond the occlusion was visible and following demonstration of resistance to crossing with conventional guidewire techniques. The inclusion/exclusion criteria were the same between the two studies as were the study endpoints and results.
.
The RVT CTO Device is substantially equivalent to the predicate device for intended use, physical characteristics, anatomical sites, performance, safety characteristics, and labeling. The RVT CTO Device incorporates similar design features, procedural steps, and performance characteristics when compared to the predicate devices. Safety and performance testing, combined with the technical equivalency assessment and clinical study results, demonstrate that the RVT CTO Device performs as intended and does not raise any new issues of safety and effectiveness when compared with the predicate devices.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is a stylized image of an eagle or bird-like figure with three overlapping profiles, creating a sense of depth and movement. The overall design is simple and conveys a sense of authority and professionalism.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
ReVascular Therapies Ms. Nancy Lincé 766 Palomar Avenue Sunnyvale, CA 94085
SEP 1 8 2013
Re: K101599
Trade/Device Name: ReVascular Therapies Chronic Total Occlusion (RVT CTO) Device Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: PDU Dated: January 7, 2011 Received: January 10, 2011
Dear Ms. Lincé:
This letter corrects our substantially equivalent letter of January 18, 2011.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{8}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Z.M.Z
Ar Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known): K101599
Device Name: RVT CTO Device
## Indications for Use:
The RVT CTO Device is indicated to facilitate the intra-luminal placement of conventional guidewires beyond peripheral artery chronic total occlusions. The device is contraindicated for use in carotid arteries.
The RVT Extension Wire is designed to extend the RVT CTO Device so that a catheter can be exchanged for another catheter.
| Prescription Use | X |
|------------------|---|
|------------------|---|
OR (per 21 CFR 801.109) Over-The-Counter Use_
|--|
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
vision of Cardiovascular Devices
510(k) Number
ー
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.