K101522 · Ultra Seal Corporation · KMJ · Dec 23, 2010 · General Hospital
Device Facts
Record ID
K101522
Device Name
ULTRA SEAL STERILE LUBRICATING JELLY
Applicant
Ultra Seal Corporation
Product Code
KMJ · General Hospital
Decision Date
Dec 23, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6375
Device Class
Class 1
Indications for Use
Ultra Seal Sterile Lubricating Jelly is a medical device intended for medical purposes to lubricate body orifices to facilitate the entry of diagnostic and therapeutic devices and gloved fingers. For lubrication to provide easy insertion of catheters, endoscopes, or gloved fingers into bodily orifices. The device is intended for use on order of a physician. Non-prescription over-the-counter use.
Device Story
Ultra Seal Sterile Lubricating Jelly is a water-based, high-viscosity, sterile gel used to lubricate body orifices for the insertion of diagnostic/therapeutic devices or gloved fingers. The device is supplied in aluminum tubes or laminated film packets, terminally sterilized via gamma radiation. It is intended for use in clinical settings or by patients (OTC) to reduce friction during medical procedures. The gel is non-sticky, non-greasy, and non-staining. By providing lubrication, it facilitates easier, more comfortable device insertion, potentially reducing patient discomfort during procedures involving catheters or endoscopes. The product does not contain spermicides or contraceptives.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including biocompatibility studies (Delayed Contact Dermal Sensitization/Buehler Method, In Vitro Cytotoxicity L929 Agar Overlay, Primary Dermal Irritation in Rabbits) and stability testing. All ingredients are GRAS and have high LD50 values, supporting safety for intended use.
Technological Characteristics
Water-based, high-viscosity gel. Ingredients: Water, Glycerin, Carbomer, Disodium EDTA, Sodium Hydroxide, Methylparaben, Propylparaben. Packaging: Aluminum tubes or foil/polyethylene/paper laminated packets. Sterilization: Gamma radiation. No electronic or software components.
Indications for Use
Indicated for lubrication to facilitate insertion of catheters, endoscopes, or gloved fingers into bodily orifices. Intended for use on physician order; available for over-the-counter use.
Regulatory Classification
Identification
A patient lubricant is a device intended for medical purposes that is used to lubricate a body orifice to facilitate entry of a diagnostic or therapeutic device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a handwritten date, "12/03/2010". The numbers are written in a cursive style, with the month and day separated by a forward slash. The year is also separated from the day by a forward slash. The handwriting is clear and legible.
DEC 2 3 2010
# Section 1
In Support of 510(k) K101522 - Ultra Seal Corporation Sterile Lubricating Jelly
## 510(k) Summary (Amended-Version 3) Ultra Seal Sterile Lubricating Jelly K101522 I. General Information on Submitter
Company: Ultra Seal Corporation 521 Main Street New Paltz, N.Y. 12561 Registration Number: 1317759 Contact: James Davis Executive Vice President, Regulatory Affairs Telephone: 845-691-8361 ext. 117 Fax: 845-691-8360 Date: July 12, 2010 .
Manufacturing Site: Ultra Seal Corporation 521 Main Street New Paltz, NY 12561 Registration Number: 1317759
I ILTRASEAL
Device Class: Class I (reserved) Trade Name: Ultra Seal Sterile Lubricating Jelly Common Name: Personal Lubricant Classification Name: Patient Lubricant C.F.R. section: 21 CFR section 880.6375. Classification Panel: General Hospital
New Device's Name: Ultra Seal Sterile Lubricating Jelly Predicated Device(s):
Triad Lubricating Jelly, Manufactured by HP Industries Originally, (K871169) Sheffield Pharmaceuticals Sterile Lubrigel, Manufactured by Sheffield Pharmaceuticals, Inc. (K073684).
Information supporting claims of substantial equivalence, as defined under the Federal Food Drug and Cosmetic Act, with respect to safety and effectiveness is summarized below. For the convenience of the reviewer, this summary is formatted in accordance with the Agency's final rule, "510(k) Summaries and 510(k) Statements" (21 CFR 807).
501(k) Summary creative contract packaging™ p. 1 of 3
521 Main Street New Paltz, NY 12561 • 1609 914 • 255 • 2490 Fax 914 • 255 • 3553
{1}------------------------------------------------
#### Intended Use:
Ultra Seal Sterile Lubricating Jelly is a medical device intended for medical purposes to lubricate body orifices to facilitate the entry of diagnostic and therapeutic devices and gloved fingers.
#### Device Description:
Ultra Seal Sterile Lubricating Jelly patient lubricant is a water- based, clear, colorless, non-sticky, non-greasy, non-staining, nonirritating patient lubricant. It is a water soluble, high viscosity gel-like liquid for use as patient lubricant when a sterile field is required. Each tube or packet is terminally sterilized by gamma radiation. The product is sterile unless package is opened, damaged, or the seal for the packet product is not intact, with label directions to discard after use.
The product is packaged in a convenient 2 oz., 4.0 oz., and 4.2 oz aluminum tube with a cap and a puncture seal blind, or a 2 oz., 4.0 oz., and 4.2 oz foil laminate tube with a flip top and a peel seal, and 3 and 5 gram laminated film packets, the lamination being paper, polyethylene, foil, polyethylene. All tube configurations contain an aluminum barrier to preserve the product's properties and sterility. Stability has been performed on all tubes and packets and data from 3 months of accelerated studies supports an initial expiration term of 24 months from the date of manufacture. Expiration dating will be confirmed through concurrent room temperature stability studies.
This product is not a contraceptive and does not contain a spermicide.
#### Regulatory Status:
As per 21CFR, 880.6375, Patient Lubricant is defined as a Class I medical device intended for medical purposes that is used to lubricate a body orifice to facilitate entry of a diagnostic or therapeutic device. Patient lubricants are not exempt from 510(k) clearance.
### Technological Characteristics:
Ultra Seal Sterile Lubricating Jelly formula is proprietary. The product has no exceptional technological characteristics and consists of safe water soluble ingredients commonly used in topical and ingested products and which are similar or the same as those used in similar to PDI (Sterile) Lubricating Jelly, E-Z Lubricating Jelly, and Sheffield Pharmaceuticals Sterile Lubrigel (K073684) currently on the market.
Ultra Seal Sterile Lubricating Jelly contains all GRAS ingredients: Water, Glycerin, Carbomer, Disodium EDTA, and Sodium Hydroxide (for pH adjustment), with Methylparaben and Propylbaraben as preservatives.
#### Biocompatibility
Biocompatibility Studies on Ultraseal Sterile Lubricating Jelly was conducted by outside laboratories, in compliance with Good Laboratory Practices (GLPs) demonstrated: In Delayed Contact Dermal Sensitization Test (Buehler Method) the product was considered a non-sensitizer.
In Vitro Cytotoxicity L929 Agar Overlay Test the product meet the requirements for the test.
Primary Dermal Irritation in Rabbits determined that the product is not a dermal irritant.
All ingredients have high LD50s as summarized in the table below which converts the Dermal Toxicity numbers (when available) based on the percent of inclusion in the formulation: Carbomer >2gm/Kg Rabbit Dermal X 1% inclusion=200gm of gel/Kg
Methyl Paraben >500 mg/Kg rat subcutaneous X .05% inclusion= 1000 gm of gel/Kg
{2}------------------------------------------------
Propyl Paraben >1.65 gm/Kg mouse subcutaneous X .05% inclusion= 3300 mg of gel/Kg Disodium EDTA > Not Available Sodium Hydroxide>Used as a neutralizer to adjust pH to ~6.8-Neutralized
> ー・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ . -- . --
In a worst case scenario, an average person of 70 Kg would have to apply 14 Kg of gel to approach the LD50 of the most dermally-irritating ingredient. The average application of the gel is 3-5 gm or about 0.3% of this amount.
These ingredients are also used in many OTC and prescription oral remedies.
Ultra Seal Sterile Lubricating Jelly is, therefore, safe for its intended use and substantially equivalent to Sheffield Pharmaceuticals Sterile Lubrigel (K073684) currently on the market.
#### Sterility:
Ultra Seal Sterile Lubricating Jelly is undergoing the Sterility testing validation mandated by FDA, guided by the Association for the Advancement of Medical Instrumentation (AAMI), and detailed in the attached protocol for sterilization of Ultra Seal (Sterile) Lubricating Jelly by Gamma Radiation and generated by Steris Isomedix, Inc.
510(k) INDICATIONS FOR USE FORM (Replica of FDA Form)
510 (k) Number (if known): K101522 Device Name: Ultra Seal Sterile Lubricating Jelly Indications for Use: For lubrication to provide easy insertion of catheters, endoscopes, or gloved fingers into bodily orifices. The device is intended for use on order of a physician.
Non-prescription over-the-counter use.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three wing-like shapes above a wavy line, representing water. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. James D. Davis Executive Vice President Ultra Seal Corporation 521 Main Street New Paltz, New York 12561
PEC 2 3 2"
Re: K101522
Trade/Device Name: Ultra Seal Sterile Lubricating Jelly Regulation Number: 21 CFR 880.6375 Regulation Name: Patient Lubricant Regulatory Class: I Product Code: KMJ Dated: December 7, 2010 Received: December 8, 2010
Dear Mr. Davis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2- Mr. Davis
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffice s/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
James J. Koszarek
for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology. General Hospital, Infection Control and Dental Devices Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Ultra Seal Corporation. The words "ULTRASEAL" are in large, bold, black letters. Underneath the word "ULTRASEAL" is the word "CORPORATION" in smaller letters. To the right of the word "ULTRASEAL" is a square made of horizontal lines.
.521. Main Street, New Paltz, NY 12561. 1609 . . 845.255.2490 . . Fax: 845.255-3553
## Indications for Use
510(k) Number (if known): K 1015 22
. .
DEC 2 3 2010
Device Name: Ultra Seal Sterile Lubricating Jelly
Indications for Use: For lubrication to provide easy insertion of catheters, endoscopes, or gloved fingers into bodily orifices.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Richard C. Chapman 12/22/10
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: k101522
creative contract packaging™
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.