← Product Code [LZO](/productcode/LZO) · K101472

# MULTIFIT TOTAL HIP SYSTEM (K101472)

_Otis Biotech Co., Ltd. · LZO · Jan 5, 2011 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K101472

## Device Facts

- **Applicant:** Otis Biotech Co., Ltd.
- **Product Code:** [LZO](/productcode/LZO.md)
- **Decision Date:** Jan 5, 2011
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 888.3353
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The MultiFit Total Hip System is intended to be cemented stem or uncemented stem use. The MultiFit Total Hip System is to replace a defective hip joint in the following cases: 1) Patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head and nonunion of previous fractures of the femur. 2) Patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital femoral epiphysis 3) Patients suffering from disability due to previous fusion 4) Patients with acute femoral neck fractures.

## Device Story

Total hip joint prosthesis; replaces defective hip joints. Components: femoral stems (Ti6Al4V for cementless; CoCrMo for cemented), metal femoral heads (CoCrMo), acetabular cups (Ti6Al4V ELI), UHMWPE liners, and bipolar cups (CoCrMo). Used in surgical settings by orthopedic surgeons. Implants restore joint function; alleviate pain; improve mobility. Single-use only; provided sterile.

## Clinical Evidence

Bench testing only. Testing included: ASTM F1820 (locking mechanism), ASTM F2009 (ball pull-off), ASTM F543 (bone screw pullout/torsion), ISO 21535 (range of motion), ISO 14242 (wear), ISO 7206 (fatigue), and ASTM F2102/F2003 (UHMWPE oxidation/aging). No clinical data provided.

## Technological Characteristics

Materials: Ti6Al4V, Ti6Al4V ELI, CoCrMo, UHMWPE. Standards: ASTM F1820, F2009, F543, F2102, F2003; ISO 21535, 14242, 7206. Components: Femoral stems, acetabular cups, bipolar cups, femoral heads, liners. Sterilization: Provided sterile. Connectivity: None (mechanical implant).

## Regulatory Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

## Predicate Devices

- Summit™ DuoFix™ Hip Prosthesis ([K011489](/device/K011489.md))
- Osteonics Imnifit HA Hip stem series ([K011489](/device/K011489.md))
- PERFECTA® IMC Hip System ([K972641](/device/K972641.md))
- Synergy Cemented Hip stem ([K972641](/device/K972641.md))
- Trilogy Acetabular System ([K934765](/device/K934765.md))
- Reflection Cross-linked UHMWPE acetabular component ([K934765](/device/K934765.md))
- Self-centering Hip prosthesis ([K812672](/device/K812672.md))
- Ringloc Bi-polar acetabular component ([K051569](/device/K051569.md))
- Total Hip Femoral Head - 12/14 Taper ([K021673](/device/K021673.md))
- Biomet Cobalt-chrome femoral components ([K911684](/device/K911684.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KI01472

Premarket Notification 510(k) 5. 510(k) Summary

# 5. 510(k) Summary

[As Required by 21 CFR 807.87(h) & 21 CFR 807.92]

#### 1. Submission information

Name of Company:

OTIS Biotech Co., Ltd FDA registration No. 3005140381 #514 Sihwa industrial complex 2Ba Block Jeongwang-Dong, Siheung-si Gyounggi-Do, Republic of Korea 429-450 Tel: 82-31-319-0406 Fax: 82-31-319-0414

Contact: srinivasareddy krotha

E-mail: srinivasareddy@otisbiotech.com Tel: 82-31-414-4618 Fax: 82-31-411-1076

Date Prepared: 10-05-2010

## 2. Device Identification

| Trade Name:     | MultiFit <sup>TM</sup> Total Hip System            |
|-----------------|----------------------------------------------------|
| Common Name:    | Total Hip Joint Prosthesis                         |
| Classification: | Class II (Special Control)                         |
|                 | JDI 888.3350 - Prosthesis, Hip, Semi-Constrained,  |
|                 | Metal/Polymer, Cemented.                           |
|                 | LZO 888.3353 -Hip joint metal/ceramic/polymer semi |
|                 | -constrained cemented or nonporous un cemented     |
|                 | prosthesis                                         |
|                 | KWY 888.3390 - Prosthesis, Hip, Hemi-, Femoral,    |
|                 | Metal/Polymer, Cemented or Uncemented              |
|                 | KWZ 888.3310 - Prosthesis, Hip, Constrained,       |
|                 | Cemented or uncemented, Metal/Polymer              |

### 3. Substantial Equivalence Predicate_Legally Marketed Devices

The substantial equivalence of this device is based on equivalence in

5 - 1

| Artificial Joint Research Center |
|----------------------------------|
| OTISBiotech co., Ltd.            |

{1}------------------------------------------------

MultiFit ™ Total Hip System

intended use, materials, designs and operational principles to the below listed predicate devices.

| OTIS Biotech Co., Ltd<br>MultiFit™ Total Hip System | Predicate Device                                                                                                                       | 510(k)<br>Approval<br>Number | Product<br>Code |
|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|------------------------------|-----------------|
| Cementless Tapered Stem                             | i ) Depuy Orthopaedicss, Inc.<br>Summit™ DuoFix™ Hip Prosthesis<br>ii) Osteonics corporation.<br>Osteonics Imnifit HA Hip stem series  | K011489                      | LPH             |
| Cement Collared Stem                                | i ) Wright Medical Technology INC.<br>PERFECTA® IMC Hip System<br>ii) Smith & Nephew, Inc.<br>Synergy Cemented Hip stem                | K972641                      | JDI             |
| Acetabular System                                   | i ) Zimmer, Inc.<br>Trilogy Acetabular System<br>ii) Smith & Nephew, Inc.<br>Reflection Cross-linked UHMWPE<br>acetabular component    | K934765                      | LPH             |
| Bipolar System                                      | i ) Depuy, Inc.<br>Self-centering Hip prosthesis<br>ii) Biomet, Inc.<br>Ringloc Bi-polar acetabular<br>component                       | K812672                      | KWY             |
|                                                     |                                                                                                                                        | K051569                      | JID             |
| Femoral Ball Head                                   | i ) Smith & Nephew, Inc.<br>Total Hip Femoral Head - 12/14<br>Taper<br>ii ) Biomet, Inc.<br>Biomet Cobalt-chrome femoral<br>components | K021673                      | LZO             |
|                                                     |                                                                                                                                        | K911684                      | JDI             |
| Bone Screw                                          | Zimmer, Inc. Trilogy Acetabular System                                                                                                 | K934765                      | LPH             |

Artificial Joint Research Center © OTISBiotech co., Ltd.

Page 1 of 5

{2}------------------------------------------------

### 4. Device Description

The MultiFit™ Total Hip System is total hip joint prosthesis which consists of stems, acetabular system, bipolar system and femoral head. Stems are available with two femoral designs. One is manufactured from Ti6A14V alloy which intended for non-cemented use. The other femoral component design is manufactured from CoCrMo and is intended for cemented use. All implant devices are designed for single use only and provided with separated sterilized package. Various sizes are available for each component.

Femoral stems - Cementless stem (sizes 130 mm to 150 mm) and cement collared stem (sizes 110 mm to 128 mm) are intended to be used with the other components of the The MultiFit™ Total Hip

Metal femoral heads - Metal femoral heads (22 mm and 28 mm in diameter) which is fabricated from Cobalt Chromium Molybdenum (CoCrMo) are intended to be used in conjunction with the commercially available press-fit Ti6Al4V or Co-Cr-Mo alloy Hip Stems cement type and cementless.

Acetabular cups and UHMWPE liners - Acetabular cup is manufactured from Ti-6Al-4V ELI and Liner is manufactured from Ultra-High Molecular Weight Polyethylene (UHMWPE). The UHMWPE insert (sizes 22 mm and 28 mm) are intended to be used in conjunction with MultiFit™ Total Hip System acetabular cups (sizes 42 mm to 60 mm in 2 mm increments) in cementless applications. UHMWPE liners are available in two types of Flat and 10° elevated.

Bipolar cups and liners - Bipolar cup is made of Cobalt Chromium Molybdenum (CoCrMo) and liner is made of Ultra-High Molecular Weight Polyethylene (UHMWPE) same as acetabular liner. Bipolar cups (sizes 42 mm to 55 mm 1 mm increments) are intended to be used with the bipolar liners (22 mm and 28 mm in diameter) of the MultiFitTM Total Hip System.

Artificial Joint Research Center COTISBiotech co., Ltd.

5 - 3

Page 3 of 5

{3}------------------------------------------------

#### 5. Indications for Use

The MultiFit Total Hip System is intended to be cemented stem or uncemented stem use.

The MultiFit Total Hip System is to replace a defective hip joint in the following cases:

1) Patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head and nonunion of previous fractures of the femur.

2) Patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital femoral epiphysis

3) Patients suffering from disability due to previous fusion

4) Patients with acute femoral neck fractures

#### 6. Statement of Technological Comparison

Bench testing as listed in Section 15 and Appendix D. was conducted in accordance with standard. It demonstrates substantial equivalence to the above listed predicate devices in terms of materials, design, indications for use and operational principles.

Artificial Joint Research Center OTISBiotech co., Ltd.

5 - 4

{4}------------------------------------------------

## 7. List of the bench tests conducted for the substantial equivalence to the predicate devices

KID 1477

- 1. Ace tabular system Push out test of Locking Mechanism according to ASTM F1820
- 2. Ball pull off test according to ASTM F2009
- 3. Bone screw axial pullout strength test according to ASTM F543
- 4. Bone screw torsion strength test according to the ASTM F543
- 5. Range of motion test according to the ISO 21535
- 6. Wear test according to the ISO 14242-1:2002-03 and ISO 14242-2:2000-10-()રે
- 7. Fatigue test according to the ISO 7206-4(2002) Implants for surgery Partial and total hip joint prostheses - Part4: Determination of endurance properties of stemmed femoral components.
- 8. Fatigue test according to the ISO 7206-6(1992)- Implants for surgery -- Partial and total hip joint prostheses -- Part 6: Determination of endurance properties of head and neck region of stemmed femoral components
- 9. Oxidation index and accelerated ageing of UHMWPE according to ASTM F2102-01el and ASTM F2003-02
- 10. Ace tabular liner torsion test
- 11. Ace tabular system Lever out test
- 12. Bipolar system Push out test between ball and UHMWPE liner
- 13. Bipolar system Lever out test between ball and UHMWPE liner

{5}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with outstretched wings, rendered in a simple, abstract design.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

OTIS Biotech Co., Ltd % Mr. Srinivasareddy Krotha 514 Sihwa Industrial Complex 2Ba Block Jeongwang-Dong, Siheung-si Gyounggi-Do, Republic of Korea 429-450

JAN 5 2011

Re: K101472 Trade/Device Name: Multifit Total Hip System Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis. Regulatory Class: Class II Product Code: LZO, JDI, KWY, KWZ Dated: December 24, 2010 Received: December 30, 2010

#### Dear Mr. Krotha

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

and the comments of the comments of

and the contraction of the country of the country of

{6}------------------------------------------------

Page 2 - Mr. Srinivasareddy Krotha

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Ar
B. rh
fr.

Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

K10472

## 4. Indications for Use

510(k) Number:

Device Name: MultiFit™ Total Hip System

The hip system is intended to be cemented stem or un cemented stem use. Indications for Use:

The MultiFit Total Hip System is to replace a defective hip joint in the following cases:

- l ) Patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head and nonunion of previous fractures of the femur;
- 2) Patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital femoral epiphysis;
- 3) Patients suffering from disability due to previous fusion and previously failed endo-prostheses;
- 4) Patients with acute femoral neck fractures.

Prescription Use Yes (Part 21 CFR 801 Subpart D) Over-The-Counter Use No (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

for M. Mikkerson

(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number_

Artificial Joint Research Center OTISBiotech co., Ltd.

---

**Source:** [https://fda.innolitics.com/device/K101472](https://fda.innolitics.com/device/K101472)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
