Vitala™ Resorbable Natural Collagen Membrane is intended for use during the process of guided bone regeneration (GBR) and guided tissue regeneration (GTR) as a biodegradable barrier for: Simultaneous use with implants; Augmentation around implants placed in immediate extraction sockets; Augmentation around implants placed in delayed extraction sockets; Localized ridge augmentation for later implantation; Alveolar ridge reconstruction for prosthetic treatment; Alveolar ridge preservation consequent to tooth extraction; Filling of bone defects after root resection, cystectomy, removal of retained teeth; Over the window in lateral window sinus elevation procedures; Furcation defects in multi-rooted teeth; Treatment of recession defects, together with coronally positioned flap; In implants with vertical bone loss due to infection, only with satisfactory debridement and implant surface disinfection; Guided bone regeneration in dehiscence defects; and Guided tissue regeneration in periodontal defects.
Device Story
Vitala™ is a resorbable natural collagen membrane derived from porcine sources; used as a physical barrier in oral surgery to facilitate guided bone regeneration (GBR) and guided tissue regeneration (GTR). The membrane is cell-occlusive, implantable, and hemostatic; it prevents soft tissue ingrowth into bone defects, allowing bone regeneration. It is applied by dental surgeons during procedures such as implant placement, ridge augmentation, and sinus elevation. The device is provided sterile in a double-tray configuration. By maintaining space and excluding non-osteogenic cells, the membrane supports the healing of bone and periodontal defects, potentially improving clinical outcomes for patients undergoing dental reconstruction.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and material similarity to legally marketed predicate devices.
Technological Characteristics
Resorbable natural collagen membrane sourced from porcine tissue. Cell-occlusive, hemostatic, and implantable. Sterilized via E-beam irradiation. Supplied in various dimensions (e.g., 13mm x 25mm to 40mm x 50mm).
Indications for Use
Indicated for patients requiring guided bone or tissue regeneration in oral surgical procedures, including implant placement, ridge augmentation, ridge preservation, and treatment of periodontal or bone defects.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{1}------------------------------------------------
K101453
### JUL 2 1 2011
..
# 510(k) Summary
| Date Prepared: | May 19, 2010 |
|--------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Sponsor : | OBI Biologics, Inc.<br>4620 71st Street, Bldg 79<br>Lubbock, TX 79424 |
| FDA Establishment Registration #: | N/A (registration after 510(k) submission approval) |
| Contact Person: | Dustyn Webb<br>Director of Regulatory Affairs<br>806-796-1923 (phone)<br>806-796-0059 (fax)<br>dustyn@cytoplast.com (email) |
| Proprietary Name: | Vitala™ |
| Common Name: | Resorbable Natural Collagen Membrane |
| Regulatory Classification Name:<br>(Regulation Number)<br>[Product Code] | Bone Grafting Material<br>(872.3930)<br>[NPL] |
| Device Classification: | Class II<br>(Class II Special Controls Guidance Document: Dental Bone Grafting Material<br>Device) |
| Predicate Devices: | BIO-GIDE® (K050446)<br>CopiOs™ Pericardium Membrane (K073097)<br>Ossix™-Plus (K053260) |
#### Device Description
Vitala™ Resorbable Natural Collagen Membrane is a natural collagen membrane sourced from veterinary certified pigs. Vitala™ Resorbable Natural Collagen Membrane, minimally manipulated without further cross-linking, is sterilized in a double tray package configuration via E-beam irradiation.
#### Indications for Use
Vitala''' Resorbable Natural Collagen Membrane is intended for use during the process of guided bone regeneration (GBR) and guided tissue regeneration (GTR) as a biodegradable barrier for:
- Simultaneous use with implants; .
- Augmentation around implants placed in immediate extraction sockets; .
- Augmentation around implants placed in delayed extraction sockets; .
- Localized ridge augmentation for later implantation; .
- Alveolar ridge reconstruction for prosthetic treatment; .
- * Alveolar ridge preservation consequent to tooth extraction;
- . Filling of bone defects after root resection, cystectomy, removal of retained teeth;
- � Over the window in lateral window sinus elevation procedures;
- . Furcation defects in multi-rooted teeth;
- . Treatment of recession defects, together with coronally positioned flap;
- In implants with vertical bone loss due to infection, only with satisfactory debridement and implant surface . disinfection;
- Guided bone regeneration in dehiscence defects; and .
- Guided tissue regeneration in periodontal defects. t
OBI Biologics, Inc. Vitolo™ Resorbable Natural Collagen Membrane 510{k) Premarket Notification Submission
{2}------------------------------------------------
#### Substantial Equivalence Comparison Chart
| Property | Vitala™ Resorbable Natural Collagen Membrane<br>(K101453)<br>OBI Biologics, Inc. | BIO-GIDE®<br>(K050446)<br>Geistlich-Pharma | CopiOs™ Pericardium Membrane<br>(K073097)<br>RTI Biologics, Inc. | Ossix™-Plus<br>(K053260)<br>Colbar Life Science, Ltd. |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SIMILAR<br>MODELS/SIZES | 13mm x 25mm, 15mm x 20mm, 20mm x 30mm,<br>30mm x 40mm, 40mm x 50mm | 13mm x 25mm, 25mm x 25mm,<br>40mm x 50mm | 15mm x 20mm, 20mm x 30mm,<br>30mm x 40mm | 15mm x 25mm, 25mm x 30mm,<br>30mm x 40mm |
| SIMILAR<br>INTENDED USE | Vitala™ Resorbable Natural Collagen Membrane<br>is intended for use during the process of guided<br>bone regeneration (GBR) and guided tissue<br>regeneration (GTR) as a biodegradable barrier<br>for:<br>• Simultaneous use with implants;<br>• Augmentation around implants placed in<br>immediate extraction sockets;<br>• Augmentation around implants placed in<br>delayed extraction sockets;<br>• Localized ridge augmentation for later<br>implantation;<br>• Alveolar ridge reconstruction for prosthetic<br>treatment;<br>• Alveolar ridge preservation consequent to<br>tooth extraction;<br>• Filling of bone defects after root resection,<br>cystectomy, removal of retained teeth;<br>• Over the window in lateral window sinus<br>elevation procedures;<br>• Furcation defects in multi-rooted teeth;<br>• Treatment of recession defects, together<br>with coronally positioned flap;<br>• In implants with vertical bone loss due to<br>infection, only with satisfactory<br>debridement and implant surface<br>disinfection;<br>• Guided bone regeneration in dehiscence<br>defects; and<br>• Guided tissue regeneration in periodontal<br>defects. | BIO-GIDE® is recommended<br>for:<br>• Simultaneous use of GBR-<br>membrane (BIO-GIDE)<br>and implants;<br>• Augmentation around<br>implants placed in<br>immediate extraction<br>sockets;<br>• Augmentation around<br>Implants placed in<br>delayed extraction<br>sockets;<br>• Localized ridge<br>augmentation for later<br>implantation;<br>• Alveolar ridge<br>reconstruction for<br>prosthetic treatment;<br>• Filling of bone defects<br>after root resection,<br>cystectomy, removal of<br>retained teeth;<br>• Guided bone regeneration<br>in dehiscence defects; and<br>• Guided tissue<br>regeneration in<br>periodontal defects. | This membrane is intended for use<br>in oral surgical procedures as a<br>resorbable material for<br>augmentation around implants<br>placed in delayed extraction<br>sockets; localized ridge<br>augmentation for later<br>implantation; alveolar ridge<br>reconstruction for prosthetic<br>treatment; filling of bone defects<br>after root resection, cystectomy,<br>removal of retained teeth; guided<br>bone regeneration in dehiscence<br>defects and guided tissue<br>regeneration procedures in<br>periodontal defects. | Ossix™-Plus biodegradable<br>Collagen membrane is intended<br>For use during the process of<br>guided bone regeneration (GBR)<br>and guided tissue (GTR) as a<br>biodegradable barrier for:<br>• Ridge augmentation for later<br>implant insertions.<br>• Simultaneous ridge<br>augmentation and implant<br>insertions.<br>• Ridge augmentation around<br>implants inserted in delayed<br>extraction sites.<br>• Ridge augmentation around<br>implants inserted in<br>immediate extraction sites.<br>• Alveolar ridge preservation<br>consequent to tooth (teeth)<br>extraction(s).<br>• Over the window in lateral<br>window sinus elevation<br>procedures.<br>• In implants with vertical<br>bone loss due to infection,<br>only in cases where<br>satisfactory debridement<br>and implant surface<br>disinfection can be achieved.<br>• In intrabony defects around<br>teeth.<br>• For treatment of recession<br>defects, together with<br>coronally positioned flap.<br>• In furcation defects in multi-<br>rooted teeth. |
| INCORPORATES<br>SAME BASIC DESIGN | Yes | Yes | Yes | Yes |
| UTILIZES THE SAME<br>OPERATING<br>PRINCIPLE | Cell-occlusive<br>Implantable<br>Resorbable<br>Hemostatic | Cell-occlusive<br>Implantable<br>Resorbable<br>Hemostatic | Cell-occlusive<br>Implantable<br>Resorbable<br>Hemostatic | Cell-occlusive<br>Implantable<br>Resorbable<br>Hemostatic |
| INCORPORATES<br>SIMILAR MATERIALS | Yes, porcine collagen | Yes, porcine collagen | Yes, bovine collagen | Yes, porcine collagen |
| STERILIZATION<br>PROCESS | Irradiation | Irradiation | Irradiation | Irradiation |
| BIOCOMPATIBLE | Yes | Yes | Yes | Yes |
| NON-PYROGENIC | Yes | Yes | Yes | Yes |
#### Basis for Substantial Equivalence
The Vitala™ Resorbable Natural Collagen Membrane consists of material (porcine collagen) that is very similar in material composition to the predicate devices, BIO-GIDE® (porcine collagen), contection Membrane (bovine collagen) and Ossix™ -Plus (porcine collagen). Design, function and intended use are substantially equivalent to the corresponding characteristics of the predicate devices. Although minor differences exist in the concessing, medical device packaging, and handling characteristics among Vitala™ Resorbable Natural Collagen Membrane and the three predicate devices, these minor differences raise no new issues of safety and efficacy of Vitala" Resorbable Natural Collagen Mentrane.
OBI Biologics, Inc. Vitala''' Resorbable Natural Collagen Membrane 510(k) Premarket Notification Submission
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image contains a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring a bird-like symbol with three curved lines representing its wings or feathers. The emblem is black, and the text is also in a dark color, providing a clear contrast against the likely white or light background of the seal.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Dustyn Webb Director of Regulatory Affairs OBI Biologics, Incorporated 4620 71st Street, Building 79 Lubbock, Texas 79424
JUL 2 1 2011
Re: K101453
Trade/Device Name: Vitala Resorbable Natural Collagen Membrane Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: NPL Dated: July 1, 2011 Received: July 13, 2011
Dear Mr. Webb:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2- Mr. Webb
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Anthony D. minton
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{5}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known): K101453
#### Device Name: Vitala™ Resorbable Natural Collagen Membrane
Indications for Use:
Vitala™ Resorbable Natural Collagen Membrane is intended for use during the process of guided bone regeneration (GBR) and guided tissue regeneration (GTR) as a biodegradable barrier for:
- . Simultaneous use with implants;
- Augmentation around implants placed in immediate extraction sockets;
- . Augmentation around implants placed in delayed extraction sockets;
- . Localized ridge augmentation for later implantation;
- . Alveolar ridge reconstruction for prosthetic treatment;
- . Alveolar ridge preservation consequent to tooth extraction;
- . Filling of bone defects after root resection, cystectomy, removal of retained teeth;
- . Over the window in lateral window sinus elevation procedures;
- . Furcation defects in multi-rooted teeth;
- . Treatment of recession defects, together with coronally positioned flap;
- In implants with vertical bone loss due to infection, only with satisfactory debridement and . implant surface disinfection;
- . Guided bone regeneration in dehiscence defects; and
- Guided tissue regeneration in periodontal defects.
Prescription Use
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
### Concurrence of CDRH, Office of Device Evaluation (ODE)
Sagar Grover
(Division Sign-Off) Division of Anesthesiology, General Hospital nfection Control, Dental Devices
Page 1 of
510(k) Number: K101453
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.