← Product Code [FLL](/productcode/FLL) · K101387

# DIGITAL THERMOMETER, MODELS BT-A11CN, BT-A21CN, BT-A41CN (K101387)

_Fudakang Industries Co., Ltd. · FLL · Aug 3, 2010 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K101387

## Device Facts

- **Applicant:** Fudakang Industries Co., Ltd.
- **Product Code:** [FLL](/productcode/FLL.md)
- **Decision Date:** Aug 3, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2910
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Pediatric, 3rd-Party Reviewed

## Indications for Use

Fudakang Digital Thermometers are intended for the measurement and monitoring of human body temperature by doctor or consumers in the hospital or home. BT-A11CN, BT-A21CN, BT-A41CN can be used for axillary measurement, oral measurement and rectal measurement.

## Device Story

Hand-held, battery-operated digital thermometers (models BT-A11CN, BT-A21CN, BT-A41CN) for clinical or home use. Device utilizes thermistor sensor to detect body temperature; ASIC processes analog signal from thermistor by comparing against reference resistor; digital output displayed on LCD. Includes buzzer for audible feedback. Operated by clinicians or patients for axillary, oral, or rectal temperature monitoring. Provides rapid temperature readings to assist in clinical decision-making regarding patient health status.

## Clinical Evidence

Bench testing only. Compliance with ASTM E1112-2006 for accuracy, precision, and repeatability confirmed. Biocompatibility testing per ISO 10993-1, ISO 10993-5, and ISO 10993-10 passed. Electrical safety and electromagnetic compatibility verified per IEC 60601-1 and IEC 60601-1-2.

## Technological Characteristics

Hand-held electronic thermometer; thermistor sensing; ASIC signal processing; LCD display; buzzer. Materials: stainless steel, ABS plastic, thermoplastic rubber. Standards: ASTM E1112-2006, IEC 60601-1, IEC 60601-1-2, ISO 10993. Battery-powered.

## Regulatory Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

## Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; 
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and 
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

## Predicate Devices

- Famidoc Digital Clinical Thermometer FDTH-V0-1, FDTH-V0-2, FDTH-V0-3, FDTH-V0-4 ([K072641](/device/K072641.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(K) SUMMARY

AUG D& 2010

This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR 807.92.

#### 1.0 Submitter's Information

Establishment Registration Name:

Fudakang Industrial Co., Ltd

No.8 Yinghe Road, Yuanjiangyuan Management Zone, Changping Town, Dongguan, Guangdong, China . .

### Contact Person of applicant:

Mr. Bob Yu

No.8 Yinghe Road, Yuanjiangyuan Management Zone, Changping Town, Dongguan, Guangdong, China

TEL: +86-769-81098181 FAX: +86-769-81098187 Email: bobyu@fudakang.com

#### Contact Person of the Submission:

Mr. Stephen Ling

MEDLAB (Shenzhen) Information Service Co., LTD

Room 2706, Block A, ZhongFang JinYuan Buiding, Xinwen Road Shenzhen, Guangdong, P.R. China, 518034

TEL: +86-755-83089699 FAX: +86-755-83089760

Email: lingbensuo(@gmail.com, please CC sabrinawei(@hotmail.com

Section 5, Page 5-1

{1}------------------------------------------------

## 2.0 Device Information

| Type of 510(k) submission: | Traditional                       |
|----------------------------|-----------------------------------|
| Device Common Name:        | Thermometer, electronic           |
| Trade Name:                | Digital Thermometer               |
| Model:                     | BT-A11CN, BT-A21CN, BT-A41CN      |
| Classification name:       | Thermometer, electronic, clinical |
| Review Panel:              | General Hospital                  |
| Product Code:              | FLL                               |
| Regulation Class:          | II                                |
| Regulation Number:         | 880.2910                          |

#### 3.0 Predicate Device Information

| Sponsor:       | Famidoc Technology Co., Ltd.                                                        |
|----------------|-------------------------------------------------------------------------------------|
| Device:        | Famidoc Digital Clinical Thermometer FDTH-V0-1,<br>FDTH-V0-2, FDTH-V0-3, FDTH-V0-4. |
| 510(K) Number: | K072641                                                                             |

### 4.0 Device description

Fudakang digital thermometers are hand-held, reusable, battery operated. The models include BT-A11CN, BT-A21CN, and BT-A41CN. All the models have the same indication for use. The difference points are their shapes, colors, voice function, and measurement speed.

The operation principle is based on thermosensor and ASIC technology. A thermistor is used as thermosensor. The ASIC gets the sensor's signal from human body, then processes the signal and calculates the result, after that displays the temperature result by a LCD. The digital thermometer comprises a thermistor for getting temperature signal, a reference resistor for comparing the resistance of the thermistor, a buzzer for sounding effect, an ASIC for processing the target temperature digitally, and a LCD for displaying the temperature result.

#### 5.0 Intended Use

Fudakang Digital The mometers are intended for the measurement and monitoring of human body temperature by doctor or consumers in the hospital or home.

#### 6.0 Performance Summary

The devices meet the ASTM Standard Specification for Electronic Thermometer for intermittent Determination of Patient Temperature ASTM E1112-2006, as well

Section 5, Page 5-2

{2}------------------------------------------------

as IEC 60601-1 and IEC 60601-1-2 requirements. The bench test confirmed that the accuracy, precision and repeatability of device are compliance with ASTM E1112-2006 requirements.

The thermometers can be used for axillary, oral and rectal measurement. For axillary use, it directly contacts body skin. For oral or rectum use, it directly contacts membrane tissue. The patient contact materials in the device are made of stainless steel, ABS plastic, and thermoplastic rubber. They are all compliance with ISO10993-1 biocompatibility requirements, and pass the ISO10993-5 and ISO10993-10 tests.

#### 7.0 Comparison to predicate device and conclusion

Compare with predicate device, they are very similar in design principle, intended use, functions, material and the adopting applicable standards. The differences between applicant device and predicate device do not raise and new questions of safety or effectiveness.

l ). Indication for use

Famidoc Digital Clinical Thermometer can be used for axillary measurement, oral measurement and rectal measurement. Fudakang Digital Thermometers, BT-A11CN, BT-A21CN, BT-A41CN have the same measurement application.

2). Environment conditions

The environment conditions of Fudakang Digital Thermometers are a little different from Famidoc Digital Clinical Thermometer. But they are both compliance with IEC 60601-1 and ASTM E1112 requirements. So the difference of environment conditions will not raise any safety or effectiveness issue.

3). Accuracy

The accuracy of subject device is: 95.0 - 102.0 F : +/- 0.2 F 35.0 - 39.0℃: +/- 0.1℃ The rest: +/- 0.2°C - - -

The accuracy of predicate device is: +/- 0.1℃ (0.2F ) for whole range

Although there is a little difference, but the subject device accuracy is compliance with ASTM E 1112 requirements. So the difference of accuracy will not raise any safety or effectiveness issue.

4). Response time

{3}------------------------------------------------

The Subject Device is 60s. The Predicate Device is 30s.

The Response time of Fudakang Thermometers is a little different from Famidoc Digital Clinical Thermometer. They are all compliance with IEC 60601-1 and ASTM E 1112 requirements. The difference of Response time will not raise any safety or effectiveness issue.

Conclusions: Fudakang Digital Clinical Thermometers, model: BT-A11CN, BT-A21CN, BT-A41CN are substantially equivalent to the predicate devices.

Submission prepared date: March 23, 2010 8.0

{4}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Fudakang Industries Company, Limited C/O Mr. Marc M. Mouser Responsible Third Party Official Underwriters Laboratories, Incorporated 2600 NW Lake Road Camas, Washington 98607-9526

## AUG 03 2010

Re: K101387

Trade/Device Name: Digital Thermometer, Models BT-A11CN, BT-A21CN, BT-A41CN Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: July 9, 2010 Received: July 19, 2010

Dear Mr. Mouser:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2-Mr. Mouser

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Susan Buono

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# Statement of Indications for Use

510(k) Number (if known):

K101387

Device Name: Digital Thermometer Models: BT-A11CN, BT-A21CN, BT-A41CN

AUG-0 3 2010

Indications For Use:

Fudakang Digital Thermometers are intended for the measurement and monitoring of human body temperature by doctor or consumers in the hospital or home.

BT-A11CN, BT-A21CN, BT-A41CN can be used for axillary measurement, oral measurement and rectal measurement.

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

8/3/10
Page 1 of 1

(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesiology, General Powises Division of Anton, Dental Devices

510(k) Number: k101387

---

**Source:** [https://fda.innolitics.com/device/K101387](https://fda.innolitics.com/device/K101387)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
