K101305 · Ark Diagnostics, Inc. · ORH · Oct 29, 2010 · Clinical Toxicology
Device Facts
Record ID
K101305
Device Name
ARK LAMOTRIGINE ASSAY, CALIBRATOR & CONTROL
Applicant
Ark Diagnostics, Inc.
Product Code
ORH · Clinical Toxicology
Decision Date
Oct 29, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3350
Device Class
Class 2
Indications for Use
The ARK™ Lamotrigine Assay is a homogeneous enzyme immunoassay intended for the quantitative determination of lamotrigine in human serum or plasma on automated clinical chemistry analyzers. Lamotrigine concentrations can be used as an aid in management of patients treated with lamotrigine. The ARK™ Lamotrigine Calibrator is intended for use in calibration of the ARK Lamotrigine Assay. The ARK™ Lamotrigine Control is intended for use in quality control of the ARK Lamotrigine Assay.
Device Story
ARK Lamotrigine Assay is a homogeneous enzyme immunoassay (EIA) for quantitative measurement of lamotrigine in human serum or plasma. It utilizes a competitive binding mechanism between drug in specimen and lamotrigine labeled with bacterial glucose-6-phosphate dehydrogenase (G6PDH) for binding to anti-lamotrigine polyclonal antibody. In presence of specimen drug, enzyme activity increases, converting NAD to NADH; this rate of change in absorbance is measured spectrophotometrically. Device is used in routine clinical laboratories on automated clinical chemistry analyzers. Results assist clinicians in managing patients undergoing lamotrigine therapy. System includes two-reagent set (R1 antibody/substrate, R2 enzyme reagent), six-level calibrators, and three-level controls.
Clinical Evidence
No clinical trials performed. Evidence consists of analytical bench testing: precision (CLSI EP5-A2), linearity (CLSI EP6), interference (CLSI EP7), and method comparison (CLSI EP9-A2). Method comparison (n=77 vs. predicate; n=193 vs. HPLC) showed high correlation (r²=0.96-0.97). LOQ established at 0.85 µg/mL. Matrix equivalence confirmed for serum and plasma.
Indicated for the quantitative determination of lamotrigine in human serum or plasma to aid in the management of patients treated with lamotrigine. No specific age or gender restrictions are stated; however, the device is intended for use in clinical laboratory settings for therapeutic drug monitoring.
Regulatory Classification
Identification
A diphenylhydantoin test system is a device intended to measure diphenylhydantoin, an antiepileptic drug, in human specimens. Measurements obtained by this device are used in the diagnosis and treatment of diphenylhydantoin overdose and in monitoring levels of diphenylhydantoin to ensure appropriate therapy.
Predicate Devices
QMS® Lamotrigine, calibrators and controls (K062966)
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k101305
B. Purpose for Submission:
New assay, calibrator and control
C. Analyte:
Lamotrigine
D. Type of Test:
Homogeneous Enzyme Immunoassay (EIA)
E. Applicant:
ARK Diagnostics, Inc.
F. Proprietary and Established Names:
ARK™ Lamotrigine Assay, Calibrator and Control
G. Regulatory Information:
1. Regulation section:
21 CFR 862.3350 – Diphenylhydantoin test system
21 CFR 862.3200 – Clinical toxicology calibrator
21 CFR 862.3280 – Clinical toxicology control material
2. Classification:
Class II (assay), Class II (calibrator), Class I, reserved (control)
3. Product code:
ORH (assay), DLJ (calibrator), LAS (control)
4. Panel:
91 Toxicology
H. Intended Use:
1. Intended use(s):
See indications for use, below.
2. Indication(s) for use:
{1}
Page 2 of 14
The ARK™ Lamotrigine Assay is a homogeneous enzyme immunoassay intended for the quantitative determination of lamotrigine in human serum or plasma on automated clinical chemistry analyzers.
Lamotrigine concentrations can be used as an aid in management of patients treated with lamotrigine.
The ARK™ Lamotrigine Calibrator is intended for use in calibration of the ARK Lamotrigine Assay.
The ARK™ Lamotrigine Control is intended for use in quality control of the ARK Lamotrigine Assay.
3. Special conditions for use statement(s):
This assay is designed for use with serum or plasma only. The sponsor notes in the labeling that it is generally good practice to use the same method (as well as matrix) consistently for individual patient care due to the potential for method-to-method variability. Also see precautions in the Expected Range Section, below.
4. Special instrument requirements:
The Lamotrigine assay has been validated on the Roche/Hitachi 917 analyzer.
I. Device Description:
The ARK Lamotrigine Assay consists of reagents R1 anti-lamotrigine rabbit polyclonal antibody with substrate and R2 lamotrigine labeled with bacterial G6PDH enzyme. The ARK Lamotrigine Calibrator consists of a six-level (1.0, 2.5, 5.0, 11.0, 15.0, 30.0 and 40.0 µg/mL) set to calibrate the assay, and the ARK Lamotrigine Control consists of a three-level set (2.0, 12.0 and 25.0 µg/mL) used for quality control of the assay. The ARK Lamotrigine Calibrator and the ARK Lamotrigine Control matrix are comprised of a synthetic proteinaceous matrix with buffer, bovine serum albumin and preservatives.
J. Substantial Equivalence Information:
1. Predicate device name(s):
QMS Lamotrigine, Calibrators and Controls
2. Predicate 510(k) number(s):
k062966
3. Comparison with predicate:
{2}
Page 3 of 14
| Characteristic | Device
ARK™ Lamotrigine Assay | Predicate
QMS® Lamotrigine
K062966 |
| --- | --- | --- |
| Intended Use/
Indications for Use | Intended for the quantitative determination of lamotrigine in human serum or plasma on automated clinical chemistry analyzers.
Lamotrigine concentrations can be used as an aid in management of patients treated with lamotrigine. | Same |
| Sample | Serum or plasma | Same |
| Methodology | Homogenous enzyme immunoassay (EIA) | Homogeneous particle-enhanced turbidimetric immunoassay (particle agglutination) |
| Reagent Components | Two (2) reagent system:
• Anti-Lamotrigine Antibody/Substrate Reagent (R1) containing rabbit polyclonal antibodies to lamotrigine, glucose-6-phosphate, nicotinamide adenine dinucleotide, bovine serum albumin, preservatives, & stabilizers
• Enzyme Reagent (R2) containing lamotrigine labeled with bacterial G6PDH, buffer, bovine serum albumin, preservatives, and stabilizers | Two (2) reagent system:
• Anti-Lamotrigine Antibody Reagent (R1) in buffers containing stabilizers with sodium azide
• Lamotrigine-coated Microparticle Reagent (R2) in buffer containing stabilizers with sodium azide |
| Platform required | Roche/Hitachi 917 analyzer | Same |
| Accessory reagents | Calibrators (six levels) and controls (three levels) | Same |
| Testing environment | Routine clinical laboratory | Same |
K. Standard/Guidance Document Referenced (if applicable):
CLSI documents: "Evaluation of Precision Performance of Clinical Chemistry Devices", EP5; "Evaluation of the Linearity of Quantitative Measurement", EP6;
{3}
Page 4 of 14
"Interference Testing in Clinical Chemistry", EP7; "Method Comparison and Bias Estimation Using Patient Samples", EP9, "Protocols for Determination of Limits of Detection and Limits of Quantitation", EP 17-A.
## L. Test Principle:
ARK Lamotrigine Assay is a homogeneous immunoassay based on competition between drug in the specimen and lamotrigine labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for binding to the antibody reagent. As the latter binds antibody, enzyme activity decreases. In the presence of drug from the specimen, enzyme activity increases and is directly proportional to the drug concentration. Active enzyme converts the coenzyme nicotinamide adenine dinucleotide (NAD) to NADH that is measured spectrophotometrically as a rate of change in absorbance. Endogenous serum G6PDH does not interfere with the results because the coenzyme NAD functions only with the bacterial enzyme used in the assay.
## M. Performance Characteristics (if/when applicable):
Performance was validated on the Roche/Hitachi 917 clinical chemistry analyzer.
### 1. Analytical performance:
#### a. Precision
Precision was determined as described in CLSI Protocol EP5-A2. Data were collected on a Hitachi 917 analyzer over twenty (20) non-consecutive days. Multiple calibrations were performed during this interval to provide variation, although each calibration was performing in a stable manner. Three levels of both control materials and human serum samples were tested in each run. Each level was assayed in quadruplicate twice a day for 20 days. Calibrator/Control matrix and pooled human serum were also tested in quadruplicate twice a day for 5 days. Each of the runs per day was separated by at least two hours. The within run, between day, total SD, and percent CVs were calculated. Results are shown below.
| Sample | N | Mean (μg/mL) | Within Run | | Between Day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | CV (%) | SD | CV (%) | SD | CV (%) |
| ARK Lamotrigine Control | | | | | | | | |
| LOW | 160 | 2.08 | 0.07 | 3.4 | 0.05 | 2.5 | 0.08 | 4.1 |
| MID | 160 | 11.70 | 0.42 | 3.6 | 0.28 | 2.4 | 0.49 | 4.2 |
| HIGH | 160 | 24.23 | 0.99 | 4.1 | 1.06 | 4.4 | 1.47 | 6.1 |
| Calibrator / Control Matrix | 40 | 38.04 | 2.05 | 5.4 | 0.95 | 2.5 | 2.27 | 6.0 |
| Human Serum | | | | | | | | |
| LOW | 160 | 2.41 | 0.08 | 3.5 | 0.09 | 3.7 | 0.12 | 5.2 |
{4}
Page 5 of 14
| MID | 160 | 10.75 | 0.41 | 3.8 | 0.42 | 3.9 | 0.59 | 5.5 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| HIGH | 160 | 25.84 | 1.33 | 5.2 | 1.12 | 4.3 | 1.88 | 7.3 |
| Pooled | 40 | 38.24 | 2.78 | 7.3 | 0.61 | 1.6 | 3.38 | 8.8 |
| Human | | | | | | | | |
| Serum | | | | | | | | |
## b. Linearity/assay reportable range:
The manufacturer's claimed assay reportable range is 0.85 to 40.0 µg/mL. The assay range was established based on the linearity, together with limit of quantitation (LOQ) and recovery/linearity studies. Method comparison evaluated specimens near the upper quantitative limit. (See respective sections below for specific information on performance).
Samples for the recovery/linearity study were prepared by gravimetric addition of lamotrigine (USP > 99.9% purity) stock solution to human serum negative for lamotrigine. Drug concentrations across the assay range (0.85, 1.0, 2.5, 5.0, 11.0, 15.0, 30.0 and 40.0 µg/mL) were tested. Each sample was assayed in triplicate in each of two separately calibrated runs for a total of six replicates. The results at each separate level were averaged and compared to the target concentration (based on spiked concentrations of USP materials) and the percentage recovery was calculated.
$$
\% \text{Recovery} = [100 \times \text{Mean recovered concentration}] / \text{Theoretical concentration}
$$
| Theoretical Concentration (µg/mL) | Mean Recovered Concentration (µg/mL) | Recovery (%) |
| --- | --- | --- |
| 0.85 | 0.84 | 98.2 |
| 1.00 | 0.99 | 99.2 |
| 2.50 | 2.48 | 99.3 |
| 5.00 | 5.25 | 105.1 |
| 11.00 | 10.97 | 99.7 |
| 15.00 | 14.80 | 98.7 |
| 30.00 | 29.16 | 97.2 |
| 40.00 | 38.33 | 95.8 |
Regression analysis of the data yields the equation, $y(\text{observed result}) = 0.96 \times (\text{target concentration}) + 0.22$, and regression coefficient $r = 0.99$.
## High Sample Carryover
The impact of high lamotrigine concentration specimens on the measurement of Lamotrigine in specimens with lower concentrations (high sample carryover) was evaluated by assaying a series of high 100.0 µg/mL spiked serum sample followed by a series of low 2.0 µg/mL spiked serum sample. No carryover was observed from the High 100.0 µg/mL sample to the Low 2.0 µg/mL sample.
5
{5}
Page 6 of 14
## Manual Dilution Protocol
The manufacturer recommends that samples with concentrations exceeding 40.0 µg/mL should be diluted 4-fold with zero calibrator. Samples spiked with lamotrigine and patient samples containing lamotrigine (with initial concentrations ranging from 47.0 to 100 µg/mL) were diluted 4-fold with zero calibrator and results using the ARK method were compared to the target level (for spiked samples) and the HPLC-determined level (for neat samples). In some cases natural patient samples containing lower levels of lamotrigine were spiked to achieve higher concentrations for this evaluation. Concentrations tested ranged from 47-100 µg/mL. Recoveries versus expected values using the dilution procedure ranged from 93% to 96%.
## c. Traceability, Stability, Expected values (controls, calibrators, reagents, sample or methods):
### Traceability and Value Assignment of Calibrators and Controls
ARK Lamotrigine Calibrators and ARK Lamotrigine Controls are prepared by gravimetric dilution of high purity lamotrigine (USP > 99.9 % purity) into a synthetic proteinaceous matrix free of lamotrigine. Each calibrator and control level is then qualified versus a master calibrator lot. Calibrator and control production lots must demonstrate comparison to the master lot within 5% at each positive calibrator and control level.
Quality control (QC) ranges were established using three runs with four replicates tested per run (n=12 for each control level) and the mean lamotrigine level of each control was calculated. Control ranges were set at ±15% around the mean level tested. The package insert notes that each laboratory should establish its own ranges for each new lot of controls.
### Specimen Stability
Stability of stored specimens (frozen storage and refrigerated exposure) and the effect of freezing/thawing on the measurement of lamotrigine by the ARK™ Lamotrigine Assay were evaluated. Clarified serum specimens were shown to be stable for at least four weeks frozen at ≤ -10 °C, and for a week when refrigerated (2-8 °C). The manufacturer's acceptance criteria for this evaluation are recovery within ±10%. No trend was observed in the stability data. The manufacturer notes that care should be taken to limit the number of freeze-thaw cycles. Specimens were shown to withstand 3 freeze-thaw cycles when stored at -20 °C.
The ARK Lamotrigine Calibrator and the ARK Lamotrigine Control are comprised of a synthetic proteinaceous matrix with buffer, bovine serum albumin and preservatives. To evaluate recovery and matrix equivalence, synthetic calibrator/control matrix (Calibrator A) and pooled human serum were supplemented with lamotrigine to achieve 1.0, 2.5, 5.0, 11.0, 15.0, 30.0 and 40.0 µg/mL levels. Multiple replicates and runs were performed and the mean
6
{6}
Page 7 of 14
recovery (of replicates) was calculated for each level and each matrix. All levels tested in the calibrator matrix recovered within 96% of the serum lamotrigine levels. Recoveries (percentage of nominal level) ranged from 96% to 104% for the calibrator/control matrix and from 96% to 105% for the serum matrix (95.8% to 105.1%).
## Calibration Stability on the analyzer
ARK provided a study on calibration stability on the analyzer. This study supports the stable operation of the assay over an extended period based on one calibration. Up to 30 days of calibration curve stability and in-use stability of reagents, calibrators and controls were observed for the study as tested.
## Calibrator and Control material Stability
1. The calibrators and controls are stable until the expiration date printed on the vial when stored unopened and opened at 2-8 °C.
2. Real time stability studies are ongoing for both unopened and opened calibrators and controls. Stability testing protocols and acceptance criteria were reviewed and found to be acceptable.
## d. Detection Limit:
Accuracy and precision studies near the low range of the assay were conducted to determine the manufacturer's claimed lower limit of quantitation (LOQ) according to CLSI Guideline EP-17. Three levels were tested below the lowest positive calibrator concentration (1.0 µg/mL). Samples were prepared by gravimetric addition of lamotrigine (USP > 99.9% purity) to human serum negative for lamotrigine to give concentrations of 0.5, 0.75 and 0.85 µg/mL. Eight replicates of each sample were tested in each of five runs on five separate days to give a minimum of 40 replicates of each sample per reagent lot. A total of three reagent lots were used for the study. The LOQ of the ARK Lamotrigine Assay is defined as the lowest concentration for which acceptable inter-assay precision (≤ 20% CV) and recovery (± 15%) is observed. The criteria of LOQ were met at 0.85 µg/mL; the precision was 2.9 %CV and the recovery was 90.1%.
## e. Analytical Specificity:
Studies included testing for interference from endogenous compounds, metabolite, and commonly co-administered and other anti-epileptic drugs.
## Endogenous Interfering Substances
Serum samples with clinically high concentrations of the potential endogenous interfering substances were tested by ARK Lamotrigine assay in the presence of varying amounts of lamotrigine. Specifically, serum samples tested contained lamotrigine at concentrations of 3.0 µg/mL and 15.0 µg/mL. Each sample containing interferent was assayed, along with a serum control of lamotrigine. No significant interference (defined by the manufacturer as ≤ 10% differences
{7}
in detecting lamotrigine) was observed. Results for endogenous compounds are shown below:
| Percentage Recovery | | | |
| --- | --- | --- | --- |
| Interfering Substance | Interferent Concentration | 3 μg/mL Lamotrigine | 15 μg/mL Lamotrigine |
| Albumin | 12 g/dL | 101.5 % | 103.4 % |
| Bilirubin - conjugated | 70 mg/dL | 93.6 % | 102.6 % |
| Bilirubin - unconjugated | 70 mg/dL | 97.1 % | 105.0 % |
| Cholesterol | 342 mg/dL | 105.2 % | 95.0 % |
| Gamma-Globulin | 12 g/dL | 106.8 % | 104.4 % |
| Hemoglobin | 1000 mg/dL | 98.2 % | 97.0 % |
| Intralipid® | 1000 mg/dL | 94.5 % | 94.3 % |
| Rheumatoid Factor | 1100 IU/mL | 107.3 % | 108.9 % |
| Triglycerides | 618 mg/dL | 101.7 % | 104.0 % |
| Uric Acid | 30 mg/dL | 101.0 % | 99.6 % |
## Metabolite Interference
Lamotrigine-2-N-glucuronide, Lamotrigine-2-N-methyl and Lamotrigine-2-N-oxide were tested for cross-reactivity. These metabolites were spiked into two separate samples each containing low and high lamotrigine concentrations of 3 and 15 μg/mL, respectively. The samples were analyzed and the lamotrigine concentrations of samples containing interferent were compared to the serum control. Cross-reactivity to Lamotrigine-2-N-glucuronide ranged from 1.09 to 2.91%, cross-reactivity to Lamotrigine-2-N-methyl ranged from 0.02 to 0.24% and cross-reactivity to Lamotrigine-2-N-oxide ranged from 1.3 to 3.94%.
| Metabolite* | Metabolite Concentration (μg/mL) | Percentage Cross-Reactivity | |
| --- | --- | --- | --- |
| | | Lamotrigine (3 μg/mL) | Lamotrigine (15 μg/mL) |
| Lamotrigine-2-N-glucuronide | 50.0 | 2.41 % | 1.86 % |
| | 25.0 | 2.57 % | 1.09 % |
| | 12.5 | 2.91 % | 1.92 % |
| | 9.0 | 2.15 % | 1.57 % |
| Lamotrigine-2-N-methyl | 400.0 | 0.04 % | 0.21 % |
| | 200.0 | 0.07 % | 0.02 % |
| | 80.0 | 0.10 % | 0.24 % |
| Lamotrigine-2-N-oxide | 80 | 3.69 % | 3.63 % |
| | 40 | 3.94 % | 3.64 % |
| | 20 | 3.72 % | 3.14 % |
| | 10 | 3.88 % | 1.30 % |
{8}
Page 9 of 14
## Drugs that Cross-React - Trimethoprim
Cross-reactivity of the antibody to trimethoprim was analyzed from 5.0 to 40.0 $\mu \mathrm{g} / \mathrm{mL}$ in the presence of either Low (3 $\mu \mathrm{g} / \mathrm{mL}$ ) or High (15 $\mu \mathrm{g} / \mathrm{mL}$ ) concentration of lamotrigine and assayed along with a serum control of lamotrigine. Interference was observed only at the low concentration of lamotrigine. Recovery of lamotrigine ranged 111.2 to $156.0\%$ with increasing concentration of trimethoprim. The manufacturer notes in the package insert that care should be taken when interpreting ARK Lamotrigine results if trimethoprim is also being administered to the patient. The results are shown below.
| Trimethoprim (μg/mL) | Percent Cross-Reactivity | | Percent Recovery | |
| --- | --- | --- | --- | --- |
| | Lamotrigine (3 μg/mL) | Lamotrigine (15 μg/mL) | Lamotrigine (3 μg/mL) | Lamotrigine (15 μg/mL) |
| 40.0 | 4.4 % | 3.0 % | 156.0 % | 108.0 % |
| 20.0 | 5.5 % | 4.0 % | 134.6 % | 105.4 % |
| 10.0 | 6.5 % | -0.8 % | 120.9 % | 99.4 % |
| 5.0 | 6.9 % | 8.6 % | 111.2 % | 103.0 % |
## Drug Interference
Lamotrigine-selective antibody did not cross-react with most other anti-epileptic or co-administered drugs tested. Due to structural similarities with lamotrigine, high trimethoprim levels may interfere. A high concentration of each compound shown below was spiked into normal human serum with known levels of lamotrigine (approximately 3 and $15\mu \mathrm{g / mL}$ ) and assayed along with a serum control of lamotrigine. Measurement of lamotrigine resulted in $\leq 10\%$ error in the presence of drug compounds at the levels tested
| Compound | Concentration (μg/mL) | Percentage Recovery (%) | |
| --- | --- | --- | --- |
| | | Lamotrigine (3 μg/mL) | Lamotrigine (15 μg/mL) |
| Acetaminophen | 200 | 103.7 | 99.1 |
| Acetazolamide | 100 | 101.2 | 99.2 |
| Acetylsalicylic acid | 1000 | 100.8 | 100.7 |
| Amikacin | 100 | 95.7 | 97.0 |
| Amitriptyline | 20 | 99.0 | 97.9 |
| Amoxapine | 40 | 104.7 | 101.2 |
| Amphotericin B | 100 | 94.0 | 91.6 |
| Ampicillin | 100 | 97.7 | 94.1 |
| Ascorbic Acid | 100 | 98.5 | 94.4 |
| Baclofen | 100 | 95.8 | 90.9 |
{9}
Page 10 of 14
| Bupropion | 40 | 98.8 | 106.2 |
| --- | --- | --- | --- |
| Caffeine | 100 | 101.3 | 103.2 |
| Carbamazepine | 120 | 104.3 | 103.2 |
| Carbamazepine-10, 11 epoxide | 120 | 101.7 | 99.0 |
| 10-Hydroxy carbamazepine | 100 | 96.2 | 94.3 |
| Chloramphenicol | 250 | 103.7 | 98.4 |
| Chlorpromazine | 20 | 97.2 | 95.0 |
| Citalopram | 20 | 98.0 | 97.5 |
| Clobazam | 100 | 103.4 | 105.6 |
| Clonazepam | 20 | 97.6 | 96.4 |
| Cyclosporin A | 40 | 101.7 | 99.4 |
| Diazepam | 20 | 101.1 | 97.7 |
| Digoxin | 80 | 103.4 | 97.6 |
| Doxepin | 20 | 101.6 | 103.1 |
| Erythromycin | 200 | 103.6 | 103.9 |
| Ethanol | 4000 | 94.0 | 98.2 |
| Ethotoin | 100 | 101.3 | 101.9 |
| Ethosuximide | 250 | 101.0 | 96.4 |
| Felbamate | 250 | 103.0 | 101.4 |
| Fluoxetine | 20 | 102.2 | 97.0 |
| Furosemide | 100 | 99.8 | 97.1 |
| Gentamicin | 100 | 99.8 | 98.6 |
| Haloperidol | 20 | 104.1 | 100.3 |
| Heparin | 200 U/mL | 99.0 | 100.5 |
| Ibuprofen | 500 | 101.6 | 96.2 |
| Imipramine | 20 | 99.6 | 97.7 |
| Gabapentin | 200 | 103.8 | 98.1 |
| --- | --- | --- | --- |
| Levetiracetam | 400 | 103.6 | 101.9 |
| Lidocaine | 100 | 101.6 | 101.8 |
| Lincomycin | 1000 | 106.0 | 99.7 |
| Mephenytoin | 100 | 95.7 | 103.9 |
| Mesoridazine | 40 | 97.6 | 101.7 |
| Methicillin | 250 | 95.2 | 99.4 |
| Naproxen | 600 | 97.3 | 104.8 |
| Neomycin | 1000 | 100.8 | 101.6 |
| Niacin | 100 | 97.8 | 105.8 |
| Nitrazepam | 20 | 101.5 | 103.9 |
| Nortriptyline | 20 | 96.6 | 104.9 |
| Olanzapine | 20 | 99.5 | 102.2 |
10
{10}
Page 11 of 14
| Oxcarbazepine | 200 | 97.3 | 100.5 |
| --- | --- | --- | --- |
| Paroxetine | 40 | 101.6 | 100.0 |
| 2-phenyl-ethyl-malonamide (PEMA) | 1000 | 100.1 | 100.9 |
| Penicillin V | 100 | 100.4 | 101.4 |
| Perphenazine | 100 | 99.5 | 103.2 |
| Phenobarbital | 200 | 101.0 | 98.9 |
| Phenytoin | 200 | 100.0 | 100.8 |
| Pregabalin | 200 | 99.6 | 98.4 |
| Primidone | 100 | 98.7 | 102.5 |
| Procainamide | 100 | 100.6 | 101.9 |
| Prochlorperazine | 40 | 99.4 | 90.3 |
| Ranitidine | 100 | 104.0 | 97.8 |
| Rifampin | 100 | 101.6 | 97.7 |
| Risperidone | 20 | 98.0 | 100.2 |
| Sertraline | 100 | 101.5 | 101.9 |
| Spectinomycin | 100 | 97.7 | 103.1 |
| Stiripentol | 100 | 102.3 | 101.6 |
| Sulfamethoxazole | 400 | 99.2 | 99.2 |
| Theophylline | 200 | 98.7 | 97.9 |
| Thioridazine | 20 | 102.9 | 101.3 |
| Tobramycin | 100 | 98.8 | 96.9 |
| Tiagabine | 200 | 100.9 | 97.8 |
| Topiramate | 250 | 100.3 | 96.7 |
| Valproic Acid | 600 | 100.8 | 96.8 |
| Vancomycin | 250 | 96.5 | 95.0 |
| Vigabatrin | 150 | 97.8 | 101.0 |
| Zonisamide | 400 | 97.9 | 99.6 |
**f. Assay cut-off:**
Not applicable.
**2. Comparison studies:**
**a. Method comparison with predicate device:**
Method comparison studies were performed according to CLSI Guideline EP9-A2. Banked samples from patients tested for lamotrigine concentrations were used. Inclusion criteria for the samples measured were based on lamotrigine concentration. No exclusion criteria were used for selection of specimens. The large majority of specimens were in a serum matrix. (See also matrix comparison, below.)
11
{11}
Page 12 of 14
Results from the ARK Lamotrigine Assay were compared with results from the predicate device (QMS) at an external clinical laboratory site. Results compared to high performance liquid chromatography (HPLC) were also included. Assay descriptions and summary validation for the comparative methods were included in the 510(k).
The results are shown below:
**Comparison to the predicate device:**
Lamotrigine concentrations by the QMS Lamotrigine turbidimetric immunoassay ranged from 2.28 to 37.70 µg/mL. ARK lamotrigine values ranged 2.51 to 36.32 µg/mL. Results of the Passing-Bablok regression analysis for the study are shown below (with 95% confidence limits).
| | | 95% confidence limits |
| --- | --- | --- |
| Slope | 0.93 | (0.89 to 0.97) |
| y-intercept | 0.41 µg/mL | (0.07 to 0.74) |
| Correlation Coefficient (r²) | 0.96 | (0.94 to 0.97) |
| Number of Samples | | 77 |
Results of the Pearson Bland Altman analysis gave a mean bias of -0.559, SE of 0.1876, and SD of difference as 1.646.
**Comparison to an HPLC method**
Lamotrigine concentrations by HPLC ranged 1.0 to 36.70 µg/mL. ARK lamotrigine values ranged 0.97 to 36.32 µg/mL. Results of the Passing-Bablok regression analysis for the study are shown below (with 95% confidence limits).
| | | 95% confidence limits |
| --- | --- | --- |
| Slope | 1.01 | (0.99 to 1.03) |
| y-intercept | 0.37 µg/mL | (0.22 to 0.55) |
| Correlation Coefficient (r²) | 0.97 | (0.96 to 0.98) |
| Number of Samples | | 193 |
Results of the Pearson Bland Altman analysis gave a mean bias of 0.42, SE of 0.0934, and SD of difference as 1.298.
**b. Matrix Comparison-Serum versus Plasma:**
Anticoagulated plasma and serum were evaluated to evaluate equivalency of these matrices for measurement of lamotrigine with the ARK™ Lamotrigine
12
{12}
Page 13 of 14
Assay. Matched samples for serum and plasma from eight (8) subjects (four subjects at 35.0 µg/mL) were evaluated. Blood was collected in three different anticoagulant tubes viz. lithium heparin, potassium EDTA, sodium heparin, and a serum tube to produce a matched set. A lamotrigine stock solution was spiked in each matrix to give a lamotrigine concentration of 35.0, 30.0, 15.0, 3.0 and 0.85 µg/mL. The mean, standard deviation and %CV for six replicates were calculated for each sample. Also, correlation analysis was performed. Percentage recovery of lamotrigine in anticoagulated samples compared to the serum control was calculated for each subject. No significant differences were observed between matrices. The overall percentage of serum levels of lamotrigine in plasma ranged 92.9 to 106.3%, and the slope was 1.0, intercept was -0.02 µg/mL, and correlation was r² = 1.00.
In addition, the method comparison study (Section a, above) included five plasma samples. In general, no significant differences were observed between the results of these samples and the averaged serum samples.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable. Not typically submitted for this type of assay.
b. Clinical specificity:
Not applicable. Not typical for this type of assay.
c. Other clinical supportive data (when a. and b. are not applicable):
The sponsor provided a discussion with balanced and representative literature discussing clinical use of lamotrigine measurements.
4. Clinical cut-off:
See expected values below.
5. Expected values/Reference range:
In the labeling the manufacturer provides the following information:
The reference range of drug concentrations which is quoted should only imply a lower limit below which a therapeutic response is relatively unlikely to occur, and an upper limit above which toxicity is relatively likely to occur in the specific patient populations studied. Generally, clinicians using reference ranges such as these should be aware that, because of individual variation, patients may achieve therapeutic benefit with serum drug concentrations outside of these ranges and may experience toxicity with levels below the lower limit of the reference range. Sampling time should be standardized such that trough serum concentrations are measured just before the next dosage, preferably in the morning.
13
{13}
Page 14 of 14
A therapeutic range for lamotrigine has not been well established. Some reports in the literature suggest a target range for steady-state concentrations of 3.0 to 15 µg/mL. However, there is not a clear relationship between lamotrigine serum concentrations and clinical response. Due to individual patient differences and other co-administered medications, considerable overlap in lamotrigine concentrations has been observed between serum responders and non-responders as well as between serum levels associated with seizure control and adverse effects. In one study, the highest mean serum level (trough) reported was 8.8 µg/mL, and less than 15% of patients reported an adverse event at serum concentrations less than 10.0 µg/mL. Mild to moderate adverse effects are more commonly associated with patients with lamotrigine concentrations above 15.0 µg/mL. Co-medications affect clearance of lamotrigine with enzyme-inducers increasing and valproic acid decreasing clearance. Lamotrigine clearance is higher in children than in adults and moderately reduced in the elderly. Clearance may be increased during pregnancy, but such increase is attenuated in women co-medicated with valproic acid. Acute overdoses associated with serum levels above 40.0 µg/mL (156 µmol/L) have been reported.
Lamotrigine drug concentrations should not be the only means of therapeutic drug management. The assay should be used in conjunction with information available from clinical evaluations and other diagnostic procedures. Clinicians should carefully monitor patients during therapy initiation and dosage adjustments. Multiple measurements of lamotrigine may be needed.
**N. Proposed Labeling:**
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
**O. Conclusion:**
The submitted information in this premarket notification is complete and supports substantial equivalence decision.
14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.