← Product Code [LZA](/productcode/LZA) · K101293

# POWDER FREE NITRILE EXAMINATION GLOVES, TESTED FOR USE WITH CHEMOTHERAPY DRUGS (BLUE) (K101293)

_Pt. Smartglove Indonesia · LZA · Jul 1, 2010 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K101293

## Device Facts

- **Applicant:** Pt. Smartglove Indonesia
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Jul 1, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

This glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. This device also provides protection against exposure to certain chemotherapy drugs.

## Device Story

Disposable nitrile examination glove; worn by healthcare professionals during patient examinations to prevent cross-contamination; provides barrier protection against specific chemotherapy drugs. Device is non-sterile; blue color. User dons glove manually. Output is physical barrier protection. Benefits include reduced risk of pathogen transmission and protection from hazardous chemotherapy agents during handling. Warning: Do not use with Carmustine.

## Clinical Evidence

Bench testing only. Permeation testing performed for 14 chemotherapy drugs (Carmustine, Thiotepa, Fluorouracil, Etoposide, Cyclophosphamide, Mitomycin C, Methotrexate, Paclitaxel, Doxorubicin Hydrochloride, Dacarbazine, Cisplatin, Mitoxantrone, Ifosfamide, Vincristine Sulfate) to determine breakthrough detection times.

## Technological Characteristics

Material: Nitrile. Form factor: Disposable examination glove. Color: Blue. Non-sterile. Tested for chemotherapy drug permeation resistance.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with three wavy lines extending from its body, representing the department's mission to protect and promote the health and well-being of Americans. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

## JUL - 1 2010

Mr. Cho S. Fong Quality Assurance Regulatory Affairs Manager PT. Smart Glove Indonesia JL Raya Medan-Lubuk Pakam KM19, Tanjong Morawa, Deli Serdang, Sumatera U Indonesia 20362

Re: K101293

Trade/Device Name: Powder Free Nitrile Examination Gloves, Tested for use with Chemotherapy Drugs (Blue) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA, LZC Dated: May 3, 2010 Received: May 10, 2010

Dear Mr. Fong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In

{1}------------------------------------------------

Page 2- Mr. Fong

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/Reportal/roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Susan Quare

4Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

K/01293

# Indications for Use

#### 510(k) Number (if known):

Device Name:

1. Powder Free Nitrile Examination Gloves, Tested For Use With Chemotherapy Drugs (Blue)

Indication For Use:

This glove is a disposable device intended for medical purposes that is worn on the examiner's

hand to prevent contamination between patient and examiner. This device also provides protection against exposure to certain chemotherapy drugs.

Chemotherapy Drugs Permeation (Breakthrough detection time in minutes)

| Carmustine (BCNU) (3.3 mg / ml)           | 1.21  |
|-------------------------------------------|-------|
| Thiotepa (10.0 mg / ml)                   | 90.7  |
| Fluorouracil (50.0 mg / ml)               | > 240 |
| Etoposide (Toposar) (20.0 mg / ml)        | > 240 |
| Cyclophosphamide (Cytoxan) (20.0 mg / ml) | > 240 |
| Mitomycin C (0.5 mg / ml)                 | > 240 |
| Methotrexate (25.0 mg / ml)               | > 240 |
| Paclitaxel (Taxol) (6.0 mg / ml)          | > 240 |
| Doxorubicin Hydrochloride (2.0 mg / ml)   | > 240 |
| Dacarbazine (DTIC) (10.0 mg / ml)         | > 240 |
| Cisplatin (1.0 mg / ml)                   | > 240 |
| Mitoxantrone (2.0 mg / ml)                | > 240 |
| Ifosfamide (50.0 mg / ml)                 | > 240 |
| Vincristine Sulfate (1.0mg / ml)          | > 240 |

### Warning: Do Not Use with Carmustine

| Prescription Use _____________________________________________________________________________________________________________________________________________________________ | AND/OR | Over-The-Counter Use   |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|------------------------|
| (Part 21 CFR 801 Subpart D)                                                                                                                                                    |        | (21 CFR 801 Subpart C) |

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office Of Device Evaluation (ODE)

Page 1 of

(Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital
Production of Anesthesion Paulose Division of Anton Dental Devices

510(k) Number: __ k101293

---

**Source:** [https://fda.innolitics.com/device/K101293](https://fda.innolitics.com/device/K101293)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).