This product is indicated for the administration into or withdrawal of fluids from implanted ports. It is designed to help protect against accidental needlestick injuries.
Device Story
GRIPPER Micro needle system consists of inserter, infusion site, extension tubing, and luer fitting; optional needless access connector y-site. Inserter features sharp trocar needle and retractor arm; infusion site includes septum and blunt cannula. Operation: trocar needle inserted through septum/cannula into implanted port; retractor arm activated to remove trocar needle into safety capture zone; blunt cannula remains in port for fluid administration/withdrawal. Used in clinical settings by healthcare professionals. Safety mechanism prevents needle stick injury during insertion, use, and extraction via rebounding action prevention. Device facilitates safe port access and fluid management.
Clinical Evidence
No human clinical studies performed. Simulated use study conducted per FDA Guidance for Industry and FDA Staff - Medical Devices with Sharps Injury Prevention Features (August 2005). Study involved healthcare professionals experienced in port access. 500 activations performed; 100% success rate in capturing trocar tip in safety zone.
Technological Characteristics
Non-coring safety needle with blunt cannula. Components: inserter, infusion site, extension tubing, luer fitting. Materials biocompatible and consistent with predicate. Safety feature: trocar needle retraction into capture zone. Bench testing verified force to attach/detach, flow rates, and joint/bond strength per FDA sharps injury prevention guidance.
Indications for Use
Indicated for administration into or withdrawal of fluids from implanted ports in patients requiring such access. Designed to protect healthcare professionals against accidental needlestick injuries.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Klollso.
心
TAB 6
AUG. 3 - 2010
# 510(k) Summary
## GRIPPER® Micro 1.25 Inch Needle
Traditional 510(k) - GRIPPER Micro 1.25 Smiths Medical ASD, Inc.
Confidential
{1}------------------------------------------------
## smiths medical
bringing technology to life
## 510(k) Number:____
Page 1 of 3
April 23, 2010 Updated June 28, 2010 Date Prepared
## Submitter Information
| Submitter's Name: | Smiths Medical ASD, Inc. |
|-------------------|------------------------------------------|
| Address: | 1265 Grey Fox Road<br>St. Paul, MN 55112 |
Establishment Registration: 2183502
| Contact Person: | Rachelle Parsons, RAC<br>Sr. Regulatory Affairs Specialist |
|-----------------|------------------------------------------------------------|
| Phone:<br>Fax: | (651) 628-7018<br>(651) 628-7457 |
### Device Information
| Trade Name: | GRIPPER® Micro Blunt Cannula, Non-Coring Safety Needle; 1.25 inch |
|----------------------|-------------------------------------------------------------------|
| Common Name: | GRIPPER® Micro Needle |
| Classification Name: | Intravascular Administration Set |
| Product Code: | FPA |
| Regulation: | 21 CFR §880.5440 |
### Predicate Device(s)
The predicate devices are the currently marketed GRIPPER Micro Needles. The reference 510(k) number for these devices:
| Device | 510(k) |
|--------------------------------------------------------|---------|
| GRIPPER® Micro Blunt Cannula, Non-Coring Safety Needle | K072059 |
{2}------------------------------------------------
## smiths medical
bringing technology to life
Page 2 of 3
### Device Description
The GRIPPER® Micro needle is comprised of the inserter and the infusion site with extension tubing and a standard luer fitting; there are versions either with or without needless access connector y-site. The inserter incorporates a sharp trocar needle and retractor arm. The infusion site incorporates a small septum and an attached blunt cannula. When fully assembled, the inserter and infusion site are combined with the trocar needle inserted through the septum and blunt cannula. After insertion of the cannula and trocar into the implanted port, the inserter retractor arm is activated removing the trocar needle from the cannula and infusion site septum leaving the blunt cannula in the implanted port. The trocar needle tip is captured in the inserter to prevent needle stick injury, and the inserter is discarded. Upon removal of the infusion site from the implanted port, the blunt cannula is designed to further prevent needle stick injury that may result from rebounding action during infusion site extraction.
#### Intended Use
The GRIPPER® Micro needle is designed for the administration into or withdrawal of fluids from implanted ports.
#### Indications for Use
The GRIPPER® Micro needle is indicated for the administration into or withdrawal of fluids from implanted ports. It is designed to help protect against accidental needlestick injuries.
#### Summary of Non-Clinical Testing
The non-clinical testing included assessment of the physical properties of the GRIPPER® Micro 1.25 inch needle and its ability to achieve its intended use. The GRIPPER Micro 1.25 inch products meet the same specifications as set for the predicate device. Testing performed based on FDA Guidance: Guidance for Industry and FDA Staff - Medical Devices with Sharps Iniurv Prevention Features. August 2005 included force to attach and detach connections, force to activate and deactivate the safety features, rate of fluid flow simulating extremes of pressure and strength of joints and bonds. The Bench testing of the device confirmed the suitability of the device for its intended use.
Biocompatibility assessment of the device was performed. The purpose of the biocompatibility assessment was to ensure that biocompatibility had been established for the device. The device is biocompatible based on the similarity of the materials of construction to the predicate devices commercially marketed by Smiths Medical ASD, Inc.
Traditional 510(k) -- GRIPPER Micro 1.25 Smiths Medical ASD, Inc.
Confidential
{3}------------------------------------------------
## smiths medical
bringing technology to life
Page 3 of 3
#### Summary of Clinical Testing
Human clinical studies were deemed not necessary to evaluate the safety or effectiveness of the GRIPPER® Micro Needle. A Simulated Use Study was condicted according to the FDA Guidance: Guidance for Industry and FDA Staff - Medical Devices with Sharps Injury Prevention Features, August 2005. Health care professionals were selected to participate in the simulated use evaluation from a cross section of users or potential users of the device. These participants were experienced in accessing and de-accessing implantable portals.
The purpose of the study was to simulate actual clinical use of the GRIPPER Micro Needle System 1.25 inch by health care professionals, in terms of installation, activation of sharps protection and deaccess of the infusion site. The sample size was based on the statistical rationale noted in the FDA guidance. All 500 activations resulted in successful captures of the trocar tip in the capture zone.
#### Statement of Equivalence
The GRIPPER ® Micro 1.25 inch Needle is substantially equivalent to the currently marketed GRIPPER Micro Needles based on a comparison of the indications for use and the technological characteristics of the device.
#### Conclusion
The GRIPPER® Micro 1.25 inch Needle is substantially equivalent to the currently marketed GRIPPER Micro Needles based on the indications for use, technological characteristics, materials of construction and principles of operation of the device. Bench tests confirmed the suitability of the device for its intended use.
Traditional 510(k) - GRIPPER Micro 1.25 Smiths Medical ASD, Inc.
Confidential
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles a stylized caduceus or a representation of human figures. There are also a series of dots on the left side of the seal.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Rachelle Parsons Senior Regulatory Affairs Specialist Smiths Medical ASD, Incorporated 1265 Grey Fox Road St. Paul, Minnesota 55112
AUG 3 - 2010
Re: K101180
Trade/Device Name: GRIPPER® Micro Blunt Cannula Non- Coring Safety Needle Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: July 2, 2010 Received: July 6, 2010
Dear Ms. Parsons:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
### Page 2- Ms. Parsons
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours
Susan Munro
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology. General Hospital. Infection Control and Dental Devices Office of Device Evaluation . Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K10/1180
## SMITHS MEDICAL ASD, INC. 510(k) Premarket Notification
## GRIPPER® Micro Blunt Cannula Non-Coring Safety Needle Indications for Use
AUG 3 - 2010
510(k) Number: K101180
Device Name: GRIPPER® Micro Blunt Cannula Non-Coring Safety Needle
Indications for Use:
This product is indicated for the administration into or withdrawal of fluids from implanted ports. It is designed to help protect against accidental needlestick injuries.
Prescription Use X (Per 21 CFR 801 . 109)
AND/OR
Over-The Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ri C. C
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K1 01180
Traditional 510(k) - GRIPPER Micro 1.25 Smiths Medical ASD, Inc.
Confidential
K101180
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.