SPINESMITH CYNCH SPINAL SYSTEM

K101085 · Spine Smith Partners L.P. · MAX · Jul 1, 2010 · Orthopedic

Device Facts

Record IDK101085
Device NameSPINESMITH CYNCH SPINAL SYSTEM
ApplicantSpine Smith Partners L.P.
Product CodeMAX · Orthopedic
Decision DateJul 1, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Cynch System is indicated intervertebral body fusion of the fumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion.

Device Story

SpineSmith Cynch Spinal System is an intervertebral body fusion device; used for lumbar spine fusion (L2-S1). Configurations include TLIF, PLIF, and T-PLIF approaches. Implants are hollow PEEK Optima LT1 cages with large anterior windows for autograft bone placement; feature superior/inferior teeth for stability and horizontal grooves for instrument positioning. Three radiopaque tantalum markers assist surgeon with placement. Used in surgical setting by physicians; requires supplemental fixation. Benefits include stabilization of spinal segments and facilitation of fusion in patients with degenerative disc disease.

Clinical Evidence

Bench testing only. Mechanical testing performed per ASTM F2077-03 (static/dynamic compression), ASTM F2267-04 (subsidence), and ASTM Draft Standard F-04.25.02.02 (expulsion). No clinical data provided.

Technological Characteristics

Material: PEEK Optima LT1 (ASTM F2026) with tantalum markers (ASTM F-560-05). Design: Hollow-body cage with teeth and grooves. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) and up to Grade I spondylolisthesis requiring intervertebral body fusion of the lumbar spine (L2-S1) after 6 months of failed non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## SpineSmith Cynch Spinal System Kioloss III - 1 2018 JUL - 1 2010 | 510(k) Summary of Safety and Effectiveness | |--------------------------------------------| |--------------------------------------------| | SUBMITTED BY | SpineSmith Partners, LP<br>93 Red River<br>Austin, TX 78701 | |--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------| | ESTABLISHMENT<br>REGISTRATION NUMBER | 3006404071 | | CONTACT PERSON | Laura LeBoeuf<br>Vice President - Quality and Regulatory Affairs<br>Phone: 512-206-0770<br>Fax: 512-637-6750<br>Email: lleboeuf@spinesmithusa.com) | | SUBMISSION PREPARED BY | Lisa Peterson<br>Kaedon Consulting, LLC<br>Phone: 512-507-0746 | | DATE PREPARED | June 28, 2010 | | CLASSIFICATION | MAX 888.3080- Intervertebral Fusion Device with Bone<br>Graft, Lumbar | | COMMON NAME | Intervertebral Body Fusion Device | | PROPRIETARY NAME | SpineSmith Cynch Spinal System | #### IDENTIFICATION OF PREDICATE DEVICES: The SpineSmith Cynch System was determined to be substantially equivalent to the previously cleared Cynch System (K090376, SpineSmith; Cleared 2/16/2009) and the down classified Lumbar I/F cage (P960025, Depuy; Approved 3/4/2005). ### DEVICE DESCRIPTION: The SpineSmith Cynch System is offered in two (2) configurations of various sizes. The configurations are designed based on indicated spinal implant level and surgical approach. The system consists of the Cynch TLIF device for transforaminal lumbar approach and the Cynch PLIF/T-PLIF device, which may be implanted via a posterior lumbar bi-lateral approach or transverse lumbar approach. The Cynch System implants are manufactured from PEEK Optima LT1 and contain three (3) radiopaque tantalum markers to assist the surgeon with proper placement of the device. The Cynch System implants are hollow to allow for the placement of autograft bone, and have large anterior graft windows to facilitate fusion. There are teeth on the superior and inferior surface of the implants to provide increased stability and inhibit movement of the implant. The implants have horizontal grooves to locate the implant holder superior/inferior and rotationally. Additionally, a cross-bar is provided to minimize buckling during insertion. {1}------------------------------------------------ ## INDICATIONS: The Cynch System is indicated for intervertebral body fusion of the lumbar spine, from L2 to SI, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. # SUMMARY OF TECHNOLOGICAL CHARACTERISTICS: The purpose of this submission is to add the Cynch PLIF/T-PLIF device, which may be implanted yia a posterior lumbar bi-lateral approach or transverse lumbar approach. The Cynch System implants are manufactured from PEEK Optima LT1 and contain three (3) radiopaque tantalum markers to assist the surgeon with proper placement of the subject device (Cynch PLIF / T-PLIF) has similar technological characteristics as the predicate devices identified above (SpineSmith's Cynch System per K090376, and Depuy's Lumbar I/F cage per P960025). Specifically, the following characteristics support this conclusion: - Intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. - Hollow-body implant design to allow for placement of autograft bone - - -Implant material: PEEK Optima LT1 per ASTM F2026 with radiopaque tantalum markers per ASTM F-560-05 - Substantially equivalent results of non-clinical testing relative to static and dynamic compression (per ASTM F2077-03), subsidence (per ASTM F2267-04), and expulsion (per ASTM Draft Standard F-04.25.02.02) # DISCUSSION OF NON-CLINICAL TESTING: The following non-clinical tests were conducted: - Static and dynamic compression testing, conducted in accordance with ASTM F2077-03 - - Subsidence testing, conducted in accordance with ASTM F2267-04 . - . Expulsion testing, conducted in accordance with ASTM Draft Standard F-04.25.02.02 # CONCLUSIONS: The subject and predicate devices share the same intended use, primary implant design and material of manufacture. The non-clinical mechanical test results demonstrate that any minor differences do not impact device performance as compared to the predicates and demonstrate that the Cynch System is substantially equivalent to the predicate devices. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles a stylized caduceus or a representation of human services. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 JUL - 1 2010 SpineSmith Partners, LP % Ms. Laura LeBoeuf 93 Red River Austin, Texas 78701 Re: K101085 Trade/Device Name: SpineSmith Cynch Spinal System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: June 08, 2010 Received: June 09, 2010 Dear Ms. LeBoeuf: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {3}------------------------------------------------ Page 2 - Ms. Laura LeBoeuf comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Barbara Spiehm Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### INDICATIONS FOR USE 510(k) Number (if known): #### Device Name: SpineSmith Cynch Spinal System Indications for Use: The Cynch System is indicated intervertebral body fusion of the fumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) signature (Division Sign-Oil) for Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number k101085 93 Red River Austin. Texas 78701 SpineSmith Partners, LP Phone: 512-637-2068 Fax: 512-302-4920 لدا
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...